AD EVAL-AD7327CB

500 kSPS, 8-Channel, Software-Selectable,
True Bipolar Input, 12-Bit Plus Sign ADC
AD7327
FEATURES
GENERAL DESCRIPTION
1
FUNCTIONAL BLOCK DIAGRAM
VDD
REFIN/OUT
VCC
AD7327
VIN0
2.5V
VREF
VIN1
VIN2
VIN3
VIN4
VIN5
VIN6
VIN7
I/P
MUX
T/H
13-BIT
SUCCESSIVE
APPROXIMATION
ADC
TEMPERATURE
INDICATOR
DOUT
CONTROL LOGIC
AND REGISTERS
CHANNEL
SEQUENCER
SCLK
CS
DIN
VDRIVE
AGND
VSS
05401-001
12-bit plus sign SAR ADC
True bipolar input ranges
Software-selectable input ranges
±10 V, ±5 V, ±2.5 V, 0 V to +10 V
500 kSPS throughput rate
Eight analog input channels with channel sequencer
Single-ended, true differential, and pseudo differential
analog input capability
High analog input impedance
Low power: 17 mW
Temperature indicator
Full power signal bandwidth: 22 MHz
Internal 2.5 V reference
High speed serial interface
Power-down modes
20-lead TSSOP package
iCMOS™ process technology
DGND
Figure 1.
PRODUCT HIGHLIGHTS
The AD7327 is an 8-channel, 12-bit plus sign successive
approximation ADC designed on the iCMOS (industrial
CMOS) process. iCMOS is a process combining high voltage
silicon with submicron CMOS and complementary bipolar
technologies. It enables the development of a wide range of high
performance analog ICs capable of 33 V operation in a footprint
that no previous generation of high voltage parts could achieve.
Unlike analog ICs using conventional CMOS processes, iCMOS
components can accept bipolar input signals while providing
increased performance, dramatically reduced power
consumption, and reduced package size.
1.
The AD7327 can accept true bipolar analog input signals,
±10 V, ±5 V, ±2.5 V, and 0 V to +10 V unipolar signals.
2.
The eight analog inputs can be configured as eight singleended inputs, four true differential inputs, four pseudo
differential inputs, or seven pseudo differential inputs.
3.
500 kSPS serial interface. SPI®-/QSPI™-/DSP-/MICROWIRE™compatible interface.
4.
Low power, 17 mW, at a maximum throughput rate of
500 kSPS.
The AD7327 can accept true bipolar analog input signals. The
AD7327 has four software-selectable input ranges, ±10 V, ±5 V,
±2.5 V, and 0 V to +10 V. Each analog input channel can be
independently programmed to one of the four input ranges.
The analog input channels on the AD7327 can be programmed
to be single-ended, true differential, or pseudo differential.
5.
Channel sequencer.
The ADC contains a 2.5 V internal reference. The AD7327 also
allows for external reference operation. If a 3 V reference is
applied to the REFIN/OUT pin, the AD7327 can accept a true
bipolar ±12 V analog input. Minimum ±12 V VDD and VSS
supplies are required for the ±12 V input range. The ADC has a
high speed serial interface that can operate at throughput rates
up to 500 kSPS.
Table 1. Similar Devices
Device
Number
AD7329
AD7328
AD7324
AD7323
AD7322
AD7321
1
Throughput
Rate
1000 kSPS
1000 kSPS
1000 kSPS
500 kSPS
1000 kSPS
500 kSPS
Number of bits
12-bit plus sign
12-bit plus sign
12-bit plus sign
12-bit plus sign
12-bit plus sign
12-bit plus sign
Number of
Channels
8
8
4
4
2
2
Protected by U.S. Patent No. 6,731,232.
Rev. 0
Information furnished by Analog Devices is believed to be accurate and reliable. However, no
responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other
rights of third parties that may result from its use. Specifications subject to change without notice. No
license is granted by implication or otherwise under any patent or patent rights of Analog Devices.
Trademarks and registered trademarks are the property of their respective owners.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.
Tel: 781.329.4700
www.analog.com
Fax: 781.461.3113
©2006 Analog Devices, Inc. All rights reserved.
AD7327
TABLE OF CONTENTS
Features .............................................................................................. 1
Control Register ......................................................................... 23
Functional Block Diagram .............................................................. 1
Sequence Register....................................................................... 25
General Description ......................................................................... 1
Range Registers........................................................................... 25
Product Highlights ........................................................................... 1
Sequencer Operation ..................................................................... 26
Revision History ............................................................................... 2
Reference ..................................................................................... 28
Specifications..................................................................................... 3
VDRIVE ............................................................................................ 28
Timing Specifications .................................................................. 7
Temperature Indicator............................................................... 28
Absolute Maximum Ratings............................................................ 8
Modes of Operation ....................................................................... 29
ESD Caution.................................................................................. 8
Normal Mode.............................................................................. 29
Pin Configuration and Function Descriptions............................. 9
Full Shutdown Mode.................................................................. 29
Typical Performance Characteristics ........................................... 10
Autoshutdown Mode ................................................................. 30
Terminology .................................................................................... 14
Autostandby Mode..................................................................... 30
Theory of Operation ...................................................................... 16
Power vs. Throughput Rate....................................................... 31
Circuit Information.................................................................... 16
Serial Interface ................................................................................ 32
Converter Operation.................................................................. 16
Microprocessor Interfacing........................................................... 33
Analog Input Structure.............................................................. 17
AD7327 to ADSP-21xx.............................................................. 33
Typical Connection Diagram ................................................... 19
AD7327 to ADSP-BF53x ........................................................... 33
Analog Input ............................................................................... 19
Application Hints ........................................................................... 34
Driver Amplifier Choice............................................................ 21
Layout and Grounding .............................................................. 34
Registers ........................................................................................... 22
Outline Dimensions ....................................................................... 35
Addressing Registers .................................................................. 22
Ordering Guide .......................................................................... 35
REVISION HISTORY
1/06—Revision 0: Initial Version
Rev. 0 | Page 2 of 36
AD7327
SPECIFICATIONS
VDD = 12 V to 16.5 V, VSS = −12 V to −16.5 V, VCC = 2.7 V to 5.25 V, VDRIVE = 2.7 V to 5.25 V, VREF = 2.5 V to 3.0 V internal/external,
fSCLK = 10 MHz, fS = 500 kSPS, TA = TMAX to TMIN, unless otherwise noted.
Table 2.
Parameter 1
DYNAMIC PERFORMANCE
Signal-to-Noise Ratio (SNR) 2
Signal-to-Noise + Distortion
(SINAD)2
Min
B Version
Typ
Max
Unit
76
75.5
72.5
dB
dB
dB
72
dB
75
dB
74
76
dB
dB
72.5
dB
72
Total Harmonic Distortion
(THD)2
−80
dB
−79
dB
dB
dB
dB
dB
−82
−77
−79
−80
Peak Harmonic or Spurious
Noise (SFDR)2
−81
dB
−80
dB
dB
dB
−82
−78
−80
Intermodulation Distortion
(IMD)2
Second-Order Terms
Third-Order Terms
Aperture Delay 3
Aperture Jitter3
Common-Mode Rejection
(CMRR)2
Channel-to-Channel Isolation2
Full Power Bandwidth
−79
dB
−88
−90
7
50
−79
dB
dB
ns
ps
dB
−72
dB
22
5
MHz
MHz
Rev. 0 | Page 3 of 36
Test Conditions/Comments
FIN = 50 kHz sine wave
Differential mode, VCC = 4.75 V to 5.25 V
Differential mode, VCC < 4.75 V
Single-ended/pseudo differential mode; ±10 V, ±2.5 V
and ±5 V ranges, VCC = 4.75 V to 5.25 V
Single-ended/pseudo differential mode ; 0 V to 10 V
VCC = 4.75 V to 5.25 V and all ranges at VCC < 4.75 V
Differential mode; ±2.5 V and ±5 V ranges
Differential mode; 0 V to 10 V
Differential mode; ±10 V range
Single-ended/pseudo differential mode; ±2.5 V and
±5 V ranges
Single-ended/pseudo differential mode; 0 V to +10 V
and ±10 V ranges
Differential mode; ±2.5 V and ±5 V ranges
Differential mode; 0 V to 10 V ranges
Differential mode; ±10 V range
Single-ended/pseudo differential mode; ±5 V range
Single-ended/pseudo differential mode; ±2.5 V range
Single-ended/pseudo differential mode; 0 V to +10 V
and ±10 V ranges
Differential mode; ±2.5 V and ±5 V ranges
Differential mode; 0 V to 10 V ranges
Differential mode; ±10 V ranges
Single-ended/pseudo differential mode; ±5 V range
Single-ended/pseudo differential mode; ±2.5 V range
Single-ended/pseudo differential mode; 0 V to +10 V
and ±10 V ranges
fa = 50 kHz, fb = 30 kHz
Up to 100 kHz ripple frequency; see Figure 17
FIN on unselected channels up to 100 kHz; see Figure
14
At 3 dB
At 0.1 dB
AD7327
Parameter 1
DC ACCURACY 4
Resolution
No Missing Codes
Min
B Version
Typ
Max
13
12-bit
plus sign
11-bit
plus sign
Integral Nonlinearity2
±1.1
±1
−0.7/+1.2
Differential Nonlinearity2
Offset Error Match2, 5
Gain Error2, 5
Gain Error Match2, 5
Positive Full-Scale Error2, 6
Positive Full-Scale Error Match2, 6
Bipolar Zero Error2, 6
Bipolar Zero Error Match2, 6
Negative Full-Scale Error2, 6
Negative Full-Scale Error Match2, 6
Test Conditions/Comments
Bits
Bits
Differential mode
Bits
Single-ended/pseudo differential mode
LSB
LSB
Differential mode; VCC = 3 V to 5.25 V, typ for VCC = 2.7 V
Single-ended/pseudo differential mode, VCC = 3 V to
5.25 V, typ for VCC = 2.7 V
Single-ended/pseudo differential mode
(LSB = FSR/8192)
Differential mode; guaranteed no missing codes to
13 bits
Single-ended mode; guaranteed no missing codes to
12 bits
Single-ended/psuedo differential mode
(LSB = FSR/8192)
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
Single-ended/pseudo differential mode
Differential mode
LSB
−0.9/+1.2
LSB
±0.9
LSB
−0.7/+1
Offset Error2, 5
Unit
LSB
−4/+9
−7/+10
±0.6
±0.5
±8
±14
±0.5
±0.5
±4
±7
±0.5
±0.5
±8.5
±7.5
±0.5
±0.5
±4
±6
±0.5
±0.5
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
LSB
Rev. 0 | Page 4 of 36
AD7327
Parameter 1
ANALOG INPUT
Input Voltage Ranges
(Programmed via Range
Registers)
B Version
Typ
Min
Max
Unit
Test Conditions/Comments
±10
V
Reference = 2.5 V; see Table 6
VDD = 10 V min, VSS = −10 V min, VCC = 2.7 V to 5.25 V
±5
±2.5
0 to 10
V
V
V
±3.5
±6
±5
+3/−5
V
V
V
V
nA
nA
pF
pF
pF
pF
Pseudo Differential VIN(−)
Input Range
DC Leakage Current
±80
3
13.5
16.5
21.5
3
Input Capacitance3
REFERENCE INPUT/OUTPUT
Input Voltage Range
Input DC Leakage Current
Input Capacitance
Reference Output Voltage
Reference Output Voltage Error
@ 25°C
Reference Output Voltage
TMIN to TMAX
Reference Temperature
Coefficient
Reference Output Impedance
LOGIC INPUTS
Input High Voltage, VINH
Input Low Voltage, VINL
Input Current, IIN
Input Capacitance, CIN3
LOGIC OUTPUTS
Output High Voltage, VOH
Output Low Voltage, VOL
Floating-State Leakage Current
Floating-State Output
Capacitance3
Output Coding
CONVERSION RATE
Conversion Time
Track-and-Hold Acquisition
Time2, 3
Throughput Rate
2.5
3
±1
±5
V
μA
pF
V
mV
±10
mV
25
ppm/°C
10
2.5
3
7
VDD = 5 V min, VSS = −5 V min, VCC = 2.7 V to 5.25 V
VDD = 5 V min, VSS = −5 V min, VCC = 2.7 V to 5.25 V
VDD = 10 V min, VSS = AGND min, VCC = 2.7 V to 5.25 V
VDD = 16.5 V, VSS = −16.5 V, VCC = 5 V; see Figure 40 and
Figure 41
Reference = 2.5 V; range = ±10 V
Reference = 2.5 V; range = ±5 V
Reference = 2.5 V; range = ±2.5 V
Reference = 2.5 V; range = 0 V to +10 V
VIN = VDD or VSS
Per input channel, VIN = VDD or VSS
When in track, ±10 V range
When in track, ±5 V and 0 V to +10 V ranges
When in track, ±2.5 V range
When in hold, all ranges
ppm/°C
Ω
2.4
0.8
0.4
±1
10
VDRIVE −
0.2 V
0.4
±1
5
V
V
V
μA
pF
VCC = 4.75 V to 5.25 V
VCC = 2.7 to 3.6 V
VIN = 0 V or VDRIVE
V
ISOURCE = 200 μA
V
μA
pF
ISINK = 200 μA
Straight natural binary
Twos complement
Coding bit set to 1 in control register
Coding bit set to 0 in control register
1.6
305
μs
ns
16 SCLK cycles with SCLK = 10 MHz
Full-scale step input; see the Terminology section
500
kSPS
See the Serial Interface section
Rev. 0 | Page 5 of 36
AD7327
Parameter 1
POWER REQUIREMENTS
VDD
VSS
VCC
VDRIVE
Normal Mode (Static)
Normal Mode (Operational)
IDD
ISS
ICC and IDRIVE
Autostandby Mode (Dynamic)
IDD
ISS
ICC and IDRIVE
Autoshutdown Mode (Static)
IDD
ISS
ICC and IDRIVE
Full Shutdown Mode
IDD
ISS
ICC and IDRIVE
POWER DISSIPATION
Normal Mode (Operational)
Full Shutdown Mode
Min
B Version
Typ
12
−12
2.7
2.7
Max
Unit
16.5
−16.5
5.25
5.25
V
V
V
V
mA
0.9
Test Conditions/Comments
Digital inputs = 0 V or VDRIVE
See Table 6
See Table 6
See Table 6
180
205
2
μA
μA
mA
100
110
0.75
μA
μA
mA
1
1
1
μA
μA
μA
1
1
1
μA
μA
μA
VDD/VSS = ±16.5 V, VCC/VDRIVE = 5.25 V
fSAMPLE = 500 kSPS
VDD = 16.5 V
VSS = −16.5 V
VCC/VDRIVE = 5.25 V
fSAMPLE = 250 kSPS
VDD = 16.5 V
VSS = −16.5 V
VCC/VDRIVE = 5.25 V
SCLK on or off
VDD = 16.5 V
VSS = −16.5 V
VCC/VDRIVE = 5.25 V
SCLK on or off
VDD = 16.5 V
VSS = −16.5 V
VCC/VDRIVE = 5.25 V
17
38.25
mW
μW
VDD = 16.5 V, VSS = −16.5 V, VCC = 5.25 V
VDD = 16.5 V, VSS = −16.5 V, VCC = 5.25 V
1
Temperature range is −40°C to +85°C.
See the Terminology section.
Sample tested during initial release to ensure compliance.
4
For dc accuracy specifications, the LSB size for differential mode is FSR/8192. For single-ended mode/pseudo differential mode, the LSB size is FSR/4096, unless
otherwise noted.
5
Unipolar 0 V to 10 V range with straight binary output coding.
6
Bipolar range with twos complement output coding.
2
3
Rev. 0 | Page 6 of 36
AD7327
TIMING SPECIFICATIONS
VDD = 12 V to 16.5 V, VSS = −12 V to −16.5 V, VCC = 2.7 V to 5.25 V, VDRIVE = 2.7 V to 5.25 V, VREF = 2.5 V to 3.0 V internal/external,
TA = TMAX to TMIN. Timing specifications apply with a 32 pF load, unless otherwise noted. 1
Table 3.
Parameter
fSCLK
tCONVERT
tQUIET
t1
t2 2
t3
t4
t5
t6
t7
t8
t9
t10
tPOWER-UP
2
Unit
kHz min
MHz max
ns max
ns min
ns min
ns min
ns min
ns max
ns max
ns min
ns min
ns min
ns max
ns min
ns min
ns min
ns max
μs max
25
μs typ
25
Description
VDRIVE ≤ VCC
tSCLK = 1/fSCLK
Minimum time between end of serial read and next falling edge of CS
Minimum CS pulse width
CS to SCLK set-up time; bipolar input ranges (±10 V, ±5 V, ±2.5 V)
Unipolar input range (0 V to 10 V)
Delay from CS until DOUT three-state disabled
Data access time after SCLK falling edge
SCLK low pulse width
SCLK high pulse width
SCLK to data valid hold time
SCLK falling edge to DOUT high impedance
SCLK falling edge to DOUT high impedance
DIN set-up time prior to SCLK falling edge
DIN hold time after SCLK falling edge
Power-up from autostandby
Power-up from full shutdown/autoshutdown mode, internal
reference
Power-up from full shutdown/autoshutdown mode, external
reference
Sample tested during initial release to ensure compliance. All input signals are specified with tr = tf = 5 ns (10% to 90% of VDRIVE) and timed from a voltage level of 1.6 V.
When using the 0 V to 10 V unipolar range, running at 500 kSPS throughput rate with t2 at 20 ns, the mark space ratio needs to be limited to 50:50.
t1
CS
tCONVERT
t2
SCLK
t6
1
2
3
4
3 IDENTIFICATION BITS
t3
ADD1
DOUT
THREE- ADD2
t9
STATE
DIN
WRITE
REG
SEL1
ADD0
SIGN
5
t4
13
14
DB11
15
16
t5
t7
DB10
DB2
t8
DB1
DB0
t10
REG
SEL2
tQUIET
THREE-STATE
MSB
LSB
Figure 2. Serial Interface Timing Diagram
Rev. 0 | Page 7 of 36
DON’T
CARE
05401-002
1
Limit at TMIN, TMAX
VCC < 4.75 V
VCC = 4.75 V to 5.25 V
50
50
10
10
16 × tSCLK
16 × tSCLK
75
60
12
5
25
20
45
35
26
14
57
43
0.4 × tSCLK
0.4 × tSCLK
0.4 × tSCLK
0.4 × tSCLK
13
8
40
22
10
9
4
4
2
2
750
750
500
500
AD7327
ABSOLUTE MAXIMUM RATINGS
TA = 25°C, unless otherwise noted
Table 4.
Parameter
VDD to AGND, DGND
VSS to AGND, DGND
VDD to VCC
VCC to AGND, DGND
VDRIVE to AGND, DGND
AGND to DGND
Analog Input Voltage to AGND 1
Digital Input Voltage to DGND
Digital Output Voltage to GND
REFIN to AGND
Input Current to Any Pin
Except Supplies 2
Operating Temperature Range
Storage Temperature Range
Junction Temperature
TSSOP Package
θJA Thermal Impedance
θJC Thermal Impedance
Pb-Free Temperature, Soldering
Reflow
ESD
Rating
−0.3 V to +16.5 V
+0.3 V to −16.5 V
VCC − 0.3 V to 16.5 V
−0.3 V to +7 V
−0.3 V to +7 V
−0.3 V to +0.3 V
VSS − 0.3 V to VDD + 0.3 V
−0.3 V to +7 V
−0.3 V to VDRIVE + 0.3 V
−0.3 V to VCC + 0.3 V
±10 mA
Stresses above those listed under Absolute Maximum Ratings
may cause permanent damage to the device. This is a stress
rating only; functional operation of the device at these or any
other conditions above those indicated in the operational
section of this specification is not implied. Exposure to absolute
maximum rating conditions for extended periods may affect
device reliability.
−40°C to +85°C
−65°C to +150°C
150°C
143°C/W
45°C/W
260(0)°C
2.5 kV
1
If the analog inputs are driven from alternative VDD and VSS supply circuitry,
Schottky diodes should be placed in series with the AD7327’s VDD and VSS
supplies.
2
Transient currents of up to 100 mA do not cause SCR latch-up.
ESD CAUTION
ESD (electrostatic discharge) sensitive device. Electrostatic charges as high as 4000 V readily accumulate on
the human body and test equipment and can discharge without detection. Although this product features
proprietary ESD protection circuitry, permanent damage may occur on devices subjected to high energy
electrostatic discharges. Therefore, proper ESD precautions are recommended to avoid performance
degradation or loss of functionality.
Rev. 0 | Page 8 of 36
AD7327
CS 1
20
SCLK
2
19
DGND
DGND 3
18
DOUT
DIN
AGND 4
REFIN/OUT 5
AD7327
TOP VIEW
(Not to Scale)
17
VDRIVE
16
VCC
VSS 6
15
VDD
VIN0 7
14
VIN2
VIN1 8
13
VIN3
VIN4 9
12
VIN6
VIN5 10
11
VIN7
05401-003
PIN CONFIGURATION AND FUNCTION DESCRIPTIONS
Figure 3. TSSOP Pin Configuration
Table 5. Pin Function Descriptions
Pin No.
1
Mnemonic
CS
2
DIN
3, 19
DGND
4
AGND
5
REFIN/OUT
6
7, 8, 14, 13, 9,
10, 12, 11
VSS
VIN0 to VIN7
15
16
VDD
VCC
17
VDRIVE
18
DOUT
20
SCLK
Description
Chip Select. Active low logic input. This input provides the dual function of initiating conversions on
the AD7327 and frames the serial data transfer.
Data In. Data to be written to the on-chip registers is provided on this input and is clocked into the
AD7327 on the falling edge of SCLK (see the Registers section).
Digital Ground. Ground reference point for all digital circuitry on the AD7327. The DGND and AGND
voltages ideally should be at the same potential and must not be more than 0.3 V apart, even on a
transient basis.
Analog Ground. Ground reference point for all analog circuitry on the AD7327. All analog input signals
and any external reference signal should be referred to this AGND voltage. The AGND and DGND
voltages ideally should be at the same potential and must not be more than 0.3 V apart, even on a
transient basis.
Reference Input/Reference Output. The on-chip reference is available on this pin for use external to the
AD7327. The nominal internal reference voltage is 2.5 V, which appears at this pin. A 680 nF capacitor
should be placed on the reference pin (see the Reference section). Alternatively, the internal reference
can be disabled and an external reference applied to this input. On power-up, the external reference
mode is the default condition.
Negative Power Supply Voltage. This is the negative supply voltage for the analog input section.
Analog Input 0 to Analog Input 7. The analog inputs are multiplexed into the on-chip track-and-hold.
The analog input channel for conversion is selected by programming the Channel Address Bit ADD2
through Bit ADD0 in the control register. The inputs can be configured as eight single-ended inputs,
four true differential input pairs, four pseudo differential inputs, or seven pseudo differential inputs.
The configuration of the analog inputs is selected by programming the mode bits, Bit Mode 1 and
Bit Mode 0, in the control register. The input range on each input channel is controlled by programming the range registers. Input ranges of ±10 V, ±5 V, ±2.5 V, and 0 V to +10 V can be selected on each
analog input channel when a +2.5 V reference voltage is used (see the Registers section).
Positive Power Supply Voltage. This is the positive supply voltage for the analog input section.
Analog Supply Voltage, 2.7 V to 5.25 V. This is the supply voltage for the ADC core on the AD7327.
This supply should be decoupled to AGND.
Logic Power Supply Input. The voltage supplied at this pin determines at what voltage the interface
operates. This pin should be decoupled to DGND. The voltage at this pin may be different to that at VCC,
but it should not exceed VCC by more than 0.3 V.
Serial Data Output. The conversion output data is supplied to this pin as a serial data stream. The bits
are clocked out on the falling edge of the SCLK input, and 16 SCLKs are required to access the data. The
data stream consists of three channel identification bits, the sign bit, and 12 bits of conversion data.
The data is provided MSB first (see the Serial Interface section).
Serial Clock, Logic Input. A serial clock input provides the SCLK used for accessing the data from the
AD7327. This clock is also used as the clock source for the conversion process.
Rev. 0 | Page 9 of 36
AD7327
TYPICAL PERFORMANCE CHARACTERISTICS
1.0
0
4096 POINT FFT
VCC = VDRIVE = 5V
VDD, VSS = ±15V
TA = 25°C
INT/EXT 2.5V REFERENCE
±10V RANGE
FIN = 50kHz
SNR = 77.30dB
SINAD = 76.85dB
THD = –86.96dB
SFDR = –88.22dB
SNR (dB)
–40
–60
–80
0.6
0.4
INL ERROR (LSB)
–20
VCC = VDRIVE = 5V INT/EXT 2.5V REFERENCE
TA = 25°C
±10V RANGE
VDD, VSS = ±15V
+INL = +0.55LSB
–INL = –0.68LSB
0.8
0.2
0
–0.2
–0.4
–100
–0.6
–120
50
100
150
200
250
FREQUENCY (kHz)
0
8192
1024
2048
3072
4096
5120
6144
7168
512
1536
2560
3584
4608
5632
6656
7680
CODE
05401-007
–1.0
0
05401-004
–140
–0.8
Figure 7. Typical INL True Differential Mode
Figure 4. FFT True Differential Mode
1.0
0
4096 POINT FFT
VCC = VDRIVE = 5V
VDD, VSS = ±15V
TA = 25°C
INT/EXT 2.5V REFERENCE
±10V RANGE
FIN = 50kHz
SNR = 74.67dB
SINAD = 74.03dB
THD = –82.68dB
SFDR = –85.40dB
–80
–100
0.4
0.2
0
–0.2
–0.4
–0.8
–120
0
50
100
150
200
250
FREQUENCY (kHz)
05401-005
–1.0
–140
Figure 5. FFT Single-Ended Mode
8192
1024
2048
3072
4096
5120
6144
7168
512
1536
2560
3584
4608
5632
6656
7680
CODE
1.0
0.8
0.8
0.6
0.6
0.4
INL ERROR (LSB)
0.4
0.2
0
–0.2
VCC = VDRIVE = 5V
TA = 25°C
VDD, VSS = ±15V
INT/EXT 2.5V REFERENCE
±10V RANGE
+DNL = +0.72LSB
–DNL = –0.22LSB
–0.4
–0.6
–0.8
0
8192
1024
2048
3072
4096
5120
6144
7168
512
1536
2560
3584
4608
5632
6656
7680
CODE
0.2
0
–0.2
VCC = VDRIVE = 5V
TA = 25°C
VDD, VSS = ±15V
–0.6
INT/EXT 2.5V REFERENCE
±10V RANGE
–0.8
+INL = +0.87LSB
–INL = –0.49LSB
–1.0
0
8192
1024
2048
3072
4096
5120
6144
7168
512
1536
2560
3584
4608
5632
6656
7680
CODE
–0.4
05401-006
DNL ERROR (LSB)
0
Figure 8. Typical DNL Single-Ended Mode
1.0
–1.0
VCC = VDRIVE = 5V
±10V RANGE
TA = 25°C
+DNL = +0.79LSB
–DNL = –0.38LSB
VDD, VSS = ±15V
INT/EXT 2.5V REFERENCE
–0.6
05401-043
–60
0.6
Figure 6. Typical DNL True Differential Mode
Figure 9. Typical INL Single-Ended Mode
Rev. 0 | Page 10 of 36
05401-044
SNR (dB)
–40
0.8
DNL ERROR (LSB)
–20
AD7327
–50
75
±2.5V SE
70
±10V SE
±10V DIFF
–75
0V TO +10V DIFF
–80
±5V SE
–100
10
1000
100
ANALOG INPUT FREQUENCY (kHz)
50
10
–50
0V TO +10V SE
±10V SE
–70
±10V DIFF
–75
0V TO +10V DIFF
–80
±5V SE
–85
±5V DIFF
–90
±2.5V SE
–95
–55
VCC = 3V
–60
VCC = 5V
–65
–70
–75
–80
VDD/VSS = ±12V
SINGLE-ENDED MODE
fS = 500kSPS
TA = 25°C
50kHz ON SELECTED CHANNEL
–85
–90
±2.5V DIFF
1000
ANALOG INPUT FREQUENCY (kHz)
–95
05401-061
100
100
0
200
80
10k
±5V DIFF
±2.5V DIFF
±5V SE
±2.5V SE
NUMBER OF OCCURRENCES
±10V DIFF
±10V SE
65
0V TO +10V SE
60
VCC = VDRIVE = 3V
VDD/VSS = ±12V
TA = 25°C
fS = 500kSPS
INTERNAL REFERENCE
100
1000
ANALOG INPUT FREQUENCY (kHz)
8k
7k
500
600
Figure 12. SINAD vs. Analog Input Frequency
for Single-Ended (SE) and True Differential Mode (Diff) at 3 V VCC
VCC = 5V
VDD/VSS = ±12V
RANGE = ±10V
10k SAMPLES
TA = 25°C
6k
5k
4k
3k
2k
1k
05401-062
55
9469
9k
0V TO +10V DIFF
70
400
Figure 14. Channel-to-Channel Isolation
Figure 11. THD vs. Analog Input Frequency
for Single-Ended (SE) and True Differential Mode (Diff) at 5 V VCC
75
300
FREQUENCY OF INPUT NOISE (kHz)
05401-012
THD (dB)
CHANNEL-TO-CHANNEL ISOLATION (dB)
VCC = VDRIVE = 5V
VDD/VSS = ±12V
TA = 25°C
fS = 500kSPS
INTERNAL REFERENCE
–65
SINAD (dB)
1000
Figure 13. SINAD vs. Analog Input Frequency
for Single-Ended (SE) and True Differential Mode (Diff) at 5 V VCC
–50
50
10
100
ANALOG INPUT FREQUENCY (kHz)
Figure 10. THD vs. Analog Input Frequency
for Single-Ended (SE) and True Differential Mode (Diff) at 3 V VCC
–100
10
VCC = VDRIVE = 5V
VDD/VSS = ±12V
TA = 25°C
fS = 500kSPS
INTERNAL REFERENCE
55
±2.5V SE
–95
–60
0V TO +10V SE
±2.5V DIFF
–90
0V TO +10V DIFF
±10V SE
65
60
±5V DIFF
–85
±10V DIFF
05401-063
–70
0
0
–2
228
–1
303
0
1
0
2
CODE
Figure 15. Histogram of Codes, True Differential Mode
Rev. 0 | Page 11 of 36
05401-013
THD (dB)
–65
–55
±5V SE
0V TO +10V SE
SINAD (dB)
–60
±5V DIFF
±2.5V DIFF
05401-060
–55
80
VCC = VDRIVE = 3V
VDD/VSS = ±12V
TA = 25°C
fS = 500kSPS
INTERNAL REFERENCE
AD7327
8k
2.0
7600
VCC = 5V
VDD/VSS = ±12V
RANGE = ±10V
10k SAMPLES
TA = 25°C
6k
1.5
1.0
INL ERROR (LSB)
5k
4k
3k
0.5
INL = 500kSPS
0
–0.5
–1.0
1165
0
23
–3
–2
–1
0
11
0
2
3
1
–2.0
05401-014
0
±5V RANGE
VCC = VDRIVE = 5V
INTERNAL REFERENCE
SINGLE-ENDED MODE
–1.5
1k
CODE
5
–50
–55
–55
–60
–60
–65
–65
CMRR (dB)
PSRR (dB)
VCC = 5V
–75
VCC = 3V
–85
200
0
400
600
800
1000
1200
VCC = 5V
VCC = 3V
–75
VDD = 12V
–80
–100
VSS = –12V
0
200
400
600
800
–50
1.5
–55
–60
–65
THD (dB)
0.5
DNL = 500kSPS
0
–0.5
VCC = VDRIVE = 5V
VDD/VSS = ±12V
TA = 25°C
INTERNAL REF
RANGE = ±10V AND ±2.5V
fS = 500kSPS
DIFFERENTIAL MODE
±5V RANGE
VCC = VDRIVE = 5V
INTERNAL REFERENCE
SINGLE-ENDED MODE
–70
–75
±2.5V RANGE
RIN = 9000Ω
RIN = 5500Ω
RIN = 2000Ω
RIN = 100Ω
RIN = 12Ω
–80
–90
–95
11
13
15
17
±VDD/VSS SUPPLY VOLTAGE (V)
Figure 18. DNL Error vs. Supply Voltage at 500 kSPS
19
–100
10
05401-049
9
1200
±10V RANGE
RIN = 4000Ω
RIN = 3000Ω
RIN = 2000Ω
RIN = 1000Ω
RIN = 100Ω
RIN = 12Ω
–85
–1.0
7
1000
SUPPLY RIPPLE FREQUENCY (kHz)
1.0
DNL ERROR (LSB)
100mV p-p SINE WAVE ON EACH SUPPLY
NO DECOUPLING
SINGLE-ENDED MODE
fS = 500kSPS
–70
2.0
5
19
Figure 20. PSRR vs. Supply Ripple Frequency Without Supply Decoupling
Figure 17. CMRR vs. Common-Mode Ripple Frequency
–1.5
17
–95
RIPPLE FREQUENCY (kHz)
–2.0
15
–90
05401-055
–95
–100
13
–85
DIFFERENTIAL MODE
FIN = 50kHz
VDD/VSS = ±12V
fS = 500kSPS
TA = 25°C
–90
11
Figure 19. INL Error vs. Supply Voltage at 500 kSPS
–50
–80
9
±VDD/VSS SUPPLY VOLTAGE (V)
Figure 16. Histogram of Codes, Single-Ended Mode
–70
7
05401-054
1201
05401-050
2k
100
INPUT FREQUENCY (kHz)
1000
05401-064
NUMBER OF OCCURENCES
7k
Figure 21. THD vs. Analog Input Frequency for Various Source Impedances,
True Differential Mode
Rev. 0 | Page 12 of 36
AD7327
–50
THD (dB)
VCC = VDRIVE = 5V
–55 VDD/VSS = ±12V
TA = 25°C
–60 INTERNAL REF
RANGE = ±10V AND ±2.5V
–65 fS = 500kSPS
SINGLE-ENDED MODE
±10V RANGE
RIN = 4000Ω
RIN = 2000Ω
RIN = 1000Ω
RIN = 100Ω
RIN = 50Ω
–70
–75
±2.5V RANGE
RIN = 4700Ω
RIN = 3000Ω
RIN = 1000Ω
RIN = 100Ω
RIN = 50Ω
–80
–85
–90
100
INPUT FREQUENCY (kHz)
1000
05401-065
–95
–100
10
Figure 22. THD vs. Analog Input Frequency for Various Source Impedances,
Single-Ended Mode
Rev. 0 | Page 13 of 36
AD7327
TERMINOLOGY
Differential Nonlinearity
This is the difference between the measured and the ideal 1 LSB
change between any two adjacent codes in the ADC.
Integral Nonlinearity
This is the maximum deviation from a straight line passing
through the endpoints of the ADC transfer function. The
endpoints of the transfer function are zero scale (a point 1 LSB
below the first code transition) and full scale (a point 1 LSB
above the last code transition).
Offset Code Error
This applies to straight binary output coding. It is the deviation
of the first code transition (00 ... 000) to (00 ... 001) from the
ideal, that is, AGND + 1 LSB.
Offset Error Match
This is the difference in offset error between any two input
channels.
Gain Error
This applies to straight binary output coding. It is the deviation
of the last code transition (111 ... 110) to (111 ... 111) from the
ideal (that is, 4 × VREF − 1 LSB, 2 × VREF − 1 LSB, VREF −1 LSB)
after adjusting for the offset error.
Gain Error Match
This is the difference in gain error between any two input
channels.
Bipolar Zero Code Error
This applies when using twos complement output coding and a
bipolar analog input. It is the deviation of the midscale
transition (all 1s to all 0s) from the ideal input voltage, that is,
AGND − 1 LSB.
Bipolar Zero Code Error Match
This refers to the difference in bipolar zero code error between
any two input channels.
Positive Full-Scale Error
This applies when using twos complement output coding and
any of the bipolar analog input ranges. It is the deviation of the
last code transition (011…110) to (011…111) from the ideal
(4 × VREF − 1 LSB, 2 × VREF − 1 LSB, VREF − 1 LSB) after
adjusting for the bipolar zero code error.
Positive Full-Scale Error Match
This is the difference in positive full-scale error between any
two input channels.
Negative Full-Scale Error
This applies when using twos complement output coding and
any of the bipolar analog input ranges. This is the deviation of
the first code transition (10 ... 000) to (10 ... 001) from the ideal
(that is, −4 × VREF + 1 LSB, −2 × VREF + 1 LSB, −VREF + 1 LSB)
after adjusting for the bipolar zero code error.
Negative Full-Scale Error Match
This is the difference in negative full-scale error between any
two input channels.
Track-and-Hold Acquisition Time
The track-and-hold amplifier returns into track mode after the
14th SCLK rising edge. Track-and-hold acquisition time is the
time required for the output of the track-and-hold amplifier to
reach its final value, within ±½ LSB, after the end of a
conversion. For the ±2.5 V range, the specified acquisition time
is the time required for the track-and-hold amplifier to settle to
within ±1 LSB.
Signal to (Noise + Distortion) Ratio
This is the measured ratio of signal to (noise + distortion) at the
output of the A/D converter. The signal is the rms amplitude of
the fundamental. Noise is the sum of all non-fundamental signals
up to half the sampling frequency (fS/2), excluding dc. The ratio
is dependent on the number of quantization levels in the digitization process. The more levels, the smaller the quantization
noise. Theoretically, the signal to (noise + distortion) ratio for
an ideal N-bit converter with a sine wave input is given by
Signal to (Noise + Distortion) = (6.02 N + 1.76) dB
For a 13-bit converter, this is 80.02 dB.
Total Harmonic Distortion
Total harmonic distortion (THD) is the ratio of the rms sum of
harmonics to the fundamental. For the AD7327, it is defined as
THD(dB) = 20 log
V2 2 + V3 2 + V 4 2 + V5 2 + V6 2
V1
where V1 is the rms amplitude of the fundamental, and V2, V3,
V4, V5, and V6 are the rms amplitudes of the second through the
sixth harmonics.
Peak Harmonic or Spurious Noise
Peak harmonic or spurious noise is defined as the ratio of the
rms value of the next largest component in the ADC output
spectrum (up to fS/2, excluding dc) to the rms value of the
fundamental. Normally, the value of this specification is
determined by the largest harmonic in the spectrum, but for
ADCs where the harmonics are buried in the noise floor, the
largest harmonic could be a noise peak.
Rev. 0 | Page 14 of 36
AD7327
Channel-to-Channel Isolation
Channel-to-channel isolation is a measure of the level of crosstalk
between any two channels. It is measured by applying a full-scale,
100 kHz sine wave signal to all unselected input channels and
determining the degree to which the signal attenuates in the
selected channel with a 50 kHz signal. Figure 14 shows the worstcase across all eight channels for the AD7327. The analog input
range is programmed to be the same on all channels.
Intermodulation Distortion
With inputs consisting of sine waves at two frequencies, fa and
fb, any active device with nonlinearities creates distortion products
at sum and difference frequencies of mfa ± nfb, where m, n = 0,
1, 2, 3, and so on. Intermodulation distortion terms are those
for which neither m nor n are equal to 0. For example, the
second-order terms include (fa + fb) and (fa − fb), whereas the
third-order terms include (2fa + fb), (2fa − fb), (fa + 2fb), and
(fa − 2fb).
The AD7327 is tested using the CCIF standard where two input
frequencies near the top end of the input bandwidth are used.
In this case, the second-order terms are usually distanced in
frequency from the original sine waves, whereas the third-order
terms are usually at a frequency close to the input frequencies.
As a result, the second- and third-order terms are specified
separately. The calculation of the intermodulation distortion is
per the THD specification, where it is the ratio of the rms sum
of the individual distortion products to the rms amplitude of
the sum of the fundamentals expressed in decibels.
PSR (Power Supply Rejection)
Variations in power supply affect the full-scale transition but
not the linearity of the converter. Power supply rejection is the
maximum change in the full-scale transition point due to a
change in power supply voltage from the nominal value (see the
Typical Performance Characteristics section).
CMRR (Common-Mode Rejection Ratio)
CMRR is defined as the ratio of the power in the ADC output at
full-scale frequency, f, to the power of a 100 mV sine wave
applied to the common-mode voltage of the VIN+ and VIN−
frequency, fS, as
CMRR (dB) = 10 log (Pf/PfS)
where Pf is the power at frequency f in the ADC output, and PfS
is the power at frequency fS in the ADC output (see Figure 17).
Rev. 0 | Page 15 of 36
AD7327
THEORY OF OPERATION
CIRCUIT INFORMATION
The AD7327 is a fast, 8-channel, 12-bit plus sign, bipolar input,
serial A/D converter. The AD7327 can accept bipolar input
ranges that include ±10 V, ±5 V, and ±2.5 V; it can also accept a
0 V to +10 V unipolar input range. A different analog input
range can be programmed on each analog input channel via the
on-chip registers. The AD7327 has a high speed serial interface
that can operate at throughput rates up to 500 kSPS.
The AD7327 requires VDD and VSS dual supplies for the high
voltage analog input structures. These supplies must be equal to
or greater than the analog input range. See Table 6 for the
requirements of these supplies for each analog input range. The
AD7327 requires a low voltage 2.7 V to 5.25 V VCC supply to
power the ADC core.
Table 6. Reference and Supply Requirements for Each
Analog Input Range
±5
±2.5
0 to +10
Reference
Voltage (V)
2.5
3.0
2.5
3.0
2.5
3.0
2.5
3.0
FullScale
Input
Range
(V)
±10
±12
±5
±6
±2.5
±3
0 to +10
0 to +12
The serial clock input accesses data from the part and provides
the clock source for the successive approximation ADC. The
AD7327 has an on-chip 2.5 V reference. However, the AD7327
can also work with an external reference. On power-up, the
external reference operation is the default option. If the internal
reference is the preferred option, the user must write to the
reference bit in the control register to select the internal
reference operation.
The AD7327 also features power-down options to allow power
savings between conversions. The power-down modes are
selected by programming the on-chip control register, as
described in the Modes of Operation section.
CONVERTER OPERATION
AVCC
(V)
3/5
3/5
3/5
3/5
3/5
3/5
3/5
3/5
Minimum
VDD/VSS (V)
±10
±12
±5
±6
±5
±5
+10/AGND
+12/AGND
It may be necessary to decrease the throughput rate when the
AD7327 is configured with the minimum VDD and VSS supplies
in order to meet the performance specifications (see the Typical
Performance Characteristics section). Figure 31 shows the
change in THD as the VDD and VSS supplies are reduced. For ac
performance at the maximum throughput rate, the THD degrades
slightly as VDD and VSS are reduced. It might therefore be necessary
to reduce the throughput rate when using minimum VDD and
VSS supplies so that there is less degradation of THD and the
specified performance can be maintained. The degradation is
due to an increase in the on resistance of the input multiplexer
when the VDD and VSS supplies are reduced. Figure 18 and
Figure 19 show the change in INL and DNL as the VDD and VSS
voltages are varied. For dc performance when operating at the
maximum throughput rate, as the VDD and VSS supply voltages
are reduced, the typical INL and DNL error remains constant.
The AD7327 is a successive approximation analog-to-digital
converter built around two capacitive DACs. Figure 23 and
Figure 24 show simplified schematics of the ADC in singleended mode during the acquisition and conversion phases,
respectively. Figure 25 and Figure 26 show simplified
schematics of the ADC in differential mode during acquisition
and conversion phases, respectively. The ADC is composed of
control logic, a SAR, and capacitive DACs. In Figure 23 (the
acquisition phase), SW2 is closed and SW1 is in Position A, the
comparator is held in a balanced condition, and the sampling
capacitor array acquires the signal on the input.
CAPACITIVE
DAC
VIN0
B
COMPARATOR
CS
A SW1
SW2
CONTROL
LOGIC
AGND
05401-017
Selected
Analog
Input Range
(V)
±10
The analog inputs can be configured as eight single-ended
inputs, four true differential inputs, four pseudo differential
inputs, or seven pseudo differential inputs. Selection can be
made by programming the mode bits, Mode 0 and Mode 1, in
the control register.
Figure 23. ADC Acquisition Phase (Single-Ended)
When the ADC starts a conversion (Figure 24), SW2 opens and
SW1 moves to Position B, causing the comparator to become
unbalanced. The control logic and the charge redistribution
DAC are used to add and subtract fixed amounts of charge from
the capacitive DAC to bring the comparator back into a
balanced condition. When the comparator is rebalanced, the
conversion is complete. The control logic generates the ADC
output code.
Rev. 0 | Page 16 of 36
AD7327
The ideal transfer characteristic for the AD7327 when twos
complement coding is selected is shown in Figure 27. The ideal
transfer characteristic for the AD7327 when straight binary
coding is selected is shown in Figure 28.
CAPACITIVE
DAC
AGND
011...111
011...110
Figure 25 shows the differential configuration during the
acquisition phase. For the conversion phase, SW3 opens and
SW1 and SW2 move to Position B (see Figure 26). The output
impedances of the source driving the VIN+ and VIN− pins must
match; otherwise, the two inputs have different settling times,
resulting in errors.
ADC CODE
Figure 24. ADC Conversion Phase (Single-Ended)
100...010
100...001
100...000
–FSR/2 + 1LSB
AGND + 1LSB
CAPACITIVE
DAC
B
CONTROL
LOGIC
SW3
111...111
111...110
CS
VREF
CAPACITIVE
DAC
Figure 25. ADC Differential Configuration During Acquisition Phase
A SW1
A SW2
B
SW3
Figure 28. Straight Binary Transfer Characteristic
CS
VREF
CAPACITIVE
DAC
ANALOG INPUT STRUCTURE
Figure 26. ADC Differential Configuration During Conversion Phase
Output Coding
The AD7327 default output coding is set to twos complement.
The output coding is controlled by the coding bit in the control
register. To change the output coding to straight binary coding,
the coding bit in the control register must be set. When
operating in sequence mode, the output coding for each
channel in the sequence is the value written to the coding bit
during the last write to the control register.
Transfer Functions
The analog inputs of the AD7327 can be configured as singleended, true differential, or pseudo differential via the control
register mode bits (see Table 9). The AD7327 can accept true
bipolar input signals. On power-up, the analog inputs operate as
eight single-ended analog input channels. If true differential or
pseudo differential is required, a write to the control register is
necessary after power-up to change this configuration.
Figure 29 shows the equivalent analog input circuit of the
AD7327 in single-ended mode. Figure 30 shows the equivalent
analog input structure in differential mode. The two diodes
provide ESD protection for the analog inputs.
VDD
The designed code transitions occur at successive integer
LSB values (that is, 1 LSB, 2 LSB, and so on). The LSB size is
dependent on the analog input range selected.
D
VIN0
C1
Table 7. LSB Sizes for Each Analog Input Range
Input Range
±10 V
±5 V
±2.5 V
0 V to +10 V
011...111
–FSR/2 + 1LSB
+FSR/2 – 1LSB BIPOLAR RANGES
AGND + 1LSB
+FSR – 1LSB
UNIPOLAR RANGE
ANALOG INPUT
CONTROL
LOGIC
05401-020
VIN–
COMPARATOR
CS
B
111...000
000...010
000...001
000...000
CAPACITIVE
DAC
VIN+
+FSR/2 – 1LSB BIPOLAR RANGES
+FSR – 1LSB
UNIPOLAR RANGE
ANALOG INPUT
05401-022
A SW1
A SW2
ADC CODE
VIN–
AGND – 1LSB
Figure 27. Twos Complement Transfer Characteristic
COMPARATOR
CS
B
05401-019
VIN+
000...001
000...000
111...111
05401-021
CONTROL
LOGIC
SW2
Full-Scale Range/8192 Codes
20 V
10 V
5V
10 V
LSB Size
2.441 mV
1.22 mV
0.61 mV
1.22 mV
Rev. 0 | Page 17 of 36
D
VSS
R1
C2
05401-023
A SW1
05401-018
COMPARATOR
CS
B
VIN0
Figure 29. Equivalent Analog Input Circuit (Single-Ended)
AD7327
VDD
VIN+
C1
R1
The AD7327 enters track on the 14th SCLK rising edge. When
running the AD7327 at a throughput rate of 500 kSPS with a
10 MHz SCLK signal, the ADC has approximately
C2
D
1.5 SCLK + t8 + tQUIET
VSS
to acquire the analog input signal. The ADC goes back into
hold mode on the CS falling edge.
C1
D
VSS
R1
C2
05401-024
D
VIN–
Figure 30. Equivalent Analog Input Circuit (Differential)
Care should be taken to ensure that the analog input does not
exceed the VDD and VSS supply rails by more than 300 mV.
Exceeding this value causes the diodes to become forward
biased and to start conducting into either the VDD supply rail or
VSS supply rail. These diodes can conduct up to 10 mA without
causing irreversible damage to the part.
In Figure 29 and Figure 30, Capacitor C1 is typically 4 pF and
can primarily be attributed to pin capacitance. Resistor R1 is a
lumped component made up of the on resistance of the input
multiplexer and the track-and-hold switch. Capacitor C2 is the
sampling capacitor; its capacitance varies depending on the
analog input range selected (see the Specifications section).
As the VDD/VSS supply voltage is reduced, the on resistance of
the input multiplexer increases. Therefore, based on the
equation for tACQ, it is necessary to increase the amount of
acquisition time provided to the AD7327, and hence decrease
the overall throughput rate. Figure 31 shows that as the VDD and
VSS supplies are reduced, the specified THD performance
degrades slightly. If the throughput rate is reduced when
operating with the minimum VDD and VSS supplies, the specified
THD performance is maintained.
–75
VCC = VDRIVE = 5V
INTERNAL REFERENCE
TA = 25°C
FIN = 10kHz
±5V RANGE
SE MODE
–80
THD (dB)
VDD
–85
–90
Track-and-Hold Section
500kSPS
The track-and-hold on the analog input of the AD7327 allows
the ADC to accurately convert an input sine wave of full-scale
amplitude to 13-bit accuracy. The input bandwidth of the trackand-hold is greater than the Nyquist rate of the ADC. The
AD7327 can handle frequencies up to 22 MHz.
The track-and-hold enters its tracking mode on the 14th SCLK
rising edge after the CS falling edge. The time required to
acquire an input signal depends on how quickly the sampling
capacitor is charged. With 0 source impedance, 305 ns is
sufficient to acquire the signal to the 13-bit level. The
acquisition time required is calculated using the following
formula:
tACQ = 10 × ((RSOURCE + R) C)
where C is the sampling capacitance, and R is the resistance
seen by the track-and-hold amplifier looking back on the input.
For the AD7327, the value of R includes the on resistance of the
input multiplexer and is typically 300 Ω. RSOURCE should include
any extra source impedance on the analog input.
–95
5
7
9
11
13
15
17
19
±VDD/VSS SUPPLIES (V)
05401-051
D
Figure 31. THD vs. ±VDD/VSS Supply Voltage at 500 kSPS
Unlike other bipolar ADCs, the AD7327 does not have a
resistive analog input structure. On the AD7327, the bipolar
analog signal is sampled directly onto the sampling capacitor.
This gives the AD7327 high analog input impedance. An
approximation for the analog input impedance can be
calculated from the following formula:
Z = 1/(fS × CS)
where fS is the sampling frequency, and CS is the sampling
capacitor value.
CS depends on the analog input range chosen (see the
Specifications section). When operating at 500 kSPS, the analog
input impedance is typically 145 kΩ for the ±10 V range. As the
sampling frequency is reduced, the analog input impedance
further increases. As the analog input impedance increases, the
current required to drive the analog input therefore decreases.
Rev. 0 | Page 18 of 36
AD7327
V+
TYPICAL CONNECTION DIAGRAM
Figure 32 shows a typical connection diagram for the AD7327.
In this configuration, the AGND pin is connected to the analog
ground plane of the system, and the DGND pin is connected to
the digital ground plane of the system. The analog inputs on the
AD7327 can be configured to operate in single-ended, true
differential, or pseudo differential mode. The AD7327 can operate
with either an internal or external reference. In Figure 32, the
AD7327 is configured to operate with the internal 2.5 V reference.
A 680 nF decoupling capacitor is required when operating with
the internal reference.
The VCC pin can be connected to either a 3 V supply voltage or a
5 V supply voltage. The VDD and VSS are the dual supplies for the
high voltage analog input structures. The voltage on these pins
must be equal to or greater than the highest analog input range
selected on the analog input channels (see Table 6). The VDRIVE
pin is connected to the supply voltage of the microprocessor.
The voltage applied to the VDRIVE input controls the voltage of
the serial interface. VDRIVE can be set to 3 V or 5 V.
+15V
+
10µF
10µF
VDD1
ANALOG INPUTS
±10V, ±5V, ±2.5V
0V TO +10V
V–
1ADDITIONAL PINS OMITTED FOR CLARITY.
Figure 33. Single-Ended Mode Typical Connection Diagram
True Differential Mode
The AD7327 can have a total of four true differential analog
input pairs. Differential signals have some benefits over singleended signals, including better noise immunity based on the
device’s common-mode rejection and improvements in
distortion performance. Figure 34 defines the configuration of
the true differential analog inputs of the AD7327.
VIN+
AD73271
+3V SUPPLY
10µF +
0.1µF
1ADDITIONAL
CS
µC/µP
SCLK
DIN
DGND
PINS OMITTED FOR CLARITY.
Figure 34. True Differential Inputs
DOUT
SERIAL
INTERFACE
AGND
1MINIMUM
VDD AND VSS SUPPLY VOLTAGES
DEPEND ON THE HIGHEST ANALOG INPUT
RANGE SELECTED.
05401-025
10µF
05401-026
VSS
VCC + 2.7V TO 5.25V
–15V
+
0.1µF
AD73271
AD7327
VSS1
VDD VCC
VIN–
REFIN/OUT
680nF
VIN+
0.1µF
VCC
VDRIVE
VIN0
VIN1
VIN2
VIN3
VIN4
VIN5
VIN6
VIN7
+
AGND
05401-027
0.1µF
5V
The amplitude of the differential signal is the difference
between the signals applied to the VIN+ and VIN− pins in
each differential pair (VIN+ − VIN−). VIN+ and VIN− should
be simultaneously driven by two signals each of amplitude
±4 × VREF (depending on the input range selected) that are 180°
out of phase. Assuming the ±4 × VREF mode, the amplitude of
the differential signal is −20 V to +20 V p-p (2 × 4 × VREF),
regardless of the common mode.
The common mode is the average of the two signals
Figure 32. Typical Connection Diagram
(VIN+ + VIN−)/2
ANALOG INPUT
Single-Ended Inputs
The AD7327 has a total of eight analog inputs when operating
the AD7327 in single-ended mode. Each analog input can be
independently programmed to one of the four analog input
ranges. In applications where the signal source is high
impedance, it is recommended to buffer the signal before
applying it to the ADC analog inputs. Figure 33 shows the
configuration of the AD7327 in single-ended mode.
and is therefore the voltage on which the two input signals are
centered.
This voltage is set up externally, and its range varies with
reference voltage. As the reference voltage increases, the
common-mode range decreases. When driving the differential
inputs with an amplifier, the actual common-mode range is
determined by the amplifier’s output swing. If the differential
inputs are not driven from an amplifier, the common-mode
range is determined by the supply voltage on the VDD and the
VSS supply pins.
When a conversion takes place, the common mode is rejected,
resulting in a noise-free signal of amplitude −2 × (4 × VREF) to
+2 × (4 × VREF) corresponding to Digital Codes −4096 to +4095.
Rev. 0 | Page 19 of 36
AD7327
5
8
±5V RANGE
±2.5V
RANGE
VCOM RANGE (V)
2
6
±5V RANGE
0
–2
–3
±10V
RANGE
±2.5V
RANGE
±10V
RANGE
±10V
RANGE
2
0
–2
–4
±5V RANGE
–4
–6
VCC = 3V
VREF = 3V
05401-045
–5
–6
±16.5V VDD/VSS
±12V VDD/VSS
Figure 35. Common-Mode Range for VCC = 3 V and REFIN/OUT = 3 V
±2.5V
RANGE
VCC = 5V
VREF = 2.5V
±16.5V VDD/VSS
±12V VDD/VSS
Pseudo Differential Inputs
±5V RANGE
6
VCOM RANGE (V)
–8
Figure 38. Common-Mode Range for VCC = 5 V and REFIN/OUT = 2.5 V
8
4
±5V RANGE
±2.5V
RANGE
4
1
–1
±10V
RANGE
05401-048
3
VCOM RANGE (V)
4
±10V
RANGE
The AD7327 can have four pseudo differential pairs or seven
pseudo differential inputs referenced to a common VIN− pin.
The VIN+ inputs are coupled to the signal source and must have
an amplitude within the selected range for that channel as
programmed in the range registers. A dc input is applied to the
VIN− pin. The voltage applied to this input provides an offset for
the VIN+ input from ground or a pseudo ground. Pseudo
differential inputs separate the analog input signal ground from
the ADC ground, allowing cancellation of dc common mode
voltages.
±5V RANGE
±2.5V
RANGE
±2.5V
RANGE
2
±10V
RANGE
0
–2
05401-046
VCC = 5V
VREF = 3V
–4
±16.5V VDD/VSS
±12V VDD/VSS
Figure 36. Common-Mode Range for VCC = 5 V and REFIN/OUT = 3 V
When a conversion takes place, the pseudo ground corresponds
to Code −4096 and the maximum amplitude corresponds to
Code +4095.
6
V+
4
±5V RANGE
±5V RANGE
VIN+
2
AD73271
0
VIN–
–2
–4
VDD VCC
±10V
RANGE
±10V
±2.5V
RANGE RANGE
VSS
±2.5V
RANGE
V–
1ADDITIONAL PINS OMITTED FOR CLARITY.
VCC = 3V
VREF = 2.5V
–8
±16.5V VDD/VSS
±12V VDD/VSS
05401-047
–6
Figure 37. Common-Mode Range for VCC = 3 V and REFIN/OUT = 2.5 V
05401-028
VCOM RANGE (V)
5V
Figure 39. Pseudo Differential Inputs
Figure 40 and Figure 41 show the typical voltage range on the
VIN− pin for the different analog input ranges when configured
in the pseudo differential mode.
For example, when the AD7327 is configured to operate in
pseudo differential mode and the ±5 V range is selected, with
±16.5 V VDD/VSS supplies and 5 V VCC, the voltage on the VIN−
pin can vary from −6.5 V to +6.5 V.
Rev. 0 | Page 20 of 36
AD7327
8
6
4
The driver amplifier must be able to settle for a full-scale step to
a 13-bit level, 0.0122%, in less than the specified acquisition
time of the AD7327. An op amp such as the AD8021 meets this
requirement when operating in single-ended mode. The AD8021
needs an external compensating NPO type of capacitor. The
AD8022 can also be used in high frequency applications where
a dual version is required. For lower frequency applications, op
amps such as the AD797, AD845, and AD8610 can be used with
the AD7327 in single-ended mode configuration.
±5V RANGE
±5V RANGE
±2.5V
RANGE
±10V
RANGE
±2.5V
RANGE
2
0
–2
±10V
RANGE
–6
–8
0V TO +10V
RANGE
VCC = 5V
VREF = 2.5V
±16.5V VDD/VSS
0V TO +10V
RANGE
±12V VDD/VSS
Differential operation requires that VIN+ and VIN− be simultaneously driven with two signals of equal amplitude that are 180°
out of phase. The common mode must be set up externally to the
AD7327. The common-mode range is determined by the REFIN/
OUT voltage, the VCC supply voltage, and the particular amplifier
used to drive the analog inputs. Differential mode with either an
ac input or a dc input provides the best THD performance over a
wide frequency range. Because not all applications have a signal
preconditioned for differential operation, there is often a need to
perform the single-ended-to-differential conversion.
05401-039
–4
Figure 40. Pseudo Input Range with VCC = 5 V
4
±5V RANGE
±5V RANGE
2
±2.5V
RANGE
0
–4
±10V
RANGE
±2.5V
RANGE
0V TO +10V
RANGE
–6
This single-ended-to-differential conversion can be performed
using an op amp pair. Typical connection diagrams for an op
amp pair are shown in Figure 42 and Figure 43. In Figure 42,
the common-mode signal is applied to the noninverting input
of the second amplifier.
±10V
RANGE
0V TO +10V
RANGE
VCC = 3V
VREF = 2.5V
–8
±16.5V VDD/VSS
±12V VDD/VSS
1.5kΩ
05401-040
–2
VIN
Figure 41. Pseudo Input Range with VCC = 3 V
3kΩ
AD845
V+
DRIVER AMPLIFIER CHOICE
In applications where the harmonic distortion and signal-tonoise ratio are critical specifications, the analog input of the
AD7327 should be driven from a low impedance source. Large
source impedances significantly affect the ac performance of the
ADC and can necessitate the use of an input buffer amplifier.
1.5kΩ
1.5kΩ
1.5kΩ
V–
10kΩ
AD845
05401-029
VCOM
20kΩ
When no amplifier is used to drive the analog input, the source
impedance should be limited to low values. The maximum source
impedance depends on the amount of THD that can be tolerated
in the application. The THD increases as the source impedance
increases and performance degrades. Figure 21 and Figure 22
show graphs of the THD vs. the analog input frequency for various
source impedances. Depending on the input range and analog
input configuration selected, the AD7327 can handle source
impedances of up to 5.5 kΩ before the THD starts to degrade.
Figure 42. Single-Ended-to-Differential Configuration with the AD845
442Ω
VIN
442Ω
AD8021
V+
442Ω
442Ω
442Ω
Due to the programmable nature of the analog inputs on the
AD7327, the choice of op amp used to drive the inputs is a
function of the particular application and depends on the input
configuration and the analog input voltage ranges selected.
442Ω
V–
05401-030
AD8021
100Ω
Figure 43. Single-Ended-to-Differential Configuration with the AD8021
Rev. 0 | Page 21 of 36
AD7327
REGISTERS
The AD7327 has four programmable registers: the control register, sequence register, Range Register 1, and Range Register 2.
These registers are write-only registers.
ADDRESSING REGISTERS
A serial transfer on the AD7327 consists of 16 SCLK cycles. The three MSBs on the DIN line during the 16 SCLK transfer are decoded to
determine which register is addressed. The three MSBs consist of the write bit, Register Select 1 bit, and Register Select 2 bit. The register
select bits are used to determine which of the four on-board registers is selected. The write bit determines if the data on the DIN line
following the register select bits loads into the addressed register. If the write bit is 1, the bits load into the register addressed by the
register select bits. If the write bit is 0, the data on the DIN line does not load into any register.
Table 8. Decoding Register Select Bits and Write Bit
Write
0
1
Register Select 1
0
0
Register Select 2
0
0
1
0
1
1
1
0
1
1
1
Description
Data on the DIN line during this serial transfer is ignored.
This combination selects the control register. The subsequent 12 bits are loaded into
the control register.
This combination selects Range Register 1. The subsequent 8 bits are loaded into
Range Register 1.
This combination selects Range Register 2. The subsequent 8 bits are loaded into
Range Register 2.
This combination selects the sequence register. The subsequent 8 bits are loaded into
the sequence register.
Rev. 0 | Page 22 of 36
AD7327
CONTROL REGISTER
The control register is used to select the analog input channel, analog input configuration, reference, coding, and power mode. The
control register is a write-only, 12-bit register. Data loaded on the DIN line corresponds to the AD7327 configuration for the next
conversion. If the sequence register is being used, data should be loaded into the control register after the range registers and the sequence
register have been initialized. The bit functions of the control register are shown in Table 9 (the power-up status of all bits is 0).
MSB
15
Write
14
Register
Select 1
13
Register
Select 2
12
ADD2
11
ADD1
10
ADD0
9
Mode 1
8
Mode 0
7
PM1
6
PM0
5
Coding
4
Ref
3
Seq1
2
Seq2
1
ZERO
LSB
0
0
Table 9. Control Register Details
Bit
12, 11, 10
Mnemonic
ADD2, ADD1,
ADD0
9, 8
Mode 1, Mode 0
7, 6
5
PM1, PM0
Coding
4
Ref
3, 2
1
Seq1, Seq2
ZERO
Description
These three channel address bits are used to select the analog input channel for the next conversion if the
sequencer is not being used. If the sequencer is being used, the three channel address bits are used to
select the final channel in a consecutive sequence.
These two mode bits are used to select the configuration of the eight analog input pins, VIN0 to VIN7. These
pins are used in conjunction with the channel address bits. On the AD7327, the analog inputs can be
configured as eight single-ended inputs, four fully differential inputs, four pseudo differential inputs, or
seven pseudo differential inputs (see Table 10).
The power management bits are used to select different power mode options on the AD7327 (see Table 11).
This bit is used to select the type of output coding the AD7327 uses for the next conversion result. If
coding = 0, the output coding is twos complement. If coding = 1, the output coding is straight binary.
When operating in sequence mode, the output coding for each channel is the value written to the coding
bit during the last write to the control register.
The reference bit is used to enable or disable the internal reference. If Ref = 0, the external reference is
enabled and used for the next conversion and the internal reference is disabled. If Ref = 1, the internal
reference is used for the next conversion. When operating in sequence mode, the reference used for each
channel is the value written to the Ref bit during the last write to the control register.
The Sequence 1 and Sequence 2 bits are used to control the operation of the sequencer (see Table 12).
A 0 should be written to this bit at all times.
The eight analog input channels can be configured as seven pseudo differential analog inputs, four pseudo differential inputs, four true
differential inputs, or eight single-ended analog inputs.
Table 10. Analog Input Configuration Selection
Channel Address Bits
ADD2 ADD1 ADD0
0
0
0
0
0
1
0
1
0
0
1
1
1
0
0
1
0
1
1
1
0
1
1
1
Mode 1 = 1, Mode 0 = 1
7 Pseudo Differential I/Ps
VIN+
VIN−
VIN0
VIN7
VIN1
VIN7
VIN2
VIN7
VIN3
VIN7
VIN4
VIN7
VIN5
VIN7
VIN6
VIN7
Temperature indicator
Mode 1 = 1, Mode 0 = 0
4 Fully Differential I/Ps
VIN+
VIN−
VIN0
VIN1
VIN0
VIN1
VIN2
VIN3
VIN2
VIN3
VIN4
VIN5
VIN4
VIN5
VIN6
VIN7
VIN6
VIN7
Rev. 0 | Page 23 of 36
Mode 1 = 0, Mode 0 =1
4 Pseudo Differential I/Ps
VIN+
VIN−
VIN0
VIN1
VIN0
VIN1
VIN2
VIN3
VIN2
VIN3
VIN4
VIN5
VIN4
VIN5
VIN6
VIN7
VIN6
VIN7
Mode 1 = 0, Mode 0 = 0
8 Single-Ended I/Ps
VIN+
VIN−
VIN0
AGND
VIN1
AGND
VIN2
AGND
VIN3
AGND
VIN4
AGND
VIN5
AGND
VIN6
AGND
VIN7
AGND
AD7327
Table 11. Power Mode Selection
PM1
1
PM0
1
1
0
0
1
0
0
Description
Full Shutdown Mode. In this mode, all internal circuitry on the AD7327 is powered down. Information in the control register
is retained when the AD7327 is in full shutdown mode.
Autoshutdown Mode. The AD7327 enters autoshutdown on the 15th SCLK rising edge when the control register is updated.
All internal circuitry is powered down in autoshutdown.
Autostandby Mode. In this mode, all internal circuitry is powered down, excluding the internal reference. The AD7327 enters
autostandby mode on the 15th SCLK rising edge after the control register is updated.
Normal Mode. All internal circuitry is powered up at all times.
Table 12. Sequencer Selection
Seq1
0
Seq2
0
0
1
1
0
1
1
Description
The channel sequencer is not used. The analog input channel, selected by programming the ADD2 bit to ADD0 bit in the
control register, selects the next channel for conversion.
Uses the sequence of channels previously programmed into the sequence register for conversion. The AD7327 starts
converting on the lowest channel in the sequence. The channels are converted in ascending order. If uninterrupted, the
AD7327 keeps converting the sequence. The range for each channel defaults to the range previously written into the
corresponding range register.
Used in conjunction with the channel address bits in the control register. This allows continuous conversions on a
consecutive sequence of channels, from Channel 0 up to and including a final channel selected by the channel address bits
in the control register. The range for each channel defaults to the range previously written into the corresponding range
register.
The channel sequencer is not used. The analog channel, selected by programming the ADD2 bit to ADD0 bit in the control
register, selects the next channel for conversion.
Rev. 0 | Page 24 of 36
AD7327
SEQUENCE REGISTER
The sequence register on the AD7327 is an 8-bit, write-only register. Each of the eight analog input channels has one corresponding bit in
the sequence register. To select an analog input channel for inclusion in the sequence, set the corresponding channel bit to 1 in the
sequence register.
MSB
16
Write
15
Register Select 1
14
Register Select 2
13
VIN0
12
VIN1
11
VIN2
10
VIN3
9
VIN4
8
VIN5
7
VIN6
6
VIN7
5
0
4
0
3
0
2
0
LSB
1
0
RANGE REGISTERS
The range registers are used to select one analog input range per analog input channel. Range Register 1 is used to set the ranges for
Channel 0 to Channel 3. It is an 8-bit, write-only register with two dedicated range bits for each of the analog input channels from
Channel 0 to Channel 3. There are four analog input ranges, ±10 V, ±5 V, ±2.5 V, and 0 V to +10 V. A write to Range Register 1 is selected
by setting the write bit to 1 and the register select bits to 0 and 1. After the initial write to Range Register 1 occurs, each time an analog
input is selected, the AD7327 automatically configures the analog input to the appropriate range, as indicated by Range Register 1.
The ±10 V input range is selected by default on each analog input channel (see Table 13).
MSB
16
Write
15
Register
Select 1
14
Register
Select 2
13
VIN0A
12
VIN0B
11
VIN1A
10
VIN1B
9
VIN2A
8
VIN2B
7
VIN3A
6
VIN3B
5
0
4
0
3
0
2
0
LSB
1
0
Range Register 2 is used to set the ranges for Channel 4 to Channel 7. It is an 8-bit, write-only register with two dedicated range bits for
each of the analog input channels from Channel 4 to Channel 7. There are four analog input ranges, ±10 V, ±5 V, ±2.5 V, and 0 V to +10 V.
After the initial write to Range Register 2 occurs, each time an analog input is selected, the AD7327 automatically configures the analog
input to the appropriate range, as indicated by Range Register 2. The ±10 V input range is selected by default on each analog input
channel (see Table 13).
MSB
16
Write
15
Register
Select 1
14
Register
Select 2
13
VIN4A
12
VIN4B
11
VIN5A
10
VIN5B
9
VIN6A
8
VIN6B
Table 13. Range Selection
VINxA
0
0
1
1
VINxB
0
1
0
1
Description
This combination selects the ±10 V input range on VINx.
This combination selects the ±5 V input range on VINx.
This combination selects the ±2.5 V input range on VINx.
This combination selects the 0 V to +10 V input range on VINx.
Rev. 0 | Page 25 of 36
7
VIN7A
6
VIN7B
5
0
4
0
3
0
2
0
LSB
1
0
AD7327
SEQUENCER OPERATION
POWER ON.
CS
DIN: WRITE TO RANGE REGISTER 1 TO SELECT THE RANGE
FOR EACH ANALOG INPUT CHANNEL.
DOUT: CONVERSION RESULT FROM CHANNEL 0, ±10V
RANGE, SINGLE-ENDED MODE.
CS
DIN: WRITE TO RANGE REGISTER 2 TO SELECT THE RANGE
FOR EACH ANALOG INPUT CHANNEL.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
SINGLE-ENDED MODE, RANGE SELECTED IN
RANGE REGISTER 1.
CS
DIN: WRITE TO SEQUENCE REGISTER TO SELECT THE
ANALOG INPUT CHANNELS TO BE INCLUDED IN
THE SEQUENCE.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
SINGLE-ENDED MODE, RANGE SELECTED IN
RANGE REGISTER 1.
CS
DIN: WRITE TO CONTROL REGISTER TO START THE
SEQUENCE, Seq1 = 0, Seq2 = 1.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
SINGLE-ENDED MODE, RANGE SELECTED IN
RANGE REGISTER 1.
CS
DIN: TIE DIN LOW/WRITE BIT = 0 TO CONTINUE TO CONVERT
THROUGH THE SEQUENCE OF CHANNELS.
CS
DOUT: CONVERSION RESULT FROM FIRST CHANNEL IN
THE SEQUENCE.
DIN: WRITE TO CONTROL
REGISTER TO STOP THE
SEQUENCE, Seq1 = 0, Seq2 = 0.
STOPPING
A SEQUENCE.
DOUT: CONVERSION RESULT
FROM CHANNEL IN SEQUENCE.
DIN TIED LOW/WRITE BIT = 0.
CONTINUOUSLY CONVERT
ON THE SELECTED SEQUENCE
OF CHANNELS.
SELECTING A NEW SEQUENCE.
CS
DIN: WRITE TO SEQUENCE REGISTER TO SELECT THE
NEW SEQUENCE.
05401-031
DOUT: CONVERSION RESULT FROM CHANNEL X IN
THE FIRST SEQUENCE.
Figure 44. Programmable Sequence Flowchart
The AD7327 can be configured to automatically cycle through a
number of selected channels using the on-chip sequence register
with the Seq1 bit and the Seq2 bit in the control register. Figure 44
shows how to program the AD7327 register to operate in
sequence mode.
After power-up, all of the four on-chip registers contain default
values. Each analog input has a default input range of ±10 V. If
different analog input ranges are required, a write to the range
registers is required. This is shown in the first two serial
transfers of Figure 44.
These two initial serial transfers are only necessary if input
ranges other than the default ranges are required. After the
analog input ranges are configured, a write to the sequence
register is necessary to select the channels to be included in the
sequence. Once the channels for the sequence have been
selected, the sequence can be initiated by writing to the control
register and setting Seq1 to 0 and Seq2 to 1. The AD7327
continues to convert through the selected sequence without
interruption provided the sequence register remains unchanged
and Seq1 = 0 and Seq2 = 1 in the control register.
Rev. 0 | Page 26 of 36
AD7327
If a change to one of the range registers is required during a
sequence, it is necessary to first stop the sequence by writing to
the control register and setting Seq1 to 0 and Seq2 to 0. Next,
the write to the range register should be completed to change
the required range. The previously selected sequence should
then be initiated again by writing to the control register and
setting Seq1 to 0 and Seq2 to 1. The ADC converts the first
channel in the sequence.
The AD7327 can be configured to convert a sequence of
consecutive channels (see Figure 45). This sequence begins by
converting on Channel 0 and ends with a final channel as
selected by Bit ADD2 to Bit ADD0 in the control register. In
this configuration, there is no need for a write to the sequence
register. To operate the AD7327 in this mode, set Seq1 to 1 and
Seq2 to 0, and then select the final channel in the sequence by
programming Bit ADD2 to Bit ADD0 in the control register.
Once the control register is configured to operate the AD7327
in this mode, the DIN line can be held low or the write bit can
be set to 0. To return to traditional multichannel operation, a
write to the control register to set Seq1 to 0 and Seq2 to 0 is
necessary.
When Seq1 and Seq2 are both set to 0, or when both are set
to 1, the AD7327 is configured to operate in traditional multichannel mode, where a write to the Channel Address Bit ADD2
to Bit ADD0 in the control register selects the next channel for
conversion.
POWER ON.
CS
DIN: WRITE TO RANGE REGISTER 1 TO SELECT THE RANGE
FOR ANALOG INPUT CHANNELS.
DOUT: CONVERSION RESULT FROM CHANNEL 0, ±10V
RANGE, SINGLE-ENDED MODE.
CS
DIN: WRITE TO RANGE REGISTER 2 TO SELECT THE RANGE
FOR ANALOG INPUT CHANNELS.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
RANGE SELECTED IN RANGE REGISTER 1,
SINGLE-ENDED MODE.
CS
DIN: WRITE TO CONTROL REGISTER TO SELECT THE FINAL
CHANNEL IN THE CONSECUTIVE SEQUENCE, SET Seq1 = 1
AND Seq2 = 0. SELECT OUTPUT CODING FOR SEQUENCE.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
RANGE SELECTED IN RANGE REGISTER 1,
SINGLE-ENDED MODE.
CS
DIN: WRITE BIT = 0 OR DIN LINE HELD LOW TO CONTINUE
TO CONVERT THROUGH THE SEQUENCE OF
CONSECUTIVE CHANNELS.
DOUT: CONVERSION RESULT FROM CHANNEL 0,
RANGE SELECTED IN RANGE REGISTER 1.
CS
DIN: WRITE BIT = 0 OR DIN LINE HELD LOW TO CONTINUE
THROUGH SEQUENCE OF CONSECUTIVE CHANNELS.
DOUT: CONVERSION RESULT FROM CHANNEL 1,
RANGE SELECTED IN RANGE REGISTER 1.
STOPPING
A SEQUENCE.
DIN TIED LOW/WRITE BIT = 0.
CONTINUOUSLY CONVERT
ON CONSECUTIVE SEQUENCE
OF CHANNELS.
CS
DOUT: CONVERSION RESULT
FROM CHANNEL IN SEQUENCE.
Figure 45. Flowchart for Consecutive Sequence of Channels
Rev. 0 | Page 27 of 36
05401-032
DIN: WRITE TO CONTROL
REGISTER TO STOP THE
SEQUENCE, Seq1 = 0, Seq2 = 0.
AD7327
REFERENCE
TEMPERATURE INDICATOR
The AD7327 can operate with either the internal 2.5 V on-chip
reference or an externally applied reference. The internal
reference is selected by setting the Ref bit in the control register
to 1. On power-up, the Ref bit is 0, which selects the external
reference for the AD7327 conversion. Suitable reference sources
for the AD7327 include AD780, AD1582, ADR431, REF193,
and ADR391.
The AD7327 has an on-chip temperature indicator. The
temperature indicator can be used to give local temperature
measurements on the AD7327. To access the temperature
indicator, the ADC should be configured in pseudo differential
mode, Mode 1 = Mode 0 = 1, and Channel Bits ADD2, ADD1,
and ADD0 should be set to 1. VIN7 must be tied to AGND or to
a small dc voltage within the specified pseudo input range for
the selected analog input range. When a conversion is initiated
in this configuration, the output code represents the temperature
(see Figure 46 and Figure 47). When using the temperature
indicator on the AD7327, the part should be operated at low
throughput rates, such as approximately 50 kSPS for the ±10 V
range and 30 kSPS for the ±2.5 V range. The throughput rate is
reduced for the temperature indicator mode because the
AD7327 requires more acquisition time for this mode.
The AD7327 is specified for a 2.5 V to 3 V reference range.
When a 3 V reference is selected, the ranges are ±12 V, ±6 V,
±3 V, and 0 V to +12 V. For these ranges, the VDD and VSS supply
must be equal to or greater than the maximum analog input
range selected (see Table 6).
4420
VCC = VDRIVE = 5V
VDD/VSS = ±12V
50kSPS
4410
4400
±10V RANGE, INT REF
4390
4380
4370
4360
4350
4340
–40
VDRIVE
–20
0
20
40
60
80
100
TEMPERATURE (°C)
The AD7327 has a VDRIVE feature to control the voltage at which
the serial interface operates. VDRIVE allows the ADC to easily
interface to both 3 V and 5 V processors. For example, if the
AD7327 is operated with a VCC of 5 V, the VDRIVE pin can be
powered from a 3 V supply. This allows the AD7327 to accept
large bipolar input signals with low voltage digital processing.
05401-033
On power-up, if the internal reference operation is required for
the ADC conversion, a write to the control register is necessary
to set the Ref bit to 1. During the control register write, the
conversion result from the first initial conversion is invalid. The
reference buffer requires 500 μs to power up and charge the
680 nF decoupling capacitor during the power-up time.
ADC OUTPUT CODE
The internal reference circuitry consists of a 2.5 V band gap
reference and a reference buffer. When operating the AD7327
in internal reference mode, the 2.5 V internal reference is available
at the REFIN/OUT pin, which should be decoupled to AGND
using a 680 nF capacitor. It is recommended that the internal
reference be buffered before applying it elsewhere in the system.
The internal reference is capable of sourcing up to 90 μA.
Figure 46. Temperature vs. ADC Output Code for ±10 V Range
5450
VCC = VDRIVE = 5V
VDD/VSS = ±12V
±2.5V RANGE
INT REFERENCE
30kSPS
ADC OUTPUT CODE
5400
5350
5300
5250
5200
5100
–40
–20
0
20
40
60
80
TEMPERATURE (°C)
Figure 47. Temperature vs. ADC Output Code for ±2.5 V Range
Rev. 0 | Page 28 of 36
05401-034
5150
AD7327
MODES OF OPERATION
The AD7327 remains fully powered up at the end of the
conversion if both PM1 and PM0 contain 0 in the control
register.
The AD7327 has several modes of operation that are designed
to provide flexible power management options. These options
can be chosen to optimize the power dissipation/throughput
rate ratio for different application requirements. The mode of
operation of the AD7327 is controlled by the power management
bits, Bit PM1 and Bit PM0, in the control register as shown in
Table 11. The default mode is normal mode, where all internal
circuitry is fully powered up.
To complete the conversion and access the conversion result,
16 serial clock cycles are required. At the end of the conversion,
CS can idle either high or low until the next conversion.
NORMAL MODE
Once the data transfer is complete, another conversion can be
initiated after the quiet time, tQUIET, has elapsed.
(PM1 = PM0 = 0)
FULL SHUTDOWN MODE
This mode is intended for the fastest throughput rate performance
with the AD7327 being fully powered up at all times. Figure 48
shows the general operation of the AD7327 in normal mode.
(PM1 = PM0 = 1)
In this mode, all internal circuitry on the AD7327 is powered
down. The part retains information in the registers during full
shutdown. The AD7327 remains in full shutdown mode until
the power management bits, Bit PM1 and Bit PM0, in the
control register are changed.
The conversion is initiated on the falling edge of CS, and the
track-and-hold section enters hold mode, as described in the
Serial Interface section. Data on the DIN line during the
16 SCLK transfer is loaded into one of the on-chip registers if the
write bit is set. The register is selected by programming the
register select bits (see Table 8).
A write to the control register with PM1 = 1 and PM0 = 1 places
the part into full shutdown mode. The AD7327 enters full shutdown mode on the 15th SCLK rising edge once the control register
is updated.
CS
1
If a write to the control register occurs while the part is in full
shutdown mode with the power management bits, Bit PM1 and
Bit PM0, set to 0 (normal mode), the part begins to power up
on the 15th SCLK rising edge once the control register is
updated. Figure 49 shows how the AD7327 is configured to exit
full shutdown mode. To ensure the AD7327 is fully powered up,
tPOWER-UP should elapse before the next CS falling edge.
16
SCLK
DOUT
05401-035
3 CHANNEL I.D. BITS, SIGN BIT + CONVERSION RESULT
DATA INTO CONTROL/SEQUENCE/RANGE1/RANGE2
REGISTER
DIN
Figure 48. Normal Mode
PART IS IN FULL
SHUTDOWN
THE PART IS FULLY POWERED UP
ONCE tPOWER-UP HAS ELAPSED
PART BEGINS TO POWER UP ON THE 15TH
SCLK RISING EDGE AS PM1 = PM0 = 0
tPOWER-UP
CS
1
16
16
1
SDATA
DIN
INVALID DATA
CHANNEL IDENTIFIER BITS + CONVERSION RESULT
DATA INTO CONTROL REGISTER
DATA INTO CONTROL REGISTER
CONTROL REGISTER IS LOADED ON THE FIRST 15 CLOCKS,
PM1 = 0, PM0 = 0
TO KEEP THE PART IN NORMAL MODE, LOAD PM1 = PM0 = 0
IN CONTROL REGISTER
Figure 49. Exiting Full Shutdown Mode
Rev. 0 | Page 29 of 36
05401-041
SCLK
AD7327
AUTOSHUTDOWN MODE
(PM1 = 1, PM0 = 0)
Once the autoshutdown mode is selected, the AD7327 automatically enters shutdown on the 15th SCLK rising edge. In
autoshutdown mode, all internal circuitry is powered down.
The AD7327 retains information in the registers during
autoshutdown. The track-and-hold is in hold mode during
autoshutdown. On the rising CS edge, the track-and-hold, which
was in hold during autoshutdown, returns to track as the AD7327
begins to power up. The power-up from autoshutdown is 500 μs.
When the control register is programmed to transition to
autoshutdown mode, it does so on the 15th SCLK rising edge.
Figure 50 shows the part entering autoshutdown mode. The
AD7327 automatically begins to power up on the CS rising
edge. The tPOWER-UP is required before a valid conversion, initiated
by bringing the CS signal low, can take place. Once this valid
conversion is complete, the AD7327 powers down again on the
15th SCLK rising edge. The CS signal must remain low again to
keep the part in autoshutdown mode.
As is the case with autoshutdown mode, the AD7327 enters
standby on the 15th SCLK rising edge once the control register is
updated (see Figure 50). The part retains information in the
registers during standby. The AD7327 remains in standby until
it receives a CS rising edge. The ADC begins to power up on the
CS rising edge. On the CS rising edge, the track-and-hold,
which was in hold mode while the part was in standby, returns
to track.
The power-up time from standby is 700 ns. The user should
ensure that 700 ns have elapsed before bringing CS low to
attempt a valid conversion. Once this valid conversion is
complete, the AD7327 again returns to standby on the 15th
SCLK rising edge. The CS signal must remain low to keep the
part in standby mode.
Figure 50 shows the part entering autoshutdown mode. The
sequence of events is the same when entering autostandby
mode. In Figure 50, the power management bits are configured
for autoshutdown. For autostandby mode, the power
management bits, PM1 and PM0, should be set to 0 and 1,
respectively.
AUTOSTANDBY MODE
(PM1 = 0, PM0 =1)
In autostandby mode, portions of the AD7327 are powered
down, but the on-chip reference remains powered up. The
reference bit in the control register should be 1 to ensure that
the on-chip reference is enabled. This mode is similar to autoshutdown but allows the AD7327 to power up much faster,
which allows faster throughput rates.
PART BEGINS TO POWER
UP ON CS RISING EDGE
PART ENTERS SHUTDOWN MODE
ON THE 15TH RISING SCLK EDGE
AS PM1 = 1, PM0 = 0
CS
1
THE PART IS FULLY POWERED UP
ONCE tPOWER-UP HAS ELAPSED
tPOWER-UP
15 16
1
15 16
SCLK
DIN
VALID DATA
VALID DATA
DATA INTO CONTROL REGISTER
DATA INTO CONTROL REGISTER
05401-042
SDATA
CONTROL REGISTER IS LOADED ON THE FIRST 15 CLOCKS,
PM1 = 1, PM0 = 0
Figure 50. Entering Autoshutdown/Autostandby Mode
Rev. 0 | Page 30 of 36
AD7327
POWER VS. THROUGHPUT RATE
20
16
6
VARIABLE SCLK
4
2
100
200
10
8
6
VARIABLE SCLK
4
2
0
0
100
200
300
400
Figure 52. Power vs. Throughput Rate with 5 V VCC
8
0
12
THROUGHPUT RATE (kSPS)
300
400
THROUGHPUT RATE (kSPS)
500
05401-052
AVERAGE POWER (mW)
VCC = 3V
VDD/VSS = ±12V
TA = 25°C
10 INTERNAL REFERENCE
14
Figure 51. Power vs. Throughput Rate with 3 V VCC
Rev. 0 | Page 31 of 36
500
05401-053
12
0
VCC = 5V
VDD/VSS = ±12V
TA = 25°C
INTERNAL REFERENCE
18
AVERAGE POWER (mW)
The power consumption of the AD7327 varies with throughput
rate. The static power consumed by the AD7327 is very low, and
significant power savings can be achieved as the throughput
rate is reduced. Figure 51 and Figure 52 shows the power vs.
throughput rate for the AD7327 at a VCC of 3 V and 5 V,
respectively. Both plots clearly show that the average power
consumed by the AD7327 is greatly reduced as the sample
frequency is reduced. This is true whether a fixed SCLK value is
used or if it is scaled with the sampling frequency. Figure 51 and
Figure 52 show the power consumption when operating in
normal mode for a variable SCLK that scales with the sampling
frequency.
AD7327
SERIAL INTERFACE
Figure 53 shows the timing diagram for the serial interface of
the AD7327. The serial clock applied to the SCLK pin provides
the conversion clock and controls the transfer of information to
and from the AD7327 during a conversion.
Data is clocked into the AD7327 on the SCLK falling edge. The
three MSBs on the DIN line are decoded to select which register
is addressed. The control register is a 12-bit register. If the
control register is addressed by the three MSBs, the data on the
DIN line is loaded into the control on the 15th SCLK falling
edge. If the sequence register or either of the range registers is
addressed, the data on the DIN line is loaded into the addressed
register on the 11th SCLK falling edge.
The CS signal initiates the data transfer and the conversion
process. The falling edge of CS puts the track-and-hold into
hold mode and takes the bus out of three-state. Then the analog
input signal is sampled. Once the conversion is initiated, it
requires 16 SCLK cycles to complete.
Conversion data is clocked out of the AD7327 on each SCLK
falling edge. Data on the DOUT line consists of three channel
identifier bits, a sign bit, and a 12-bit conversion result. The
channel identifier bits are used to indicate which channel
corresponds to the conversion result. The ADD2 bit is clocked
out on the CS falling edge, and the ADD1 bit is clocked out on
the first SCLK falling edge.
The track-and-hold goes back into track mode on the 14th SCLK
rising edge. On the 16th SCLK falling edge, the DOUT line
returns to three-state. If the rising edge of CS occurs before
16 SCLK cycles have elapsed, the conversion is terminated,
and the DOUT line returns to three-state. Depending on where
the CS signal is brought high, the addressed register may be
updated.
t1
CS
tCONVERT
t2
1
2
3
4
3 IDENTIFICATION BITS
t3
DOUT
THREE- ADD2
STATE
DIN
ADD1
WRITE
ADD0
SIGN
t4
13
14
DB11
15
16
t5
t7
DB10
DB2
t8
DB1
DB0
t10
t9
REG
SEL1
5
REG
SEL2
tQUIET
THREE-STATE
MSB
LSB
DON’T
CARE
Figure 53. Serial Interface Timing Diagram (Control Register Write)
Rev. 0 | Page 32 of 36
05401-036
SCLK
t6
AD7327
MICROPROCESSOR INTERFACING
The serial interface on the AD7327 allows the part to be directly
connected to a range of different microprocessors. This section
explains how to interface the AD7327 with some common
microcontroller and DSP serial interface protocols.
AD7327 TO ADSP-21xx
The ADSP-21xx family of DSPs interface directly to the AD7327
without requiring glue logic. The VDRIVE pin of the AD7327 takes
the same supply voltage as that of the ADSP-21xx. This allows
the ADC to operate at a higher supply voltage than its serial
interface. The SPORT0 on the ADSP-21xx should be configured
as shown in Table 14.
Table 14. SPORT0 Control Register Setup
Setting
TFSW = RFSW = 1
INVRFS = INVTFS = 1
DTYPE = 00
SLEN = 1111
ISCLK = 1
TFSR = RFSR = 1
IRFS = 0
ITFS = 1
Description
Alternative framing
Active low frame signal
Right justify data
16-bit data-word
Internal serial clock
Frame every word
The frequency of the serial clock is set in the SCLKDIV register.
When the instruction to transmit with TFS is given (AX0 =
TX0), the state of the serial clock is checked. The DSP waits
until the SCLK has gone high, low, and high again before
starting the transmission. If the timer and SCLK are chosen so
that the instruction to transmit occurs on or near the rising
edge of SCLK, data can be transmitted immediately or at the
next clock edge.
For example, the ADSP-2111 has a master clock frequency of
16 MHz. If the SCLKDIV register is loaded with the value 3, an
SCLK of 2 MHz is obtained, and eight master clock periods elapse
for every one SCLK period. If the timer registers are loaded with
the value 803, 100.5 SCLKs occur between interrupts and, subsequently, between transmit instructions. This situation leads to
nonequidistant sampling because the transmit instruction occurs
on an SCLK edge. If the number of SCLKs between interrupts is
an integer of N, equidistant sampling is implemented by the DSP.
AD7327 TO ADSP-BF53x
The connection diagram is shown in Figure 54. The ADSP-21xx
has TFS0 and RFS0 tied together. TFS0 is set as an output, and
RFS0 is set as an input. The DSP operates in alternative framing
mode, and the SPORT0 control register is set up as described in
Table 14. The frame synchronization signal generated on the TFS
is tied to CS, and, as with all signal processing applications, requires
equidistant sampling. However, as in this example, the timer
interrupt is used to control the sampling rate of the ADC, and
under certain conditions equidistant sampling cannot be achieved.
The ADSP-BF53x family of DSPs interfaces directly to the
AD7327 without requiring glue logic, as shown in Figure 55.
The SPORT0 Receive Configuration 1 register should be set up
as outlined in Table 15.
ADSP-BF53x1
AD73271
SCLK
RSCLK0
CS
RFS0
DIN
DT0
DOUT
DR0
VDRIVE
SCLK
VDD
SCLK0
CS
TFS0
RFS0
DIN
DT0
DOUT
DR0
1ADDITIONAL PINS OMITTED FOR CLARITY.
05401-038
ADSP-21xx1
AD73271
Figure 55. Interfacing the AD7327 to the ADSP-BF53x
Table 15. SPORT0 Receive Configuration 1 Register
VDD
1ADDITIONAL PINS OMITTED FOR CLARITY.
05401-037
VDRIVE
Figure 54. Interfacing the AD7327 to the ADSP-21xx
The timer registers are loaded with a value that provides an
interrupt at the required sampling interval. When an interrupt
is received, a value is transmitted with TFS/DT (ADC control
word). The TFS is used to control the RFS, and hence the
reading of data.
Setting
RCKFE = 1
LRFS = 1
RFSR = 1
IRFS = 1
RLSBIT = 0
RDTYPE = 00
IRCLK = 1
RSPEN = 1
SLEN = 1111
TFSR = RFSR = 1
Rev. 0 | Page 33 of 36
Description
Sample data with falling edge of RSCLK
Active low frame signal
Frame every word
Internal RFS used
Receive MSB first
Zero fill
Internal receive clock
Receive enable
16-bit data-word
AD7327
APPLICATION HINTS
LAYOUT AND GROUNDING
The printed circuit board that houses the AD7327 should be
designed so that the analog and digital sections are confined to
certain areas of the board. This design facilitates the use of
ground planes that can easily be separated.
To provide optimum shielding for ground planes, a minimum
etch technique is generally best. All AGND pins on the AD7327
should be connected to the AGND plane. Digital and analog
ground pins should be joined in only one place. If the AD7327
is in a system where multiple devices require an AGND and
DGND connection, the connection should still be made at only
one point. A star point should be established as close as possible
to the ground pins on the AD7327.
Good connections should be made to the power and ground
planes. This can be done with a single via or multiple vias for
each supply and ground pin.
Avoid running digital lines under the AD7327 device because
this couples noise onto the die. However, the analog ground
plane should be allowed to run under the AD7327 to avoid
noise coupling. The power supply lines to the AD7327 device
should use as large a trace as possible to provide low impedance
paths and reduce the effects of glitches on the power supply line.
To avoid radiating noise to other sections of the board, components, such as clocks, with fast switching signals should be
shielded with digital ground and never run near the analog inputs.
Avoid crossover of digital and analog signals. To reduce the effects
of feedthrough within the board, traces should be run at right
angles to each other. A microstrip technique is the best method,
but its use may not be possible with a double-sided board. In
this technique, the component side of the board is dedicated to
ground planes, and signals are placed on the other side.
Good decoupling is also important. All analog supplies should
be decoupled with 10 μF tantalum capacitors in parallel with
0.1 μF capacitors to AGND. To achieve the best results from
these decoupling components, they must be placed as close as
possible to the device, ideally right up against the device. The
0.1 μF capacitors should have a low effective series resistance
(ESR) and low effective series inductance (ESI), such as is
typical of common ceramic and surface mount types of
capacitors. These low ESR, low ESI capacitors provide a low
impedance path to ground at high frequencies to handle
transient currents due to internal logic switching.
Rev. 0 | Page 34 of 36
AD7327
OUTLINE DIMENSIONS
6.60
6.50
6.40
20
11
4.50
4.40
4.30
6.40 BSC
1
10
PIN 1
0.65
BSC
1.20 MAX
0.15
0.05
COPLANARITY
0.10
0.30
0.19
0.20
0.09
8°
0°
SEATING
PLANE
0.75
0.60
0.45
COMPLIANT TO JEDEC STANDARDS MO-153-AC
Figure 56. 20-Lead Thin Shrink Small Outline Package [TSSOP]
(RU-20)
Dimensions show in millimeters
ORDERING GUIDE
Model
AD7327BRUZ 1
AD7327BRUZ-REEL1
AD7327BRUZ-REEL71
EVAL-AD7327CB 2
EVAL-CONTROL BRD2 3
Temperature Range
–40°C to +85°C
–40°C to +85°C
–40°C to +85°C
Package Description
20-Lead TSSOP
20-Lead TSSOP
20-Lead TSSOP
Evaluation Board
Controller Board
1
Package Option
RU-20
RU-20
RU-20
Z = Pb-free part.
This can be used as a standalone evaluation board or in conjunction with the EVAL-CONTROL board for evaluation/demonstration purposes.
3
This board is a complete unit allowing a PC to control and communicate with all Analog Devices evaluation boards ending in the CB designators. To order a complete
evaluation kit, the particular ADC evaluation board (for example, EVAL-AD7327CB), the EVAL-CONTROL BRD2, and a 12 V transformer must be ordered. See the relevant
evaluation board technical note for more information.
2
Rev. 0 | Page 35 of 36
AD7327
NOTES
©2006 Analog Devices, Inc. All rights reserved. Trademarks and
registered trademarks are the property of their respective owners.
D05401–0–1/06(0)
Rev. 0 | Page 36 of 36