EXTECH MA120

User’s Guide
Model MA120
200A AC/DC Mini Clamp-on Meter
• 3-2/3 digit (2400 count) LCD display
• White LED backlight
• Built-in non-contact AC voltage detector plus flashlight
• Frequency measurement 40Hz to 1kHz
• Autoranging with auto power off
• 0.7” (18mm) Jaw opening
Introduction
Congratulations on your purchase of Extech’s MA120 AC/DC Mini Clamp Meter.
This professional meter, with proper care, will provide years of safe reliable service.
Specifications
General Specifications
Display
3-2/3 (2400 count) Digit LCD with white LED backlight
Polarity
Minus sign ( - ) indicates negative polarity
Current sensor
Hall effect sensor type
Overload indication
“OL” displayed on the LCD
DCA zero adjust
One touch zero key
Display rate
2 readings/second, nominal
Battery
Two 1.5V AAA batteries
Low Battery indication
“BATT” displayed on the LCD
Operating conditions
32 F to 86 F (0 C to 30 C) 90%RH; 86 F to 104 F (30 C to 40 C)
º
º
º
º
75%RH; 104 F to 122 F (40 C to 50 C) 45%RH
Storage conditions
- 14 F to 140 F (-30 C to 60 C); < 90% Relative Humidity
Altitude
Operate at less than 3000 meters
Weight
6.2 oz. (176g) including battery
Dimensions
6.5 x 2.6 x 1.3” (164 x 65 x 32mm) (HWD)
Jaw opening
0.7” (18mm)
Standards
IEC 1010 Category III 300V, Category II 600V,
º
º
º
º
º
º
º
º
º
º
º
º
CC
Range Specifications
Function
Range
Resolution
Accuracy (of reading)
DC Current
200A
0.1A
± (2.8% + 8 digits)
AC Current
200A
0.1A
±(3.0% + 8 digits)
Frequency
40-51Hz
0.01Hz
±(1.2% + 5 digits)
(Auto-Ranging)
51-510Hz
0.1Hz
Input Sensitivity: 2.5A AC RMS min.
0.51-1kHz
0.001kHz
Non-Contact AC Voltage
100VAC to 600VAC 50/60Hz
2
MA120 Ver. 1.2
10/2004
Meter Description
5. Data HOLD and Backlight key
6. LCD display
7. Clamp trigger
8. One-touch DCA ZERO key
1. Current sense jaw
2. Non-contact AC voltage indicator light
3. Flashlight button
4. Rotary function switch
2
1
3
4
5
7
6
8
Safety Information
!
Caution! Refer to the explanation in this Manual
Double Insulation
This meter has been designed to be safe in use, but the operator must use caution in its
operation. The rules listed below should be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
2. USE EXTREME CAUTION when working with voltages greater than 60VDC or 25VAC
rms. These voltages are considerd a shock hazard.
3. NEVER operate the meter unless the back cover and the battery/fuse door are in place
and fastened securely.
Input Limits
Function
Maximum Input
AC Current
200A
DC Current
±200A
Frequency
1kHz
3
MA120 Ver. 1.2
10/2004
Operation
AC Current Measurements
1) Set the Function switch to the ACA range.
2) Press the jaw trigger and clamp around, fully enclosing a single
conductor. Do not allow a gap between the two halves of the jaw.
Refer to the diagram at right for the correct way to enclose a single
conductor.
3) Read the ACA value on the LCD.
Yes
No
DC Current Measurements
1) Set the Function switch to the DCA range.
2) Press the ZERO key to null the meter display.
3) Press the Trigger to open the current sense Jaw.
4) Fully enclose the conductor to be measured. Do not allow a gap between the two
halves of the jaw.
5) Read the DCA value on the LCD.
Non-Contact AC Voltage Detection
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a
known live circuit to verify proper operation
1) AC Voltage detection operates on any of the four Function switch positions.
2) Touch the probe tip to the hot conductor or insert into the hot side of the electrical
outlet.
3) If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results, slowly
slide the probe tip along a length of the cord to assure placing the tip in close
proximity to the live conductor.
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of
energy may randomly turn the detector light on. This is normal operation
Frequency Measurements
1) Set the Function switch to the Hz position.
2) Fully enclose the conductor to be measured. Do not allow a gap between the two
halves of the jaw.
3) Read the Frequency value on the LCD.
Flashlight
Press and hold the top button to turn the flashlight on. Release the button to turn the
flashlight off.
Data Hold
To freeze the current reading on the LCD, press the “Hold / Backlight” key. The word HOLD
will appear on the LCD while the meter is in the Data Hold mode. To release the Data Hold
function and return the meter to normal operation, press the “Hold / Backlight” key again. The
word HOLD will switch off.
4
MA120 Ver. 1.2
10/2004
Backlight Display
Press and hold the “Hold / Backlight” key for more than 2 seconds to turn on the backlight.
This will also activate the Data Hold function. To release the Data Hold function and return
the meter to normal operation, press the Data Hold key momentarily. To turn off the backlight,
press and hold the “Hold Backlight” key for more than 2 seconds.
Maintenance
Battery Replacement
1) When the low battery symbol appears on the LCD the batteries must be replaced.
2) Power down and remove the rear battery compartment Phillips screw.
3) Lift off the battery compartment cover and replace the two 1.5V AAA cells.
4) Replace compartment cover and secure the screw.
Cleaning
Caution: Use only a dry cloth to clean the plastic case.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides
NIST certification for most products. Call the Customer Service Department for information
on calibration services available for this product. Extech recommends that annual
calibrations be performed to verify meter performance and accuracy.
Warranty
EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of
defects in parts and workmanship for one year from date of shipment (a six month limited
warranty applies on sensors and cables). If it should become necessary to return the
instrument for service during or beyond the warranty period, contact the Customer Service
Department at (781) 890-7440 EXTENSION 210 for authorization or visit www.extech.com for
more information. A Return Authorization (RA) number must be issued before any
product is returned to Extech. The sender is responsible for shipping charges, freight,
insurance and proper packaging to prevent damage in transit. This warranty does not apply
to defects resulting from action of the user such as misuse, improper wiring, operation
outside of specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set forth
above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice