ETC 2559

Data Sheet
Digital Storage Oscilloscopes
2550 Series
The 2550 series digital storage oscilloscopes
provide high performance and value in 2-channel
and 4-channel configurations. With bandwidth
from 70 MHz to 300 MHz and 2 GSa/s sample
rates, these oscilloscopes offer 24 kpts/Ch
waveform memory, 32 automatic measurements,
and advanced triggering capabilities including
math functions. Engineered to allow you to see
more of your signal under test, the 2550 series’
widescreen 7” TFT display offers a significantly
larger viewing area than typical economy
oscilloscopes (5.7”).
Educators who want to teach waveform
measurement fundamentals can benefit from
the ability to disable the Auto set button, a
function that automatically sets up the scope
to display a signal.
The 2550 series oscilloscopes are ideal for
applications in design and debug, service and
repair, and education.
2553
2555
Bandwidth
70 MHz
100 MHz
Bandwidth up to 300 MHz
■
2 GSa/s sample rate
■
4-channel acquisition (on select models)
■
Large 7” widescreen color display
■
FFT including four additional math
functions - Add, Subtract, Multiply, and
Divide
■
32 automatic measurements
■
50 Ω input coupling (200 MHz and 300
MHz models)
■
Maximize productivity with PC connectivity via
LAN and USB. The downloadable EasyScope
PC software lets you easily capture, save, and
analyze measurement results. All oscilloscope
parameters can be controlled via a PC without
the need for programming.
Model
Features & Benefits
■
device port (USBTMC compliant) for remote
PC control
■
Technical data subject to change
© B&K Precision Corp. 2012
4
Front and rear panel USB host port for saving
and recalling waveform setups, data, and
screenshots on a USB flash drive
2556
2557
2558
200 MHz
■
PC control through EasyScope software
■
Advanced tools include digital filters with
adjustable limits, pass/fail testing and
2559
waveform recorder mode
300 MHz
■
Channels
Standard LAN (supports SCPI) and USB
2
4
2
www.bkprecision.com
4
Multi language user interface and context
sensitive help
Digital Storage Oscilloscopes
Models 2553, 2555, 2556, 2557, 2558, and 2559
Front panel
Widescreen display
The 7" widescreen color
display lets you see more
of your signal.
Me nu On/Of f but ton
Configure the menu
parameters and hide the
menu with the push of a
button to view your signal
in full screen.
Cont ext sen siti ve h elp
A pop up window will
display the functionality of
a control while help mode
is active.
Auto setup
Vertical, horizontal,
and trigger controls
are automatically
adjusted for fast signal
display.
Ad van ce d trigge ri ng
Isolate the signal with
advanced triggering
including pulse width and
selectable video trigger.
Int ui tive channe l
o pe rati on
All channels in the 2550
series are clearly indicated
by their own color, labeled
on the input, knobs, and
display.
USB h ost po rt
Connect your USB flash
drive to conveniently
store and recall waveform data, setups, and
screenshots.
Prin t but ton
Simply press the Print
button to save a
screenshot in bitmap
format to a USB flash
drive.
Wave form an al ysi s wit h math
and FFT
Analyze your signals with add, subtract,
multiply, and divide functions. View the
signal’s frequency spectrum and perform
harmonic distortion analysis.
Rear panel
USB host port
Additional USB port on rear
for data storage to a USB flash
drive.
K ens ingt on se curity sl ot
Helps to secure your
oscilloscope and prevent theft.
2
Co mmunicatio n
LAN and USB ports enable
remote control from a PC.
www.bkprecision.com
Digital Storage Oscilloscopes
Models 2553, 2555, 2556, 2557, 2558, and 2559
The tools you need
Powerful measurement functions
Digital filtering
Filter OFF
Filter ON
3
Display and measure the input signal’s frequency spectrum. Select
one of the 4 FFT windows: Rectangular, Hanning, Hamming, and
Blackman. Use cursors to measure the spectral component’s
magnitude and frequency.
Filter out unwanted signal components such as various types of noise
with built-in digital filters. Choose from Low-Pass, High-Pass,
Band-Pass, and Band-Stop filters.
Waveform recorder
Delayed sweep/zoom
Monitor and analyze long-term signal behavior by recording data
continuously over an extensive period of time and playing it back for
post acquisition analysis. Data is recorded in a sequence of up to
2500 frames.
Use the oscilloscope’s delayed sweep feature to zoom in a particular
area of a signal in real time while viewing the entire captured
waveform simultaneously.
Large internal storage
Pass/Fail testing
Minimize debug time by saving and recalling setups and waveforms
from internal memory. Save and recall up to 20 different oscilloscope
setups and 20 different waveforms.
Generate user-defined pass/fail limits to quickly identify go/no go
test results.
www.bkprecision.com
Digital Storage Oscilloscopes
Models 2553, 2555, 2556, 2557, 2558, and 2559
PC connectivity
High bandwidth passive oscilloscope
probes
PR150B
The included EasyScope software provides seamless integration
between the oscilloscope and PC. Capture and transfer waveforms,
screen images, setups and measurement results to a Windows PC via
the USB device port on the back of the instrument. A USB host port
on the front and rear allows for quick and easy screen saving.
PR250B & PR500B
Avoid limiting the bandwidth of your measurement system. All 2550
series models come standard with high bandwidth, slimline passive
probes (one per channel) to help you get the most out of your scope.
Features
4
■
Slim, stylish body
■
Snap-locking sprung hook
■
Easily replaceable tip
■
Large accessory set
■
Meets IEC 61010-031 CATII
■
RoHS compliant
Model
Included Probes
2553
four 150 MHz bandwidth, x1/x10 probes (model PR150B)
2555
four 150 MHz bandwidth, x1/x10 probes (model PR150B)
2556
two 250 MHz bandwidth, x10 probes (model PR250B)
2557
four 250 MHz bandwidth, x10 probes (model PR250B)
2558
two 500 MHz bandwidth, x10 probes (model PR500B)
2559
four 500 MHz bandwidth, x10 probes (model PR500B)
www.bkprecision.com
Digital Storage Oscilloscopes
Models 2553, 2555, 2556, 2557, 2558, and 2559
Specifications
2553
2555
2556
70 MHz
100 MHz
2557
2558
2559
Performance Characteristics
Bandwidth
Channels
Rise Time
200 MHz
300 MHz
2 GSa/s (half-channel interleaved)(1), 1 GSa/s (per channel)
Real Time Sampling Rate
4
< 5 ns
2
< 3.5 ns
Ch to Ch Isolation
>100:1 at 35 MHz >100:1 at 50 MHz
(Both channels in same V/div setting)
4
2
4
< 1.8 ns
< 1.2 ns
>100:1 at 100 MHz
>100:1 at 150 MHz
24 kpts (half-channel interleaved)(1)(2), 12 kpts (per channel)
Max Memory Depth
Vertical Resolution
8 bit
Vertical Sensitivity
2 mV/div -10 V/div (1-2-5 order)
DC Gain Accuracy
<±3.0%: 5 mV/div to 5 V/div in fixed gain ranges
<±4.0%: 2 mV/div in variable gain ranges
Maximum Input Voltage
400 V (DC+AC PK-PK, 1 MΩ input impedance, X10), CAT I, 5 Vrms (50 Ω input impedance)
Position Range
2 mV–100 mV: ±800 mV
102 mV – 5 V: ±40 V
Bandwidth Limit
Horizontal Scan Range
20 MHz ±40% (Note: BW limited below 20 MHz when using probe in X1)
5 ns/div – 50 s/div
2.5 ns/div – 50 s/div
1 ns/div – 50 s/div
Timebase Accuracy
±100 ppm measured over 1 ms interval
Input Coupling
AC, DC, GND
Input Impedance
Vertical and Horizontal Zoom
1 MΩ ± 2% || 13 pF ± 3 pF
1 MΩ ± 2% || 13 pF ± 3 pF,
50 Ω ± 2%
Vertically or horizontally expand or compress a live or stopped waveform
I/O Interface
USB
Front and rear USB host ports support USB flash drives, USBTMC compliant USB device port for connecting to PC
LAN
Supports SCPI commands for remote control
Pass/Fail
Pass/Fail output
Acquisition Modes
Sampling
Display sample data only
Peak Detect
Capture the maximum and minimum values of a signal
Average
Waveform averaged, selectable from 4, 16, 32, 64, 128, 256
Trigger System
Edge, Pulse Width, Video*, Slope, Alternative
Trigger Types
*Support signal Formats: PAL/SECAM, NTSC
Trigger condition: odd field, even field, all lines, or line number
Trigger Modes
Auto, Normal, Single
Trigger Coupling
AC, DC, LF reject, HF reject
Trigger Source
CH1, CH2, CH3, CH4, EXT, EXT/5, AC Line
Pulse Width Trigger
Trigger Modes: Positive Pulse (>,<, =), Negative Pulse (>, <, =)
Slope Trigger
Positive slope (>,<, =), Negative slope (>,<, =)
Time: 20 ns-10 s
Alternate Trigger
CH1 trigger type: Edge, Pulse, Video, Slope
CH2 trigger type: Edge, Pulse, Video, Slope
CH3 trigger type: Edge, Pulse, Video, Slope
CH4 trigger type: Edge, Pulse, Video, Slope
Notes:
(1) On 4-Ch models, Ch1 and Ch2 are interleaved, and Ch3 and Ch4 are interleaved. Half channel operation means that only Ch1 or Ch2 and/or only Ch3 or Ch4 is active.
(2) When timebase is 25 ns or faster and maximum data depth mode is enabled.
5
www.bkprecision.com
Digital Storage Oscilloscopes
Models 2553, 2555, 2556, 2557, 2558, and 2559
Specifications
2553
2555
2556
2557
2558
2559
Hardware Frequency Counter
Reading Resolution
6 digits
Accuracy
± 0.01%
Range
DC couple, 10 Hz to MAX bandwidth
Signal Types
Satisfying all trigger signals (except pulse width trigger and video trigger)
Waveform Math and Measure
Math Operation
Add, Subtract, Multiply, Divide, FFT
FFT
Window mode: Hanning, Hamming, Blackman, Rectangular
Sampling points: 1024
Measure
Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean, Crms, Vrms, ROV, FOV, RPRE, FPRE,
FREQ, Period, Rise Time, Fall Time, BWid, + Wid, - Wid, + Duty, - Duty, Phase, FRR, FRF,
FFR, FFF, LRR, LRF, LFR, LFF
Types
Voltage, Time
Measurements
ΔV, ΔT, 1/ΔT (frequency)
Cursors
Display System
Display
7 in. Color TFT, 480 x 234 resolution, 64K color
Display Contrast (Typical state)
150:1
BacklightIintensity (Typical state)
300 nit
Wave Display Range
8 x 18 div
Wave Display Mode
Dots, Vector
Persistence
Off, 1 sec, 2 sec, 5 sec, Infinite
Menu Display
2 sec, 5 sec, 10 sec, 20 sec, Infinite
Screen-Saver
Off, 1 min, 2 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hr, 5 hr
Waveform Interpolation
Sin(x)/x, Linear
Color Mode
Normal, Invert
Environmental and Safety
Temperature
Operating: 50° F to 104 °F (10 °C to +40 °C)
Not operating: -4 °F to 140 °F (-20 °C to +60 °C)
Humidity
Operating: 85%RH, 104 °F (40 °C), 24 hours
Not operating: 85%RH, 149 °F (65 °C), 24 hours
Altitude
Operating: 9,842.5 ft (3,000 m)
Not operating: 50,085.3 ft (15,266 m)
Electromagnetic Compatibility
EMC Directive 2004/108/EC,
EN61326:2006
Safety
Low voltage directive 2006/95/EC, EN61010-1:2001
Power Requirements
100-240 VAC, CAT II, 50 VA max, 45 Hz to 440 Hz
Dimensions (W x H x D)
14.1” x 6.14” x 4.65” (358 x 156 x 118 mm)
Weight
4-channel models: Approx. 9.9 lbs (4.5 kg)
2-channel models: Approx. 9.5 lbs (4.3 kg)
Supplied Accessories
User manual, passive probes (one per channel), power cord, certificate of calibration, USB (Type A to B) communication cable
General
Three-Year Warranty
6