ETC 5492B

Data Sheet
True RMS Bench Multimeter
5492B
True RMS Bench Multimeter
The B&K Precision model 5492B is a versatile
5½-digit, 120,000-count bench multimeter
suitable for applications in education, service
repair, and manufacturing. The instrument
enhances your productivity with built-in math
functions and USB connectivity. Rel, Max/Min,
dBm, dB, %, and Hold functions provide
educators with a convenient tool to teach basic
math concepts. Additionally, the 5492B offers
powerful features not commonly found in other
5½- digit DMMs such as advanced triggering,
buffer storage operation, and a GPIB interface
option.
Features & benefits
The 5492B measures volts and amps with great
accuracy and stability, with basic VDC accuracy
of 0.01%. The meter is also capable of measuring frequency, period, resistance, continuity, and
performing diode tests. Readings can be taken at
a maximum rate of 57 readings/sec and stored
to an internal 512-byte buffer for post
acquisition statistical analysis.
5½-digit, 120,000-count display resolution
■
0.01% basic VDC accuracy
■
Advanced triggering options such as external,
■
Two and four-wire resistance measurement up
■
AC voltage and AC current measurement over
bus and event trigger
to 120 MΩ
wide frequency range (ACV 100 kHz/ACI 10
The 5492B was designed for cost-conscious
users requiring an accurate meter with a broad
range of features offered at a value price.
Increase Productivity with PC Connectivity
kHz)
■
AC (RMS) and DC current measurements up
to 12 A
■
Limit mode for Pass/Fail testing
■
Save/recall up to 10 instrument settings
■
The 5492B is programmable via USB, RS232,
and GPIB (5492BGPIB only) interface using
industry standard SCPI commands. Users can
remotely control and configure the instrument
from a PC and retrieve measurement results for
further analysis. The multimeter can also be
remotely controlled using application software
(via USB & RS232), which supports front panel
emulation and data logging of measurement
results.
Technical data subject to change
© B&K Precision Corp. 2012
■
Built-in math functions: Rel, Max/Min, dBm,
dB, %, Hold, Compare
w w w.bk prec ision.c om
■
CATI (1000 V)/CATII (300 V) protection
■
USB (Virtual Com) and RS232 interface
■
SCPI compatible
True RMS Bench Multimeter
Model 5492B
▲ Versatile tools
▲ Easy operation
Limit Operation
The limit operation lets you set and control the
values that determine a HI / IN / LO status of
subsequent measurements. The meter can be
configured to emit an audible alarm when
readings are outside of the configured limit.
Front panel
Input banana terminals
Bright VFD display
Advanced Triggering
A variety of advanced trigger options are
available: Manual Trigger via a front panel button
press, Bus Trigger sending a *TRG command
via PC interface, External Trigger using the EXT
TRIG input terminal on the rear panel.
Additionally, a trigger delay can be programmed
after an event detection, and a VM Comp output
is available to indicate the instrument’s
completion of a present measurement.
Fast Buffering with Built-in
Statistics
The 5492B provides an internal buffer capable
of storing up to 512 readings. Buffer storage
operation can be enabled via front panel or
remote command. The buffer fills up with the
requested number of readings at a rate of up to
57 readings/sec and then stops. The instrument
automatically calculates statistical information
based on the stored readings such as minimum,
maximum, or average reading, and standard
deviation, which can then be retrieved along with
the readings via the front panel or remote
command.
Power button
Interface keypad
Rear panel
Serial number label
External
Trigger BNC
input terminal
RS-232
(Serial) interface
USB interface
Power line
fuse holder
2
VM Comp
(Voltmeter
complete) BNC
output terminal
mA Fuse socket
GPIB (IEEE-488) interface
(model 5492BGPIB)
Chassis ground
screw terminal
w w w.bk prec ision.c om
True RMS Bench Multimeter
Model 5492B
Specifications
Specifications are based on the following conditions:
■ One year calibration cycle
■
Operating temperature between 18 °C to 28 °C
■
Accuracy expressed as: ± (% of reading + % of range) after 30 minutes warm-up and valid for 10 PLC (Slow)
■
Temperature coefficient: add ± [0.1% × (applicable accuracy)/°C] for 0 °C to 18 °C and 28 °C to 40 °C
■
Relative humidity: Up to 80% RH for 0 °C to 28 °C (75% RH for 10 MΩ and above ranges
for resistance measurement), up to 70% RH for 28 °C to 40 °C
DC Voltage
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Full Scale
Accuracy
Typical Input
Rang e
Resolu tion
Reading
(1 year)
Impe dance
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
1 µV
10 µV
100 µV
1 mV
10 mV
119.999
1.19999
11.9999
119.999
1010.00(2)
0.02+0.008(1)
0.01+0.004(1)
0.01+0.004
0.01+0.004
0.01+0.004
>10 GΩ
>10 GΩ
>10 GΩ
10 MΩ ±1%
10 MΩ ±1%
(1) under REL status
(2) 1% overrange (1010 V) is readable at 1000 V range
Input protection: 1000 V
AC Voltage (True RMS, AC coupled)
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
A ccuracy (1 year) (1 )
Range
Re solution
Fu ll Scale
Re ading
10-20 Hz
20-50 Hz
50-20 kHz
20-50 kHz
50-100 kHz
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
1 µV
10 µV
100 µV
1 mV
10 mV
119,999
1.19999
11.9999
119.999
757.50(3)
1.50+0.100
1.50+0.100
1.50+0.100
1.50+0.100
1.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.30+0.150
0.30+0.150
0.30+0.150
0.30+0.150
0.30+0.150(2)
1.0+0.150
1.0+0.100
1.0+0.100
1.0+0.100
1.0+0.100(1)
(1) Specifications are for sine wave inputs >5% of range
(2) ≤ 3×107 Volt-Hz for 750 V range
(3) 1% overrange (757.5V) is readable at 750 V range
Measurement method: AC coupled true RMS, measuring the AC component only
Maximum Crest factor: 3.0 at full scale
Input impedance: 1 MΩ ± 2% in parallel with capacitance <100 pF
Input protection: 750 Vrms, 500 Vdc
DC Current
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Fu ll Scale
Accuracy
B urden Voltage (1)
Range
Resolu tion
Re ading
(1 year)
& Shu nt Re sistor
12.0000 mA
0.1 µA
11.9999
0.05+0.008(2)
120.000 mA
1 µA
119.999
0.05+0.004(2)
1.20000 A
10 µA
1.19999
0.10+0.004
12.0000 A(3)
100 µA
11.9999
0.25+0.004
(1) Typical voltage across the input terminals at full scale reading
(2) Use REL function
(3) In 12 A range, >10~12 ADC is readable for 20 seconds maximum
<0.15 V / 10.1 Ω
<1.5 V / 10.1 Ω
<0.3 V / 0.1 Ω
<0.15 V / 10 mΩ
Input protection: Externally accessible 2A / 250V fuse protecting the low input current terminal (LO).
Internally accessible 20A / 250V fuse protecting the high current input terminal (10A)
3
w w w.bk prec ision.c om
True RMS Bench Multimeter
Model 5492B
Specifications (cont.)
AC Current (True RMS, AC Coupled)
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Accu racy (1 year) (3 )
Range
Re solution
Full Scale
Readi ng
Bu rden Vol tage (1 )
& Shunt Res istor
10-20 Hz
20-50 Hz
50-2 kHz
2-20 kHz
12.0000 mA
0.1 µA
11.9999
<0.15 V / 10.1 Ω
1.0+0.080
0.50+0.080
0.25+0.080
2.0+0.080
1.20000 A
10 µA
1.19999
<0.3 V / 0.1 Ω
1.0+0.080
0.50+0.080
0.25+0.080
2.0+0.080
12.0000 A(2)
100 µA
11.9999
<0.15 V / 10 mΩ
1.0+0.080
0.50+0.080
0.25+0.080
2.0+0.080
(1) Typical voltage across the input terminals at full scale reading
(2) In 12 A range,>10-12 A AC is readable for 20 seconds maximum
(3) Specifications are for sine wave inputs >5% of range
Measurement method: AC coupled true RMS, measuring the AC component only
Input protection: Externally accessible 2 A / 250 V fuse protecting the low input current terminal (LO). Internally accessible 20 A / 250 V fuse protecting the
high current input terminal (10A)
Max. crest factor: 3.0 at full scale
Resistance (2-Wire and 4-Wire)
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
A ccu racy
Ran ge (1)
Re solution
Full Scale Re ading
Test Curren t
(1 year)
120.000 Ω
1 mΩ
119.999
1 mA
0.05 +0.008(2)
1.20000 kΩ
10 mΩ
1.19999
1 mA
0.03 +0.004(2)
12.0000 kΩ
100 mΩ
11.9999
100 µA
0.03 +0.004(2)
120.000 kΩ
1Ω
119.999
10 µA
0.03 +0.004
1.20000 MΩ
10 Ω
1.19999
10 µA
0.03 +0.004
12.0000 MΩ
100 Ω
11.9999
7.0/(10M+Rx) µA
0.10 +0.004
120.000 MΩ
1 kΩ
119.999
7.0/(10M+Rx) µA
0.50 +0.008
(1) In order to eliminate the noise interference, which might be induced to the test leads, it is recommended to use a
shielded test cable for measuring resistance above 120 kΩ
(2) Using REL function
Input protection: 1000 Vdc or 750 Vac for all ranges
Note: Rx is the measured resistance value
Continuity
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Range
Re solution
Full Scale
Reading
Test Curren t
A ccu racy
(1 year)
1 kΩ
100 mΩ
999.9
1 mA
0.1%+0.02%
For 0.1 PLC (Fast)
Input protection: 1000 Vdc or 750 Vac
Diode
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
R ange
Res ol ution
Fu ll Scale
R eadin g
Te st Cu rrent
Accuracy
(1 year)
3.0000 V
10.0000 V
10.0000 V
100 µV
100 µV
100 µV
2.9999
10.0000
10.0000
1 mA
100 µA
10 µA
0.030+0.020
0.030+0.020
0.030+0.020
For 1 PLC (Medium)
Input protection: 1000 Vdc or 750 Vac for all ranges
4
w w w.bk prec ision.c om
True RMS Bench Multimeter
Model 5492B
Specifications (cont.)
Frequency
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
A CV
Range
Fre qu ency Range
R esolu tion
Fu ll Scale
Re ading
Accuracy (2 )
Input Sensi ti vity
(Sine Wave )
100 mV
to
750 V
5 ~ 10 Hz
10 ~ 100Hz
100 ~ 100 kHz
100 k ~ 1 MHz (1)
10 µHz
100 µHz
1 mHz
1 Hz
9.99999
99.9999
999.999
999.999
0.05
0.01
0.005
0.005
200 mVrms
40 mVrms
40 mVrms
100 mVrms
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.
Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc
Period
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
ACV Range
Period Ran ge
Res olution
Fu ll Scale
Re ading
Accuracy (2 )
I nput Se nsitiv ity
(Sin e Wave )
100 mV
to
750 V
1 ~ 10 µs (1)
10 µs ~ 10 ms
10 ms ~ 100 ms
100 ms ~ 200 ms
0.01 ns
0.1 ns
0.1 µs
1 µs
9.99999
9.99999
99.9999
199.999
0.005
0.005
0.01
0.05
100 mVrms
40 mVrms
40 mVrms
200 mVrms
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.
Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc
Supplemental
Environmental
Operating Environ men t
0 °C to 40 °C, 32 ºF to 104 ºF
Storage En vironment
-40 °C to 70 °C -40 ºF to 158 ºF
Hu midity
≤ 90 %RH
Remote Interfaces
RS232, USB (virtual COM) GPIB (model 5492BGPIB only)
General
Measu reme nt Category
CAT II 300V, CAT I 1000V, pollution degree 2
Math Fu nctions
Rel, Max/Min/Average/StdDev (for stored readings), dB, dBm, Limit Test, % and mX+b
Input Vol tage
110/220 V ± 10%, 50/60 Hz ± 5%
Powe r Cons umption
≤ 20 VA
Safety & EMC
EN 61010-1:2001, EN61326-1:2006 and
EN61326-2-1:2006
Dimen sions (W×H×D )
8.85” x 3.93 “ x 13.97”
(225 mm×100 mm×355 mm)
Ne t Weig ht
5.51 lbs (2.5 kg)
Three Year Warranty
Included Accessories: Test leads, AC power cord, spare fuse, user manual, USB cable, calibration certificate and test report
5
v052412
w w w.bk prec ision.c om