ETC TPP0850

Passive High-voltage Probes
P5100A • TPP0850 • P5122 • P5150 • P6015A Datasheet
P5122
For TPS2000 and THS3000 Series Oscilloscopes
DC to 200 MHz
1000 VRMS CAT II when DC-coupled*1
Floatable up to 600 VRMS CAT II
100X
IEC61010-031
P5150
For TPS2000 and THS3000 Series Oscilloscopes
DC to 500 MHz
P5100A
2500 VPeak, 1000 VRMS CAT II
Features & Benefits
Floatable up to 600 VRMS CAT II or 300 VRMS CAT III
P5100A
DC to 500 MHz
Preferred Probe for Ripple Measurements
50X
2500 VPeak, 1000 VRMS CAT II
10 to 25 pF Compensation Range
100X with Readout Coding
UL61010B-2-031, EN61010-2-031, IEC61010-2-031, CSA61010.2.031
UL3111-1, EN61010-1, IEC61010-2-031, CSA1010.1, CSA1010.2.031
7 to 30 pF Compensation Range
TPP0850
DC to 800 MHz
2500 VPeak, 1000 VRMS CAT II
50X with Readout Coding
Designed for use with the MSO/DPO5000 and MSO/DPO4000B Series
Oscilloscopes
P6015A
High Voltage – 20 kV DC / 40 kV Peak (100 ms Pulse Width)
High Bandwidth – DC to 75 MHz
Silicone Dielectric
Optional 1000X Readout Coding
Wide Compensation Range (7 to 49 pF)
Heavy-duty Versatile Ground Lead and Alligator Clip
Datasheet
Applications
Power Supply Design
Design Motor Drive
Electronic Ballast
Power Semiconductors
Switch Mode Control
UPS Systems
Power Converters
*1 The P5122 probe should not be used on the TPS2000 Series oscilloscopes for AC-coupled measurements
on signals with greater than 300 V DC offset. The P5150 is the recommended probe for measuring ripple
on high-voltage DC supplies.
P5100A High-voltage Probe
TPP0850
The P5100A is a low input capacitance high-voltage probe (2.5 kV)
designed for higher-frequency applications. The probe can be compensated
to match plug-ins and oscilloscopes with nominal input capacitances of
7-30 pF. A variety of screw-on accessories provide easy connection to the
device-under-test.
channels. The P5150 or P5122, coupled with the TPS2000 and
THS3000 Series digital storage oscilloscopes, deliver both the isolation
for the measurement from earth ground and full isolation between the
channels. The P5150 is the recommended probe for measuring ripple on
high-voltage DC supplies. The P5122 probe should not be used on the
TPS2000 Series oscilloscopes for AC-coupled measurements on signals
with greater than 300 V DC offset.
TPP0850 High-voltage Probe
The TPP0850 offers the industry’s highest bandwidth probe (800 MHz)
for high-voltage signals (up to 2500 Vp-p). This is ideal for testing power
semiconductors and switch-mode power supplies, which continue to
increase in switching speed to minimize power loss, creating a need
for faster rise time, higher bandwidth, and higher voltage probes. The
TPP0850 meets this need and can be used with switching transistor circuits
operating at 1200 VRMS, above the voltage operating range of standard
general-purpose probes. It will also be able to accommodate emerging
power applications that will require even faster rise time capabilities.
P6015A High-voltage Probe
For heavy-duty high-performance measurements of voltages over 2.5 kV,
the P6015A is the industry standard. You can measure DC voltages up
to 20 kVRMS and pulses up to 40 kV (peak, 100 ms duration). The 75 MHz
bandwidth enables you to capture fast, high-voltage signals.
P5150 and P5122 IsolatedChannel™
Applications
The P6015A uses an environmentally safe silicone compound for a
dielectric and never needs refilling. Other features include: A 7-49 pF
compensation range, small compensation box that fits on adjacent amplifier
inputs, and a readout option for use with most Tektronix digital scopes. With
the readout option, displayed voltage amplitude values will be the actual
signal value rather than understated by a factor of 1,000.
In many applications, it is important to be able to isolate the measurement
from earth ground and also to isolate the common voltage between
Note: Using the readout version with other than Tektronix digital scopes
may result in an erroneous readout display.
2
www.tektronix.com
Passive High-voltage Probes — P5100A • TPP0850 • P5122 • P5150 • P6015A
Characteristics
P5100A/TPP0850/P5122/P5150/P6015A
Probe
Nominal
Length
Attenuation
Bandwidth
Rise Time
(Typical)
Loading
P5100A
2m
100X
500 MHz
<700 ps
40 MΩ / 2.5 pF
TPP0850
1.3 m
50X
800 MHz
<525 ps
40 MΩ / 1.8 pF
P5122
P5150
1.2 m
2m
100X
50X
200 MHz
500 MHz
2.2 ns
<700 ps
100 MΩ / 4.0 pF
40 MΩ / 3.8 pF
P6015A
P6015A
Opt. 1R
3m
3m
1000X
1000X
75 MHz
75 MHz
4.0 ns
4.0 ns
100 MΩ / 3 pF
100 MΩ / 3 pF
Ordering Information
P5100A/TPP0850
High-voltage Probes.
Includes: Large hook tip (013-0384-xx), small hook tip (013-0386-xx), 6 in.
ground lead (196-3524-xx), 18 in. ground lead (196-3525-xx), ground spring
(214-5298-xx), crocodile clip (344-0461-xx), adjustment tool (003-1433-xx), color
bands (016-1886-xx).
Recommended Accessories
Accessory
Description
013-0291-xx
206-0060-xx
Probe Tip to BNC Adapter
0.080 in. diameter Spring Tip
P5122
100X, High-voltage Probe for TPS2000 and THS3000 Series Oscilloscopes.
Includes: Large retractable hook tip, lead with hook tip, lead with crocodile clip,
adjustment tool, instruction manual.
P5150
50X, High-voltage Probe for TPS2000 and THS3000 Series Oscilloscopes.
Includes: Large hook tip (013-0389-xx), small hook tip (013-0388-xx), 6 in.
ground lead (196-3526-xx), 18 in. ground lead (196-3527-xx), common spring
(214-5299-xx), crocodile clip (344-0461-xx), color bands (016-1886-xx), adjustment
tool (003-1433-xx).
Max. Input V Max. Float V Comp. Range
DC or RMS
in pF
2.5 kVPeak
N/A
1000 VRMS CAT II
2.5 kVPeak
N/A
1000 VRMS CAT II
1000 VRMS CAT II 600 VRMS CAT II
2.5 kVPeak
600 VRMS CAT II,
1000 VRMS CAT II 300 VRMS CAT III
20 kV
N/A
20 kV
N/A
Readout
7 to 30
Yes
N/A
Yes
10 to 22
10 to 25
No
No
7 to 49
7 to 49
No
Yes
P6015A
1000X, 3-meter High-voltage Probe.
Includes: Hook probe tip (206-0463-xx); banana plug tip (134-0016-xx); crocodile
clip – plugs onto ground lead (344-0461-xx); ground lead (196-3363-xx); carrying
case (016-1147-xx); instruction manual (070-8223-xx).
Options (P6015A only)
Option
Description
Opt. 1R
3-meter length with readout
Service (P6015A only)
Option
Description
Opt.
Opt.
Opt.
Opt.
Opt.
Opt.
Opt.
Opt.
Opt.
Calibration Service 3 Years
Calibration Service 5 Years
Calibration Data Report
Calibration Data Report 3 Years (with Opt. C3)
Calibration Data Report 5 Years (with Opt. C5)
Repair Service 3 Years
Repair Service 5 Years
Standard Warranty Extended to 5 Years (TPP0850)
Standard Warranty Extended to 5 Years (P6015A)
C3
C5
D1
D3
D5
R3
R5
SILV200
SILV600
Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.
Recommended Accessories for P5150
See P5100A/TPP0850 Recommended Accessories.
www.tektronix.com
3
Datasheet
Contact Tektronix:
ASEAN / Australasia (65) 6356 3900
Austria 00800 2255 4835*
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 00800 2255 4835*
Brazil +55 (11) 3759 7627
Canada 1 800 833 9200
Central East Europe and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark +45 80 88 1401
Finland +41 52 675 3777
France 00800 2255 4835*
Germany 00800 2255 4835*
Hong Kong 400 820 5835
India 000 800 650 1835
Italy 00800 2255 4835*
Japan 81 (3) 6714 3010
Luxembourg +41 52 675 3777
Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90
Middle East, Asia, and North Africa +41 52 675 3777
The Netherlands 00800 2255 4835*
Norway 800 16098
People’s Republic of China 400 820 5835
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 001 800 8255 2835
Russia & CIS +7 (495) 7484900
South Africa +41 52 675 3777
Spain 00800 2255 4835*
Sweden 00800 2255 4835*
Switzerland 00800 2255 4835*
Taiwan 886 (2) 2722 9622
United Kingdom & Ireland 00800 2255 4835*
USA 1 800 833 9200
* European toll-free number. If not accessible, call: +41 52 675 3777
Updated 10 February 2011
For Further Information. Tektronix maintains a comprehensive, constantly expanding
collection of application notes, technical briefs and other resources to help engineers working
on the cutting edge of technology. Please visit www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents,
issued and pending. Information in this publication supersedes that in all previously published material.
Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
of their respective companies.
26 Jul 2012
www.tektronix.com
56W-10262-10