www.analog.com/apm/adi-water-analysis-solution_en.pdf

Project Code: APM-Water-Analysis-2013
ADI Water Analysis Solution for pH Meters
and Conductivity Meters
Application Introduction
In this article, we will talk about two of the most popular applications in the water analysis field: pH meters and conductivity meters.
pH is a measure of the acidity or alkalinity of a solution. As we know, most living things depend on a proper pH level to sustain life. The measurement of pH is
critical in many fields ranging from chemical processing, pulp and paper, food, beverage, pharmaceutical, and so on.
Conductivity is a measure of how well a solution conducts electricity, which is an extremely widespread and useful method for quality control purposes.
Surveillance of water purity, control of drinking water and process water quality, and estimation of the total number of ions in a solution can all be performed
using conductivity measurements. In general, the measurement of conductivity is a rapid and inexpensive way of determining the total concentration of ions
in a solution. However, it is a nonspecific technique, unable to distinguish between different types of ions, giving instead a reading that is proportional to the
combined effect of all the ions present.
System Design Considerations
Stability: Drift with time and temperature are very important factors during pH meter and conductivity meter design. To achieve this objective, low drift and an
accurate signal chain would be required, which ADI is very good at.
Resolution: To take full advantage of sensor dynamic range, low noise and high resolution should be taken into consideration during signal chain and power
design, especially for laboratory instruments.
Low Power Consumption: Portability is one major trend in instruments recently. pH meters and conductivity meters also have portable versions, which should
work a pretty long time outdoors with limited battery resources. In this condition, low power consumption is required during the design phase.
Solutions from ADI
System Block Diagram
1.Below is the system block diagram of a pH meter including pH electrodes, low leakage input stage, gain stage (optional), microcontroller (ADC and reference
integrated), power management, and a communication interface.
5. POWER MANAGEMENT
BATTERY
DC-TO-DC
LDO
4.REF
6. INTERFACE
1. AMP
1. AMP
2. ADC
LOW LEAKAGE
BUFFER STAGE
GAIN STAGE (OPTIONAL)
RTD
TEMPERATURE
PH ELECTRODE
3. MICROCONTROLLER
4 mA TO 20 mA
RS-485
RS-232
LCD/KEY/
ALARM UNIT
PROGRAMMABLE
EXCITATION CURRENT
Note: The signal chains above are representative of the system block diagram of a system block diagram of pH meter design. The technical requirements of the blocks vary, but the
products listed in the table below are representative of ADI's solutions that meet some of those requirements.
instrumentation.analog.com
1. Amplifier
2. ADC
3. Microcontroller
4. Reference
5. Power Management
6. Interface
ADA4505-2/AD8626
ADA4665-2/ADA4692-2
AD7792/AD7793
ADuCM361/ADuC7061
ADR4525/
ADR3425/
ADR291
ADP2503/ADP2370/
ADP160/ADP7102
AD5412/AD5422/
ADM2484E/ADM3251E
pH measurement can be achieved by determining the hydrogen ion concentration in solution through measuring an electrode and reference electrode.
The purpose of the reference electrode is to provide a constant reference potential regardless of the solution in which it is immersed. The function of the
measuring electrode is to develop a potential difference against the reference electrode, which is proportional to the pH value of the solution. A low bias current
amplifier should be applied as a buffer to achieve accurate pH measurement since pH electrodes have a very large output resistance. After the low leakage
buffer stage, the signal is presented at the gain and low pass filter stage to achieve more resolution. It finally goes into the Σ-Δ ADC, which is also responsible
for temperature measurement of the solution for compensation.
Stability: ADI is committed to providing accurate signal chain products like low bias current and low offset drift amplifiers, low noise/drift/hysteresis
references, and highly accurate ADCs, all of which help designers to build accurate and stable pH measurement systems.
The amplifiers in the table above with around 1 μV/°C offset drift and around 1 pA, not only at room temperature but also from 0°C to 50°C, can make
environmental testing easy.
The reference in the table above is also not only good at low drift and low noise but also the hysteresis in a wide temperature range helps to pass
environmental tests, too.
The ADCs in the table above feature not only low noise but also temperature measurement, which is used to compensate pH measurement. The internal
current programmable excitation and potential bias are great fit to RTD and TC temperature measure.
Resolution: Most pH meters deliver 3½ digits resolution, which means a 16-bit Σ-Δ ADC is enough for this application. Some laboratory pH meters can go up
to 5½ digits resolution, which means a 24-bit Σ-Δ ADC like the AD7793 and low flicker noise amplifier ADA4505-2/AD8626 should be applied.
Low power consumption: As we know, one trend of instrumentation is smaller and portable. The products in the table above for every category get
low power consumption, like ADA4505-2/AD8626, ADR291, AD7792/3, ADuCM361, ADP2503, ADP160.
2.Below is the system block diagram of a conductivity meter including conductivity electrode, programmable AC current source, voltage sense, current sense,
microcontroller (ADC and reference integrated), power management, and communication interface.
2 | ADI Water Analysis Solution for pH Meters and Conductivity Meters
PROGRAMMABLE AC
CURRENT SOURCE
9. POWER MANAGEMENT
BATTERY
4.
M
U
X
DC-TO-DC
LDO
1. AMP
5. SWITCH
1. AMP
1. AMP
7. REF
CONDUCTIVITY
ELECTRODES
VOLTAGE SENSE
5. SWITCH
RTD
4.
M
U
X
1. AMP
CURRENT SENSE
6. ADC
5. SWITCH
8. MICROCONTROLLER
3. DIFF
AMP
2. INA
3. DIFF
AMP
10. INTERFACE
4 mA TO 20 mA
RS-485
RS-232
TEMPERATURE
LCD/KEY/
ALARM UNIT
PROGRAMMABLE EXCITATION CURRENT
Note: The signal chains above are representative of the system block diagram of a conductivity meter design. The technical requirements of the blocks vary, but the products listed
in the table below are representative of ADI's solutions that meet some of those requirements.
1. Amp
2. INA
3. Difference Amplifier
4. Mux
5. Switch
AD8626/ADA4692-2
AD8603/AD8605
AD8220/AD8228/AD8421
AD8271/AD8278
ADG704/ADG708/ADG1609
ADG733/ADG1636
6. ADC
7. Ref
8. MCU
9. Power Management
10. Interface
AD7792/AD7793
ADR4525/ADR3425
ADR291
ADuCM361/ADuC7061
ADP2503/ADP2370
ADP160/ADP7102
AD5412/AD5422
ADM2484E/ADM3251E
Conductivity is a measure of the total concentration of ions in solution which can be achieved by applying alternating current. This current forces the ions to flow forth
and back to produce alternating potential between conductivity electrodes. The conductivity meter measures the voltage with two nodes mode and four nodes mode
depending on system accuracy requirements. With Ohm’s law, the conductance of the solution can be calculated with measured voltage and current. As shown in the
block diagram, the current is generated in the programmable AC current source with reference, op amp and multiplexer. The voltage across the electrodes is sensed by
an instrumentation amplifier followed by switch and difference amplifier. The current through the electrodes is sensed by an I/V circuit followed by a switch and difference
amplifiers, too. The switch and difference amplifier demodulation circuit can transfer AC signals to DC signals that present at the input of Σ-Δ ADCs.
Stability: Like pH meters, ADI precision signal chain products can help designers to build an accurate and stable conductivity measurement system. Precision
references, low bias drift op amp, and low on resistance multiplexers compose a precision programmable current source. Low bias current instrumentation
amplifiers, low on resistance switches, and difference amplifiers make accurate voltage measurement. To get accurate conductivity results, temperature
measurement based RTD is done which can be facilitated by an ADI ADC with internal current source.
Low power consumption: Like pH meters, the products in the table above for almost every category get low power consumption, which can compose a low
power consumption system.
instrumentation.analog.com 
| 3
Main Products
Part Number
Description
Benefits
Operational Amplifier
ADA4505-2
0.5 pA bias current @ typ room temperature, less than 1 pA bias current
@ typ 50°C, low offset drift 2 μV/°C, 10 μA quiescent current, rail-to-rail
both input/output, zero input crossover distortion amplifiers;
PSRR: 100 dB minimum; CMRR: 105 dB typical
Low bias current @ 0°C to 50°C, low offset drift, suitable for pH meter,
low power consumption, suitable for portable instrumentation,
higher PSRR brings little affection to accuracy even without regulator
AD8626
0.25 pA bias current @ typ room temperature, less than 2 pA bias
current @ typ 50°C, low offset drift 2 μV/°C, up to ±13 V power supply,
high bandwidth 5 MHz, rail-to-rail output
Wider power supply range, low bias current @ 0°C to 50°C, low offset drift,
suitable for pH meter, high bandwidth for AC current source
ADA4665-2
0.1 pA bias current @ typ room temperature, 0.2 pA bias current @ typ 50°C,
low offset drift 3 μV/°C, up to ±8 V power supply, rail-to-rail input/output
Wider power supply range, low bias current @ 0°C to 50°C, low offset
drift, suitable for pH meter
ADA4692-2
0.5 pA bias current @ typ room temperature, 0.5 pA bias current @ typ 50°C, Low bias current @ 0°C to 50°C, low offset drift, low noise and high
low offset drift 1 μV/°C, 3.6 MHz bandwidth, rail-to-rail output
bandwidth, suitable for pH meter and AC current source.
AD8603
1 pA max @ 25°C, micropower: 50 μA, low offset voltage: 50 μV max,
rail-to-rail input/output
Low bias current at room temperature, low power and low offset opamp
AD8605
1 pA max @ 25°C, low offset voltage: 65 μV max, 10 MHz bandwidth,
high output current: 80 mA, rail-to-rail input/output
Low bias current at room temperature, high speed, high output current,
low offset opamp
Instrumentation Amplifier
AD8220
JFET input, low bias current 10 pA @ typ, high bandwidth 1.5 MHz @
G = 1, gain range 1 to 1000
Low bias current, enough bandwidth, suitable for conductivity meter
AD8228
Low bias current 0.5 nA, low gain drift 1 ppm/°C, low noise 15 nV/√Hz
Fixed gain with internal resistor save cost and improve gain accuracy
AD8421
Low bias current 0.1 nA, low noise 3 nV/√Hz, high bandwidth
10 MHz @ G = 1, low offset drift 0.2 μV/°C, Slew rate 35 V/μs
Low noise and low bias current, high speed instrumentation amplifier
AD8271
Gain = ½, 1, 2; gain drift 10 ppm/°C; 15 MHz; 30 V/μs slew rate
Low gain drift and high speed, suitable for the drive ADC in conductivity
application
AD8278
Low power consumption 100 μA, G = ½ or 2, bandwidth 1 MHz,
Low power consumption, enough bandwidth for conductivity meter
ADG704
4-channel multiplexer, low on resistance 2.5 ohm @ typ, low leakage
current 10 pA @ typ, low power consumption 1 μA
Low leakage and low on resistance help to build high accurate system
ADG708
8-channel multiplexer, low on resistance 3 ohm @ typ, low leakage current
10 pA @ typ, low power consumption 1 μa
Low leakage and low on resistance help to build high accurate system
ADG1609
4-channel multiplexer, ±8 V power supply, low on resistance 4.5 ohm @ typ,
low leakage current 20 pA @ typ, low power consumption 1 μA
Wider power supply range, low leakage and low on resistance help to build
high accurate system
ADG733
Double SPDT switch, low on resistance 2.5 ohm @ typ, low leakage current
10 pA @ typ, low power consumption 1 μA
Low leakage and low on resistance help to build high accurate system
ADG1636
Double SPDT switch, ±8 V power supply low on resistance 2.5 ohm @ typ,
low leakage current 10 pA @ typ, low power consumption 1 μA
Wider power supply range, low leakage and low on resistance help to build
high accurate system
AD7792
400 μA quiescent current, 3-channel 16-bit peak-to-peak resolution, up
to 470 Hz output update rate, on chip reference, internal bias voltage,
internal current excitation
Low power consumption and high integrated Σ-Δ ADC, high resolution and
high accuracy, suitable for precision measurement especially temperature
AD7793
400 μA quiescent current, 3-channel 24-bit Σ-Δ ADC, up to 470 Hz output
Low power consumption and high integrated Σ-Δ ADC, high resolution and
update rate, on-chip reference, internal bias voltage, internal current
high accuracy, suitable for precision measurement especially temperature
excitation
Difference amplifier
Mux
Switch
ADC
4 | ADI Water Analysis Solution for pH Meters and Conductivity Meters
Main Products (Continued)
Part Number
Description
Benefits
Reference
ADR4525
2.5 V reference, very low drift: 2 ppm/°C (max), low noise: 1.25 μV pp @
0.1 Hz to 10 Hz, long time stability: 25 ppm/√1000hr, hysteresis: 50 ppm
ADR3425
2.5 V reference, low drift 8 ppm/°C (max), long time stability 30 ppm/√1000hr, Low drift, good stability, many other choices for output voltage in
100 μA max quiescent current, small size SOT-236 package
ADR34xx family
ADR291
2.5 V reference,12 μA quiescent current
Low power consumption, pretty good drift and stability
ADuCM361
Precision analog microcontrollers, ARM Cortex™ -M3 32-bit processor,
6 differential channels, single (24-bit) ADCs, single 12-bit DAC, power
consumption 1.0 mA, 290 μA/MHz, 19-pin GPIO, 128k bytes flash/EE
memory, 8k bytes SRAM. Small package, low drift internal reference
5 ppm typical, integrated programmable current source
Low power consumption, high precision 24-bit Σ-Δ ADC, 4 mA to 20 mA
loop applications, small package
ADuC7061
A precision analog microcontroller based on a 10 MHz ARM7 and a highly
precise dual sigma-delta ADC front-end, 24 bits of resolution and 16-bit
ENOB and sub-100 Hz output rates; memory footprint includes a 32 kB
flash and 4 kB SRAM; other key specs includes sub-3 mA operation (with
MCU core at 1 MHz) making the part suitable for 4 mA to 20 mA loop
applications, a 12-bit DAC and small packaging, 5 mm × 5 mm 32-lead LFCSP
Low power consumption, low cost 24-bit Σ-Δ ADC, 4 mA to 20 mA loop
applications, small package
ADP2503
38 μA quiescent current; 2.5 MHz buck-boost dc-to-dc converters, has
ability to operate at input voltages greater than, less than, or equal to the
regulated output voltage
Low power consumption to achieve long battery life, small package and
few external parts around cost small PCB space
ADP2370
3.0 V to 15 V input buck regulator, 800 mA output current, 1.2 MHz or 600 KHz
PWM frequency, low quiescent current 14 μA, high efficiency larger than
90%, current-mode control architecture
Small 3 mm × 3 mm LFCSP package, few peripHeral components, and
small solution size
ADP160
2.2 V to 5.5 V input LDO, 150 mA maximum output current, ultralow
quiescent current: 10 μA when output 10 mA, up to 15 fixed-output
voltage options available from 1.2 V to 4.2 V
Low power consumption, integrated output discharge resistor, small
package with only two 1 μF external capacitor
ADP7102
20 V input LDO, 300 mA output current, low noise 15 μV rms,
7 fixed version and adjustable version
High input voltage, low noise LDO
AD5422
Current output ranges: 0 mA to 24 mA, voltage output range: 0 V to 5 V,
0 V to 10 V, ±5 V, ±10 V, 16-bit resolution, 0.01% FSR typical total
unadjusted error; 3 ppm/C typical output drift; on-chip reference
(10 ppm/°C maximum)
16-bit resolution and monotonicity, supports HART communication
AD5412
Current output ranges: 0 mA to 24 mA, voltage output range: 0 V to 5 V,
0 V to 10 V, ±5 V, ±10 V, 12-bit resolution, 0.01% FSR typical total
unadjusted error; 3 ppm/°C typical output drift; on-chip reference
(10 ppm/°C maximum)
12-bit resolution and monotonicity, supports HART communication
ADM2484E
Full/half-duplex isolated RS-485/RS-422 transceiver, 500 kbps data rate,
256 nodes, 5 V or 3.3 V operations, 15 kV ESD protection 5 kV isolation
Highly integrated isolated RS-485 transceiver
ADM3251E
Isolated RS-232 transceiver, 460 kbps data rate, 5 V or 3.3 V operations,
15 kV ESD protection, 2.5 kV isolation
Highly integrated isolated RS-232 transceiver
Low drift, very good stability and low noise reference, low hysteresis,
many other choices for output voltage in ADR45xx family
Microcontroller
Power Management
Interface
instrumentation.analog.com 
| 5
Design Resources
Application Notes/Articles
• Dual-Channel Colorimeter with Programmable Gain Transimpedance Amplifiers—www.analog.com/CN0312
• Programmable-Gain Transimpedance Amplifiers Maximize Dynamic Range in Spectroscopy Systems
—www.analog.com/library/analogdialogue/archives/47-05/pgtia
Design Tools/Forums
• ADuCM361 Design Tools—ftp://ftp.analog.com/pub/MicroConverter
• Analog Photodiode Wizard—www.analog.com/en/content/photodiode_wizard/fca
• Analog Filter Wizard™: ADI Active Filter Design Tool—www.analog.com/en/segment/dt-input-stage-filter-design/Filter_Wizard/resources/fca
• ADIsimPower™: ADI Voltage Regulator Design Tool—www.analog.com/adisimpower
• ADIsimOpAmp™: ADI OpAmp Design Tool—www.analog.com/adisimopamp
• EngineerZone®: Online Technical Support Community—ez.analog.com
To view additional gas detector resources, tools, and product information, please visit: instrumentation.analog.com
Customer Interaction Center
Analog Devices China
Asia Pacific Headquarters
22/F One Corporate Ave.
222 Hu Bin Road
Shanghai, 200021
China
Tel: 86.21.2320.8000
Fax: 86.21.2320.8222
Technical Hotline 1-800-419-0108 (India)
1-800-225-5234 (Singapore)
0800-055-085 (Taiwan)
82-2-2155-4200 (Korea)
Email [email protected]
EngineerZone ez.analog.com
Free Sample www.analog.com/sample
Analog Devices, Inc.
Korea Headquarters
6F Hibrand Living Tower
215 Yangjae-Dong
Seocho-Gu
Seoul, 137-924
South Korea
Tel: 82.2.2155.4200
Fax: 82.2.2155.4290
Analog Devices, Inc.
Taiwan Headquarters
5F-1 No.408
Rui Guang Road, Neihou
Taipei, 11492
Taiwan
Tel: 886.2.2650.2888
Fax: 886.2.2650.2899
Analog Devices, Inc.
India Headquarters
Rmz - Infinity
#3, Old Madras Road
Tower D, Level 6
Bangalore, 560 016
India
Tel: 91.80.4300.2000
Fax: 91.80.4300.2333
Analog Devices, Inc.
Singapore Headquarters
1 Kim Seng Promenade
Great World City
EastTower, #11-01
Singapore, 237994
Singapore
Tel: 65.6427.8430
Fax: 65.6427.8436
©2013 Analog Devices, Inc. All rights reserved.
Trademarks and registered trademarks are the
property of their respective owners.
BR11684-0-7/13
instrumentation.analog.com