AGILENT B2200A

Agilent B2200A
fA Leakage Switch Mainframe
Agilent B2201A
14ch Low Leakage Switch Mainframe
Data Sheet
Introduction
Basic functions
Agilent B2200A fA leakage switch
mainframe and B2201A 14ch low
leakage mainframe functions:
• Switches DC current, DC voltage,
capacitance and LCR meters
• Set and monitor connection status
on front panel
• Configurated status display with
LED matrix display
• Connection setup by light pen
• Controls through GP-IB interface
• Auto ground function
• Self-test, relay function test
• Relay cleaning
Agilent B2200A fA leakage switch mainframe
Configuration
General specifications
Temperature range
Input (B2200A/B2201A)
• IV port: 8 triaxial ports
(or 4 Kelvin triaxial ports) with
low leakage performance
• AUX port: 6 BNC ports including
dedicated 2 CV ports
• All input ports can use concurrently
Output (B2210A/B2211A)
• 12 triaxial ports per switch module.
Frame can have 4 modules
maximum. (Total 48 ports)
Software
The B2200A and B2201A include
a CD-ROM with VXI Plug&Play driver
which contains capacitance compensation routine.
Operation:
Storage:
Operation:
Storage:
+5° C to 35° C
–20° C to 70° C
Humidity range
+5% to 70% R.H. no condensation
(B2200A, B2210A)
< 80% @35° C, < 60% @ 65° C
no condensation
(B2201A, B2211A)
< 80% @ 65° C no condensation
Altitude
Operation:
0 m to 2,000 m
Storage:
0 m to 15,240 m
Regulatory
Safety:
CSA C22.2 No.1010.1/IEC 1010-1
Compliance
EMC:
CISPR 11 Group 1 class A&EN50082-1
Power requirement 90 to 264 V (continuous), 47 to 63 Hz, 2 A/200 VA max
Number of slots
4 slots for 48 mm height switch module.
Size
B2200A/B2201A
430 mm W x 320 mm H x 600 mm D
B2210A/B2211A
395 mm W x 48 mm H x 500 mm D
Weight (approx.)
B2200A/B2201A
14.0 kg
B2210A
4.5 kg
B2211A
3.5 kg
Number of ports (B2210A, B2211A use with B2200A, B2201A)
I-V port:
8 triaxial ports (with guard)
AUX port:
6 BNC ports (2 of CV port)
Output channel:
12 triaxial ports (with guard),
max. 48 ports
Switch module specifications (Used with mainframe*1)
Condition
Max current rating (A):
Max voltage rating (V):
Close channel residual resistance (Ω):
Channel isolation resistance (Ω):
I-V port
AUX port
I-V: (to common)
I-V: (to other ch.)
AUX: (to common)
AUX: (to other ch.)
I-V port
AUX port
I-V port
AUX port
Condition: 23° C ± 5° C, 5% to 60% R.H.
1. B2200A only supports B2210A fA leakage switch module.
B2201A only supports B2211 14ch low leakage switch module.
Mixed configuration is not supported.
2
B2210A
B2211A
1.0
0.5
200
300
100
100
0.6
1.5
1x1014
1x109
1.0
0.5
200
300
100
100
0.6
1.5
5x1013
1x109
Specification condition
Supplemental characteristics (B2200A/B2210A)*1
Offset current
IM noise (RMS)
Channel crosstalk capacitance
Offset voltage
Settling time*4
Band width (at –3 dB)
Guard capacitance
Additional C measurement error
2
10 fA*
0.6 fA*3
< 1 pF/ch
< 3 pF/ch
< 50 µV
< 80 µV
2.0 sec at 50 fA
30 MHz
< 145 pF*5
< ± 1 % + 0.2 pF*6
I-V port
I-V port
I-V port
AUX port
I-V port
AUX port
AUX port
I-V port
AUX port
Offset voltage
Settling time*4
Band width (at –3 dB)
Guard capacitance
Additional C measurement error
100 fA*7
5 fA*3
< 0.5 pF/ch
< 3 pF/ch
< 80 µV
< 100 µV
2.0 sec at 300 fA
30 MHz
< 145 pF*5
< ± 1 % + 0.2 pF*6
I-V port
I-V port
I-V port
AUX port
I-V port
AUX port
AUX port
I-V port
AUX port
*1. B2200A only supports B2210A fA leakage switch module.
B2201A only supports B2211 14ch low leakage switch module.
Mixed configuration is not supported.
*2. The offset current when zero volts is applied to all input and output channels.
This measurement is made on a port, 5 seconds after a switching a relay.
*3. Measured 100 PLC by 4156C when zero volts are applied to all other paths.
*4. After 10 V applied.
*5. The guard capacitance of the closed port on input and output ports when
4 modules per frame are installed.
*6. The additional error using the C-compensation algorithm for Agilent 4284 at
1 kHz to 1 MHz, <1000 pF and 3 m cable.
*7. The offset current when zero volts is applied to all input and out channels.
This measurement is made on a port, 60 seconds after a switching a relay.”
Agilent B2200A
fA leakage switch
mainframe (back
panel)
3
The supplemental characteristics
entries in the following specification are not warranted but they
provided useful information about the
functions and performances of the
instruments.
Accessories & cables
Supplemental characteristics (B2201A/B2211A)*1
Offset current
IM Noise (RMS)
Channel crosstalk capacitance
Specification and supplemental
characteristics are defined at 23° C ±
5° C < 60% relative humidity (R.H.)
Agilent 16443A
Agilent 16494A
Agilent 16493K
Agilent 16494B
Agilent 16494F
Agilent 16493N
Light pen for B2200A/
B2201A
Triaxial cable
Kelvin triaxial cable
(for input port)
(Between 4155/56
series or E5270 series
and B2200/2201A
input port)
Kelvin triaxial cable
(for output port)
CMU input cable
GND cable
(Between GND of
E5270/41501 and
B2201A/B2200A)
www.agilent.com
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
For information regarding self
maintenance of this product, please
contact your Agilent office.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
Product specifications and descriptions
in this document subject to change
without notice.
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 1, 2008
© Agilent Technologies, Inc. 2009
Printed in USA, February 25, 2009
5989-1354EN