ROITHNER ERYF

Special UV-Index Sensor
EryF
Features of the EryF special UV-Index
Sensor
The UVI Sensor EryF is based on our approved broad band UV-Sensor TW30SX and
uses a Filter leading to excellent accordance with the erythema action curve of the
human skin. The EryF is designed for use as a erythema sensor according to ISO
17166 CIE S 007/E (2000) – DIN 5050. It is optimally suited to feature high quality
instruments for exact measurements of the UV Index. Overview on the advantages:
•
UVI precision is possible up to ± ½ UVI
•
the Sensor’s current is directly proportional to the UV-Index
•
also suited for sun tanning bank dosimetry
•
Based on approved TW30SX technology
•
Schottky-type photodiode
•
Intrinsic visible blindness due to
wide-bandgap semiconductor material
•
large photoactive area
•
designed to operate in photovoltaic mode
•
hermetically sealed metal TO18 housing and UV-glass window
•
we are able to manufacture up to 2.000.000 pcs. per year.
Rev. 1.4
Page 1 [4]
Special UV-Index Sensor
EryF
Maximum Ratings
Parameter
Symbol
Value
Unit
Topt
-20 ... +80
°C
Reverse voltage
VRmax
3
V
Forward current
IFmax
1
mA
Total power dissipation at 25°C
Ptot
1
mW
Operating temperature range
General Characteristics
(Ta = 25 °C)
Parameter
Symbol
Value
Unit
A
4,18
mm2
LxW
2.2 x 1.9
mm
Max. viewing angle
α
70
degree
Shunt resistance
(dark)
Rs
300
MΩ
Dark current at
10mV reverse bias
Id
30
pA
Open circuit voltage
(200µW/cm2, λ=300nm)
V0
>250
mV
min. Short circuit current
(200µW/cm2, λ=300nm)
I0
160
nA
VBR
>3
V
Active area
Active area dimensions
Breakdown voltage
(dark)
Rev. 1.4
2
Page 2 [4]
Special UV-Index Sensor
EryF
Spectral Characteristics
(Ta = 25 °C)
Symbol
garanteed
Value
Unit
min. spectral sensitivity at peak
Smax
19
mA W -1
Wavelength of peak spectral sensitivity
λSmax
300
nm
-
215 - 325
nm
Smax
S 400nm
10000
Parameter
Range of spectral sensitivity
(S=0.1*Smax)
Visible blindness
Spectral Response
Rev. 1.4
Page 3 [4]
Special UV-Index Sensor
EryF
Pin Layout