MEANWELL MSP-200

MSP-200
200W Single Output Medical Type
¡¡E¡½
E
¡E
¡E
¡E
¡E
¡E
¡E
¡E
¡E
¡E
¡E
¡E
series
Features :
Universal AC input / Full range
Built-in active PFC function, PF>0.95
High efficiency up to 89%
Withstand 300VAC surge input for 5 seconds
Protections: Short circuit / Overload / Over voltage / Over temperature
Cooling by free air convection
Built-in constant current limiting circuit
1U low profile 38mm
Built-in remote ON-OFF control
Standby [email protected]
Built-in remote sense function
No load power consumption<0.5W (Note.6)
5 years warranty
SPECIFICATION
MODEL
OUTPUT
MSP-200-3.3
MSP-200-5
MSP-200-7.5
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
36V
48V
RATED CURRENT
40A
35A
26.7A
16.7A
13.4A
8.4A
5.7A
4.3A
CURRENT RANGE
0 ~ 40A
0 ~ 35A
0 ~ 26.7A
0 ~ 16.7A
0 ~ 13.4A
0 ~ 8.4A
0 ~ 5.7A
0 ~ 4.3A
RATED POWER
132W
175W
200.3W
200.4W
201W
201.6W
205.2W
206.4W
MSP-200-24
MSP-200-36
MSP-200-48
RIPPLE & NOISE (max.) Note.2 80mVp-p
90mVp-p
100mVp-p
120mVp-p
150mVp-p
150mVp-p
250mVp-p
250mVp-p
4.3 ~ 5.8V
2.0%
6.8 ~ 9V
2.0%
10.2 ~ 13.8V
1.0%
13.5 ~ 18V
1.0%
21.6 ~ 28.8V
1.0%
¡Ó
Ó¡ 0.2%
¡Ó0.5%
28.8 ~ 39.6V
1.0%
¡Ó
Ó¡ 0.2%
¡Ó0.5%
40.8 ~ 55.2V
1.0%
88%
89%
89%
2.8 ~ 3.8V
VOLTAGE TOLERANCE Note.3 2.0%
LOAD REGULATION
¡Ó
Ó¡ 0.5%
¡Ó1.5%
¡Ó
Ó¡ 0.5%
¡Ó1.0%
SETUP, RISE TIME
1000ms, 50ms/230VAC
HOLD UP TIME (Typ.)
16ms/230VAC
¡Ó
Ó¡ 0.3%
¡Ó0.5%
¡Ó
Ó¡ 0.3%
¡Ó0.5%
¡Ó
Ó¡ 0.2%
¡Ó0.5%
2500ms, 50ms/115VAC at full load
120 ~ 370VDC
POWER FACTOR (Typ.)
PF>0.95/230VAC
EFFICIENCY (Typ.)
80%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
2.2A/115VAC
35A/115VAC
LEAKAGE CURRENT
¡Ó
Ó¡ 0.5%
¡Ó1.0%
16ms/115VAC at full load
VOLTAGE RANGE
Note.5 85 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
PF>0.99/115VAC at full load
84%
86%
88%
88%
1.1A/230VAC
70A/230VAC
£g
£g
Note.7 Earth leakage current < 300 A/264VAC , Touch leakage current < 100 A/264VAC
OVERLOAD
PROTECTION
MSP-200-15
VOLTAGE ADJ. RANGE
LINE REGULATION
INPUT
MSP-200-12
OVER VOLTAGE
105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
6 ~ 7V
9.4 ~ 10.9V
14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V
41.4 ~ 48.6V
3.96 ~ 4.62V
57.6 ~ 67.2V
Protection type : Shut down o/p voltage, re-power on to recover
5V STANDBY
¢J ¡Ó5¢J(TSW1) detect on heatsink of power transistor
¢J ¡Ó5¢J (TSW2) detect on main power output choke
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
5VSB : [email protected] ; tolerance ¡Ó 5%, ripple : 50mVp-p(max.)
REMOTE CONTROL
RC+/RC- : 4 ~10V or open = power on ; 0 ~ 0.8V or short = power off
WORKING TEMP.
-40 ~ +70
95
OVER TEMPERATURE
FUNCTION
105
¢J (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85 , 10 ~ 95% RH
TEMP. COEFFICIENT
0.03%/ (0 ~ 50
VIBRATION
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
¡Ó
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note 4)
OTHERS
¢J)
ANSI/AAMI ES60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
¢J
EMC EMISSION
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2
MTBF
209.4K hrs min.
DIMENSION
199*98*38mm (L*W*H)
0.77Kg; 18pcs/14.9Kg/0.9CUFT
PACKING
NOTE
¢J
¢J
¢J
MIL-HDBK-217F (25 )
¢J
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. No load power consumption<0.5W when RC+ & RC- (CN100 pin1,2) 0 ~ 8V or short.
7. Touch current was measured from primary input to DC output.
File Name:MSP-200-SPEC 2012-06-29
MSP-200
200W Single Output Medical Type
Mechanical Specification
Case No.902E
series
Unit:mm
3.5
9
7
197
1
9.5
2
8.2
80
5
Pin No.
Assignment
4,5
DC OUTPUT -V
DC OUTPUT +V
6,7
£r3.
7
LED
CN100
SVR1
5
4-
M
3
57.5
4.5
120
199
6.5
9.5
.5
18
18.5
Terminal
£r3
Mating Housing
Pin No. Assignment
RC+
1
RC2
AUX
3
AUXG
4
+S
5
-S
6
190
9
Connector Pin No. Assignment (CN100) :
HRS DF11-6DP-2DS or equivalent
13max.
150.75
M3
28
3-
HRS DF11-6DS HRS DF11-**SC
or equivalent
or equivalent
Block Diagram
fosc :70KHz
ACTIVE
INRUSH
CURRENT
LIMITING
EMI
FILTER
I/P
38
Assignment
AC/L
AC/N
FG
6
3.5
28.5
Pin No.
1
2
3
1 5
2 6
Terminal Pin No. Assignment
85.5
4
98
3
AUX
POWER
RECTIFIERS
&
FILTER
RECTIFIERS
&
PFC
POWER
SWITCHING
RECTIFIERS
&
FILTER
O.T.P.
O.L.P.
Derating Curve
+S
+V
-V
-S
DETECTION
CIRCUIT
FG
PFC
CONTROL
AUX
AUXG
O.V.P.
PWM
CONTROL
RC+
REMOTE
CONTROL
RC-
Output Derating VS Input Voltage
100
125
100
90
80
80
70
LOAD(%)
LOAD (%)
60
40
20
-40
0
10
20
30
40
50
¢J
AMBIENT TEMPERATURE ( )
60
70 (HORIZONTAL)
60
50
40
85
100
125
135
155
264
INPUT VOLTAGE (V) 60Hz
File Name: MSP -200-SPEC 2012-06-29
MSP-200
200W Single Output Medical Type
series
Function Description of CN100
Pin No. Function Description
1
RC+ Turns the output on and off by electrical or dry contact between pin 2 (RC-). Short: Power OFF, Open: Power ON.
RC- Remote control ground.
2
Auxiliary voltage output, 4.75~5.25V, reference to pin 4(AUXG). The maximum load current is 0.3A. This output has the built-in oring
3
AUX diodes and is not controlled by the "remote ON/OFF control".
4
5
AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V).
Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to
+S
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
6
-S
Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
Function Manual
1.Remote Control
The PSU can be turned ON/OFF by using the "Remote
ON/OFF" function
Between RC-(pin2) and RC+(pin1)
Output Status
SW ON (Short)
OFF
SW OFF (Open)
ON
CN100
AUX
1
RC+
2
RC- AUXG
+S
5
-S
6
+S
5
-S
6
SW
-V
-V +V
+V
CN100
135
246
LOAD
-
+
Fig 1.1
2.Remote Sense
The remote sensing compensates voltage drop on the
load wiring up to 0.5V.
-V
-V +V
+V
CN100
135
246
-S
(Pin6)
LOAD
-
+
+S
(Pin5)
Sense lines should
be twisted in pairs
CN100
AUX
1
RC+
2
RC- AUXG
Fig 2.1
File Name: MSP -200-SPEC 2012-06-29