MSP-600 series - Total Power International

Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-600
600W Single Output Medical Type
series
■ Features :
•Universal AC input / Full range
•Built-in active PFC function, PF>0.94
•High efficiency up to 89%
•Withstand 300VAC surge input for 5 seconds
•Protections: Short circuit / Overload / Over voltage / Over temperature
•Medical safety approved (MOOP level)
•Built-in cooling fan ON-OFF control
•Built-in DC OK signal
•Built-in remote ON-OFF control
•Standby [email protected]
•Built-in remote sense function
•No load power consumption<0.8W (Note.7)
•Current sharing up to 2400W (3+1) (24V,36V,48V)
•5 years warranty
SPECIFICATION
MODEL
OUTPUT
MSP-600-3.3
MSP-600-5
MSP-600-7.5
MSP-600-12
MSP-600-15
MSP-600-24
MSP-600-36
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
36V
48V
RATED CURRENT
120A
120A
80A
53A
43A
27A
17.5A
13A
CURRENT RANGE
0 ~ 120A
0 ~ 120A
0 ~ 80A
0 ~ 53A
0 ~ 43A
0 ~ 27A
0 ~ 17.5A
0 ~ 13A
RATED POWER
396W
600W
600W
636W
645W
648W
630W
624W
RIPPLE & NOISE (max.) Note.2 100mVp-p
100mVp-p
100mVp-p
120mVp-p
150mVp-p
150mVp-p
200mVp-p
240mVp-p
VOLTAGE ADJ. RANGE
4.3 ~ 5.8V
±2.0%
6.8 ~ 9V
±2.0%
10.2 ~ 13.8V
±1.0%
13.5 ~ 18V
±1.0%
21.6 ~ 28.8V
±1.0%
28.8 ~ 39.6V
±1.0%
40.8 ~ 55.2V
±1.0%
2.8 ~ 3.8V
VOLTAGE TOLERANCE Note.3 ±2.0%
LINE REGULATION
±0.5%
±0.5%
±0.5%
±0.3%
±0.3%
±0.2%
±0.2%
±0.2%
LOAD REGULATION
±1.0%
±1.0%
±1.0%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
SETUP, RISE TIME
1000ms, 50ms/230VAC
HOLD UP TIME (Typ.)
16ms/230VAC
88%
89%
89%
41.4 ~ 48.6V
57.6 ~ 67.2V
PF>0.94/230VAC
EFFICIENCY (Typ.)
78.5%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
8.5A/115VAC
35A/115VAC
OVERLOAD
PROTECTION
FUNCTION
120 ~ 370VDC
POWER FACTOR (Typ.)
LEAKAGE CURRENT
OVER VOLTAGE
2500ms, 50ms/115VAC at full load
16ms/115VAC at full load
Note.5 85 ~ 264VAC
VOLTAGE RANGE
FREQUENCY RANGE
47 ~ 63Hz
INPUT
MSP-600-48
PF>0.99/115VAC at full load
82%
86%
88%
88%
5A/230VAC
80A/230VAC
Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC
105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
6 ~ 7V
9.4 ~ 10.9V
14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V
3.96 ~ 4.62V
OVER TEMPERATURE
5V STANDBY
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, recovers automatically after temperature goes down
5VSB : [email protected] ; tolerance±5%, ripple : 50mVp-p(max.)
DC OK SIGNAL
PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V
REMOTE CONTROL
RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off
FAN CONTROL (Typ.)
Load 35±15% or RTH2≧50℃ Fan on
WORKING TEMP.
-40 ~ +70℃ (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note 4)
EMC EMISSION
OTHERS
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2
MTBF
138.7K hrs min.
DIMENSION
218*105*63.5mm (L*W*H)
1.57Kg;8pcs/13.6Kg/1.34CUFT
PACKING
NOTE
ANSI/AAMI ES60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
MIL-HDBK-217F (25℃)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. No load power consumption<0.8W when RC+ & RC- (CN100 pin3,4) 0 ~ 0.8V or short.
8. When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this
deviation that does not affect basic safety or essential performance.
File Name:MSP-600-SPEC 2013-12-30
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-600
600W Single Output Medical Type
Mechanical Specification
Case No. 977A
O/P
2
105
4
5
9
2
10
CN100
218
152.4
32.8
8-M4 L=4 (Both Sides)
12.5
63.5
18max.
38.3
12max.
1
LED
6
1
2
3
(Vo ADJ.)SVR1
8.2
9.2
63.5
3
I/P
10
11
1
4-M4 L=6
Air flow
direction
Unit:mm
32.8
20.75
152.4
series
AC Input Terminal Pin No.
Assignment
Pin No.
1
2
3
Connector Pin No. Assignment(CN100) : HRS DF11-10DP-2DS or equivalent
DC Output Terminal Pin No.
Assignment
Assignment
AC/L
AC/N
FG
Pin No.
1~3
4~6
Pin No.
1
2
3
4
5
Assignment
-V
+V
Assignment
AUXG
AUX
RC+
RCCS
Pin No.
6,8
7
9
10
Terminal
Assignment Mating Housing
GND
DC-OK
HRS DF11-10DS HRS DF11-**SC
+S
or equivalent
or equivalent
-S
Block Diagram
PWM fosc : 70KHz
ACTIVE
INRUSH
CURRENT
LIMITING
EMI
FILTER
I/P
RECTIFIERS
&
PFC
AUX
POWER
RECTIFIERS
&
FILTER
AUX
AUXG
+S
POWER
SWITCHING
RECTIFIERS
&
FILTER
+V
-V
-S
DETECTION
CIRCUIT
O.L.P.
O.T.P.
FG
DC OK
O.T.P.
PFC
CONTROL
PWM
CONTROL
Derating Curve
CURRENT
SHARING
O.V.P.
CS
RC+
REMOTE
CONTROL
RC-
Output Derating VS Input Voltage
100
125
90
100
80
70
LOAD(%)
LOAD (%)
80
50
20
-40
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
60
70
(HORIZONTAL)
60
50
40
85
100
125
135
155
264
INPUT VOLTAGE (V) 60Hz
File Name:MSP-600-SPEC 2013-12-30
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-600
600W Single Output Medical Type
series
Function Description of CN100
Pin No. Function Description
AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V).
1
Auxiliary voltage output, 4.75~5.25V, referenced to pin 1(AUXG). The maximum load current is 0.3A. This output has the built-in oring
2
AUX diodes and is not controlled by the "remote ON/OFF control".
3
4
RC+
RC-
5
CS
6,8
7
9
10
Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power OFF, Open: Power ON.
Remote control ground.
Current sharing signal. When units are connected in parallel, the CS pins of the units should be connected to allow current balance
between units.
GND This pin connects to the negative terminal(-V). Return for DC-OK signal output.
DC-OK DC-OK signal is a TTL level signal, referenced to pin8(DC-OK GND). High when PSU turns on.
Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to
+S
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
-S
Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
Function Manual
1.Remote Sense
The remote sensing compensates voltage drop on
the load wiring up to 0.5V.
CN100
CN100
2
1
10
9
2
AUX AUXG 1
RC- RC+
GND
CS
GND DC-OK
10
+S
(Pin9)
-
+
LOAD
-S
+S
9
-S
(Pin10)
Sense lines should
be twisted in pairs
Fig 1.1
2.DC-OK Signal
DC-OK signal is a TTL level signal. High when PSU turns on.
CN100
Between DC-OK(pin7) and GND(pin6,8) Output Status
CN100
3.3 ~ 5.6V
ON
2
1
0 ~ 1V
OFF
10
9
2
AUX AUXG 1
RC- RC+
GND
CS
GND DC-OK
10
-S
+S
9
Fig 2.1
File Name:MSP-600-SPEC 2013-12-30
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-600
600W Single Output Medical Type
3.Remote Control
The PSU can be turned ON/OFF by using the
"Remote Control" function.
Between RC+(pin3) and RC-(pin4)
CN100
CN100
Output Status
SW ON (Short)
OFF
SW OFF (Open)
ON
series
2
1
10
9
2
AUX AUXG 1
RC- RC+
GND
CS
GND DC-OK
10
-S
+S
9
SW
Fig 3.1
4.Current Sharing with Remote Sensing (Only for 24V, 36V and 48V)
MSP-600 has the built-in active current sharing function and can be connected in parallel to provide higher output power :
(1)Parallel operation is available by connecting the units shown as below.
(+S,-S,CS and GND are connected mutually in parallel).
(2)Difference of output voltages among parallel units should be less than 2%.
(3)The total output current must not exceed the value determined by the following equation.
(output current at parallel operation)=(Rated current per unit)×(Number of unit)×0.9
(4)In parallel operation 4 units is the maximum, please consult the manufacturer for applications of more connecting in parallel.
(5)The power supplies should be paralleled using short and large diameter wiring and then connected to the load.
PSU1
PSU2
PSU3
-V
PSU4
-V
+V
-V
+V
CN100
-V
+V
CN100
+V
CN100
CN100
PSU1
PSU2
PSU3
PSU4
CN100
CN100
CN100
CN100
1 AUXG RC+
CS DC-OK +S
2 AUX RC- GND GND
-S
9
1 AUXG RC+
CS DC-OK +S
10
2 AUX RC- GND GND
-S
9
1 AUXG RC+
10
2 AUX RC- GND GND
+V
-V
CS DC-OK +S
-S
9
1 AUXG RC+
CS DC-OK +S
10
2 AUX RC- GND GND
-S
9
10
Sense lines should
be twisted in pairs
LOAD
Fig 4.1
Note : 1.In parallel connection, maybe only one unit (master) operate if the total output load is less than 2% of rated load condition.
The other PSU (slave) may go into standby mode and its output LED and relay will not turn on.
2.2% min. of dummy load is required.
File Name:MSP-600-SPEC 2013-12-30