Get new life out of your 33120A

Agilent Technologies 33120A Option 001 External Clock Reference
Can my 33120A be retrofit with the Option 001 External Clock Reference?
All instruments shipped after August, 1994 may be updated. To be certain, find the serial number
on the rear panel of the unit. If it is less than serial number US34004078, your unit is not able to
accept the retrofit.
What do I order?
Ask for the 33120-80001 timebase retrofit kit.
Can I install the retrofit myself?
The actual installation can be done by most people, however the 33120A requires a frequency
adjustment to complete the procedure. The 1 kHz frequency measurement should be done with
an accuracy of at least 8 digits (0.01 ppm). Retrofit instructions
What are the key specifications?
The option will allow the 33120A timebase to lock to a 10 MHz signal (±50 Hz) with a level range
of –10 dBm to +15 dBm into 50Ω. For more complete specifications, see the 33120A data sheet.
How do I use the option?
The retrofit for the option comes with a User’s and Service Guide. Here are some steps to help
get you started.
•
To phase lock to an external clock signal
Attach the 10 MHz clock signal to the rear-panel Ext Ref In BNC terminal. The 33120A’s
internal 40 MHz timebase is now synchronized to the 10 MHz clock signal and will not drift.
•
To phase lock multiple function generators
Attach a BNC from the 10 MHz clock source of one generator (labled Ref Out on the 33120A)
to the to the reference input on the second (labled Ext Ref In on the 33120A). These are now
your Master and Slave units, respectively. The units are synchronized and will not drift,
however the initial phase offset relationship is random.
•
To adjust phase
Real-time phase adjustments can be made on sine, square, ramp, or triangle waveforms
only. Adjustments made on arbitrary waveforms will only be implemented when the value is
entered.
1. Turn on the menu by pressing Shift and Menu On/Off
2. Press the right or left arrow keys to move to G: PHASE MENU
3. Press the down arrow key to move down a level.
4. Press the right or left arrow keys to move to 1: ADJUST
5. Press the down arrow key to move down a level to adjust the phase offset.
6. Using the arrow keys and/or the knob or the numeric entry method, adjust the phase to
any value between –360 degrees and +360 degrees.
7. Press Enter to turn off the menu and save your changes.
•
To begin the 33120A and another unit (either a 33120A or 33250A) simultaneously
1. Connect the Ext Trig between the two units.
2. Setup both units with the desired waveform, frequency, amplitude, and offset.
3. Enable the burst mode on both instruments. On the 33120A, press Shift and then Burst
4. Set the burst count to the desired number of cycles or infinite to output the desired
waveform until the waveform key is pressed. On the 33120A Master unit, Press Shift and
then Recall Menu to be brought immediately to the Burst Count (4: BURST CNT)
command in the modulation menu. Press the down arrow key and set the number of
cycles, or press the right arrow key until the CYC units are flashing. Then, press the
down arrow key to display INFINITE. Press Enter to save the change and turn off the
menu. Repeat on the slave unit.
5. Set the burst phase on each waveform. On the 33120A Master unit, press Shift and then
Recall Menu to be brought to the Burst Count command again (4: BURST CNT). Press
the right arrow to reach the Burst Phase command (6: BURST PHAS). Press the down
arrow then adjust the phase to any value between –360 degrees and +360 degrees using
the arrow keys and/or the knob or the numeric entry method. Press the Enter key to save
the change and exit the menu. Repeat on the slave unit.
6. Set up the trigger to output on the Master unit. On the 33120A, turn on the menu by
pressing Shift and Menu On/Off. Press the right or left arrow keys to move to G: PHASE
MENU. Press the down arrow key to move down a level. Press the right or left arrow
keys to move to 3: TRIG OUT. Move down a level by pressing the down arrow, then
enable the trigger output by pressing the right or left arrow keys until the display reads
ENABLE. Press Enter to save the change and exit the menu.
7. Execute a single trigger on the Master unit to initiate the triggered burst. On the 33120A,
press the Single trigger key on the front panel of the unit. The 33120A will trigger itself
and also output a trigger pulse from the rear-panel Ext Trig terminal. At this point, the
function generators should be synchronized with the specified phase relationship.
•
The 33120A can be configured to generate an error condition whenever the clock reference
is lost. To be notified when the unit is unlocked
1. Turn on the menu by pressing Shift and Menu On/Off
2. Press the right or left arrow keys to move to G: PHASE MENU
3. Press the down arrow key to move down a level.
4. Press the right or left arrow keys to move to 4: UNLOCK ERR
5. Press the down arrow key to move down a level and then press the right or left arrow
keys until the display reads ENABLE
6. Press Enter to turn off the menu and save your changes.