OPTICAL UV POWER METER POWER C9536 / H9535 SERIES MONITORING For absolute optical power measurement Left: Controller C9536 Right: Sensor Head H9535 Series OVERVIEW The C9536/H9535 series are UV power meters that directly measure UV light at specific wavelengths. The H9535 (sensor head) uses a UV sensor (phototube) that ensures stable UV measurement over a long period of operation and is calibrated in combination with the C9536 (controller) based on national standards. The detachable sensor head contains a rechargeable battery for UV detection and data memory, allowing remote measurement such as on a belt conveyor or in locations where cable routing is difficult. In addition to external control via the RS-232C interface, an analog control is available for recording on a data recorder, etc. FEATURES ●Absolute optical power monitoring (mW/cm2) Sensitivity-calibrated data at each wavelength (126 nm, 146 nm, 172 nm, 222 nm, 185 nm, 254 nm) is supplied with the power meter. We also append a Certificate of Calibration. ●Detachable sensor head / remote measurement The detachable sensor head allows remote measurement on a conveyer belt of UV oven, etc. ●Thin, flat sensor head: 14 mm thickness Allows the sensor head to be used closer to a light source. ●External control: RS-232C The main unit has RS-232C interface for external control. ●Stable performance at deep UV since no optical filters are used Specific UV light can be directly detected without any optical filter. ●Wide light incident angle Acceptable light incident angle is close to cos θ, even without using diffusers. APPLICATIONS ●Excimer lamp monitoring ●Low pressure mercury lamp monitoring ●UV light monitor • UV furnace (coating, adhesive, printing) • UV cleaning, UV sterilization • Improving surface characteristics SPECIFICATIONS Controller: C9536 Display Function Integration Time Setting Analog Output External Interface Power Requirement Operating Ambient Temperature / Humidity Storage Temperature / Humidity Weight Liquid crystal digital display [irradiance (mW/cm2) / integrated optical power (mJ/cm2): Max.effective 4 digits] Real time / Integration / Peak hold / Temperature 1 s to 5 min (1 s step) 0 V to +2 V RS-232C Rechargeable battery (built-in) / AC adapter (AC100 V to 240 V) (supplied) 0 °C to +40 °C / Below 80 % (no condensation) -20 °C to +50 °C / Below 80 % (no condensation) Approx. 285 g Sensor Head : H9535 Series Type H9535-126 Calibrated Wavelength 126 Applicable light Source Sensitive Area Optical Power Measurement Range Operating Ambient Temperature / Humidity Storage Temperature / Humidity Weight H9535-146 H9535-172 146 172 Excimer lamp H9535-222 222 H9535-185 H9535-254 185 254 Low-pressure mercury lamp 6 Irradiance [0.001 mW/cm2 to 200 mW/cm2] 0 °C to +60 °C / Below 80 % (no condensation) -20 °C to +60 °C / Below 80 % (no condensation) Sensor Head: Approx. 60 g / Cable: Approx. 110 g C9536 and H9535 series are not available to vacuum operation. Unit nm — mm — — — — SENSOR CHARACTERISTICS H9535-172, -185 Spectral Response H9535-126, -146 Spectral Response 1.2 TPT B0018EA 1.2 1.0 RELATIVE SENSITIVITY RELATIVE SENSITIVITY 1.0 TPT B0015EB 0.8 0.6 0.4 0.2 0.8 0.6 0.4 0.2 0.0 100 150 200 250 300 350 0 100 400 150 WAVELENGTH (nm) 1.2 RELATIVE SENSITIVITY RELATIVE SENSITIVITY 350 400 350 400 TPT B0016EB 1.0 0.8 0.6 0.4 0.8 0.6 0.4 0.2 0.2 0 100 150 200 250 300 350 0 100 400 Angular Response TPT B0017EB 1.0 0.8 cos θ 0.6 0.4 0.2 0 -90 -60 -30 0 150 200 250 300 WAVELENGTH (nm) WAVELENGTH (nm) RELATIVE SENSITIVITY 300 H9535-254 Spectral Response TPT B0019EA 1.0 1.2 250 WAVELENGTH (nm) H9535-222 Spectral Response 1.2 200 30 ANGLE (DEGREE) 60 90 DIMENSIONAL OUTLINES (Unit: mm) 2000 SENSOR HEAD CABLE (SUPPLIED WITH H9535 SERIES) LCD 65 MADE IN JAPAN CONTRAST ADJUSTER 3 25 10 14 45 CONTRAST 110 UV POWER METER 15 S/N HAMAMATSU PHOTONICS K.K. 58 UV POWER METER HEAD MODEL 4-M3 DEPTH 8 BATT SELECT SENSITIVE AREA ENTER START STOP POWER BATTERY CHARGE LAMP 6 [SENSOR HEAD H9535 SERIES] 75 20 20 ANALOG OUTPUT CONNECTOR RS232C RS-232C CONNECTOR ANALOG OUT DC IN DC INLET [CONTROLLER C9536] [RS-232C CABLE] (SUPPLIED WITH C9536) [ANALOG OUTPUT CABLE] (SUPPLIED WITH C9536) 2000 2000 [AC ADAPTER] (SUPPLIED WITH C9536) 68.5 50 27 1800 TPT A0032ED Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2010 Hamamatsu Photonics K.K. HAMAMATSU PHOTONICS K.K. www.hamamatsu.com HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected] Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected] France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected] United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected] North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected] TPT 1017E05 Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected] DEC. 2010 IP