ECHOTEL® 940/941 Installation and Operating Manual Ultrasonic Level Switches ® UNPACKING Unpack the instrument carefully. Make sure all components have been removed from the foam protection. Inspect all components for damage. Report any concealed damage to the carrier within 24 hours. Check the contents of the carton/crates against the packing slip and report any discrepancies to Magnetrol. Check the nameplate model number to be sure it agrees with the packing slip and purchase order. Check and record the serial number for future reference when ordering parts. 0038 These units are in compliance with: 1. The EMC directive 2004/108/EC. The units have been tested to EN 61326: 1997 + A1 + A2 2. The PED directive 97/23/EC (pressure equipment directive). Safety accessories per category IV module H1. Model and serial number High alarm ▼ Filling line Nozzle mounting: gap must be free min. 25 mm ▼ MOUNTING Low alarm High alarm Front view of gap Low alarm Pump protection Pump Gap indication 2 Threaded Welded flange Tri-Clamp® Wrench flats are aligned with sensor gap WIRING – RELAY VERSION: ECHOTEL 940 green black red brown (NC) + orange (COM) violet (NO) OUTPUT Level Relay De-energized Energized Wiring NO (Violet) COM (Orange) NC (Brown) NO (Violet) COM (Orange) NC (Brown) IMPORTANT: When power is lost or gap is dry; the relay de-energizes. 3 WIRING – CURRENT SHIFT VERSION: ECHOTEL 941 green black red OUTPUT Level + Current 8 mA (± 1 mA) 16 mA (± 1 mA) mA Wet gap ▼ 4 Normal operation ▼ Dry gap SPECIFICATIONS PERFORMANCE SPECIFICATIONS Description Response time Repeatability Ambient Temperature Humidity Specification 0,5 s typical ± 2 mm (0.078") -40 °C up to +85 °C (-40 °F up to +185 °F) 0-99 %, non condensing Meets CE requirements (EN-61326: 1997 + A1 + A2) and in compliance with Directive 2004/108/EC Electromagnetic Compatibility PHYSICAL/FUNCTIONAL SPECIFICATIONS Description Input voltage Power consumption Current Shift Signal output Relay Potted cable Net Weight Specification 12 - 35 V DC <1W 8 mA (± 1 mA) - dry gap or 16 mA (± 1 mA) - wet gap SPDT: 1 A @ 30 V DC 45 cm (18") flying leads of 18 AWG wires 510 g (unit with 3/4" NPT transducer and shortest length) TRANSDUCER SPECIFICATIONS Description Material Mounting Actuation length Operating process temp. Specification 316/316L SST (1.4401/1.4404) Threaded (NPT/BSP) – Flanged (ANSI) – Hygienic (Tri-Clamp®) From 3 cm up to 250 cm (1.2" up to 98.4") -40 °C up to +85 °C (-40 °F up to +185 °F) 103 bar @ +85 °C (1500 psi @ +185 °F) - transducers > 3 cm 138 bar @ +85 °C (2000 psi @ +185 °F) - transducers = 3 cm Flanged models are downrated to the design pressure of the selected flange Max pressure TEMPERATURE/PRESSURE RATING bar (psi) Process pressure 200 (2900) 150 (2175) 100 (1450) 138 bar (transducer = 3 cm) 103 bar (transducer > 3 cm) 50 (725) -40 (-40) 0 (32) 40 (100) 80 (176) 120 (250) Process temperature 150 C (300) (F) 5 MODEL IDENTIFICATION A complete measuring system consists of: 1. One code for electronics and transducer 2. OPTION: Standard cable length is 45 cm (18"). Specify cable length separately when a longer cable is required. 3. OPTION: 052-7012-002: code for Echotel 941 cast aluminium housing with terminal block, 3/4" NPT cable entry, IP 65 052-7012-003: code for Echotel 940 cast aluminium housing with terminal block, 3/4" NPT cable entry, IP 65 BASIC MODEL NUMBER 9 4 0 – N 9 4 1 – N Echotel® 940 OEM switch with integrated 1 A SPDT relay Echotel® 941 OEM switch with current shift output 8/16 mA PROCESS CONNECTION Threaded 1 1 M 2 2 M 3/4" NPT 1" BSP (G 1") 2 3 M 3 3 M 4 3 M 1" 150 lbs ANSI RF 1 1/2" 150 lbs ANSI RF 2" 150 lbs ANSI RF 3 T M 4 T M 1 1/2" 2" ANSI Flanges Hygienic Tri-Clamp® Tri-Clamp® ACTUATION LENGTH – Specify per cm (0.39") increment Total insertion length = actuation length + 6 mm (0.25") 0 0 3 0 0 6 2 5 0 9 4 6 N M Minimum 3 cm (1.2") – for all connections except 1" BSP (G 1") Minimum 6 cm (2.4") – for 1" BSP (G 1") Maximum 250 cm (98.4") complete code for Echotel 940/941 X = product with a specific customer requirement DIMENSIONS IN mm (inches) 3/4" NPT 84 (3.31) 84 1) .3 (3 3/4" NPT 105 (4.13) 65 (2.56) 89 (3.50) 3/4" NPT cable entry Cast aluminium housing 6 (0.25) 6 (0.25) Ø 22,9 (0.90) Ø 22,9 (0.90) Model 940/941 with 1" BSP Model 940/941 with 3/4" NPT 3/4" NPT 3/4" NPT 6 (0.25) Actuation length Actuation length Ø 22,9 (0.90) 99 (3.9) 99 (3.9) Actuation length Actuation length Model 940/941 with ANSI flange 6 (0.25) Ø 22,9 (0.90) Model 940/941 with Tri-Clamp® 7 IMPORTANT SERVICE POLICY Owners of Magnetrol products may request the return of a control; or, any part of a control for complete rebuilding or replacement. They will be rebuilt or replaced promptly. Magnetrol International will repair or replace the control, at no cost to the purchaser, (or owner) other than transportation cost if: a. Returned within the warranty period; and, b. The factory inspection finds the cause of the malfunction to be defective material or workmanship. If the trouble is the result of conditions beyond our control; or, is NOT covered by the warranty, there will be charges for labour and the parts required to rebuild or replace the equipment. In some cases, it may be expedient to ship replacement parts; or, in extreme cases a complete new control, to replace the original equipment before it is returned. If this is desired, notify the factory of both the model and serial numbers of the control to be replaced. In such cases, credit for the materials returned, will be determined on the basis of the applicability of our warranty. No claims for misapplication, labour, direct or consequential damage will be allowed. RETURNED MATERIAL PROCEDURE So that we may efficiently process any materials that are returned, it is essential that a “Return Material Authorisation” (RMA) form will be obtained from the factory. It is mandatory that this form will be attached to each material returned. This form is available through Magnetrolʼs local representative or by contacting the factory. Please supply the following information: 1. Purchaser Name 2. Description of Material 3. Serial Number and Ref Number 4. Desired Action 5. Reason for Return 6. Process details Any unit that was used in a process must be properly cleaned in accordance with the proper health and safety standards applicable by the owner, before it is returned to the factory. A material Safety Data Sheet (MSDS) must be attached at the outside of the transport crate or box. All shipments returned to the factory must be by prepaid transportation. Magnetrol will not accept collect shipments. All replacements will be shipped Ex Works. BULLETIN N°: EFFECTIVE: SUPERSEDES: UNDER RESERVE OF MODIFICATIONS BE 51-647.4 MAY 2012 January 2011 www.magnetrol.com BENELUX FRANCE Heikensstraat 6, 9240 Zele, België -Belgique Tel. +32 (0)52.45.11.11 • Fax. +32 (0)52.45.09.93 • E-Mail: [email protected] DEUTSCHLAND Alte Ziegelei 2-4, D-51491 Overath Tel. +49 (0)2204 / 9536-0 • Fax. +49 (0)2204 / 9536-53 • E-Mail: [email protected] INDIA C-20 Community Centre, Janakpuri, New Delhi - 110 058 Tel. +91 (11) 41661840 • Fax +91 (11) 41661843 • E-Mail: [email protected] ITALIA Via Arese 12, I-20159 Milano Tel. +39 02 607.22.98 • Fax. +39 02 668.66.52 • E-Mail: [email protected] U.A.E. DAFZA Office 5EA 722 • PO Box 293671 • Dubai Tel. +971-4-6091735 • Fax +971-4-6091736 • E-Mail: [email protected] UNITED KINGDOM Unit 1 Regent Business Centre, Jubilee Road Burgess Hill West Sussex RH 15 9TL Tel. +44 (0)1444 871313 • Fax +44 (0)1444 871317 • E-Mail: [email protected]