NI PXI-5122 Specifications

NI PXI-5122 Specifications
14-Bit 100 MS/s Digitizer
These specifications are valid over the full operating temperature range,
for all filter settings, and for all impedance selections with a sample rate
of 100 MS/s, unless otherwise noted.
Specifications are subject to change without notice. For the most current product
specifications, visit ni.com/manuals.
Note
Contents
Vertical.................................................................................................... 2
Analog Input (Channel 0 and Channel 1) ........................................ 2
Horizontal (Timing Characteristics) ....................................................... 11
Sample Clock ................................................................................... 11
Phase-Locked Loop (PLL) Reference Clock................................... 13
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector) ................................................................. 13
CLK OUT (Sample Clock and Reference Clock Output,
Front Panel Connector) ................................................................. 14
Trigger..................................................................................................... 14
Reference (Stop) Trigger ................................................................. 14
TRIG (External Trigger, Front Panel Connector)............................ 16
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors) ............................................................. 17
Waveform Specifications........................................................................ 18
Calibration............................................................................................... 19
Power ...................................................................................................... 19
Software .................................................................................................. 20
Environment............................................................................................ 21
Safety, Electromagnetic Compatibility, and CE Compliance................. 22
Physical ................................................................................................... 23
Vertical
Analog Input (Channel 0 and Channel 1)
Table 1.
Specification
Value
Comments
Number of
Channels
Two (simultaneously sampled)
—
Connector
BNC
—
Impedance and Coupling
Input
Impedance
50 Ω ±2.0%
Input Coupling
AC, DC, GND
1 MΩ ±0.75% in parallel
with a typical capacitance
of 27 pF ±2 pF
Software
selectable
AC coupling
available on
1 MΩ only
Voltage Levels
Full Scale (FS)
Input Range and
Programmable
Vertical Offset
Maximum Input
Overload
NI PXI-5122 Specifications
50 Ω
1 MΩ
—
Range
(Vpk-pk)
Vertical Offset
Range (V)
Range
(Vpk-pk)
Vertical Offset
Range (V)
0.2
±0.1
0.2
±0.1
0.4
±0.2
0.4
±0.2
1
±0.5
1
±0.5
2
±1
2
±1
4
±2
4
±2
10
—
10
±5
20
—
50 Ω
1 MΩ
7 Vrms with |Peaks| ≤ 10 V
|Peaks| ≤ 42 V
2
—
ni.com
Table 1. (Continued)
Specification
Value
Comments
Accuracy
Resolution
DC Accuracy
(Programmable
Vertical
Offset = 0 V)
14 bits
—
Range
(Vpk-pk)
50 Ω and 1 MΩ
0.2, 0.4, 1, 2
±(0.65% of Input + 1.0 mV)
4, 10
±(0.65% of Input + 8.0 mV)
20
(1 MΩ only)
±(0.65% of Input + 10.0 mV)
Programmable
Vertical Offset
Accuracy
±0.4% of offset setting
DC Drift
±(0.057% of Input + 0.006% of FS + 100 µV) per °C
AC Amplitude
Accuracy
Crosstalk,
Typical
Within ±5 °C of
self-calibration
temperature
Within ±5 °C of
self-calibration
temperature
50 Ω
1 MΩ
±0.06 dB
(±0.7%) at 50 kHz
±0.09 dB
(±1.0%) at 50 kHz
≤ –100 dB at 10 MHz
—
Within ±5 °C of
self-calibration
temperature
CH 0 to/from
CH 1, External
Trigger to CH 0
or CH 1
Bandwidth and Transient Response
Bandwidth
(–3 dB)
Rise/Fall Time,
Typical
Range
(Vpk-pk)
Filters off
50 Ω and 1 MΩ
All ranges
except 0.2
100 MHz
0.2
80 MHz up to 40 ºC*
Range
(Vpk-pk)
50 Ω and 1 MΩ
MHz above
40 ºC
—
All ranges
except 0.2
3.5 ns
0.2
4.2 ns
© National Instruments Corporation
* 78
3
NI PXI-5122 Specifications
Table 1. (Continued)
Specification
Value
Bandwidth
Limit Filters
AC-Coupling
Cutoff
(–3 dB)
Comments
Noise Filter
Antialias Filter
20 MHz
2-pole Bessel filter
40 MHz (–6 dB, typical)
35 MHz (–3 dB)
6-pole Chebyshev filter
12 Hz
Passband
Flatness
NI PXI-5122 Specifications
AC coupling
available on
1 MΩ only
Filter
Settings
Range
(Vpk-pk)
Filters Off
All
ranges
except
0.2
±0.4 dB DC to 20 MHz
±1 dB 20 MHz to 50 MHz
0.2
±0.4 dB DC to 20 MHz
±1 dB 20 MHz to 40 MHz
All
ranges
±1.2 dB DC to 16 MHz
±1.6 dB 16 MHz to 32 MHz
Antialias
Filter On
Only one filter
can be enabled
at any given
time. The
antialias filter
is enabled by
default.
50 Ω and 1 MΩ
4
Referenced
to 50 kHz
ni.com
Figure 1. NI 5122 Frequency Response (Typical)
Table 1. (Continued)
Specification
Value
Comments
Spectral Characteristics
Spurious Free
Dynamic Range
with Harmonics
(SFDR), Typical
Range
(Vpk-pk)
50 Ω
1 MΩ
0.2
75 dBc
70 dBc
0.4
75 dBc
70 dBc
1
75 dBc
70 dBc
2
75 dBc
70 dBc
4
65 dBc
70 dBc
10
65 dBc
60 dBc
20
(1 MΩ only)
N/A
60 dBc
© National Instruments Corporation
5
10 MHz,
–1 dBFS input
signal.
Includes the
2nd through
the 5th
harmonics.
Measured
from DC to
50 MHz.
NI PXI-5122 Specifications
Table 1. (Continued)
Specification
Total Harmonic
Distortion
(THD),
Typical
Intermodulation
Distortion,
Typical
Signal-to-Noise
Ratio (SNR),
Typical
NI PXI-5122 Specifications
Value
Comments
Range
(Vpk-pk)
50 Ω
1 MΩ
0.2
–75 dBc
–68 dBc
0.4
–75 dBc
–68 dBc
1
–75 dBc
–68 dBc
2
–73 dBc
–68 dBc
4
–63 dBc
–68 dBc
10
–63 dBc
–58 dBc
20
(1 MΩ only)
N/A
–58 dBc
Ranges up to 2 Vpk-pk on 50 Ω Input
1 MΩ
Range
(Vpk-pk)
Filters
Off
Antialias
Filter On
Filters
Off
Antialias
Filter On
0.2
60 dB
60 dB
56 dB
60 dB
0.4
62 dB
62 dB
61 dB
62 dB
1
62 dB
62 dB
62 dB
62 dB
2
62 dB
62 dB
62 dB
62 dB
4
—
—
61 dB
62 dB
6
Includes the
2nd through
the 5th
harmonics.
Two tones at
10.2 MHz and
11.2 MHz.
Each tone is
–7 dBFS.
–75 dBc
50 Ω
10 MHz,
–1 dBFS input
signal.
10 MHz,
–1 dBFS input
signal.
Excludes
harmonics.
Measured
from DC to
50 MHz.
ni.com
Table 1. (Continued)
Specification
Signal to Noise
and Distortion
(SINAD),
Typical
Value
Comments
50 Ω
1 MΩ
Range
(Vpk-pk)
Filters
Off
Antialias
Filter On
Filters
Off
Antialias
Filter On
0.2
60 dB
60 dB
56 dB
59 dB
0.4
62 dB
62 dB
60 dB
61 dB
1
62 dB
62 dB
61 dB
61 dB
2
62 dB
62 dB
61 dB
61 dB
4
—
—
60 dB
61 dB
10 MHz,
–1 dBFS input
signal.
Includes
harmonics.
Measured
from DC to
50 MHz.
Figure 2. NI 5122 Dynamic Performance, 50 Ω, 1 Vpk-pk Range (Typical)
© National Instruments Corporation
7
NI PXI-5122 Specifications
Table 1. (Continued)
Specification
RMS Noise
(Noise Filter On)
RMS Noise
(Antialias
Filter On)
NI PXI-5122 Specifications
Value
Comments
Range
(Vpk-pk)
50 Ω
1 MΩ
0.2
46 µVrms
(0.023% FS)
60 µVrms
(0.030% FS)
0.4
92 µVrms
(0.023% FS)
92 µVrms
(0.023% FS)
1
230 µVrms
(0.023% FS)
230 µVrms
(0.023% FS)
2
460 µVrms
(0.023% FS)
460 µVrms
(0.023% FS)
4
920 µVrms
(0.023% FS)
920 µVrms
(0.023% FS)
10
2.3 mVrms
(0.023% FS)
2.3 mVrms
(0.023% FS)
20
(1 MΩ only)
N/A
4.6 mVrms
(0.023% FS)
Range
(Vpk-pk)
50 Ω
1 MΩ
0.2
66 µVrms
(0.033% FS)
80 µVrms
(0.040% FS)
0.4
100 µVrms
(0.025% FS)
120 µVrms
(0.030% FS)
1
250 µVrms
(0.025% FS)
300 µVrms
(0.030% FS)
2
500 µVrms
(0.025% FS)
600 µVrms
(0.030% FS)
4
1 mVrms
(0.025% FS)
1.2 mVrms
(0.030% FS)
10
2.5 mVrms
(0.025% FS)
3 mVrms
(0.030% FS)
20
(1 MΩ only)
N/A
6 mVrms
(0.030% FS)
8
Inputs 50 Ω
terminated
Inputs 50 Ω
terminated
ni.com
Table 1. (Continued)
Specification
Value
RMS Noise
(Filters Off)
Comments
Inputs 50 Ω
terminated
Range
(Vpk-pk)
50 Ω
1 MΩ
0.2
66 µVrms
(0.033% FS)
110 µVrms
(0.055% FS)
0.4
100 µVrms
(0.025% FS)
160 µVrms
(0.040% FS)
1
250 µVrms
(0.025% FS)
300 µVrms
(0.030% FS)
2
500 µVrms
(0.025% FS)
600 µVrms
(0.030% FS)
4
1 mVrms
(0.025% FS)
1.6 mVrms
(0.040% FS)
10
2.5 mVrms
(0.025% FS)
3 mVrms
(0.030% FS)
20
(1 MΩ only)
N/A
6 mVrms
0.030% FS)
60 n
Noise Density (nV/root Hz)
55 n
50 n
45 n
40 n
35 n
30 n
25 n
20 n
15 n
10 n
5n
1n
1k
10 k
100 k
1M
10 M
24 M
Frequency (Hz)
Figure 3. Representation of NI 5122 Spectral Noise Density at 0.2 V Range,
Noise Filter Enabled, 1 MΩ Input Impedance
© National Instruments Corporation
9
NI PXI-5122 Specifications
Noise Density (dBm/Hz)
–145.000
–147.500
–150.000
–152.500
1.0 k
10.0 k
100.0 k
1.0 M
10.0 M
50.0 M
Frequency (Hz)
Figure 4. Representation of NI 5122 Spectral Noise Density at 0.2 V Range,
Full Bandwidth, 50 Ω Input Impedance (Does Not Include System Spurs.
All Spurs ≤135 dBm/Hz)
NI PXI-5122 Specifications
10
ni.com
Horizontal (Timing Characteristics)
Sample Clock
Table 2.
Specification
Sources
Value
1. Internal, Onboard Clock (internal VCXO)*
2. External, CLK IN (front panel
SMB connector)
3. External, PXI Star Trigger
(backplane connector)
Comments
* Internal
Sample Clock is
locked to the Reference
Clock or derived from the
onboard VCXO.
Onboard Clock (Internal VCXO)
Sample Rate
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
1.526 kS/s to
100 MS/s*
200 MS/s to 2 GS/s in
multiples of 100 MS/s
* Divide
by n decimation
used for all rates less than
100 MS/s. For more
information about Sample
Clock and decimation,
refer to the NI High-Speed
Digitizers Help.
Phase Noise
Density, Typical
< –100 dBc/Hz at 100 Hz
< –120 dBc/Hz at 1 kHz
< –130 dBc/Hz at 10 kHz
10 MHz input signal
Sample Clock
Jitter, Typical
≤1 ps rms (100 Hz to 100 kHz)
≤2 ps rms (100 Hz to 1 MHz)
Includes the effects of
the converter aperture
uncertainty and the clock
circuitry jitter. Excludes
trigger jitter.
Timebase
Frequency and
Accuracy
100 MHz ±25 ppm
When not phase-locked
to Reference Clock
Sample Clock
Delay Range
±1 Sample Clock period
—
Sample Clock
Delay Resolution
10 ps
—
© National Instruments Corporation
11
NI PXI-5122 Specifications
Table 2. (Continued)
Specification
Value
Comments
External Sample Clock
Sources
1. CLK IN (front panel SMB connector)
—
2. PXI Star Trigger (backplane connector)
Frequency Range
30 MHz to 105 MHz
Duty Cycle
Tolerance
45% to 55%
Divide by n decimation
available where
1 ≤ n ≤ 65,535.
For more information
about Sample Clock and
decimation, refer to the
NI High-Speed Digitizers
Help.
—
Sample Clock Exporting
Exported Sample
Clock
Destinations
NI PXI-5122 Specifications
Maximum
Frequency
Destination
CLK OUT (front panel
SMB connector)
105 MHz
PXI_TRIG <0..6>
(backplane connector)*
20 MHz
PFI <0..1> (front panel 9-pin
mini-circular DIN connector)*
25 MHz
12
* Decimated
Sample
Clock only
ni.com
Phase-Locked Loop (PLL) Reference Clock
Table 3.
Specification
Sources
Value
1. PXI_CLK10 (backplane connector)
Comments
—
2. CLK IN (front panel SMB connector)
Frequency
Range
1 MHz to 20 MHz in 1 MHz increments. Default of 10 MHz.
The PLL Reference Clock frequency has to be accurate
to ±50 ppm.
—
Duty Cycle
Tolerance
45% to 55%
—
Exported
Reference
Clock
Destinations
1. CLK OUT (front panel SMB connector)
—
2. PXI_TRIG <0..6> (backplane connector)
3. PFI <0..1> (front panel 9-pin mini-circular
DIN connector)
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector)
Table 4.
Specification
Value
Comments
Input Voltage
Range
Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)
Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk
—
Maximum
Input
Overload
7 Vrms with |Peaks| ≤ 10 V
—
Impedance
50 Ω
—
Coupling
AC
—
© National Instruments Corporation
13
NI PXI-5122 Specifications
CLK OUT (Sample Clock and Reference Clock Output,
Front Panel Connector)
Table 5.
Specification
Value
Comments
Output
Impedance
50 Ω
—
Logic Type
3.3 V CMOS
—
Maximum
Drive Current
±48 mA
—
Trigger
Reference (Stop) Trigger
Table 6.
Specification
Trigger Types
and Sources
Time
Resolution
Holdoff
Value
Types
Edge, Window, Hysteresis,
Video, Digital, Immediate,
and Software
Comments
Sources
CH 0, CH 1, TRIG,
PXI_TRIG <0..6>,
PFI <0..1>, PXI Star
Trigger, and Software
TDC
Onboard Clock
External Clock
On
100 ps
N/A
Off
10 ns
External Clock Period
TDC
Onboard Clock
External Clock
On
10 µs to 171.79 s
N/A
Off
2 µs to 171.79 s
NI PXI-5122 Specifications
Refer to the
following
sections and
NI High-Speed
Digitizers Help
for more
information about
what sources are
available for each
trigger type.
TDC = Time to
Digital
Conversion
Circuit
—
200 ∗ (External Clock
Period) to
(232 – 1) ∗ (External Clock
Period)
14
ni.com
Table 6. (Continued)
Specification
Value
Comments
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources
1. CH 0 (front panel BNC connector)
—
2. CH 1 (front panel BNC connector)
3. TRIG (front panel BNC connector)
Trigger Level
Range
Trigger Level
Resolution
Edge Trigger
Sensitivity
CH 0, CH 1
TRIG (External Trigger)
—
100% FS
±5 V
—
10 bits (1 in 1,024)
CH 0, CH 1
2.5% FS up to 50 MHz,
increasing to 5% FS at
100 MHz
Level
Accuracy,
Typical
±3.5% FS up to 10 MHz
Jitter
80 ps rms
Trigger
Filters
—
TRIG (External Trigger)
—
0.25 Vpk-pk up to 100 MHz,
increasing to 1 Vpk-pk at
200 MHz
CH 0, CH 1
TRIG (External Trigger)
—
±0.35 V (±3.5% FS) up to
10 MHz
Within ±5 °C of
self-calibration
temperature
Low-Frequency
(LF) Reject
High-Frequency
(HF) Reject
50 kHz
50 kHz
—
Digital Trigger (Digital Trigger Type)
Sources
1. PXI_Trig <0..6> (backplane connector)
—
2. PFI <0..1> (front panel SMB connector)
3. PXI Star Trigger (backplane connector)
Video Trigger (Video Trigger Type)
Sources
1. CH 0 (front panel BNC connector)
—
2. CH 1 (front panel BNC connector)
3. TRIG (front panel BNC connector)
© National Instruments Corporation
15
NI PXI-5122 Specifications
Table 6. (Continued)
Specification
Types
Value
Comments
1. Specific Line
—
2. Any Line
3. Specific Field
Standard
Negative sync of NTSC, PAL, or SECAM signal
—
TRIG (External Trigger, Front Panel Connector)
Table 7.
Specification
Value
Comments
Connector
BNC
—
Impedance
1 MΩ in parallel with 22 pF
—
Coupling
AC, DC
—
AC-Coupling Cutoff
(–3 dB)
12 Hz
—
Input Voltage Range
±5 V
—
Maximum Input Overload
|Peaks| ≤ 42 V
—
NI PXI-5122 Specifications
16
ni.com
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors)
Table 8.
Specification
Value
Comments
Connector
9-pin mini-circular DIN
—
Direction
Bi-directional
—
As an Input (Trigger)
Destinations
1. Start Trigger (Acquisition Arm)
—
2. Reference (Stop) Trigger
3. Arm Reference
4. Advance Trigger
Input Impedance
150 kΩ
—
VIH
2.0 V
—
VIL
0.8 V
—
Maximum Input
Overload
–0.5 V to 5.5 V
—
Maximum
Frequency
25 MHz
—
As an Output (Event)
Sources
1. Start Trigger (Acquisition Arm)
—
2. Reference (Stop) Trigger
3. End of Record
4. Done (End of Acquisition)
5. Probe Compensation (1 kHz, 50% duty cycle
square wave, PFI 1 only)
Output Impedance
50 Ω
—
Logic Type
3.3 V CMOS
—
Maximum Drive
Current
±24 mA
—
Maximum
Frequency
25 MHz
—
© National Instruments Corporation
17
NI PXI-5122 Specifications
Waveform Specifications
Table 9.
Specification
Value
Onboard
Memory Size
8 MB per
channel standard
(4 megasamples
per channel)
Minimum
Record Length
1 Sample
Number of
Pretrigger
Samples
Zero up to full Record Length
True for both
single-record
mode and
multiple-record
mode
Number of
Posttrigger
Samples
Zero up to full Record Length
True for both
single-record
mode and
multiple-record
mode
Maximum
Number of
Records in
Onboard
Memory
Allocated
Onboard
Memory Per
Record
32 MB per
channel option
(16 megasamples
per channel)
Comments
256 MB per
channel option
(128 megasamples per
channel)
—
—
8 MB per
Channel
Standard
32 MB per
Channel Option
256 MB per
Channel Option
—
32,768
131,072
1,048,576
Record Length in samples + 100 samples. Round the sum
up to the next multiple of 64 samples.
—
1 sample = 2 bytes
NI PXI-5122 Specifications
18
ni.com
Calibration
Table 10.
Specification
Value
Comments
Self-Calibration
Self-calibration is done on software command.
The calibration corrects for gain, offset, frequency
response, triggering, and timing adjustment errors
for all input ranges.
—
External Calibration
(Factory Calibration)
The external calibration calibrates the VCXO and the
voltage reference. Appropriate constants are stored
in nonvolatile memory.
—
Interval for External
Calibration
2 years
—
Warm-Up Time
15 minutes
—
Power
Table 11.
Specification
Typical Value
Comments
+3.3 VDC
1.4 A
—
+5 VDC
1.6 A
+12 VDC
0.10 A
–12 VDC
0.27 A
Total Power
17.1 W
© National Instruments Corporation
19
NI PXI-5122 Specifications
Software
Table 12.
Specification
Value
Comments
Driver
Software
NI-SCOPE 2.5 or later. NI-SCOPE is an IVI-compliant
driver that allows you to configure, control, and calibrate the
NI 5122. NI-SCOPE provides application programming
interfaces for many development environments.
—
Application
Software
NI-SCOPE provides programming interfaces,
documentation, and examples for the following application
development environments:
—
• LabVIEW
• LabWindows™/CVI™
• Measurement Studio
• Microsoft Visual C/C++
• Microsoft Visual Basic
Interactive
Soft Front
Panel and
Configuration
The Scope Soft Front Panel 2.0 or later supports interactive
control of the NI 5122. The Scope Soft Front Panel is
included on the NI-SCOPE CD.
—
National Instruments Measurement & Automation Explorer
(MAX) also provides interactive configuration and test tools
for the NI 5122. MAX is included on the NI-SCOPE CD.
NI PXI-5122 Specifications
20
ni.com
Environment
Table 13.
Specification
Operating
Temperature
Value
Comments
0 ºC to +55 ºC in all NI PXI chassis except the following:
—
0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis
(Meets IEC-60068-2-1 and IEC-60068-2-2)
Storage
Temperature
–20 ºC to +70 ºC
—
Operating Relative
Humidity
10% to 90%, noncondensing
Storage Relative
Humidity
5% to 95%, noncondensing
Operating Shock
30 g, half-sine, 11 ms pulse
(Meets IEC-60068-2-1 and IEC-60068-2-2)
—
(Meets IEC-60068-2-56)
—
(Meets IEC-60068-2-56)
—
(Meets IEC-60068-2-27. Test profile developed in
accordance with MIL-PRF-28800F)
Storage Shock
50 g, half-sine, 11 ms pulse
—
(Meets IEC-60068-2-27. Test profile developed in
accordance with MIL-PRF-28800F)
Operating Vibration
5 Hz to 500 Hz, 0.31 grms
—
(Meets IEC-60068-2-64)
Storage Vibration
5 Hz to 500 Hz, 2.46 grms
—
(Meets IEC-60068-2-64. Test profile exceeds requirements
of MIL-PRF-28800F, Class B)
Altitude
2,000 m maximum (at 25 °C ambient temperature)
—
Pollution Degree
2
—
© National Instruments Corporation
21
NI PXI-5122 Specifications
Safety, Electromagnetic Compatibility, and
CE Compliance
Table 14.
Specification
Safety
Value
The NI 5122 meets the requirements of the
following standards for safety and electrical
equipment for measurement, control, and
laboratory use:
Comments
For UL and other
safety certifications,
refer to the product
label or visit ni.com.
• IEC 61010-1, EN 61010-1
• UL 3111-1, UL 61010B-1
• CAN/CSA C22.2 No. 1010.1
Emissions
EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
—
Immunity
EN 61326:1997 + A2:2001, Table 1
—
EMC/EMI
CE, C-Tick, and FCC Part 15 (Class A) Compliant.
For EMC compliance, you must operate this device
with shielded cabling.
—
This product meets the essential requirements of applicable European Directives, as amended
for CE marking, as follows:
Low-Voltage
Directive (safety)
73/23/EEC
—
Electromagnetic
Compatibility
Directive (EMC)
89/336/EEC
—
Note: Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory
compliance information. To obtain the DoC for this product, visit ni.com/hardref.nsf/.
NI PXI-5122 Specifications
22
ni.com
Physical
Table 15.
Specification
Dimensions
Value
Comments
Single 3U PXI Slot. CompactPCI compatible.
—
Front Panel Connectors
Label
Functions
Connector Type
CH 0
Analog Input
BNC female
CH 1
Analog Input
BNC female
TRIG
External Trigger
BNC female
CLK IN
Sample Clock Input and
Reference Clock Input
SMB jack
CLK OUT
Sample Clock Output and
Reference Clock Output
SMB jack
AUX I/O
PFI 0, PFI 1
9-pin mini-circular DIN
Front Panel Indicators
LEDs
Access LED
Active LED
The Access LED indicates the
status of the PCI bus and the
interface from the NI 5122 to
the controller.
© National Instruments Corporation
The Active LED
indicates the status of
the onboard acquisition
hardware of the
NI 5122.
23
For more
information,
refer to the
NI High-Speed
Digitizers Help.
NI PXI-5122 Specifications
CVI™, IVI™, LabVIEW™, Measurement Studio™, National Instruments™, NI™, and ni.com™ are
trademarks of National Instruments Corporation. Product and company names mentioned herein are
trademarks or trade names of their respective companies. For patents covering National Instruments
products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on
your CD, or ni.com/patents.
© 2003 National Instruments Corp. All rights reserved.
*323497A-01*
323497A-01
Aug03