NI 6608 Timing I/O Device Calibration Procedure

CALIBRATION PROCEDURE
NI 6608 TIMING I/O DEVICE
Introduction
This document contains information and step-by-step instructions for
calibrating the oven-controlled crystal oscillator (OCXO) on the NI 6608
timing I/O device.
What Is Calibration?
Calibration consists of two steps: verifying the measurement accuracy of a
device and correcting for any measurement error. Verification is measuring
the performance of a device and comparing the results to the factory
specifications of the device. If the device is not operating within its
specifications, use the second step of the calibration procedure to correct
the measurement error of the NI 6608.
Why Should You Calibrate?
The accuracy of timing components can drift with time and temperature,
which could cause them to be less accurate than specified. Calibration
ensures that the NI 6608 still meets National Instruments specifications.
How Often Should You Calibrate?
The measurement requirements of the application determine how often you
should calibrate the device to ensure its accuracy. NI recommends that you
calibrate the NI 6608 at least once every year. You can shorten this interval
to six months or 90 days based on the demands of the application.
LabVIEW™, National Instruments™, NI™, ni.com™, NI-DAQ™, and PXI ™ are trademarks of National Instruments Corporation. Product and company
names mentioned herein are trademarks or trade names of their respective companies.
370275A-01
Copyright © 2001 National Instruments Corp. All rights reserved.
June 2001
Equipment and Other Test Requirements
This section describes the equipment, software, documentation, and test
conditions required for calibrating the NI 6608.
Test Equipment
Calibrating the NI 6608 requires the following equipment:
•
A highly stable 10 MHz clock source that is accurate to within
0.75 parts-per-billion (ppb). NI recommends a rubidium frequency
standard clock source.
•
An SH68-68-D1 cable or a standard 68-pin cable connector
•
A 68-pin breakout box
Software and Documentation
You need the following software and documentation to calibrate
the NI 6608. You can download these from the NI Web site at
ni.com\manuals:
•
The latest version of the NI-DAQ driver
•
NI-DAQ Function Reference Help file
•
NI-DAQ User Manual for PC Compatibles
Software
The NI 6608 calibration procedure requires using the latest version
of the NI-DAQ driver software. NI-DAQ supports a number of application
development environments (ADEs) and programming languages,
including LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft
Visual Basic, and Borland C++. When you install the driver, you only need
to install support for the ADE or programming language that you are using.
You can find the calibration software for adjusting the frequency of the
10 MHz OCXO on the NI 6608 on the NI calibration web page at
ni.com/support/calibrat. Click the Manual Calibration Procedures link.
Note
Documentation
The NI-DAQ Function Reference Help file and NI-DAQ User Manual for
PC Compatibles contain detailed information on using the NI-DAQ driver.
The help file, which you can access by clicking Start»Programs»National
Instruments»NI-DAQ»NI-DAQ Help, includes detailed information on
the driver functions. The NI-DAQ User Manual for PC Compatibles
provides instructions on installing NI-DAQ. This manual also includes
detailed information on creating applications that use the NI-DAQ driver.
6608 Calibration Procedure
2
ni.com
These sources are your primary references for writing your calibration
utility. You can also refer to the documentation for the programming
language you are using. For further information on the device you are
calibrating, refer to the 660X User Manual.
Test Conditions
Follow these guidelines to optimize connections and test conditions during
the calibration procedure:
•
Keep connections to the device as short as possible. Long cables and
wires act as antennae by picking up extra noise that can affect
measurements.
•
Maintain a temperature of approximately 25 °C.
•
Keep relative humidity below 80%.
•
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
•
Allow a warm-up time of at least 15 minutes to ensure the
measurement circuitry is at a stable operating temperature.
Verifying the Frequency of the NI 6608
Note If you prefer, you can go directly to the Adjusting the Frequency of the 10 MHz
OCXO on the NI 6608 section.
This section contains instructions for verifying the performance of the
OCXO on the NI 6608 using a stable clock source. Verifying the
performance of the oscillator on an NI 6608 requires two counters on the
NI 6608—counters 0 and 1. Counter 0 uses the stable clock as the source
to generate a 400 s single pulse, which is applied to the GATE of counter 1.
The 10 MHz OCXO signal on the NI 6608 is the SOURCE of counter 1.
Counter 1 counts the number of clock edges of the 10 MHz signal that
occur during the 400 s pulse and calculates the frequency of the oscillator
to determine if the device is within its specifications. Figure 1 shows a
simple diagram of the verification process.
© National Instruments Corporation
3
NI 6608 Calibration Procedure
External Stable Clock Source
CTR 0
10 MHz OCXO Signal
400 s
Source
CTR 1
Source
Gate
Output
Figure 1. NI 6608 Verification Process
The frequency measurement has an uncertainty that is equal to the sum of
the stable clock error and the quantization error caused by the counter. The
clock error should be 0.75 ppb or less, as specified in the Test Equipment
section of this document. The quantization error, which is the inherent
uncertainty in digitizing an analog value, is always up to 100 ns. Using a
measurement duration of 400 s gives a quantization error as follows:
100 ns
---------------- = 0.25 ppb
400 s
The sum of the stable clock error, 0.75 ppb, and the quantization error,
0.25 ppb, gives a total measurement uncertainty of 1 ppb. You can reduce
measurement uncertainty by using a more stable clock, by increasing the
measurement duration, or both. The value returned from the Measuring the
Frequency of the 10 MHz OCXO section allows for a 1 ppb measurement
uncertainty.
Both NI-DAQ programming and LabVIEW programming instructions are
presented in the following sections. Choose the appropriate instructions for
your programming environment.
Connecting the Clock Source to the NI 6608
Note The stable clock must meet the specifications described in the Test Equipment
section of this document. Also, allow adequate time for the stable clock to warm up.
To connect a clock source to the NI 6608, complete the following steps:
6608 Calibration Procedure
1.
Connect the NI 6608 to a 68-pin breakout box using the
SH68-68-D1 cable.
2.
Connect the stable clock to pin 2 (PFI_39), which is the SOURCE
of counter 0 on the NI 6608.
3.
Connect the ground of the stable clock to pin 36, which is the
digital ground.
4
ni.com
Measuring the Frequency of the 10 MHz OCXO
Make sure you plug the NI 6608 in the star trigger slot (slot 2) of the PXI chassis
and configure it in Measurement & Automation Explorer.
Note
To measure the frequency of the 10 MHz OCXO the NI 6608 for
verification, do the following:
NI-DAQ Function Call
LabVIEW Block Diagram
Call Verify_TIO with the following
parameters:
deviceNumber: The value set by MAX
actualFrequency: A user-declared
variable.
Example:
double frequency;
Verify_TIO (deviceNumber,
&frequency);
Note
Note: You can set or view the parameters above on the
front panel of the Verify TIO VI.
The verification process for the NI 6608 takes at least 6 minutes.
After the software has finished, you have completed verifying the
frequency of the 10 MHz OCXO on the NI 6608. Refer to the next section
to ensure the NI 6608 is within specifications.
Comparing Measured Frequency to NI 6608 Device Specifications
Compare the values returned from the Measuring the Frequency of the
10 MHz OCXO section with the following values:
•
If the frequency you measure falls between the upper limit of
10,000,000.1 Hz and the lower limit of 9,999,999.9 Hz, the NI 6608 is
operating within its specifications.
•
If the frequency you measure does not fall between the upper limit of
10,000,000.1 Hz and the lower limit of 9,999,999.9 Hz, the NI 6608
needs to be adjusted.
You have completed verifying the performance of the NI 6608. If the
NI 6608 needs to be adjusted, the next section guides you through the
process.
© National Instruments Corporation
5
NI 6608 Calibration Procedure
Adjusting the Frequency of the 10 MHz OCXO
on the NI 6608
This section contains instructions for adjusting the frequency of the OCXO
on the NI 6608. The adjustment process iteratively measures the OCXO
frequency and adjusts the hardware until the measured frequency is within
specification. The final settings are stored in the device EEPROM and are
automatically loaded upon device initialization.
Both NI-DAQ programming and LabVIEW programming instructions are
presented in the following table. Choose the appropriate instructions for
your programming environment.
Note Prior to adjusting the frequency on the 10 MHz OCXO, refer to the Connecting the
Clock Source to the NI 6608 section earlier in this document.
After you have downloaded the NI 6608 calibration software and have set
up the hardware connections, you are ready to adjust the frequency of the
10 MHz OCXO. Configure the NI 6608 to do an external calibration with
a 10 MHz reference frequency as follows:
NI-DAQ Function Call
LabVIEW Block Diagram
Call C_Calibrate_TIO with the following
parameters:
deviceNumber: The value set by MAX
operation: ND_EXTERNAL_CALIBRATION
setOfCalConst: ND_USER_EEPROM_AREA
referenceFreq: 10000000.00
Note: You can set the parameters above on the front
panel of the Calibrate TIO VI.
Note
The adjustment process for the NI 6608 takes at least 95 minutes.
After the software has finished executing, you have completed adjusting the
frequency on the 10 MHz OCXO on the NI 6608. If the measured
frequency is outside the device specification, an error is returned. If you get
an error, verify that the external stable clock source is operating properly
and is connected to the NI 6608. If you still get an error, contact NI
for repair.
6608 Calibration Procedure
6