MIHW1000 SERIES - MINMAX Technology

®
MIHW1000 SERIES
DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety
FEATURES
►Reinforced Insulation rated for 300VAC Working Voltage
►I/O-isolation Voltage 4000VACrms
►Industrial & Medical Safety Approval
►Wide 2:1 Input Voltage Range
►Fully regulated Output Voltage
►Low Leakage Current
►Operating Temp. Range –40°C to +85 °C
►Input Filter meets EN 55022, class A and FCC, level A
►Overload Protection
►3 Year Product Warranty
PRODUCT OVERVIEW
The MINMAX MIHW1000 series is a range of high performance DC/DC converter modules with a reinforced insulation system .The I/O- isolation voltage is
specified for 4000VACrms.The product comes in a small DIP-24 package. All 20 models features wide 2:1 input voltage range and fully regulated output
voltage.
The MIHW1000 DC/DC converters offer an economical solution for demanding applications in industrial and medical instrumentation requesting a certified
supplementary or reinforced insulation system to comply with relative industrial or medical safety standards.
Model Selection Guide
Model
Number
MIHW1002
MIHW1003
MIHW1008
MIHW1006
MIHW1007
MIHW1012
MIHW1013
MIHW1018
MIHW1016
MIHW1017
MIHW1022
MIHW1023
MIHW1028
MIHW1026
MIHW1027
MIHW1032
MIHW1033
MIHW1038
MIHW1036
MIHW1037
Input
Voltage
(Range)
Output
Voltage
VDC
VDC
5
12
24
±12
±15
5
12
24
±12
±15
5
12
24
±12
±15
5
12
24
±12
±15
5
(4.5 ~ 9)
12
(9 ~ 18)
24
(18 ~ 36)
48
(36 ~ 75)
Output Current
Input Current
Max.
Min.
@Max. Load
@No Load
Reflected
Ripple
Current
mA
600
250
125
±125
±100
600
250
125
±125
±100
600
250
125
±125
±100
600
250
125
±125
±100
mA
90
37.5
18.8
±18.8
±15
90
37.5
18.8
±18.8
±15
90
37.5
18.8
±18.8
±15
90
37.5
18.8
±18.8
±15
mA(typ.)
857
800
800
800
800
338
313
313
313
313
160
151
151
151
151
80
75
75
75
75
mA(typ.)
mA (typ.)
40
60
30
30
20
15
10
10
Max. capacitive
Load
μ F
1000
470
470
220#
220#
1000
470
470
220#
220#
1000
470
470
220#
220#
1000
470
470
220#
220#
@Max. Load
# For each output
E-mail:[email protected]
2014/09/23 REV:5
Tel:886-6-2923150
Page 1 of 4
Efficiency
(typ.)
%
70
75
76
75
75
74
80
81
80
80
78
83
84
83
83
78
83
84
83
83
®
MIHW1000 SERIES
DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety
Input Specifications
Parameter
Model
Input Surge Voltage (1 sec. max.)
Start-Up Threshold Voltage
Under Voltage Shutdown
Min.
-0.7
Typ.
5V Input Models
---
Max.
11
12V Input Models
-0.7
---
25
24V Input Models
-0.7
---
50
48V Input Models
-0.7
---
100
5V Input Models
12V Input Models
24V Input Models
48V Input Models
5V Input Models
12V Input Models
24V Input Models
48V Input Models
3.7
4
4.5
8
8.5
9
VDC
15
17
18
30
33
36
----4
----8.5
----17
----34
----0.3
A
----2000
mW
----2500
mW
Compliance to EN 55022,class A and FCC part 15,class A
Reverse Polarity Input Current
Short Circuit Input Power
Internal Power Dissipation
Conducted EMI
All Models
Unit
Output Specifications
Parameter
Output Voltage Setting Accuracy
Output Voltage Balance
Line Regulation
Load Regulation
Conditions
Dual Output, Balanced Loads
Vin=Min. to Max.
Io=25% to 100%
5V Output Models
0-20 MHz Bandwidth
Other Output Models
Ripple & Noise (20MHz)
Transient Recovery Time
Transient Response Deviation
Temperature Coefficient
Over Current Protection
Short Circuit Protection
25% Load Step Change
Foldback
Min.
------------------120
Continuous
Typ.
--±0.5
±0.3
±0.5
75
100
150
±3
±0.02
150
Max.
±1.0
±2.0
±0.5
±1.0
100
150
500
±6
±0.05
---
Unit
%Vnom.
%
%
%
mV P-P
mV P-P
μ sec
%
%/℃
%
Max.
--2
--13
Unit
VACrms
μ A
GΩ
pF
Isolation, Safety Standards
Parameter
I/O Isolation Voltage (rated)
Leakage Current
I/O Isolation Resistance
I/O Isolation Capacitance
Conditions
60 Seconds
240VAC, 60Hz
500 VDC
100KHz, 1V
Safety Standards
Approvals
Min.
Typ.
4000
------10
----7
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
UL60601-1,CSA C22.2 No.601-1
IEC/EN 60950-1, IEC/EN 60601-1
IEC60950-1 CB report, cUL/UL 60950-1 certificate
UL60601-1 UL certificate
General Specifications
Parameter
Conditions
Switching Frequency
MIL-HDBK-217F@25℃, Ground Benign
MTBF(calculated)
E-mail:[email protected]
2014/09/23 REV:5
Tel:886-6-2923150
Page 2 of 4
Min.
Typ.
Max.
Unit
---
150
---
KHz
1,000,000
---
---
Hours
®
MIHW1000 SERIES
DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety
Environmental Specifications
Parameter
Conditions
Operating Ambient Temperature Range (See Power Derating Curve)
Case Temperature
Storage Temperature Range
Humidity (non condensing)
Cooling
Lead Temperature (1.5mm from case for 10Sec.)
Min.
-40
Natural Convection
---50
--Free-Air convection
---
Max.
+85
Unit
℃
+100
+125
95
℃
℃
% rel. H
260
℃
Power Derating Curve
100
Natural
Convection
20LFM
Output Power (%)
80
100LFM
60
200LFM
400LFM
40
20
0
~
-40
0
20
40
60
80
100 110
Ambient Temperature C
Notes
1
2
3
4
5
6
7
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these modules;
however, they may not meet all specifications listed.
We recommend to protect the converter by a slow blow fuse in the input supply line.
Other input and output voltage may be available, please contact factory.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
Package Specifications
0.2 [0.008]
Pin Connections
3.9 10.5 [0.41]
[0.15]
Mechanical Dimensions
0.60 [0.024]
12
Bottom View
15
23
20.32 [0.80]
2.54
[0.10]
31.8 [1.25]
+Vin
+Vin
11
No Pin
Common
12
-Vout
No Pin
13
+Vout
-Vout
15
No Pin
+Vout
23
-Vin
-Vin
24
-Vin
-Vin
5.08
[0.20]
2.0
[0.08]
►
All dimensions in mm
►
Tolerance: X.X±0.25
Physical Characteristics
Case Size
Case Material
:
:
31.8x20.3x10.5mm (1.25x0.8x0.41 inches)
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
Pin Material
Weight
:
:
Copper Alloy with Gold Plate Over Nickel Subplate
16.2g
2014/09/23 REV:5
Tel:886-6-2923150
Page 3 of 4
(inches)
(X.XX±0.01)
X.XX±0.13 ( X.XXX±0.005)
►
Pin
E-mail:[email protected]
Dual Output
13
2.54
[0.10]
24
Single Output
1
20.3 [0.80]
11
15.22 [0.60]
1
Pin
diameter  0.6 ±0.05 (0.024±0.002)
®
MIHW1000 SERIES
DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety
Test Setup
Input Reflected-Ripple Current Test Setup
Input reflected-ripple current is measured with a inductor Lin (4.7μ H) and Cin (220μ F, ESR < 1.0Ω at 100 KHz) to simulate source impedance. Capacitor Cin,
offsets possible battery impedance. Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz.
To Oscilloscope
+
+
Battery
+Vin
Lin
DC / DC
Converter
Current
Probe
Cin
+Out
-Vin
Load
-Out
Peak-to-Peak Output Noise Measurement Test
Use a Cout 0.47μ F ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load
between 50 mm and 75 mm from the DC/DC Converter.
+Vin
+Out
Copper Strip
Single Output
DC / DC
Converter
-Vin
Cout
-Out
Scope
+Vin
+Out
Copper Strip
Dual Output
DC / DC
Converter
Com.
Copper Strip
-Vin
-Out
Resistive
Load
Copper Strip
Scope
Cout
Cout
Resistive
Load
Scope
Copper Strip
Technical Notes
Overcurrent Protection
To provide protection in a fault (output overload) condition, the unit is equipped with internal current limiting circuitry and can endure current limiting for an unlimited
duration. At the point of current-limit inception, the unit shifts from voltage control to current control. The unit operates normally once the output current is brought back
into its specified range.
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In
applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor on the input to insure startup. Capacitor
mounted close to the power module helps ensure stability of the unit, it is recommended to use a good quality low Equivalent Series Resistance (ESR < 1.0Ω at 100
kHz) capacitor of a 10μ F for the 5V input devices and a 4.7μ F for the 12V input devices and 2.2μ F for the 24V and 48V devices.
+
DC Power
Source
+Vin
+
+Out
DC / DC
Converter
Load
Cin
-
-Vin
-Out
Output Ripple Reduction
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is
recommended to use 3.3μ F capacitors at the output.
+
+Vin
DC Power
Source
-
+Out
Single Output
DC / DC
Converter
-Vin
+
Cout
+Vin
DC Power
Source
Load
-Out
-
+Out
Dual Output
DC / DC Com.
Converter
-Vin
-Out
Cout
Load
Load
Cout
Maximum Capacitive Load
The MIHW1000 series has limitation of maximum connected capacitance on the output. The power module may operate in current limiting mode during start-up,
affecting the ramp-up and the startup time. Connect capacitors at the point of load for best performance. The maximum capacitance can be found in the data
sheet.
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the
maximum temperature rating of the components inside the power module, the case temperature must be kept below 100℃. The derating curves are determined
from measurements obtained in a test setup.
Position of air velocity
probe and thermocouple
15mm / 0.6in
50mm / 2in
Air Flow
DUT
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected]
Minmax Technology Co., Ltd.
2014/09/23 REV:5