51
CrCl3
Product Number:
NEZ020
LOT SPECIFIC INFORMATION
Stock Date
Lot Number
Specific Activity
Concentration
Radionuclidic Purity
14-Jan-2015
14M06K8
698.90 mCi/mg
25.86 GBq/mg
506.61 mCi/mL
18.74 GBq/mL
99.9 %
Values given are all as of Stock Date. Calibration is 7 days from ship date.
Use decay chart on back page to decay values to Calibration date.
PACKAGING: Solvent is 0.5M HCl.
SPECIFIC ACTIVITY RANGE: Calculated as Cr.
STORAGE CONDITIONS: Store this product at room temperature.
CONTAMINANTS: none detected
PHYSICAL PROPERTIES:
Decay Mode
Half Life
Principal Gamma Photon
Electron Capture
27.7 Days
0.320 MeV (9.8%)
Occupational Limits (based on most restrictive intake category: 10 CFR 20 U.S.NRC Regulations)
Derived Air Concentration (DAC)
Annual Limit on Intake (ALI)
-6
8x10 µCi/mL
4
2x10 µCi
DECAY CHART:
To use the decay table find the number of days in the top row and left hand column of
the chart then find the corresponding decay factor. To obtain a precalibration number,
divide by the decay factor. For a postcalibration number multiply by the decay factor.
Days
0
10
20
30
0
1.000
.7787
.6064
.4722
1
.9753
.7595
.5914
.4605
2
.9512
.7407
.5768
.4491
3
.9277
.7224
.5625
.4380
4
.9048
.7045
.5486
.4272
5
.8824
.6871
.5351
.4167
6
.8606
.6702
.5219
.4064
7
.8394
.6536
.5090
.3963
8
.8186
.6375
.4964
.3865
9
.7984
.6217
.4841
.3770
HAZARD INFORMATION:
WARNING: this product contains a chemical known to the state of California to cause cancer.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
51
Designate area for handling Cr, clearly label all containers.
51
Store Cr behind lead shielding.
Wear extremity and whole body dosimeters while handling mCi (37 MBq) quantities.
51
Use shielding to minimize exposure while handling Cr.
Use tools to indirectly handle unshielded sources and potentially contaminated vessels.
51
Prohibit eating, drinking, smoking and mouth pipetting in room where Cr is handled.
Use transfer pipettes, spill trays and absorbent coverings to confine contamination.
51
Handle Cr compounds that are potentially volatile or in powder form in ventilated enclosures.
Sample exhausted effluent and room air by continuously drawing a known volume through membrane filters.
Wear lab coat, wrist guards and disposable gloves for secondary protection.
Maintain contamination and exposure control by regularly monitoring and promptly decontaminating gloves and
surfaces.
51
Use end-window Geiger-Mueller detectors, NaI(T1) detector or liquid scintillation counter to detect Cr.
51
51
Submit urine samples for bioassay at least four hours after handling Cr to indicate uptake by personnel. Cr is
slowly eliminated from the body. Whole body counting provides a more sensitive method than urinalysis for
51
determining Cr body burdens. Whole body counting may be used occasionally to verify the urinalysis results.
Isolate waste in clearly labeled, shielded containers and hold for decay.
Establish surface contamination, air concentration and urinalysis action levels below regulatory limits. Investigate
and correct any causes that threaten these levels to be exceeded.
51
On completing an operation, secure all Cr, remove and dispose protective clothing and coverings, monitor and
decontaminate self and surfaces, wash hands and monitor them again.
PerkinElmer, Inc.
549 Albany Street
Boston, MA 02118 USA
P: (800) 762-4000 or (+1) 203-925-4602
www.perkinelmer.com/nenradiochemicals
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2010, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc.
All other trademarks are the property of their respective owners.