MINT1400A 400W Single Output Series Medical – Universal Input Power Supply 2 Year Warranty •Compact 3.94” x 7” x 1.58”, 1U compatible •9.5 W/in³ •Up to 250W Convection Rating •Efficiency up to 91% •12Vdc to 48Vdc outputs, up to 400W •Remote ON/OFF •Safety approved to Medical and ITE Standards • EN/IEC/UL60601-1 and CSA 601.1 •CSA/EN/IEC/UL60950-1, 2nd Edition •Compliant to EN55011/22 Conducted Noise, Class B •RoHS Compliant – Category 6 Specifications All Specifications are typical at nominal input, full load at 25°C unless otherwise stated. AC Input Input Current Inrush Current Turn On Time 500mSec maximum 120VAC: 4.2A, 230VAC: 2.1A Hold-up Time 20 ms at 400W Active circuit, 37A max Input Fuses F1, F2: 6.3A Power Factor Efficiency 85-270VAC, 47-63 Hz single phase 0.99 typical, 0.96 minimum Typical 88% @ 120VAC, 91% @ 240VAC Remote Sense Compensates for up to 500mV drop on output leads Output Power 400W with fan cooling. See derating chart for other conditions. Peak: 450W for 1 min., 10% duty cycle Ripple and Noise 1% pk-pk max., 20MHz BW, differential mode Output Voltages See chart. Voltages adjustable +/- 5% Total Regulation See chart Minimum Load Not required Transient Response 500µs to 1%, 0.2A/µs, 3.5% Deviation Shock and Vibration Shock: MIL-STD-810E, 516.4 part IV Vibration: 5Hz – 500Hz, 2.5 Grms Overvoltage Protection Overload Protection See chart Cycling type, 470W typical Overtemperature Protection Provided on all models, autorecovery. See page 2 for cooling information Switching Frequency Output: 170kHz ± 8kHz; PFC: 90kHz, ± 8kHz; Bias: 100kHz, ± 6 kHz Safety Standards Medical: EN/IEC/UL 60601-1, CSA 601.1 ITE: CSA/EN/IEC/UL 60950-1, 2nd Edition Isolation Input-Output: 4000VAC, Input-Ground: 1500VAC, Output-Ground: 500VAC Earth Leakage Current 100µA at 120VAC, 60Hz Normal. 400µA at 264Vac, 60Hz SFC. Operating Temperature 0° to 70°C, refer to derating curve for ratings >40°C Storage Temperature -40 to +85°C Relative Humidity 5% to 95%, non-condensing Altitude Operating: -500 to 10,000 ft. Non-Operating: -500 to 40,000 ft. Model Number Key MINT 1 400 A 12 10 L 01 01 = Standard Model, 02 and higher indicates a m odified m odel. "L" = Input Connector - EBY P/N 5002-03-N-12 "10" = Output Connections - Bus Bars w ith #10 screw s Output Voltage: "12" = 12V output, "24" = 24V output, etc. Model Configuration: "A" = Original Configuration; "C" = Optional fan cover included Output Pow er (Watts) - "400" = 400W # of Outputs Product Fam ily: "M" = Medical, "I" = Internal, "NT" = New Technology ______________________________________________________________________________________________________________________________ 1 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com MINT1400A 400W Single Output Series Medical – Universal Input Power Supply 2 Year Warranty Output Parameters Model Number Output Volts (V) MINT1400A1210L01 12V MINT1400A2410L01 MINT1400A2810L01 MINT1400A4810L01 Output Current Fan Cooled Convection Ripple & Noise Total Regulation 33.0A 21.0A 1% 2% 24V 16.7A 10.4A 1% 2% 28V 14.3A 8.9A 1% 2% 48V 8.30A 5.2A 1% 2% Signals Inhibit: TTL LOW to inhibit PS Off: TTL HIGH to inhibit Pwr OK TTL HIGH when AC is OK, 100-200 mS delay. Pwr OK PF LOW 4 mS warning before loss of output regulation. DC OK: HIGH when output is >90% of nominal Fan Output: 12Vdc @ 250 mA. 5V Standby Voltage: Always on, up to 100mA. EMI/EMC Compliance Conducted Emissions EN55011/22 Class B Radiated Emissions EN55011/22 Class A Voltage Fluctuations EN61000-3-3, complies Static Discharge Immunity EN61000-4-2, 6kV Contact Discharge, 8kV air discharge Radiated RF Immunity EN61000-4-3, 3V/m. EFT/Burst Immunity EN61000-4-4, 2kV/5kHz.. Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m Voltage Dip Immunity EN61000-4-11 criteria A @ 250 Watts or less. Line Frequency Harmonics EN61000-3-2, Class A Output Power Derating Curve Output Power (Watts) MINT1400 Output Power Derating Chart 450 400 350 300 250 200 150 100 50 0 Note: For convection cooling, the power supply is required to be mounted on a heatsink or system chassis with a minimum of 150 in² of surface area. Fan Cooled Convection 0 10 20 30 40 50 60 70 Ambient Temperature (deg C) ______________________________________________________________________________________________________________________________ 2 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com MINT1400A 400W Single Output Series Medical – Universal Input Power Supply 2 Year Warranty Outline Drawing & Connector Information Connector Information Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. Rev. 6-7-2013 However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice. ______________________________________________________________________________________________________________________________ 3 of 3 SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com