Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com MSP-200 200W Single Output Medical Type series ■ Features : ‧Universal AC input / Full range ‧Built-in active PFC function, PF>0.95 ‧High efficiency up to 89% ‧Withstand 300VAC surge input for 5 seconds ‧Protections: Short circuit / Overload / Over voltage / Over temperature ‧Cooling by free air convection ‧Built-in constant current limiting circuit ‧1U low profile 38mm ‧Medical safety approved (MOOP level) ‧Built-in remote ON-OFF control ‧Standby [email protected] ‧Built-in remote sense function ‧No load power consumption<0.5W (Note.6) ‧5 years warranty SPECIFICATION MODEL OUTPUT MSP-200-3.3 MSP-200-5 MSP-200-7.5 MSP-200-12 MSP-200-15 MSP-200-24 MSP-200-36 MSP-200-48 DC VOLTAGE 3.3V 5V 7.5V 12V 15V 24V 36V 48V RATED CURRENT 40A 35A 26.7A 16.7A 13.4A 8.4A 5.7A 4.3A CURRENT RANGE 0 ~ 40A 0 ~ 35A 0 ~ 26.7A 0 ~ 16.7A 0 ~ 13.4A 0 ~ 8.4A 0 ~ 5.7A 0 ~ 4.3A RATED POWER 132W 206.4W 175W 200.3W 200.4W 201W 201.6W 205.2W RIPPLE & NOISE (max.) Note.2 80mVp-p 90mVp-p 100mVp-p 120mVp-p 150mVp-p 150mVp-p 250mVp-p 250mVp-p VOLTAGE ADJ. RANGE 4.3 ~ 5.8V ±2.0% 6.8 ~ 9V ±2.0% 10.2 ~ 13.8V ±1.0% 13.5 ~ 18V ±1.0% 21.6 ~ 28.8V ±1.0% 28.8 ~ 39.6V ±1.0% 40.8 ~ 55.2V ±1.0% 2.8 ~ 3.8V VOLTAGE TOLERANCE Note.3 ±2.0% LINE REGULATION ±0.5% ±0.5% ±0.5% ±0.3% ±0.3% ±0.2% ±0.2% ±0.2% LOAD REGULATION ±1.5% ±1.0% ±1.0% ±0.5% ±0.5% ±0.5% ±0.5% ±0.5% SETUP, RISE TIME 1000ms, 50ms/230VAC HOLD UP TIME (Typ.) 16ms/230VAC 88% 89% 89% Note.5 85 ~ 264VAC VOLTAGE RANGE FREQUENCY RANGE 47 ~ 63Hz INPUT PF>0.95/230VAC EFFICIENCY (Typ.) 80% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 2.2A/115VAC 35A/115VAC FUNCTION PF>0.99/115VAC at full load 84% 86% 88% 88% 1.1A/230VAC 70A/230VAC Note.7 Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC OVERLOAD PROTECTION 120 ~ 370VDC POWER FACTOR (Typ.) LEAKAGE CURRENT 2500ms, 50ms/115VAC at full load 16ms/115VAC at full load OVER VOLTAGE 105 ~ 135% rated output power Protection type : Constant current limiting, recovers automatically after fault condition is removed 6 ~ 7V 9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V 41.4 ~ 48.6V 3.96 ~ 4.62V 57.6 ~ 67.2V Protection type : Shut down o/p voltage, re-power on to recover OVER TEMPERATURE 5V STANDBY Shut down o/p voltage, recovers automatically after temperature goes down 5VSB : [email protected] ; tolerance± 5%, ripple : 50mVp-p(max.) REMOTE CONTROL RC+/RC- : 4 ~10V or open = power on ; 0 ~ 0.8V or short = power off WORKING TEMP. -40 ~ +70℃ (Refer to "Derating Curve") 20 ~ 90% RH non-condensing WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃ , 10 ~ 95% RH TEMP. COEFFICIENT ±0.03%/ ℃ (0 ~ 50℃) VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes SAFETY STANDARDS ANSI/AAMI ES60601-1, IEC60601-1 approved I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC SAFETY & WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC (Note 4) OTHERS NOTE EMC EMISSION I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 ℃ / 70% RH Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 MTBF 209.4K hrs min. DIMENSION 199*98*38mm (L*W*H) PACKING 0.77Kg; 18pcs/14.9Kg/0.9CUFT MIL-HDBK-217F (25℃) All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. Tolerance : includes set up tolerance, line regulation and load regulation. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Derating may be needed under low input voltages. Please check the derating curve for more details. 6. No load power consumption<0.5W when RC+ & RC- (CN100 pin1,2) 0 ~ 8V or short. 7. Touch current was measured from primary input to DC output. 1. 2. 3. 4. File Name:MSP-200-SPEC 2014-02-12 Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com MSP-200 200W Single Output Medical Type Mechanical Specification Case No.902E series Unit:mm 197 1 2 3 4 5 6 7 LED CN100 SVR1 Terminal Pin No. Assignment Pin No. 1 2 3 Assignment AC/L AC/N FG Pin No. Assignment 4,5 DC OUTPUT -V 6,7 DC OUTPUT +V 57.5 4.5 120 199 6.5 190 Connector Pin No. Assignment (CN100) : HRS DF11-6DP-2DS or equivalent Pin No. Assignment 1 RC+ 2 RC3 AUX 4 AUXG +S 5 -S 6 Mating Housing Terminal HRS DF11-6DS HRS DF11-**SC or equivalent or equivalent 13max. 28 150.75 Block Diagram fosc :70KHz ACTIVE INRUSH CURRENT LIMITING EMI FILTER I/P AUX POWER RECTIFIERS & FILTER AUX AUXG +S RECTIFIERS & PFC POWER SWITCHING RECTIFIERS & FILTER +V -V O.T.P. O.L.P. FG PFC CONTROL Derating Curve -S DETECTION CIRCUIT O.V.P. PWM CONTROL RC+ REMOTE CONTROL RC- Output Derating VS Input Voltage 100 125 100 90 80 80 70 60 60 40 50 20 -40 40 0 10 20 30 40 50 AMBIENT TEMPERATURE (℃) 60 70 (HORIZONTAL) 85 100 125 135 155 264 INPUT VOLTAGE (V) 60Hz File Name:MSP-200-SPEC 2014-02-12 Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com MSP-200 200W Single Output Medical Type series Function Description of CN100 Pin No. Function Description 1 RC+ Turns the output on and off by electrical or dry contact between pin 2 (RC-). Short: Power OFF, Open: Power ON. RC- Remote control ground. 2 Auxiliary voltage output, 4.75~5.25V, reference to pin 4(AUXG). The maximum load current is 0.3A. This output has the built-in oring 3 AUX diodes and is not controlled by the "remote ON/OFF control". 4 5 AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V). Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to +S minimize noise pick-up effect. The maximum line drop compensation is 0.5V. 6 -S Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. Function Manual 1.Remote Control The PSU can be turned ON/OFF by using the "Remote ON/OFF" function Between RC-(pin2) and RC+(pin1) Output Status SW ON (Short) OFF SW OFF (Open) ON CN100 1 RC+ 2 RC- AUXG 1 RC+ 2 RC- AUXG AUX +S 5 -S 6 +S 5 -S 6 SW -V - V +V +V CN100 135 246 LOA D - + F ig 1. 1 2.Remote Sense The remote sensing compensates voltage drop on the load wiring up to 0.5V. -V - V +V +V CN100 135 246 -S (Pin6) LOAD - + +S (Pin5) Sense lines should be twi sted in pairs CN100 AUX F ig 2. 1 File Name:MSP-200-SPEC 2014-02-12