MSP-100 series - Total Power International

Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-100
100W Single Output Medical Type
series
■ Features :
‧Universal AC input / Full range
‧Built-in active PFC function
‧High efficiency up to 90%
‧Withstand 300VAC surge input for 5 seconds
‧Protections: Short circuit / Overload / Over voltage
‧Protections: Over temperature (optional)
‧Cooling by free air convection
‧1U low profile 38mm
‧Medical safety approved (MOOP level)
‧Built-in remote ON-OFF control
‧No load power consumption<0.5W (Note.6)
‧All using 105℃ long life electrolytic capacitors
‧5 years warranty
SPECIFICATION
MODEL
OUTPUT
MSP-100-3.3
MSP-100-5
MSP-100-7.5
MSP-100-12
MSP-100-15
MSP-100-24
MSP-100-36
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
36V
48V
RATED CURRENT
20A
17A
13.5A
8.5A
7A
4.5A
2.9A
2.2A
CURRENT RANGE
0 ~ 20A
0 ~ 17A
0 ~ 13.5A
0 ~ 8.5A
0 ~ 7A
0 ~ 4.5A
0 ~ 2.9A
0 ~ 2.2A
RATED POWER
66W
85W
101.3W
102W
105W
108W
104.4W
105.6W
RIPPLE & NOISE (max.) Note.2 80mVp-p
80mVp-p
100mVp-p
120mVp-p
150mVp-p
150mVp-p
200mVp-p
240mVp-p
VOLTAGE ADJ. RANGE
4.75 ~ 5.8V
+2.5,-3.5%
7.1 ~ 9V
±2.5%
11.4 ~ 13.8V
±1.5%
14.25 ~ 18V
±1.5%
22.8 ~ 28.8V
±1.5%
34.2 ~ 39.6V
±1.5%
45.6 ~ 55.2V
±1.5%
3.1 ~ 3.8V
VOLTAGE TOLERANCE Note.3 +2.5,-3.5%
LINE REGULATION
±0.5%
±0.5%
±0.5%
±0.3%
±0.3%
±0.2%
±0.2%
±0.2%
LOAD REGULATION
±2.0%
±2.0%
±1.5%
±0.8%
±0.8%
±0.5%
±0.5%
±0.5%
SETUP, RISE TIME
2500ms, 100ms/230VAC
HOLD UP TIME (Typ.)
50ms/230VAC
88%
88.5%
89%
90%
PF>0.95/230VAC
EFFICIENCY (Typ.)
78%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.2A/115VAC
35A/115VAC
FUNCTION
PF>0.98/115VAC at full load
83%
84%
87.5%
0.6A/230VAC
65A/230VAC
Note.7 Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC
OVERLOAD
PROTECTION
120 ~ 370VDC
POWER FACTOR (Typ.)
LEAKAGE CURRENT
2500ms, 100ms/115VAC at full load
20ms/115VAC at full load
Note.5 85 ~ 264VAC
VOLTAGE RANGE
FREQUENCY RANGE
47 ~ 63Hz
INPUT
MSP-100-48
OVER VOLTAGE
OVER TEMPERATURE
REMOTE CONTROL
105 ~ 135% rated output power
Protection type : Constant current limiting for Vo=50 ~ 100% of rated voltage, recovers automatically after fault condition is removed
6 ~ 7V
9.4 ~ 10.9V
14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V
41.4 ~ 48.6V 57.6 ~ 67.2V
3.96 ~ 4.62V
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, recovers automatically after temperature goes down
RC+/RC- : 0 ~ 0.8V= power on ; 4 ~ 10V = power off
-40 ~ +60℃ (Refer to "Derating Curve")
WORKING TEMP.
20 ~ 90% RH non-condensing
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃ , 10 ~ 95% RH
±0.04%/℃ (0 ~ 50℃)
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note 4)
EMC EMISSION
EMC IMMUNITY
OTHERS
NOTE
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, IEC60601-1 approved
I/P-O/P:4KVAC
I/P-FG:2KVAC
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2
MIL-HDBK-217F (25℃)
MTBF
295.7K hrs min.
DIMENSION
159*97*38mm (L*W*H)
PACKING
0.38Kg; 24pcs/10.1Kg/0.76CUFT
1.
2.
3.
4.
5.
6.
7.
8.
O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃ / 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
Tolerance : includes set up tolerance, line regulation and load regulation.
The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
Derating may be needed under low input voltages. Please check the derating curve for more details.
No load power consumption<0.5W when RC+ & RC- (CN100 pin1,2) 0 ~ 8V or short.
Touch current was measured from primary input to DC output.
When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this
deviation that does not affect basic safety or essential performance.
File Name:MSP-100-SPEC 2014-02-12
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MSP-100
100W Single Output Medical Type
Mechanical Specification
Case No.901I
series
Unit:mm
152.5
1
2
3
4
5
6
7
CN100
SVR1
4.5
Terminal Pin No. Assignment
Pin No.
1
2
3
Assignment
AC/L
AC/N
24
Pin No.
Assignment
4,5
DC OUTPUT -V
6,7
DC OUTPUT +V
78
159
6.5
150
FG
10 max.
Remote ON/OFF (CN100) : JST B-XH or equivalent
Pin No.
1
2
Assignment
RCRC+
Mating Housing
Terminal
JST XHP
or equivalent
JST SXH-001T
or equivalent
6-M3(Both Sides) L=6
22
117
Block Diagram
ACTIVE
INRUSH
CURRENT
LIMITING
EMI
FILTER
I/P
RECTIFIERS
&
PFC
POWER
SWITCHING
O.T.P.
O.L.P.
RECTIFIERS
&
FILTER
DETECTION
CIRCUIT
FG
PFC
CONTROL
O.V.P.
PWM
CONTROL
Derating Curve
+V
-V
RC+
REMOTE
CONTROL
RC-
Output Derating VS Input Voltage
100
100
90
80
80
70
60
60
40
50
20
-40
40
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
60
70
(HORIZONTAL)
85
100
125
135
155
264
INPUT VOLTAGE (V) 60Hz
File Name:MSP-100-SPEC 2014-02-12