VER : A_1 update : 2015.09.11 AC -DC Medical Power Module MFC15 SERIES 15 Watts KEY FEATURES Switching Power Module for PCB Mountable Fully Encapsulated Plastic Case Universal Input Range 90-264VAC Regulated Output and Low Ripple and Noise <0.1W No Load Input Power Isolation Class II Small Size CE, CB, UL, cUL Approvals 3-Year Product Warranty ELECTRICAL SPECIFICATIONS All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated. Model No. Max Output Wattage (W) Voltage MFC15-5S 15W (Note 1) Frequency (Hz) Input Output Isolation Environment Physical MFC15-12S MFC15-24S 47-440 Hz 385 mA max. (115 VAC) / 250 mA max. (230 VAC) Inrush current (<2ms,Cold Start) 20 A max. (115 VAC) / 40 A max. (230 VAC) Leakage Current < 0.1mA / 264 VAC (Touch Current) External fuse (recommend) 3.15 A slow blow type Voltage (V.DC.) 5V Voltage Accuracy ±2% Current (mA) max Maximum Capacitive Load (at 230 VAC) Line Regulation (LL-HL) (typ.) Load Regulation (5-100%) (typ.) (Note 2) Ripple (Full load) ±0.5% ±1%. 75mV max (Vp-p) Noise (Full load) 120mV max (Vp-p) 1% of Vout 79% 84% (Note 2) 9V 12V 15V 24V 3000 1666 1250 1000 625 7000uF 5000uF 1500uF 1000uF 470uF 84% 85% 1% of Vout 80% Hold-up Time(typ) 15 ms (115VAC) / 56ms (230VAC) Over Power Protection Hiccup technique, auto-recovery Over Voltage Protection Zener diode clamp Short Circuit Protection Hiccup mode, indefinite (automatic recovery) Input-Output (V.AC) 4000V Operating Temperature -40°C…+80°C (Case Temperature max. +95°C) Storage Temperature -40°C…+90°C Temperature Coefficient ±0.05%/°C Altitude During Operation 5000m Humidity up to 95% RH MTBF >350,000 h @ 25°C (MIL-HDBK-217F) Atmospheric Pressure Dimension (L x W x H) 70 kPa to 106 kPa Case Material Weight Plastic resin (flammability to UL 94V-0) 59 g Cooling Method Free air convection http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 MFC15-15S 90-264 VAC or 120-370 VDC, “ N ” to DC “ + “ ; “L ” to DC “ - “ Current (Full load) Efficiency (at 230 VAC) Protection MFC15-9S 2.07 x 1.08 x 0.93 Inches ( 52.5 x 27.5 x 23.5 mm ) Tolerance 0.5 mm -1- VER : A_1 update : 2015.09.11 AC -DC Medical Power Module MFC15 SERIES 15 Watts ELECTRICAL SPECIFICATIONS All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated. Model No. Safety Approval Conducted and radiated EMI EMC MFC15-5S MFC15-9S MFC15-12S MFC15-15S MFC15-24S cUL / UL Standard: UL 60950-1, CAN/CSA C22.2 No. 60950-1-07 ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1 (2008), 2 x MOPP CB Standard: IEC 60950-1:2005 (2nd Edition) + Am 1:2009 + Am 2:2013 IEC 60601-1:2005 (3rd Edition) + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) or IEC 60601-1 (2012 reprint), 2 x MOPP EN55011 class B ESD EN61000-4-2 air ± 8kV , Contact ± 4Kv Radiated Immunity EN61000-4-3 10V/m Fast Transient EN61000-4-4 ± 2kV Surge EN61000-4-5 ±1kV Conducted Immunity EN61000-4-6 10Vrms PFMF EN61000-4-8 30A/m Dips EN61000-4-11 30% 10ms Interruption EN61000-4-11 >95% 5000ms NOTE 1. This product is not designed for use in critical life support systems, equipment used in hazardous environment, nuclear control systems or other such applications which necessitate specific safety and regulatory standards other the ones listed in this datasheet. 2. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor. 3. Safety approvals cover frequency 47-63 Hz. 4. That “natural convection” is about 20LFM but is not equal to still air (0 LFM). 5. It's recommended to add Varistor 14S471K at L / N input side in parallel. 6. Please refer to our PDF file "AC-DC Application" on our website: www.archcorp.com.tw MECHANICAL DIMENSION PIN# Single 1 AC IN (L) 2 AC IN (N) 3 +DC OUT 4 -DC OUT Maximum Torque 12{1.21} (kgf.cm{N.m}) http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -2- VER : A_1 update : 2015.09.11 AC -DC Medical Power Module MFC15 SERIES 15 Watts DERATING BLOCK DIAGRAM http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -3-