EFM10/M series

1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
EFM10/M
series
Universal 90-264VAC Input
High Efficiency
Improved EMI Performance
4000V Isolation
Single, Dual & Triple Outputs
UL60601-1, EN60601-1
1.66”W x 2.46”L x 0.91”H
Model Number
Output Voltage
Output Amps(max)
SINGLE OUTPUT
EFM-0100/M
EFM-0101/M
EFM-0122/M
EFM-0102/M
EFM-0103/M
EFM-0118/M
EFM-0105/M
3.3 Volts(DC)
5 Volts(DC)
9 Volts(DC)
12 Volts(DC)
15 Volts(DC)
18 Volts(DC)
24 Volts(DC)
2.5Amps
2Amps
1.1Amps
0.85Amps
0.67Amps
0.55Amps
0.42Amps
DUAL OUTPUT
EFM-0106/M
EFM-0107/M
EFM-0108/M
±12 Volts(DC)
±15 Volts(DC)
5 / 12volts(DC)
±0.42Amps
±0.34Amps
1.3/0.6Amps
TRIPLE OUTPUT
EFM-0109/M
EFM-0111/M
5 /±12 Volts(DC)
5 / ±15 Volts(DC)
1.8 / ±0.13Amps
1.8 / ±0.13Amps
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current (115/230Vin)
Leakage Current
series
GENERAL SPECIFICATIONS
90-264 VAC
47-63 Hz
20/40Amps, typ.
<500UA @ 264VAC, 50Hz
OUTPUT SPECIFICATIONS
Voltage and Current (Note 5)
See Selection Chart
Load Regulation (20%-FL)(Note 6)Singles: ± 3%, Duals: ± 3% (EFM-0108/M: 0.5/3%)
Triple: 0.2 / ±0.3%
Line Regulation, typ (Note 7)
Singles ± 0.1~1%
Duals ± 0.3~1%
(except EFM-0108/M: 0.5/ 0.5%)
Triple: 0.3 / ±0.1%
Cross Regulation (Note 10)
Duals 2% (EFM-0108/M: 0.5/3%)
Triple: NA/ ±0.1%
Preset Accuracy (Note 8)
Single/Duals 1%
(except EFM-0108/M: 2 / 5%)
Triple: 1 / ±5%
DC Voltage Adjust (typ)
± 6% of FS (Primary)
Temperature Coefficient
±0.03%/°C
Ripple/Noise, typ (Note 1, 9)
Single/Dual: 100mVpk-pk (except
EFM-0108/M: 50/50mVpk-pk)
Triple: 50 / ±75mVpk-pk
Over Voltage Protection
Clamp
Over Temp. Protection
Latching, recov after cooling and
power restored
Short Circuit Protection (Note 2) Continuous, self-recovering
Hold Up Time
30 mS, typ
DC Output Power
10 Watts, max
PHYSICAL SPECIFICATIONS
Size
Construction
Weight
EFM10/M
Isolation (Note 3)
Efficiency
Switching Frequency
Safety
EMC
4KVAC Input-Output
2500VAC Inpout-Ground
1000VAC Output-Ground
Single/Dual 75% typ
Triple 70%
100Khz, (fixed, typical)
UL60601-1, EN60601-1
EN55011
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature (Note 4)
Storage Temperature
Relative Humidity
MTBF -0 to +40°C (See Derate)
-25 to +71°C *
0 to + 95%, non-cond *
(Mil Std 217, 25°C)
575,000 Hrs
NOTES
1. Ripple and Noise depend upon output voltage as specified per particular model.
2. Short Curcuit Protection is self-recovering when overcurrent condition
is removed.
3.Isolation for up to 1 minute duration.
4. Specified for free-air convection cooling.
5. Minimum load is NOT required for proper operation. However, auxiliary outputs should be reduced as a function of primary output minimum
load or load regulation will be higher.
6. Load regulation measured from 20% to full load.
7. Line regulation measured from 90VAC to 264VAC.
8. Preset accuracy measured at nominal load, 120VAC input.
9. O/P noise measured directly at pins/terminals at nom. load, 0.1uF
bypass, pk-pk @20MHz bandwidth.
10.Cross Reg-Duals: Measured at nom. load with the other output varied
between 60% and 100% inom.
1.66” x 2.46” x 0.91”
Open Frame
1.6oz (44g)
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
EFM10/M
1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
series
OUTPUT DERATING CURVE
°C
MECHANICAL DIMENSIONS - SINGLE + DUAL OUTPUTS
Pin#
Pin#
. 91
23
1
2
.07 3
1. 8
4
.04 5
1
. 15
3. 8
1
2
3
4
5
Dual Outputs
Single Single
Outputs Dual
Dual(0108)
AC Line
AC Line
ACL
AC NeACL
ut ra l
AC NeuACL
t ra l
ACN
ACN
ACN
+Output
+Output
NC +Vout +Vout
Commo+12V
n
NC
Com
+5V
-Output
-Output
-Vout -Vout
Com
Dimensions are in inches
5
E x- 0 1 0 6
REV:A
2
1.66
42.2
.3 9
10
4
ACN
.6 9
17.5
.3 9
10
3
1
.4 4
11.1
ACL
.1 4
3.6
. 17
4 .3
2.13
54
2.46
62.6
WWW.ASTRODYNE.COM
TR
:
:
: A:
ASTRODYNE USA: 1-800-823-8082
:
ASTRODYNE PACIFIC: 886-2-26983458 :
: mm
: +/-
1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
EFM10/M
series
MECHANICAL DIMENSIONS - TRIPLE OUTPUTS
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458