ASM15 series

1 5 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
ASM15
series
Universal 90-264VAC Input
High Efficiency
Regulated Outputs
4000VAC Isolation
Single Output from 3.3 to 24 VDC
3”W x 2.18”L x 0.93”H
Model Number
Output Voltage
Output Amps (max)
(pk)
ASM-1500
ASM-1501
ASM-1502
ASM-1503
ASM-1505
3.3 VDC
5 VDC
12 VDC
15VDC
24VDC
3.5A
3A
1.25A
1A
0.63A
3.8A
3.6A
1.5A
1.25A
0.75A
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
ASM15
1 5 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current, typ:
Leakage Current
PHYSICAL SPECIFICATIONS
90-264 VAC
47-63 Hz
20A @ 100VAC Input *
40A @ 200VAC Input *
<0.3mA @ 264VAC, 50Hz
OUTPUT SPECIFICATIONS
Voltage and Current
Load Regulation Line Regulation
Preset Accuracy
DC Voltage Adjust (typ)
Temperature Coefficient
Ripple/Noise
Over Voltage Protection
Over Temp. Protection
Short Circuit Protection
Hold Up Time
series
See Selection Chart
+/- 3% (20% to 100% Load)
Singles +/- 0.1~1%, typ
1%
+/- 6% of FS (Primary)
+/-0.03%/°C
100mV Pk-Pk, typ Clamp, 130-150% *
Latching, Cool and Repower
Continuous, self-recovering
20 mS, typ
Size
Construction
Weight
Connectors: Molex P/N:
3” x 2.18” x 0.93”
Open Frame
2.5oz (70g)
50-37-5043 (Input)
50-37-5043 (Output)
Contact: 08-70-1040
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature
Storage Temperature
Relative Humidity
EMC
MTBF
-10 to +50°C(FL)
-25 to +71°C *
0 to + 95%, non-cond *
EN55011 Class A 180,000 Hrs
Mil Std 217, 25°C
OUTPUT DERATING CURVE
100
90
80
GENERAL SPECIFICATIONS
I/P-O/P: 4000VAC
I/P-Ground: 2500VAC
O/P-Ground: 1000VAC
75%~78%, typ
100Khz, (fixed, typical)
ANSI/AAMI ES60601-1 3rd ed.
CSA C22.2 No. 60601-1 3rd ed.
EN60601-1 3rd ed.
IEC60601-1 3rd ed.
EN60601-1-2
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Load (%)
Isolation
Efficiency
Switching Frequency
Safety UL/cUL: UL-EU: CB: CE: 70
60
50
40
30
20
10
0
0
10
20
30
40
50
60
70
Ambient Temperature °C
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
A S M 1 5 series
15W
U l t r a m i n i a t ure Medical Grade Switching Power Supplies

MECHANICAL DIMENSIONS
����������������������������
���
����
0.93
����
max
����
2.18
����
max
 ­€‚ƒ



€‚­€‚ƒ




  WWW.ASTRODYNE.COM

ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458