DC SIGNAL SOURCE SS7012

DC SIGNAL SOURCE SS7012
Signal Source & Calibrator
Improved stabilit y and lower
calibration costs
Conduct loop testing of
instrumentation systems (4-20 mA)
Validate thermostatic devices and
distributors
Use as a calibrator and low-power
generator for production facilities
maintenance and electronic device
evaluation
DC Signal Source Supplies and
Measures Signals Simultaneously
The HIOKI DC SIGNAL SOURCE SS7012 can be used not only for instrumentation system (4-20 mA) maintenance, but also for maintaining and servicing
thermostatic devices, distributors, and industrial meters, as well as for evaluating
electronic circuits and devices, and for production facility maintenance.
The signal source also supplies thermoelectromotive force corresponding to eight
types of thermocouples. Furthermore, this compact DC signal source simultaneously sources and measures signals.
2
Use as A Hand-Held Field Calibrator for Loop
Testing of Instrumentation Systems, and for
Production Facility Maintenance
SS7012 Features
Example applications
• Simultaneously sources up to ±25 V and ±25 mA
while measuring up to ±28 V and ±28 mA (DC)
• Eight thermoelectromotive force sourcing (TC)
modes (K, E, J, T, R, S, B, and N)
• TEMP measurement from -25 to 80°C
• Improved stability and lower calibration costs
• Control from a PC via USB connection
• Instrumentation system loop testing (4-20 mA)
• Validate thermostatic devices and distributors
• Use as a calibrator and low-power generator for production facilities maintenance and
electronic device evaluation
• Use as an electronic load (max. 25 mA DC)
Loop test (two-wire transmitter sensor)
Verifies proper sensor output (two-wire transmitter sensors). The
SS7012 applies distributor power to the sensor as a constantvoltage (CV) output, and monitors current output from the sensor.
Sensor
Two-Wire
Transmitter
Loop test (distributor)
Test distributor operation. Apply the default current from the
SS7012 as a substitute for sensor current output. Set SS7012 polarity to negative (to operate as a sink).
SS7012
4-20 mA
SS7012
24 V DC Source
(using the Monitor function)
4-20 mA Current Source
4-20 mA
Distributor
Output Device
Store up to 20 steps in memory
Thermometer calibration (TC, CV, and CC source)
Substitute the output of sensors such as TC (thermocouple) voltage, current and resistance with TC, CV, CC and Rs (100 Ω) test
signals from the SS7012.
Thermometer
Output
SS7012
Copper Wire
or similar
Up to 20 levels can be stored for each source function, which can
be recalled manually or in sequence automatically (auto scan).
Use this function for meter checking, repetitive adjustments, and
linearity testing.
TC (RJ) Source
Place Temperature
Probe near thermometer terminals
Recall or Auto Scan Output
Temperature measurement
Use the optional 9184 Temperature Probe to measure temperatures from -25 to 80°C. As well as measuring ambient temperature, you can calibrate thermoelectric thermometers with reference junction compensation TC (RJ).
Time
Portable, precise and stable
Despite the small size and light weight, the high precision and
stability of ±0.03% for CV and CC sourcing, and ±0.05% ±0.5°C*
for TC (0°C) sourcing, have one-year guaranteed accuracy. The
compact hand-held design is convenient for travel, field calibration, and servicing applications.
*
For TC (0°C) sourcing of K, E, J, and T thermoelectromotive force .
3
Use as A Signal Source for Production Facility
Maintenance, plus Electronic Device Testing and
Evaluation.
Simultaneous voltage and current
sourcing and display
Source up to ±25.000 V and ±25.000 mA while measuring up to
±28.000 V and ±28.000 mA. Isolated source and measurement
circuits can increase operating efficiency for applications such
as simultaneous input and output testing of transducers using a
single instrument.
Small source applications
Use as a portable signal source for electronic device testing and
evaluation. Full external control is available via the USB interface.*
*
Only for communicating with a PC. Accessing USB memory is not supported.
Electronic Circuit or Device
Output
USB
Cable
Input
SS7012
SS7012 Sample PC Application Screen
The SS7012 PC application is available at no cost from the HIOKI
website.
Easy key operations
Intuitive operation is ensured by an up/down key for each setting digit, allowing quick output setup.
Source display
• Bar indicates selected function
• Display of thermocouple setting for TC
function
• Set level and on/off output state
Measurement display
• Bar indicates selected function
• Measured value or output value monitor
• Display of address and interval for
memory setting and output
Operation keys
[SOURCE FUNCTION]
Source function selection key
[TC TYPE] Emulated thermocouple type
selection key
[MEMORY] Memory source setting key
[SET] Source level setting keys (for each digit)
[ +/– ] Source polarity selection key
[OUTPUT] Source output on/off key
[MEASURE FUNCTION]
Measurement function selection key
[0 ADJ] Measurement zero adjustment key
[MONITOR] Measurement/Monitor value
selection key
SS7012 Specifications
SS7012 General Specifications
(At 23 ±5°C, accuracy guaranteed for one year)
Sourcing
Sourcing &
Measurement Range
Resolution
Accuracy
Constant Voltage (CV)
2.5V: 0 to ±2.5000 V
100 µV
25V: 0 to ±25.000 V
1 mV
(Sink/Source)
±0.03% of setting ±300 µV ±25 mA
±0.03% of setting ±3 mV ±25 mA
Constant Current (CC)
25mA: 0 to ±25.000 mA
±0.03% of setting ±3 µA
Thermoelectromotive
Force TC (0°C)
K:-174.0 to 1372.0 °C
E:-220.0 to 839.0 °C
J: -208.0 to 1108.0 °C
T: -169.0 to 400.0 °C
R: -50 to 100 °C
101 to 1768 °C
S: -50 to 100 °C
101 to 1768 °C
B: 300 to 600 °C
601 to 1820 °C
N:-113.0 to 1300.0°C
1 µA
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±2.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±0.5 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
1 °C
1 °C
1 °C
0.1 °C
Thermoelectromotive Force TC (RJ)
Thermocouple types, source range
and resolution are the same as for
TC (0°C)
Measurement
(Sink/Source)
Approx. ±25 V
Output
Resistance
5Ω or less
IEC 584
Add the following to
Using
TEMPERATURE
TC (0°C) accuracy
±0.5 °C (@23 ±5 °C)
PROBE 9184
±1.0 °C (from 0 to 50 °C)
±2.0 °C (from -25 to 80 °C)
Memory Sourcing (RECALL, SCAN) Max. steps: 20
One type for each function: CV2.5, Interval: 1 to 99 s
CV25, CC, TC (0°C and RJ)
Monitor
Remarks
Timing
accuracy is not
specified
Standard Resistance (Rs) 100Ω
±0.2 Ω
±5 mΩ/°C
Voltage
2.5V: 0 to ±2.8000 V
25V: 0 to ±28.000 V
100 µV
1 mV
(after zero adjustment)
±0.03% of rdg. ±300 µV
±0.03% of rdg. ±3 mV
Input Resistance
Approx. 1 MΩ
1 µA
(after zero adjustment)
±0.03% of rdg. ±3 µA
Input Resistance
Approx. 30 Ω* or less
Current
25mA: 0 to ±28.000 mA
Temperature
-25 to 80 °C
Load current monitor
Terminal voltage monitor
Reference contact
temperature monitor
10 µA
10 mV
0.1 °C
AC ADAPTER
9445 - 02
±0.5 °C (@23 ±5 °C)
Using
±1.0 °C (from 0 to 50 °C) TEMPERATURE
±2.0 °C (from -25 to 80 °C) PROBE 9184
0.1 °C
±0.1% of rdg. ±20 µA
0 to ±28.00 mA
±0.1% of rdg. ±20 mV
0 to ±28.00 V
±0.5 °C (@23 ±5 °C)
-25 to 80 °C
±1.0 °C (from 0 to 50 °C)
±2.0 °C (from -25 to 80 °C)
* Including approximately 15 Ω fuse resistance
CARRYING CASE
9782
CARRYING CASE
9380
COMMUNICATION
PACKAGE SS9000
Source output method: Bipolar sink/source
A/D conversion method: Delta-Sigma
Temperature coefficients (sourcing):
±(0.005% of setting + 30 µV) / °C (CV: 2.5V)
±(0.005% of setting + 0.2 mV) / °C (CV: 25V)
±(0.005% of setting + 0.2 µA) / °C (CC: 25mA)
±(0.005% of setting + 0.05 °C) / °C (TC(°C): K, E, J, T, and N)
±(0.005% of setting + 0.1 °C) / °C (TC(°C): R, S, and B)
±(0.01% of setting + 0.1 °C) / °C (TC(RJ): K, E, J, T, R, S, B, and N)
±0.005 Ω / °C (Rs)
Temperature coefficients (measuring):
±(0.005% rdg. + 10 µV) / °C (V: 2.5V)
±(0.005% rdg. + 0.1 mV) / °C (V: 25V)
±(0.005% rdg. + 0.1 µA) / °C (A: 25mA)
±0.05 °C / °C (TEMP)
±(0.01% rdg. + 1 µA) / °C (MONITOR CV: 2.5V and 25V)
±(0.01% rdg. + 1 mV) / °C (MONITOR CC: 25mA)
±0.05 °C / °C (MONITOR TC (RJ))
Zero adjustment function: within ±100 counts (V, A, and measuring)
Display: Five-digit LCD (“28000”)
Operating temperature and humidity:
0 to 40 °C, 80% RH or less (non-condensing)
I/O protection: 50 mA, fused (sourcing and measuring)
Withstand voltage: 370 Vrms AC @1 mA for 15 seconds
(source-to-measurement terminals)
Power supply: Four LR6 alkaline batteries or HR6 Ni-MH
batteries, or 9445 -02 / -03 AC ADAPTER
Continuous operating time: (w/fully charged 2500-mAh Ni-MH batteries)
Approx. 200 min (CC: 25mA range, sourcing 20 mA into 250 Ω load)
Approx. 170 min (CV: 25V range, sourcing 25 V into 1 kΩ load)
Approx. 280 min (V: 2.5V range, input terminals shorted)
Size and weight: 104W × 180H × 58D mm, 570 g
Accessories: One INPUT CORD 9168 set, one TEST LEAD
9170 -10 set, one spare fused (F50 mA, 250 V “mizet”
fuse), four LR6 alkaline batteries
Ordering information
DC SIGNAL SOURCE SS7012
Recommended combination examples
SS7012 (main unit) + AC ADAPTER 9445 -02 / -03 +
TEMPERATURE PROBE 9184 + CARRYING CASE 9782
Options
AC ADAPTER 9445 -02
(For Japan, US, and Canada. 100V to 240V AC, 9V DC / 1A output)
AC ADAPTER 9445 -03
(For EU. 100V to 240V AC, 9V DC / 1A output)
TEMPERATURE PROBE 9184
(reference junction compensation)
CARRYING CASE 9380 (Holds main unit only, soft case)
CARRYING CASE 9782 (Holds options, hard case)
COMMUNICATION PACKAGE SS9000
(Includes USB cable and driver software)
INPUT CORD
9168 (Provided)
TEST LEAD
9170-10 (Provided)
TEMPERATURE
PROBE 9184
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1608-1610 Shanghai Times Square Office, 93 Huai Hai Zhong
Road, Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: [email protected]
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: [email protected]
Commercially available rechargeable batteries (AA Ni-MH
batteries x 4) may also be used to power the SS7012. Using
locally purchased rechargeable batteries and dedicated
battery chargers is recommended; however, HIOKI will not
be able to guarantee operating time as different rechargeable
batteries exhibit different power specifications per charge.
The SS7012 cannot be used to recharge batteries.
DISTRIBUTED BY
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: [email protected]
Guangzhou Office :
Room A-3206, Victory PlazaServices Center, No.103,
Tiyuxi Road, Guangzhou, P.R.China POSTCODE:510620
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: [email protected]
All information correct as of Jan. 5, 2009. All specifications are subject to change without notice.
SS7012E2-01B-03K
Printed in Japan