User's Manual Heavy Duty Pressure Meter Model 407495 Introduction Congratulations on the purchase of the Extech Heavy Duty Pressure Meter, Model 407495. This meter measures and displays pressure in eight (8) measurement units. Up to three (3) interchangeable transducers (sold separately) in varying ranges can be used with this meter. Special features include RS-232 PC Interface, Min/Max Memory, push-button Zero, Data Hold, and Auto Power-Off. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Meter Description 1. Transducer cable (shown connected to meter input jack) 1 2 2. RS-232 Interface Output Jack 3 4 3. LCD Display 4. Protective rubber holster 5. Keypad 5 6. Battery compartment on rear of meter 7. Transducer shown attached to meter cable 6 7 2 407495-EN v2.3 10/13 Preparation for Use Measurement Procedure Overview The following list provides an overview of pressure measurement steps. Each step is discussed in subsequent paragraphs. Note that the first 4 steps should only be performed the first time the meter is used and each time the transducer is changed. 1. Select a Transducer model that covers the desired range 2. Plug the Transducer into the meter's input jack 3. Program the Transducer range 4. Select the measurement units 5. Zero the meter 6. Connect the Transducer to the process under test 7. Apply pressure to the process under test 8. Read pressure reading on the meter's LCD Selecting Transducer Range 1. With the meter turned OFF, simultaneously press and hold the HOLD, SENSOR, & RECORD buttons down. 2. While holding the three buttons down, press the POWER key momentarily. Release the 3 buttons when the LCD display switches on. 3. A number will appear on the LCD that represents the Transducer type. Select the Transducer type by pressing the SENSOR TYPE key until the desired transducer number appears. Select the transducer from the list below in step 5. 4. Press the ZERO button to store the new transducer type. The meter will memorize the new type even when the meter is turned OFF. 5. Select the transducer from the following list: Select 2 for the 30 psi (2 bars) transducer Select 10 for the 150 psi (10 bars) transducer Select 20 for the 300 psi (20 bars) transducer NOTE: The meter does not need to be recalibrated each time the transducer is changed. Connecting the Transducer to the Meter 1. Press the POWER button to turn the meter OFF. 2. Remove the existing transducer, if necessary, by unplugging it from the meter’s input jack (top of meter). 3. Plug the new transducer into the meter’s input jack. Selecting the Unit of Measure Press the UNIT key to select the desired unit of measure: Bar, psi, Kg/cm2, mmHg, inHg, mH2O, inch H2O, or Atmospheres (ATP). 3 407495-EN v2.3 10/13 Operation Push-button Zero 1. Attach a transducer to the meter, turn the meter ON, and select the range that matches the attached transducer as described in the 'Preparation for Use' section. 2. With the transducer attached to the meter (but not attached to the process) press the ZERO button. Note that the ZERO button should always be pressed after swapping transducers. 3. If the display reads zero after pressing the ZERO button proceed to the next section. If the display does not zero, the transducer may be defective. Connect Transducer Port to Process 1. Ensure that the process under test is not pressurized 2. Connect the Transducer's ¼” PS, 19 teeth per inch threaded port to the process under test. 3. Apply pressure to the process. 4. Read the pressure on the meter's LCD. Data Hold 1. When a reading is on the display that you wish to hold, press the Data Hold button. 2. The reading will freeze and the HOLD icon will appear on the display. 3. Press the Data Hold button again to release the reading and return the meter to normal operation. Minimum/Maximum Reading Memory 1. Press the RECORD/RECALL button to start tracking the highest (MAX) and lowest (MIN) readings. The REC indicator will appear on the LCD. 2. Take pressure measurements. 3. When pressure measurements are complete, press the RECORD/RECALL button to recall the Maximum reading recorded (the MAX indicator will appear on the LCD). 4. Press the RECORD/RECALL button again to recall the Minimum reading recorded (the MIN indicator will appear on the LCD). 5. To exit this mode and return the meter to normal operation, press and hold the RECORD/RECALL button until the REC indicator goes out. Auto Power-Off In order to conserve battery life, the meter automatically shuts off after 10 minutes. To defeat this feature put the meter in the Record mode by pressing the RECORD/RECALL button; the REC indicator will appear. As long as the REC icon is displayed, the meter will stay on indefinitely. RS-232 PC Interface This meter is equipped with an RS-232 PC Interface output port (meter top). A Data Acquisition kit that includes WindowsTM compatible software and a meter-to-PC interface cable can be obtained from Extech for use with this meter. The Part Number for the kit is 407001A. Instructions for use are supplied with the kit. Check the software download page of the website www.extech.com for the latest version of the PC software and its operating system compatibility. 4 407495-EN v2.3 10/13 Specifications General Specifications Display Multi-function LCD Transducers 30, 150 and 300 max. psi (sold separately) Transducer signal Full scale: 100mVDC Display units Bar, psi, Kg/cm2, mmHg, inches Hg, meters H2O, inches H2O, Atmosphere Zero adjust Front panel push-button Data Hold Displayed reading can be held by front panel push-button Min/Max Memory Lowest/Highest readings can be stored for later recall PC Interface kit Built-in RS-232 port for use with optional WindowsTM data acquisition software Auto Power-Off To conserve battery life, the meter turns OFF automatically after approx. 10 minutes (programmable over-ride). Sampling time 0.8 seconds approx. Power supply 9V Battery Low battery indication Rectangular battery-life box appears on the left side of LCD. Power consumption Approx. 7mA Operating Temperature 32 to 122oF (0 to 50oC) Operating Humidity Less than 80% RH Dimensions 7.1 x 2.8 x 1.3" (180 x 72 x 32mm) Weight Meter only: 0.41 lbs. (187 g) Meter and Transducer: 0.81 lbs. (370 g) Accessories Protective rubber holster and 9V battery 5 407495-EN v2.3 10/13 Transducer Range, Resolution, and Accuracy Specifications psi bar Kg/cm2 mm Hg in Hg m H2O in H2O atm 30 2 2.040 1500 59.05 20.4 802 1.974 Resolution 0.02 0.002 0.002 2 0.05 0.02 1 0.002 Range 150 10 10.19 7500 295.2 101.9 4010 9.87 Resolution 0.2 0.01 0.01 10 0.2 0.1 5 0.01 Range 300 20 20.40 15000 590.5 204.0 8020 19.74 Resolution 0.2 0.02 0.02 20 0.5 0.2 10 0.02 Range Accuracy: Meter only: ± (0.5% reading + 1 digit) Transducers only: 2bar ± (2% + 0.02bar) 10bar ± (2% + 0.1bar) 20bar ± (2% + 0.2bar) Combined: ± (2.5% full scale + 1 digit) Accuracy note: Specified accuracy applies to meter at 23oC (±5oC) Calibration note: Re-calibration is not required each time a new transducer is attached. Transducer Specifications Model Maximum Capacity PT30 30psi (2 BAR) PT150 150psi (10BAR) PT300 300psi (20 BAR) Overload protection Diaphragm Supply voltage Thread Accuracy Operating Temperature Operating Humidity Output Size Weight Cable length 150% of max capacity Ceramic 5V ¼” PS, 19 teeth per inch Transducers only: 2bar ± (2% + 0.02bar) 10bar ± (2% + 0.1bar) 20bar ± (2% + 0.2bar) 32ºF to 140ºF (0ºC to 40ºC) Max 80% RH Full capacity output = 100 mV DC 1.2” diameter x 4.0” (30mm diameter x 102mm) 6.46 oz. (183g) 39” (1m) 6 407495-EN v2.3 10/13 Warranty FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transi t. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration, Repair, and Customer Care Services FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. NIST certification for most products is also provided. Call the Customer Service Department for information on calibration services available for this product. Annual calibrations should be performed to verify meter performance and accuracy. Technical support and general customer service is also provided, refer to the contact information provided below. Support Lines: U.S. (877) 439-8324; International: +1 (603) 324-7800 Technical Support: Option 3; E-mail: [email protected] Repair & Returns: Option 4; E-mail: [email protected] Product specifications are subject to change without notice Please visit our website for the most up-to-date information www.extech.com FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA ISO 9001 Certified Copyright © 2013 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com 7 407495-EN v2.3 10/13