English User`s Manual

User's Manual
Heavy Duty Pressure Meter
Model 407495
Introduction
Congratulations on the purchase of the Extech Heavy Duty Pressure Meter, Model 407495. This
meter measures and displays pressure in eight (8) measurement units. Up to three (3)
interchangeable transducers (sold separately) in varying ranges can be used with this meter. Special
features include RS-232 PC Interface, Min/Max Memory, push-button Zero, Data Hold, and Auto
Power-Off. This meter is shipped fully tested and calibrated and, with proper use, will provide years of
reliable service.
Meter Description
1. Transducer cable (shown connected to meter
input jack)
1
2
2. RS-232 Interface Output Jack
3
4
3. LCD Display
4. Protective rubber holster
5. Keypad
5
6. Battery compartment on rear of meter
7. Transducer shown attached to meter cable
6
7
2
407495-EN v2.3 10/13
Preparation for Use
Measurement Procedure Overview
The following list provides an overview of pressure measurement steps. Each step is discussed in
subsequent paragraphs. Note that the first 4 steps should only be performed the first time the meter
is used and each time the transducer is changed.
1. Select a Transducer model that covers the desired range
2. Plug the Transducer into the meter's input jack
3. Program the Transducer range
4. Select the measurement units
5. Zero the meter
6. Connect the Transducer to the process under test
7. Apply pressure to the process under test
8. Read pressure reading on the meter's LCD
Selecting Transducer Range
1. With the meter turned OFF, simultaneously press and hold the HOLD, SENSOR, & RECORD
buttons down.
2. While holding the three buttons down, press the POWER key momentarily. Release the 3
buttons when the LCD display switches on.
3. A number will appear on the LCD that represents the Transducer type. Select the Transducer
type by pressing the SENSOR TYPE key until the desired transducer number appears. Select
the transducer from the list below in step 5.
4. Press the ZERO button to store the new transducer type. The meter will memorize the new type
even when the meter is turned OFF.
5. Select the transducer from the following list:

Select 2 for the 30 psi (2 bars) transducer

Select 10 for the 150 psi (10 bars) transducer

Select 20 for the 300 psi (20 bars) transducer
NOTE: The meter does not need to be recalibrated each time the transducer is changed.
Connecting the Transducer to the Meter
1. Press the POWER button to turn the meter OFF.
2. Remove the existing transducer, if necessary, by unplugging it from the meter’s input jack (top
of meter).
3. Plug the new transducer into the meter’s input jack.
Selecting the Unit of Measure
Press the UNIT key to select the desired unit of measure: Bar, psi, Kg/cm2, mmHg, inHg,
mH2O, inch H2O, or Atmospheres (ATP).
3
407495-EN v2.3 10/13
Operation
Push-button Zero
1. Attach a transducer to the meter, turn the meter ON, and select the range that matches the
attached transducer as described in the 'Preparation for Use' section.
2. With the transducer attached to the meter (but not attached to the process) press the ZERO
button. Note that the ZERO button should always be pressed after swapping transducers.
3. If the display reads zero after pressing the ZERO button proceed to the next section. If the
display does not zero, the transducer may be defective.
Connect Transducer Port to Process
1. Ensure that the process under test is not pressurized
2. Connect the Transducer's ¼” PS, 19 teeth per inch threaded port to the process under test.
3. Apply pressure to the process.
4. Read the pressure on the meter's LCD.
Data Hold
1. When a reading is on the display that you wish to hold, press the Data Hold button.
2. The reading will freeze and the HOLD icon will appear on the display.
3. Press the Data Hold button again to release the reading and return the meter to normal
operation.
Minimum/Maximum Reading Memory
1. Press the RECORD/RECALL button to start tracking the highest (MAX) and lowest (MIN)
readings. The REC indicator will appear on the LCD.
2. Take pressure measurements.
3. When pressure measurements are complete, press the RECORD/RECALL button to recall the
Maximum reading recorded (the MAX indicator will appear on the LCD).
4. Press the RECORD/RECALL button again to recall the Minimum reading recorded (the MIN
indicator will appear on the LCD).
5. To exit this mode and return the meter to normal operation, press and hold the
RECORD/RECALL button until the REC indicator goes out.
Auto Power-Off
In order to conserve battery life, the meter automatically shuts off after 10 minutes. To defeat this
feature put the meter in the Record mode by pressing the RECORD/RECALL button; the REC
indicator will appear. As long as the REC icon is displayed, the meter will stay on indefinitely.
RS-232 PC Interface
This meter is equipped with an RS-232 PC Interface output port (meter top). A Data Acquisition kit
that includes WindowsTM compatible software and a meter-to-PC interface cable can be obtained
from Extech for use with this meter. The Part Number for the kit is 407001A. Instructions for use are
supplied with the kit.
Check the software download page of the website www.extech.com for the latest version of the PC
software and its operating system compatibility.
4
407495-EN v2.3 10/13
Specifications
General Specifications
Display
Multi-function LCD
Transducers
30, 150 and 300 max. psi (sold separately)
Transducer signal
Full scale: 100mVDC
Display units
Bar, psi, Kg/cm2, mmHg, inches Hg, meters H2O, inches H2O,
Atmosphere
Zero adjust
Front panel push-button
Data Hold
Displayed reading can be held by front panel push-button
Min/Max Memory
Lowest/Highest readings can be stored for later recall
PC Interface kit
Built-in RS-232 port for use with optional WindowsTM data acquisition
software
Auto Power-Off
To conserve battery life, the meter turns OFF automatically after
approx. 10 minutes (programmable over-ride).
Sampling time
0.8 seconds approx.
Power supply
9V Battery
Low battery indication
Rectangular battery-life box appears on the left side of LCD.
Power consumption
Approx. 7mA
Operating Temperature
32 to 122oF (0 to 50oC)
Operating Humidity
Less than 80% RH
Dimensions
7.1 x 2.8 x 1.3" (180 x 72 x 32mm)
Weight
Meter only: 0.41 lbs. (187 g)
Meter and Transducer: 0.81 lbs. (370 g)
Accessories
Protective rubber holster and 9V battery
5
407495-EN v2.3 10/13
Transducer Range, Resolution, and Accuracy Specifications
psi
bar
Kg/cm2
mm Hg
in Hg
m H2O
in H2O
atm
30
2
2.040
1500
59.05
20.4
802
1.974
Resolution
0.02
0.002
0.002
2
0.05
0.02
1
0.002
Range
150
10
10.19
7500
295.2
101.9
4010
9.87
Resolution
0.2
0.01
0.01
10
0.2
0.1
5
0.01
Range
300
20
20.40
15000
590.5
204.0
8020
19.74
Resolution
0.2
0.02
0.02
20
0.5
0.2
10
0.02
Range
Accuracy:
Meter only: ± (0.5% reading + 1 digit)
Transducers only:
2bar ± (2% + 0.02bar)
10bar ± (2% + 0.1bar)
20bar ± (2% + 0.2bar)
Combined: ± (2.5% full scale + 1 digit)
Accuracy note:
Specified accuracy applies to meter at 23oC (±5oC)
Calibration note: Re-calibration is not required each time a new transducer is attached.
Transducer Specifications
Model
Maximum Capacity
PT30
30psi (2 BAR)
PT150
150psi (10BAR)
PT300
300psi (20 BAR)
Overload protection
Diaphragm
Supply voltage
Thread
Accuracy
Operating Temperature
Operating Humidity
Output
Size
Weight
Cable length
150% of max capacity
Ceramic
5V
¼” PS, 19 teeth per inch
Transducers only:
2bar ± (2% + 0.02bar)
10bar ± (2% + 0.1bar)
20bar ± (2% + 0.2bar)
32ºF to 140ºF (0ºC to 40ºC)
Max 80% RH
Full capacity output = 100 mV DC
1.2” diameter x 4.0” (30mm diameter x 102mm)
6.46 oz. (183g)
39” (1m)
6
407495-EN v2.3 10/13
Warranty
FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for one year
from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the
instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit
the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is
returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transi t.
This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied
warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or
consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is
inclusive and no other warranty, whether written or oral, is expressed or implied.
Calibration, Repair, and Customer Care Services
FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we
sell. NIST certification for most products is also provided. Call the Customer Service Department for
information on calibration services available for this product. Annual calibrations should be
performed to verify meter performance and accuracy. Technical support and general customer
service is also provided, refer to the contact information provided below.
Support Lines: U.S. (877) 439-8324; International: +1 (603) 324-7800
Technical Support: Option 3; E-mail: [email protected]
Repair & Returns: Option 4; E-mail: [email protected]
Product specifications are subject to change without notice
Please visit our website for the most up-to-date information
www.extech.com