Oyster pH - Conductivity - TDS - Meters

User's Guide
Oyster pH - Conductivity - TDS - Meters
Models 341350 and 341350-P
INTRODUCTION
Congratulations on your purchase of the Extech's Oyster Series meter or meter kit. These
devices measure pH, Conductivity and TDS. These meters are intended for routine laboratory
and field testing. Accurate measurements are provided in a battery operated, portable meter
with a hinged display that can be adjusted to any viewing angle. Model 341350 includes a
glass conductivity electrode and model 341350-P includes a Polymer conductivity electrode.
METER DESCRIPTION
1.
LCD display
2.
Battery compartment
3.
Conductivity cal pot
4.
pH temperature adjust
5.
pH CAL and SLOPE adjustments
6.
Function switch
7.
Electrode connectors
8.
ON/OFF switch
9.
Lanyard attachment pins
10. pH electrode
11. Glass Conductivity cell
12. Polymer Conductivity cell
OPERATING INSTRUCTIONS
pH Calibration
1.
Calibrate the meter and electrode before each use to ensure accuracy.
2.
Connect the pH electrode via the BNC connector to the meter.
3.
Open the meter to the fully extended position. Open the battery cover and install a 9V
battery. Replace the cover and turn the unit ON. The LCD indication should appear.
4.
Slide the SELECT switch to pH
5.
Adjust the TEMP C knob to approximate the solution temperature.
6.
Remove the wetting cap from the tip of the electrode. Rinse in de-ionized water, blot
dry and place the electrode in a 7.00 pH buffer solution. Allow to stand for
approximately one minute or until the display stabilizes.
o
7.
Adjust the CALIB knob until the display reads 7.00.
8.
Remove the electrode from the buffer and rinse it in distilled water and blot dry. Place
the electrode in a 4 or 10 pH buffer solution and allow to stand for approximately 1
minute or until the display stabilizes.
9.
Adjust the SLOPE knob until the display reads 4.00 or 10.00 (as required).
10. Repeat steps 6 through 9 until no further adjustment is necessary.
Conductivity / TDS Calibration
The conductivity cell calibration should be performed frequently to ensure accuracy.
1.
Connect the conductivity cell to the input socket on the side of the meter.
2.
Prepare a conductivity solution of approximately the same value as the solution to be
0
measured. The solution should be prepared at 25 C.
3.
Set the meter to the uS/cm or TDS mode.
4.
Place the cell in calibration solution and allow the reading to stabilize.
5.
Adjust the Cell knob until the display reads the value of the calibration solution.
Example: For a 12560uS/cm conductivity solution the display will read 1256. The
user must multiply the reading x 10.
2
341350 V2.0 10/06
pH Measurement
1.
Turn the meter ON and set the function slide switch to the “pH” position.
2.
Connect the pH electrode to the side BNC connector.
3.
Calibrate the electrode if required.
4.
Adjust the TEMP °C dial to the temperature of the sample in degrees C.
5.
Place the electrode in the sample solution and read the pH value on the LCD.
6.
When all measurements have been taken, unplug the electrode and rinse in distilled
water, shake and store in the wetting cap or a pH 4 buffer solution.
Notes on pH measurements and electrodes
1.
The 60120B pH electrode is recessed in the polymer body allowing it to rest against
the bottom of a beaker without damage the glass bulb.
2.
The Electrode should stored in its wetting cap until used. Use a pH 4 buffer solution or
tap water as the storage medium.
3.
If bubbles are seen in the bulb area, hold the electrode by its cap and shake
downwards until bubbles are removed.
4.
To improve speed of response, vigorously stir the electrode in the sample, buffer, or
rinse solution.
5.
After exposure to a sample, buffer, or rinse solution, shake the electrode with a snap
motion to remove residual drops of solution.
6.
Use part of the next sample/buffer to be measured as a rinse solution, when possible.
7.
When calibrating, use a buffer as close as possible to the expected pH of the sample.
8.
Keep buffers and samples at the same temperature to avoid temperature effects.
9.
pH readings stabilize faster in some solutions than others; allow time to stabilize.
10. Electrodes deteriorate over time. If accuracy falls to 10% the electrode should be
cleaned or replaced. If no improvement is observed, replace the electrode.
Conductivity/TDS Measurements (Multiply readings by 10)
1.
Turn the meter ON and set the function slide switch to the “uS/cm” or ”TDS” position
2.
Place the electrode in the sample solution.
3.
Multiply the reading in the LCD display by 10.
4.
When finished, unplug the electrode and rinse in distilled water.
Notes on the Conductivity/TDS Cell
1.
Cell Storage: On sheathed cells, replace the sheath over the cell when storing. For
non-sheathed versions, soak the cell tip in de-ionized water for storage.
2.
Cell Cleaning: After each use, the cell tip should be rinsed with de-ionized water. If
solids build up inside the cell carefully remove with a cotton swab soaked in solvent
taking care not to touch the metal parts of the inner cell.
3
341350 V2.0 10/06
pH Troubleshooting Chart
Symptom
Long response
time or reading
drift
Cause
Clogged Junction
Recommended Solution
o
Soak in 4.07 M KCL @ 60 C for 30
minutes
Strong Alkaline measurement
Soak in 0.1 M HCL overnight
Deteriorated Gel layer
Protein coating on electrode
surface
Soak in 1 gm Pepsin dissolved in
100ml of 0.1 m HCL for 30 minutes or
as needed
Oil, paint, dyes, suspended solids Rinse electrode alternately with
on sensor
materials solvent then buffer 7.00
Organic solvents are coating
sensor
Dehydrated membrane
Organic mole fraction has to be less
than 50% to assure reasonable
readings. Limit time of measurement.
Keep probe in the 7.0 buffer between
readings.
Dry Bulb
Long term storage without wetting Soak electrode tip in wetting cap filled
with 1ml 7.00 buffer for 24 to 48 hours
Static Charge
Wiping electrodes
Rinse electrode in 7.0 buffer and blot.
Do no wipe electrode.
Same readings
in different
buffers and
samples
Cracked or broken bulb
Replace electrode. Use bulb guard.
Avoid plunging electrode to bottom of
container and spinning bars. A
wetting cap will protect bulb between
measurements.
Erratic LCD
display
Samples have low ionic strength
(lacks salt); e.g. distilled, deionized, boiled, lake water (high
pressure)
For each 50 ml of sample add 1 drop
(50uL) of SAT.KCL No alteration in
pH will occur by inert KCL.
Temperature conversion table
°C to °F Conversion
°C
°F
0
32
20
68
25
77
30
86
40
104
50
122
60
140
70
158
80
176
90
194
100
212
4
341350 V2.0 10/06
SPECIFICATIONS
Range (resolution)
Accuracy
± 0.5%
Conductivity
10 - 19,990µS (10µS)
pH
0.00 to 14.00pH (0.01pH)
± 0.02pH
TDS
10 - 19,990ppm (10ppm)
± 0.5%
o
Auto/Manual Temperature Manual for pH/Auto for conductivity/TDS; 0-100 C;
o
Compensation
Correction: 2% per C
Power
9V battery
Display
3-1/2 Digit LCD (90 to 180 adjustable viewing angle)
Overrange Indication
Automatic (1---) on all ranges
Low Battery Indication
Automatic
Dimensions/Weight
3.75 x 4.25 x 2" (96 x 108 x 45mm) / 12 oz (340g)
Accessories supplied
9V battery, pH electrode, pH buffers, bottle, case, and
conductivity cell
o
MAINTENANCE
Conductivity Cell Simulator
The meter is supplied with a cell simulator which can be used in place of the conductivity
cell to provide a means to verify the operation of the meter. The simulator will produce a
reading of approximately 800 - 1200uS (~1mS). The simulator should not be used as a
calibration device.
This device is helpful in determining if the cause of calibration difficulties is related to the
meter or (more likely) an aged or defective conductivity cell.
Battery Installation / Replacement
Install a new battery when the LOW BATTERY indication appears on the LCD.
1.
Open the top cover (display) to its fully extended position.
2.
Locate the slot below the arrow on the battery cover.
3.
Use a flat bladed screwdriver or small coin to pry up and out in the slot until the
cover opens.
4.
Install the new battery and replace the cover.
Calibration / Repair Services
Extech offers complete repair and calibration services for all of the products we sell. For
periodic calibration or repair of any Extech product, call customer service for details on
services available. Extech recommends that calibration be performed on an annual basis
to insure calibration integrity.
5
341350 V2.0 10/06
WARRANTY
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and
workmanship for one year from date of shipment (a six month limited warranty applies on sensors and
cables). If it should become necessary to return the instrument for service during or beyond the warranty
period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our
website at www.extech.com (click on Contact Extech and go to Service Department to request an RA
number). A Return Authorization (RA) number must be issued before any product is returned to Extech. The
sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in
transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper
wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification.
Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and
will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited
to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty,
whether written or oral, is expressed or implied.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451
Copyright © 2006 Extech Instruments Corporation. All rights reserved
including the right of reproduction in whole or in part in any form.
6
341350 V2.0 10/06