FUJI EDS7-64

ULTRASONIC LIQUID LEVEL METER
DATA SHEET
FQH-2,FQG-2,FLY
An ultrasonic liquid level meter measures the liquid level
by the propagation time of the ultrasonic pulses emitted
out of the ultrasonic detector are reflected on the surface
of the measurement object and are received as echoes by
the detector again.
It provides high reliability and it permits easy installation, running and maintenance, as it does not make contact with the measurement object and it is free of mechanical moving parts.
It deals with uses of a broad range including water level
measurement of water, wastewater, tide level, rivers and
reservoirs and level measurement of highly viscous liquid
in chemical tanks, slurry agitation tanks, roll mills and food
tanks.
Detector (FQG)
FEATURES
1. No liquid contact
Permits easy installation and maintenance.
2. Detector of waterproof structure and of excellent corrosion resistance
It is of no problem against adhesion of water drops
and splashes, and it is optimum for wastewater, alkali,
acid and high viscosity liquid.
3. Not affected by fixed reflection noise
Not affected by obstacles such as projections of detector mounting area, walls and ladders because of the
DGL (dynamic gain limit) function that automatically
learns the noise level at all times and masks noise so
that no noise is detected.
4. Minimum influence of electrical noise
Not affected by electrical noise or noise from agitating
blades, as such noise is largely reduced by statistical
processing.
5. Use of high gain amplifier
Permits use at a pump well involving scum that could
be hardly measured with conventional meters and also
at a place of large attenuation such as sludge storage
tank.
SPECIFICATIONS
Converter (Type: FQH)
Structure:
Material grade:
Coating color:
Accuracy:
Output:
IP65
Aluminum alloy
Blue (cover), silver (case)
± 1%/FS or better
4 to 20mA DC (insulated, load resistance
0 to 1kΩ)
Converter (FQH)
Display:
LCD display (16 characters, 2 lines, with
back light)
Level measured value (m, %), detector
peripheral temperature, ultrasonic wave
received signal intensity, etc. are displayed.
Dumping:
2 to 600 seconds (once every second)
Power failure recovery processing:
Memory backup with lithium cell (valid
for about 10 years)
Power supply:
100 to 240V AC, 50/60Hz or 20 to 30A
DC
Power consumption:
35VA or less (AC power supply)
12W or less (DC power supply)
Contact output: Upper limit and lower limit alarms, error
alarms, up to 8 points (optional), relay
contact output, contact capacity AC
250V, 5A, normally open contact
Working ambient temperature: -10 to50˚C
Working ambient humidity: 90% RH or less
External dimensions: H320 X W240 X D137mm
Mounting method: Wall surface mounting
Mass:
About 7 kg
Lightning measure:
Power supply, output signals and alarm
outputs are provided with arresters
Others:
Provided with self-diagnostic function
Detector (Type: FQG)
Structure:
Composition:
IP67
With built-in preamplifier, transmitter and
EDS7-64
Date
Mar. 10, 2003
FQH-2,FQG-2,FLY
resistance thermometer bulb
Measuring distance:Up to 12 m (blind distance: 0.3 m)
Material grade: Heat resistant PVC (color: gray)
Matching flange: JIS 5K 150A or larger
External dimensions: ø140 X 180mm
Mass:
About 1.2 kg
Lightening measure: With arrester
Working ambient temperature: -20 to 80˚C
Working ambient humidity: 95% RH or less
Dedicated cable (between converter and
detector) (Type: FLY5)
Twisted paired shield cable
2 pairs x 0.5 mm2 shield (length up to 300 m)
Outside diameter: ø10.5mm (Max ø11mm)
Cable end treatment:
With M3.5 crimp-style terminal (at both
ends)
CONFIGURATION
Converter
Dedicated cable (up to 300m)
Detector
Error alarm
Upper limit, lower limit alarms
Optional
Power supply (100 to 240V AC or 24V DC)
Blind distance (0.3m)
Zero-point
distance
(up to 12m)
Output signal (DC4 to 20mA)
*1 Error alarm
An alarm is output on occurrence of set numeric
value data error, cable open circuit or reception
error. (Also indicated in LCD display screen.)
Full level output 20mA DC
Span (up to 11.7m)
Zero level output 4mA DC
USES
Level measurement of
waterworks,
wastewater, rivers,
dams and reservoirs
and tide level
Level measurement of
food, high viscosity
fluid and chemical
storage tanks
Level measurement of
dehydrated slurry and
sludge tanks
Level measurement of
agitation tanks such as
slurry agitation tanks
Level measurement of
high viscosity fluid in
tanks such as roll mills
Typical detector installation
* Detectors are of watertight structure.
Avoid exposure to direct sunlight and rain by means of a roof or the like, when they are used outdoors.
[Mounting with flange]
[Mounting with square flange]
[Mounting on wave guide]
200 mm
max.
150A flange
or larger
ø200
200 mm
max.
400 mm min.
Sound
absorbing
cartridge
ø200 or more
Wave guide pipe
of ø200 or larger
* Side wall must be free from standout
and sticking material
* Inquire us separately regarding
wave-guide pipe and sound
absorbing cartridge.
CODE SYMBOLS
<Detector>
<Converter>
1 2 3 4 5 6 7 8
FQH
0
1 2 3 4 5 6 7 8 9 10
F QG
02 -
9 10 11 12
2 -
Specification
Application
Level meter for liquid/slurry
1
A
Structure
Immersion-proof
F
Measurable distance (frequency)
12 m (30 kHz) max.
1
Power source
100 to 240 V AC, 50/60 Hz
20 to 30 V DC
3
7
Resistance bulb
Incorporated
1
Current output (4 to 20 mA DC)
1 point
1
0
1
2
Wiring port
With waterproof gland
With union (for PRICA) gland
Y
A
Y
T
Tag indication
No
Yes
Y
T
Contact closure command
None
5 points
8 points
0
5
8
Specification
Structure
Heat-resisting PVC immersionproof
Detector fixture
None
L-shaped fixture, mounted on
1 inch pipe
L-shaped fixture, suspended
from flat plate
Tag indication
No
Yes
<Cable>
12345678
FLY5
1
Specification
Applicable detector
Ultrasonic level meter detector
5
Detector-converter cable length
5m
5 m steps up to 100 m.
10 m steps for 100 to 300 m.
(Ex.: 150 for 150 m.)
005
300 m
300
OUTLINE DIAGRAM (Unit : mm)
Converter (Type: FQH)
74
Detector (Type: FQG)
2-ø12
74
2-ø12
4-ø7 mounting holes
TAG nameplate
450
420*
320
ø1
24
Main nameplate
240
Main nameplate
(SUS304)
12
12
. If provided with waterproof gland
. * Adjustable to 450 mm
12
Cable lead-in
port (nylon 66)
G1/2
. If provided with gland for
PRICA tube (JIS C8309, No. 17)
150
50
180
12
25
TAG nameplate
(SUS304 except
where noted)
28
134
ø140
25
3
120
ø90
95
70
Ball chain
(SUS305)
100
(Unit: mm)
(Unit: mm)
FQH-2,FQG-2,FLY
CONNECTION DIAGRAM
Converter
(FQH)
Precautions against obstacles
AC power connection
L
AC LINE
Source
100 to 240 V AC,
50/60 Hz
N
(1)
DC power connection
1
DC
20 to 30 V DC
2
(2)
Mount the detector 400mm
or more away from the side
wall
Get at least θ of 8° away
from shelf, step or standout,
if any
Dedicated cable (FLY 5)
(Twisted pair shield cable)
θ
A
R
B
SENSOR
C
T
D
SH
OUT1
CURRENT
OUT2
OUT3
Blue
Blue
White/blue
White/blue
Yellow
Yellow
White/yellow
White/yellow
E
A1
B1
B
C
D
400mm
or more
Detector
(FQG)
Shelf
E
+
+
+
-
AL1
A
Current output (4 to 20 mA DC)
(Terminal screw M3.5)
FAIL alarm
Relay output 1 (device error alarm)
Detector fixture (option)
A2
AL2
Relay output 2 High (low) limit alarm
B2
A3
AL3
Relay output 3 High (low) limit alarm
B3
L-shaped fixture made of SUS304
On 1 inch pipe or suspended from flat plate
A4
AL4
ALARM
(option)
Relay output 4 High (low) limit alarm
e
B4
A5
AL5
Relay output 6 High (low) limit alarm
B6
A7
AL7
4-ø7
Cable
lead-in
A6
AL6
100
42
90
Relay output 5 High (low) limit alarm
B5
Relay output 7 High (low) limit alarm
Fixture
B7
A8
Relay output 8 High (low) limit alarm
90
170
AL8
B8
60
(Terminal screw M3.5)
Scope of delivery
f
[Mounting size]
Converter main unit, detector main unit, dedicated cable (of
specified length)
4-ø7, or M6 Tap
øg
What are specified at the time of placement of an
order
1. Detector type
FQG_____________
2. Converter type
FQH_____________
3. Dedicated cable
FLY
4. Span distance
__________m
5. Zero-point distance __________m
6. Tag No. designation ______________
øh
(Boring of flanges, etc.)
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan