User`s Guide Pinless Moisture Meter Model MO257

User's Guide
Pinless Moisture Meter
Model MO257
Introduction
Congratulations on your purchase of the Extech MO257 Pinless Moisture Meter. The pinless
moisture sensor monitors the moisture in wood and other building materials without causing surface
damage. High frequency sensing technology allows the meter to take non-invasive moisture
measurements. User programmable alarms alert the user if moisture measurements exceed alarm
set-points. This meter is shipped fully tested and calibrated and, with proper use, will provide years of
reliable service.
Meter Description
1.
2.
3.
4.
5.
6.
7.
Measurement sensor
Power ON-OFF button
LCD display
Measure (MEAS) button
UP and BACKLIGHT button
SET button
DOWN button
Battery compartment located on rear of unit
LCD Display
1.
2.
3.
4.
5.
6.
7.
8.
Low Battery icon
Main LCD display
MAX reading
MIN reading
HOLD function
Measurement alert
MAX mode icon
MIN mode icon
2
MO257-EU-EN -V1.2 04/12
Battery Replacement
1.
Turn off the meter
2.
Remove the Philips head screw located on the rear of the unit
3.
Lift off the rear battery cover
4.
Replace the three (3) ‘AAA’ 1.5V batteries
5.
Secure the rear battery cover
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever batteries
/ accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
3
MO257-EU-EN -V1.2 04/12
Operation
Pinless Moisture Measurements
1.
Press the power
button to turn the meter on.
2.
As described below, you must calibrate the meter before first use.
3.
To calibrate, press the MEAS button immediately after powering on the meter. The display will
audibly beep and visually flash the CAL display icon. Keep the measurement sensor at least 3
to 4” (8 to 10 cm) from any object during calibration.
4.
When the calibration is complete the beeping and flashing stop. The calibration reading must
not exceed 0.5. If so, re-calibrate ensuring that hands or other objects are kept clear of the
measurement sensor.
5.
The meter should be calibrated whenever it is moved from one measurement area to another.
6.
To take a measurement, hold the meter and position the sensor so that it is very near or
directly touching the surface under test. The meter can sense moisture to a depth of 0.79 to
1.57” (20 to 40mm).
7.
IMPORTANT NOTE: The position of the hand can greatly affect a reading. Do not move or
reposition the hand that holds the meter while taking measurements or when calibrating the
unit.
8.
The LCD indicates a relative moisture reading in numeric digits from 0.0 to 100.0 and with
programmable alarm alerts (DRY, RISK, and WET).
9.
Press the MEAS button to HOLD (freeze) a reading on the LCD (the HOLD icon will appear on
the LCD). The meter will hold the reading for 30 seconds before automatically shutting the
meter down.
10. The MAX and MIN display areas indicate the highest and lowest readings encountered since
the measurement session was started.
11. If the battery symbol appears or if the meter does not switch ON, replace the batteries.
Alarm Settings
Two high alarm points can be set for Moisture measurements: A RISK Alarm (0 to 50 setting range
with a default setting of 30 units) and a WET Alarm (50 to 100 setting range with a default setting of
60 units).
Moisture Alarm Setup Procedure:
1.
Press the MEAS button to place the meter in HOLD mode to change the Alarm settings.
2.
Press the SET button. The RISK display icon will appear flashing on the LCD indicating that a
RISK alarm can now be programmed.
3.
Use the up/down buttons to select the desired RISK Alarm value.
4.
Press the SET button when the desired setting is displayed.
5.
The WET icon will appear flashing on the LCD indicating that the WET Alarm is ready to be
programmed.
6.
Use the up/down buttons to select the desired Wet Alarm setting.
7.
Press SET to complete the programming.
8.
Now, if the moisture measurement is higher than the RISK setting, the meter will beep once
every two seconds and if the moisture measurement is higher than the WET setting the meter
will beep four times per second.
9.
Press the MEAS button to exit HOLD mode.
4
MO257-EU-EN -V1.2 04/12
Auto Power Off
To conserve battery energy, the meter will enter a sleep mode after approx. 30 minutes of inactivity.
Press the power button to wake the meter.
Backlit LCD
Press the up/backlight button to switch the backlighting ON or OFF.
Specifications
Function
Range
Accuracy
Moisture Range
0.0 to 100.0
Relative only
Measurement Depth
20 to 40mm (0.79 to 1.57”)
Display type
Backlit LCD
Display digits
Three (3) numeric displays; each displaying from 0.0 to 100.0
Display icons
HOLD, DRY, RISK, WET, MIN, MAX and low battery icon
Operating Temperature
4 to 43°C (40 to 110°F)
Operating Humidity
90%, 0-30°C (32-86°F), 75%, 30-40°C (86-104°F),
45%, 40-50°C (104-122°F)
Power Supply
Three (3) 1.5V ‘AAA’ batteries
Auto Power Off (APO)
After 30 minutes of inactivity
Dimensions
235 x 63 x 28mm (9.2 x 2.5 x 1.1”)
Weight
218g (7.7oz)
Moisture Comparison Table (typical)
Material
Gypsum
Cement
Wood
Display units
< 30
30 - 60
> 60
< 25
25 - 50
> 50
< 50
50 - 80
> 80
Moisture status
DRY
RISK
WET
DRY
RISK
WET
DRY
RISK
WET
Copyright © 2012 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
5
MO257-EU-EN -V1.2 04/12