EXTECH EX540 Manual

User's Guide
Wireless TRMS Multimeter
Model EX540
Introduction
This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency
(electrical & electronic), Duty Cycle, Diode Test, and Continuity plus Thermocouple
Temperature. It can store and recall data. It features a waterproof, rugged design for heavy
duty use. This meter can transmit data wirelessly when linked to a PC. Proper use and care
of this meter will provide many years of reliable service.
Safety
This symbol adjacent to another symbol, terminal or operating device
indicates that the operator must refer to an explanation in the Operating
Instructions to avoid personal injury or damage to the meter.
WARNING
CAUTION
MAX
1000V
This WARNING symbol indicates a potentially hazardous situation, which
if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which
if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be
connected to a circuit point at which the voltage with respect to earth
ground exceeds (in this case) 1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being
associated with ranges that may, in normal use, be subjected to
particularly hazardous voltages. For maximum safety, the meter and its
test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double
insulation or reinforced insulation.
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be
supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial
use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
2
EX540-EU-EN V1.4-3/11
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
Maximum Input
V DC or V AC
1000VDC/AC rms
mA AC/DC
500mA 1000V fast acting fuse
A AC/DC
10A 1000V fast acting fuse (20A for 30 seconds
max every 15 minutes)
Frequency, Resistance, Capacitance,
Duty Cycle, Diode Test, Continuity
1000VDC/AC rms
Temperature
1000VDC/AC rms
Surge Protection: 8kV peak per IEC 61010
2.
USE EXTREME CAUTION when working with high voltages.
3.
DO NOT measure voltage if the voltage on the "COM" input jack exceeds 1000V above
earth ground.
4.
NEVER connect the meter leads across a voltage source while the function switch is in
the current, resistance, or diode mode. Doing so can damage the meter.
5.
ALWAYS discharge filter capacitors in power supplies and disconnect the power when
making resistance or diode tests.
6.
ALWAYS turn off the power and disconnect the test leads before opening the covers to
replace the fuse or batteries.
7.
NEVER operate the meter unless the back cover and the battery and fuse covers are in
place and fastened securely.
8.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
3
EX540-EU-EN V1.4-3/11
Controls and Jacks
1. 40,000 count LCD display
2. STORE(<RECALL) button
1
3. MAX/MIN (_)button
2
13
3
4
12
11
10
4. MODE button
5. RANGE(SETUP) button
6. Function switch
7. mA, µA and 10A input jacks
5
8. COM input jack
9. Positive input jack
10.
6
Backlight button
11. EXIT(AC+DC) button
9
7
12. HOLD(PEAKHOLD>) button
8
13. REL(+) button
Note: Tilt stand and battery compartment are on rear of unit.
Symbols and Annunciators
•)))
n
µ
m
A
k
F
M
Ω
Hz
%
AC
DC
ºF
MAX
N0.
SET
TRMS
RCL
Continuity
Diode test
Battery status
-9
nano (10 ) (capacitance)
-6
micro (10 ) (amps, cap)
-3
milli (10 ) (volts, amps)
Amps
3
kilo (10 ) (ohms)
Farads (capacitance)
6
mega (10 ) (ohms)
Ohms
Hertz (frequency)
Percent (duty ratio)
Alternating current
Direct current
Degrees Fahrenheit
Maximum
Serial number
Set up parameter
True RMS
Recall
PEAK
V
AUTO
HOLD
ºC
MIN
S
AC+DC
STO
AUTO
Auto Power off enabled
RF transmitter active
Peak Hold
Volts
Relative
Autoranging
Display hold
Degrees Centigrade
Minimum
second
Alternating current + Direct current
Store
Auto Range
Backlight
4
EX540-EU-EN V1.4-3/11
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous and should be measured with great care.
1.
ALWAYS turn the function switch to the OFF position when the meter is not in use.
2.
If “OL” appears in the display during a measurement, the value exceeds the range
you have selected. Change to a higher range.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or
OFF. Large voltage surges may occur that can damage the meter.
1.
2.
3.
4.
Set the function switch to the VDC position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
Read the voltage in the display.
AC VOLTAGE (FREQUENCY, DUTY CYCLE) MEASUREMENTS
WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the
live parts inside some 240V outlets for appliances because the contacts are recessed
deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has
voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet
before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or
OFF. Large voltage surges may occur that can damage the meter.
1.
2.
3.
4.
5.
6.
7.
8.
9.
Set the function switch to the VAC/Hz/% position.
Insert the black test lead banana plug into the negative COM jack.
Insert red test lead banana plug into the positive V jack.
Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
Read the voltage in the main display and the frequency in the right
auxiliary display
Press the MODE button to indicate “Hz”.
Read the frequency in the main display.
Press the MODE button again to indicate “%”.
Read the % of duty cycle in the main display.
With ACV in the main display, press EXIT for 2 seconds to measure AC+DC.
5
EX540-EU-EN V1.4-3/11
mV VOLTAGE MEASUREMENTS
CAUTION: Do not measure mV voltages if a motor on the circuit is being switched ON
or OFF. Large voltage surges may occur that can damage the meter.
1.
2.
3.
4.
5.
6.
Set the function switch to the mV position.
Press the MODE button to indicate “DC”or ““AC ”, or in AC range press
EXIT for two seconds and chose ”AC+DC”
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
Touch the black test probe tip to the negative side of the circuit. Touch
the red test probe tip to the positive side of the circuit.
Read the mV voltage in the main display.
With AC mV in the main display, press EXIT for 2 seconds to measure
AC+DC.
DC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1.
2.
3.
4.
5.
6.
7.
8.
9.
Insert the black test lead banana plug into the negative COM jack.
For current measurements up to 4000µA DC, set the function switch to
the µA position and insert the red test lead banana plug into the µA/mA
jack.
For current measurements up to 400mA DC, set the function switch to
the mA position and insert the red test lead banana plug into the µA/mA
jack.
For current measurements up to 20A DC, set the function switch to the
10A/HZ/% position and insert the red test lead banana plug into the 10A
jack.
Press the MODE button to indicate “DC” on the display.
Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
Apply power to the circuit.
Read the current in the display.
6
EX540-EU-EN V1.4-3/11
AC CURRENT (FREQUENCY, DUTY CYCLE) MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
10.
Insert the black test lead banana plug into the negative COM jack.
For current measurements up to 4000µA AC, set the function switch to
the µA position and insert the red test lead banana plug into the µA/mA
jack.
For current measurements up to 400mA AC, set the function switch to
the mA position and insert the red test lead banana plug into the µA/mA
jack.
For current measurements up to 20A AC, set the function switch to the
10A/HZ/% position and insert the red test lead banana plug into the 10A
jack.
Press the MODE button to indicate “AC” on the display.
Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
Apply power to the circuit.
Read the current in the display. In the 10AAC range, right auxiliary display frequency.
Press and hold the MODE button to indicate “Hz”.
Read the frequency in the display.
Momentarily press the MODE button again to indicate “%”.
Read the % duty cycle in the display.
Press and hold the MODE button to return to current measurement.
With ACA in the main display, press EXIT for 2 seconds to measure AC+DC.
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance measurements. Remove the
batteries and unplug the line cords.
1.
2.
3.
4.
5.
Set the function switch to the Ω CAP
position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Ω jack.
Press the MODE button to indicate “Ω” on the display.
Touch the test probe tips across the circuit or part under test. It is best
to disconnect one side of the part under test so the rest of the circuit will
not interfere with the resistance reading.
Read the resistance in the display.
7
EX540-EU-EN V1.4-3/11
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that
have voltage on them.
1.
2.
3.
4.
5.
Set the function switch to the Ω CAP
position.
Insert the black lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Ω jack.
Press the MODE button to indicate“ "and “Ω” on the display
Touch the test probe tips to the circuit or wire you wish to check.
If the resistance is less than approximately 35Ω, the audible signal will
sound. If the circuit is open, the display will indicate “OL”.
DIODE TEST
1.
2.
Set the function switch to the Ω CAP
position.
Insert the black test lead banana plug into the negative COM jack and
the red test lead banana plug into the positive V jack.
3.
4.
Press the MODE button to indicate“ “ and “V” on the display.
Touch the test probes to the diode under test. Forward voltage will
typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”.
Shorted devices will indicate near 0V and an open device will indicate
“OL” in both polarities.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.
1.
2.
3.
4.
5.
6.
Set the rotary function switch to the Ω CAP
position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
Press the MODE button to indicate “F”
Touch the test leads to the capacitor to be tested.
Read the capacitance value in the display
8
EX540-EU-EN V1.4-3/11
TEMPERATURE MEASUREMENTS
1.
Set the function switch to the Temp position.
2.
Insert the Temperature Probe into the input jacks, making sure to
observe the correct polarity.
3.
Press the MODE button to indicate “ºF” or “ºC”
4.
Touch the Temperature Probe head to the part whose temperature
you wish to measure. Keep the probe touching the part under test
until the reading stabilizes (about 30 seconds).
5.
Read the temperature in the display.
Note: The temperature probe is fitted with a type K mini connector. A mini connector to
banana connector adaptor is supplied for connection to the input banana jacks.
FREQUENCY (DUTY CYCLE) MEASUREMENTS (ELECTRONIC)
1. Set the rotary function switch to the Hz/% position.
2. Insert the black lead banana plug into the negative COM jack and the red test lead
banana plug into the positive Hz jack.
3. Touch the test probe tips to the circuit under test.
4. Read the frequency on the display.
5. Press the MODE button to indicate “%”.
6. Read the % duty cycle in the display.
% 4 – 20mA MEASUREMENTS
1. Set up and connect as described for DC mA measurements.
2. Set the rotary function switch to the 4-20mA% position.
3. The meter will display loop current as a % with 0mA=-25%, 4mA=0%, 20mA=100%,
and 24mA=125%.
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into AutoRanging. This
automatically selects the best range for the measurements being made and is generally
the best mode for most measurements. For measurement situations requiring that a range
be manually selected, perform the following:
1.
2.
3.
Press the RANGE key. The “AUTO” display indicator will turn off.
Press the RANGE key to step through the available ranges until you select the range
you want.
To exit the Manual Ranging mode and return to Autoranging, press EXIT
Note: Manual ranging does not apply for the Temperature functions.
MAX/MIN
1. Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon
"MAX" will appear. The meterleft auxiliary displaywill display and hold the maximum
reading and will update only when a new “max” occurs.The display icon "MIN" will
appear. The right auxiliary displaymeter will display and hold the minimum reading
and will update only when a new “min” occurs.
2.
To exit MAX/MIN mode press EXIT
9
EX540-EU-EN V1.4-3/11
RELATIVE MODE
The relative measurement feature allows you to make measurements relative to a stored
reference value. A reference voltage, current, etc. can be stored and measurements made
in comparison to that value. The displayed value is the difference between the reference
value and the measured value.
1.
2.
3.
4.
5.
Perform the measurement as described in the operating instructions.
Press the REL button to store the reading in the display and the "▲" indicator will
appear on the display.
The Right auxiliary display displays the initial reading (the stored value)
The Left auxiliary display displays the actual currently measured value.
Main display displays the Relative value (currently measured value minus the stored
value).
DISPLAY BACKLIGHT
key to turn the backlight on. The backlight will automatically turn off after
Press the
SET time. Press the EXIT button to exit the backlight on mode.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to
activate or to exit the HOLD function.
PEAK HOLD
The Peak Hold function captures the peak AC or DC voltage or current. The meter can
capture negative or positive peaks as fast as 1 millisecond in duration. Momentarily press
the PEAK button, “PEAK” and “MAX” will display in left auxiliary display. MIN” will
display in right auxiliary display. The meter will update the dispay each time a lower
negative peak occurs. Press the EXIT button to exit the PEAK HOLD mode. Auto Power
Off feature will be disabled automatically in this mode.
DATA STORAGE
1. Set the function switch to the measurement function desired.
2. Press the STORE button to enter the STORE function. The left upper aux display
indicates the current storage location (0000 to 9999). New measurements will begin
storing into the next location.
3. Press the PEAKHOLD button to change into the initial storage number 0000. (Press it
again to change back). On the right upper aux display shows XXXX which states how
many current storage locations are used.
4. Press STORE button again to enter into recording interval time set up function.
5. The left upper aux display indicates 0000 S, which is the recording sample rate; use
the + & - buttons to select the sample rate desired (0 to 255 sec)
6. Set the sample rate to 0000 S for manual recording. In this mode, each press of the
STORE button will save one measurement reading.
7. Set the sample rate to 1 to 255 S for automatic recording. In this mode, pressing the
STORE button will start data recording at the set sample rate.
8. Press the EXIT button to end the recording session.
10
EX540-EU-EN V1.4-3/11
DATA STORAGE RECALL
1. Press STORE button for two seconds to enter into RECALL function.
2. On the left upper aux display will show XXXX, which is the current storage location.
The right aux display will show XXXX, which is the number of storage locations
currently used.
3. Press the + or — button to select the storage location. The value in that location will
be indicated in the main display.
4. Press the PEAKHOLD button once to continuously scan data from 0000 to XXXX.
Press again then scan again.
5. Press the EXIT button to end the recall session.
CLEARING MEMORY
To clear the memory of all stored data, from the OFF position, hold the EXIT button and
switch the function switch to any position. Release the EXIT button and the LCD will flash
3 times and a buzzer will beep three times. Memory is now clear.
PC WIRELESS COMMUNICATION:
1. Install and launch the pc software
2. Press and Hold the backlight button for two seconds to enter RF wireless transmit
mode.
3.
4.
5.
6.
will appear on the display.
The RF icon
When communication is established, the RF icon on the display will blink and the led
indicator on the receiver will also blink.
Once per second, the data will be displayed on the pc screen, plotted and inserted
into a list.
Hold the backlight button for two seconds to exit the RF wireless transmit mode
SENDING STORED DATA TO THE PC
1. Launch the pc software
2. Press the STORE button for two seconds to enter into data RECALL function.
3. Press the HOLD button for two seconds. The RF transmit icon will flash while the
stored data is sent to the pc
SETUP
1. Press and Hold the RANGE/SETUP button for two seconds to enter the SET function.
The first of five settable functions will appear.
2. Press the RANGE button to step through the functions
A: Upper limit buzzer alarm
OFF or Value
B: Lower limit buzzer alarm
OFF or Value
C: Auto power off time
OFF, 10 to 30 sec
D: Button beeper
ON/OFF
E: Back light time
OFF, 10 to 30 sec
Use the +, - ◄, and ◄ buttons to select and change conditions and digits.
3. Press the RANGE/SETUP button until the meter returns to the normal display to exit
this mode.
AC+DC
In the measuring modes VAC, mV(AC), 10A(AC), mA(AC) and uA(AC), press the EXIT
button for 2 seconds to enter into AC+DC testing. The precision is the same as in the AC
measure modes. The LCD shows AC+DC icon. Press the EXIT button to exit the mode.
LOW BATTERY INDICATION
When the
icon appears in the display, the battery should be replaced
11
EX540-EU-EN V1.4-3/11
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and fuse
covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care
instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature
extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the
electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT
use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove
old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries
should be removed to prevent damage to the unit.
12
EX540-EU-EN V1.4-3/11
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the battery cover.
1. Turn power off and disconnect the test leads from the meter.
2. Open the rear battery cover by removing two screws (B) using a Phillips head
screwdriver.
3. Insert the battery into battery holder, observing the correct polarity.
4. Put the battery cover back in place. Secure with the screws.
5.
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over
your used batteries / accumulators at collection points in your community or
wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device
at the end of its lifecycle
WARNING: To avoid electric shock, do not operate the meter until the battery cover is in
place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and batteries to make sure that
they are still good and that they are properly inserted.
A
B
A
B
A
F1 F2
13
EX540-EU-EN V1.4-3/11
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the meter cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the battery cover (two “B” screws) and the battery.
4. Remove the six “A” screws securing the rear cover.
5. Gently remove the old fuse and install the new fuse into the holder.
6. Always use a fuse of the proper size and value (0.5A/1000V fast blow for the 400mA
range [SIBA 70-172-40], 10A/1000V fast blow for the 20A range [SIBA 50-199-06]).
7. Replace and secure the rear cover, battery and battery cover.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in
place and fastened securely.
FCC Part 15
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to which the receiver
is connected.
—Consult the dealer or an experienced radio/TV technician for help.
Warning: Changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment.
14
EX540-EU-EN V1.4-3/11
Specifications
Function
DC Voltage
Range
400mV
4V
40V
400V
1000V
Resolution
0.01mV
0.0001V
0.001V
0.01V
0.1V
Accuracy
±(0.06% reading + 2 digits)
±(0.1% reading + 2 digits)
AC Voltage 400mV
0.01mV
±(1.0% reading + 4digits)
(AC+DC)
4V
0.0001V
50 to 1000Hz 40V
0.001V
±(1.0% reading + 3digits)
400V
0.01V
1000V
0.1V
All AC voltage ranges are specified from 5% of range to 100% of range
DC Current
400μA
0.01μA
4000μA
0.1μA
±(1.0% reading + 3 digits)
40mA
0.001mA
400mA
0.01mA
10A
0.001A
(20A: 30 sec max with reduced accuracy)
AC Current
400μA
0.01μA
(AC+DC)
4000μA
0.1μA
50 to 1000Hz 40mA
±(1.5% reading + 3 digits)
0.001mA
400mA
0.01mA
10A
0.001A
(20A: 30 sec max with reduced accuracy)
All AC voltage ranges are specified from 5% of range to 100% of range
o
o
o
o
NOTE: Accuracy is stated at 18 C to 28 C (65 F to 83 F) and less than 75% RH.
15
EX540-EU-EN V1.4-3/11
Function
Resistance
Capacitance
Frequency
(electronic)
Frequency
(electrical)
Duty Cycle
Range
Resolution
400Ω
0.01Ω
4kΩ
0.0001kΩ
40kΩ
0.001kΩ
400kΩ
0.01kΩ
4MΩ
0.001MΩ
40MΩ
40nF
400nF
0.001MΩ
0.001nF
0.01nF
4μF
0.0001μF
40μF
0.001μF
Accuracy
±(0.3% reading + 9 digits)
±(0.3% reading + 4 digits)
±(2.0% reading + 10 digits)
±(3.5% reading + 40 digits)
±(3.5% reading + 10 digits)
400μF
0.01μF
4000µF
0.1µF
±(5% reading + 10 digits)
40mF
0.001mF
40Hz
0.001Hz
400Hz
0.01Hz
4kHz
0.0001kHz
40kHz
0.001kHz
±(0.1% reading + 1 digits)
400kHz
0.01kHz
4MHz
0.0001MHz
40MHz
0.001MHz
100MHz
0.01MHz
Not specified
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms
min @ 20% to 80% duty cycle and > 100kHz.
40.00HZ-4KHz
0.01HZ to
±(0.5% reading)
0.001KHz
Sensitivity:5Vrms
0.1 to 99.90% 0.01%
±(1.2% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
±(1.0% reading + 4.5°F)
±(1.0%
reading + 2.5°C)
-50 to 750°C 0.1°C
(probe accuracy not included)
4-20mA%
-25 to 125%
0.01%
±50 digits
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Note: Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.
Temp
(type-K)
-58 to 1382°F
0.1°F
16
EX540-EU-EN V1.4-3/11
Enclosure
Shock (Drop Test)
Diode Test
Double molded, waterproof (IP67)
2 meters (6.5 feet)
Test current of 0.9mA maximum, open circuit voltage 2.8V DC
typical
Storage capacity
9999 records
RF transmit distance
10 meters (approx)
Transmitter Frequency 915MHz
Continuity Check
Audible signal will sound if the resistance is less than 35Ω
(approx.), test current <0.35mA
Peak
Captures peaks >1ms
Temperature Sensor
Requires type K thermocouple
Input Impedance
>10MΩ VDC & >9MΩ VAC
AC Response
True rms
ACV Bandwidth
50Hz to 1000Hz
Crest Factor
≤3 at full scale up to 500V, decreasing linearly to ≤1.5 at 1000V
Display
40,000 count, backlit, liquid crystal with bargraph
Overrange indication
“OL” is displayed
Auto Power Off
15 minutes (approximately) with disable feature
Polarity
Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate
2 times per second, nominal
Low Battery Indication
“
” is displayed if battery voltage drops below operating
voltage
Battery
One 9 volt (NEDA 1604) battery
Fuses
mA, µA ranges; 0.5A/1000V ceramic fast blow
A range; 10A/1000V ceramic fast blow
Operating Temperature 5ºC to 40ºC (41ºF to 104ºF)
o
o
o
o
Storage Temperature
-20 C to 60 C (-4 F to 140 F)
Operating Humidity
Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC
(104ºF)
Storage Humidity
<80%
Operating Altitude
2000meters (7000ft) maximum.
Weight
342g (0.753lb) (includes holster).
Size
187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster)
Safety
This meter is intended for origin of installation use and protected,
against the users, by double insulation per EN61010-1 and
nd
IEC61010-1 2 Edition (2001) to Category IV 600V and Category
III 1000V; Pollution Degree 2.
17
EX540-EU-EN V1.4-3/11