PRC15 - Extech Instruments

USER GUIDE
Current / Voltage Calibrator
Model PRC15
Introduction
Thank you for selecting the Extech Model PRC15. This device is shipped fully tested and calibrated
and, with proper use, will provide years of reliable service. Please visit the Extech Instruments
website (www.extech.com) to check for the latest version of this User Guide. Extech Instruments is
an ISO-9001 certified company.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages
may be present
Double insulation
Safety Notes





Do not exceed the maximum allowable input range.
Turn the unit OFF when the device is not in use.
Remove the batteries if the device is to be stored for longer than 60 days.
Never dispose of batteries in a fire. Batteries may explode or leak.
Never mix battery types. Always install new batteries of the same type.
Cautions




Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
2
PRC15-EU-EN-V1.2 01/13
Meter Description
1.
AC adaptor input jack
2.
Display
3.
Power ON/OFF
4.
Source output adjustment arrow buttons
5.
Set button
6.
Unit button (mA or %)
7.
I/O MODE button
8.
MEM (memory STEP button)
9.
Backlight/ZERO button
10. Test lead input jacks
Display Layout
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
SOURCE mode icon
ZERO function status icon
MEASURE mode icon
Auto Power OFF active icon
Battery status icon
Measure mode value
Measure mode units icon
Datalogger memory location
Source mode value
Source mode units icon
3
PRC15-EU-EN-V1.2 01/13
Keypad Descriptions and Operation
POWER BUTTON and AUTO POWER OFF FEATURE
1.
Use the POWER button to turn the unit ON or OFF. When the unit is powered up, a short
self-test will ensue after which the display will stabilize.
2.
When the battery symbol flashes on the display, replace the battery as soon as possible.
Low battery power may cause inaccurate readings and erratic meter operation.
3.
This instrument is equipped with Auto Power OFF which turns the meter off after 10 minutes
of inactivity. To override this feature; press and hold the POWER button until the display
icon “ATP” turns off.
UNIT BUTTON
Press the UNIT button in the current function to select mA or % units or V or mV in the voltage
function. Voltage is autoranging in the MEASURE mode.
I/O BUTTON
Momentarily press the I/O button to select either SOURCE (output) or MEASURE (input).
MODE BUTTON
In the MEASURE mode, Press and Hold the MODE (I/O) button for 1second to select either the
current (mA/%) or voltage (mV/V) function. Release the button when the desired function appears.
(Backlight) BUTTON
Press the Backlight button to turn the backlight on or off.
ZERO (
) BUTTON
In the MEASURE or SOURCE mode, Press and Hold the ZERO (
the meter.
) button for 1 second to zero
► ◄ ▼ and ▲ BUTTONS
The arrow buttons are used to set the output value in the SOURCE mode.
1.
Select the SOURCE mode
2.
Press the ► or ◄ button select a digit for adjustment. The blinking underline curser
identifies the digit selected.
3.
Press the ▼or ▲button to adjust the value of the digit. Press and Hold the ▼or ▲ button to
rapidly adjust the value.
4
PRC15-EU-EN-V1.2 01/13
SET BUTTON
The SET button is used to manually step through the 5 stored output values.
1.
Select the SOURCE mode
2.
Press the SET button and the value stored in memory location 01 will be sourced. “MEM.01”
appears in the display.
3.
Each press of the SET button will step through the 5 memory locations.
4.
The arrow buttons can be used to adjust the value in each memory location.
STEP/MEM BUTTON
The STEP/MEM button is used to automatically step through the 5 stored output values. The meter
can be set for a single cycle of the stored values or a continuous cycle.
1.
Select the SOURCE mode
2.
Press and HOLD the STEP/MEM button. ”STEPSS” (single cycle) and “STEPSC”
(continuous cycle) will alternately appear in the display. Release the button when the
desired mode is displayed.
3.
In single cycle mode the meter will source the current or voltage displayed in MEM01 for 5
seconds. The meter will then advance to MEM02 for 5 seconds. This will continue through
MEM05 and then step down through the memory locations. The cycle will end when MEM01
has been reached.
4.
In the continuous mode the cycle will continue until manually stopped.
5.
Momentarily press the MEM button to stop the cycle. “END” will briefly appear in the display.
STORING VALUES INTO MEMORY
The default values stored in the memory locations are:
Memory
Location
mA
%
mV
V
M1
4.00mA
0.0%
0mV
0V
M2
8.00mA
25%
500mV
5V
M3
12.00mA
50%
1000mV
10V
M4
16.00mA
75%
1500mV
15V
M5
20.00mA
100%
2000mV
20V
To change the values in memory:
1.
Select the SOURCE mode
2.
Press the SET button to select the memory location to be changed.
3.
Press the arrow buttons to adjust to the new value
4.
Momentarily press the MEM button to store the value. The memory location icon will blink
while the value is being stored..
5
PRC15-EU-EN-V1.2 01/13
Modes of Operation
MEASURE (Input) Mode of Operation
In this mode, the unit will measure up to 50mADC or 20VDC.
1.
Turn the meter ON.
2.
“MEASURE” will appear in the display.
3.
Press and Hold the MODE button for 1 second to select mA or % or mV
4.
Connect the Calibration Cable to the meter.
5.
Connect the Calibration Cable to the device or circuit under test.
6.
Read the measurement on the LCD display.
SOURCE (Output) Mode of Operation
In this mode, the unit can source current up to 24mADC at 1000 ohms or voltage up to 20.00V.
The current or voltage can be output either manually or stepped from memory as described
earlier.
1.
Turn the meter ON
2.
MEASURE” will appear in the display.
3.
Press and Hold the MODE button for 1 second to select mA or % or mV
4.
Momentarily press the “I/O” button to select SOURCE.
5.
Press the UNIT button to select % / mA or mV / V.
6.
Connect the Calibration Cable to the meter
7.
Connect the Calibration Cable to the device or circuit under test
8.
Use the arrow buttons to set the desired output value in the lower display. The upper
display indicates the actual current or voltage value being sourced. If the upper display does
not match the set value, either the batteries need to be replaced or the load impedance is
beyond the specified range.
6
PRC15-EU-EN-V1.2 01/13
Tilt Stand / Hanger
The rear stand is provides two methods for convenience in viewing.
1. Pull the bottom portion of the stand out to place the unit on a flat surface for viewing.
2. Pull the bottom and top portions of the stand out, and then rotate the stand to allow the unit
to be hung.
Battery Replacement
When the battery icon appears on the display, the six AA batteries must be replaced.
The battery compartment is located on the rear of the meter.
1. Open the tilt stand, loosen the captive Philips head screw and remove the battery cover.
2. Remove and replace the batteries, observing polarity.
3. Replace and secure the battery cover.
All EU users are legally bound by the Battery Ordinance to return all used batteries to
community collection points or wherever batteries / accumulators are sold.
Disposal in household trash or refuse is prohibited.
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
Battery Safety Reminders
 Please dispose of batteries responsibly; always observe local, state, and federal regulations
with regard to battery disposal.
 Never dispose of batteries in a fire. Batteries may explode or leak.
 Never mix battery types or old and new batteries. Always install new batteries of the same
type.
7
PRC15-EU-EN-V1.2 01/13
Specifications
General Specifications
Display
Maximum Load
Meter Power
Auto Power OFF
Current output capability
Voltage input impedance
Dot matrix LCD
1000 ohms @ 24mA
6 AA batteries or AC adaptor
Meter automatically powers off after 10 minutes of inactivity
24mADC at 1000 ohms
10kohms minimum
Operating Temperature
Storage Temperature
Operating Humidity
5ºC to 40ºC (41ºF to 104ºF)
-20ºC to 60ºC (- 4ºF to 140ºF)
Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC
(104ºF)
<80%
7000ft (2000meters) maximum
159 x 80 x 44 mm (6.3 x 3.2 x 1.7")
234g (8.3 oz.) without batteries
Storage Humidity
Operating Altitude
Dimensions
Weight
Range Specifications
Mode
Function
Range (Resolution)
DC
Measure
Current
0 to 50mA (0.01mA)
Percent (%)
-25% to +230% (0.1%)
Voltage
(autorange)
0 to 1999mV (1mV)
DC Source
2 to 20V (0.01V)
Current
0 to 24mA (0.01mA)
Percent (%)
-25% to +125% (0.1%)
Voltage
Accuracy (% of reading)
± (0.01% + 1 digit)
0 to 2000mV (1mV)
0 to 20V (0.01V)
Copyright © 2013 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com
8