SERIES: ETMA 110W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
•
up to 110 W power
universal input (90~264 Vac)
single regulated output from 12~24 V
over voltage, overload and short circuit protections
full medical safety approvals
level V efficiency
custom designs available
RoHS
output
voltage
output
current
output
power
ripple
and noise1
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
ETMA120750U
12
7.5
90
120
ETMA135667U
13.5
6.67
90
135
V
ETMA150667U
15
6.67
100
150
V
ETMA190580U
19
5.8
110
190
V
ETMA200550U
20
5.5
110
200
V
ETMA240460U
24
4.6
110
240
V
MODEL
Notes:
efficiency
level
V
1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with 10 μF aluminum electrolytic and 0.1 μF ceramic capacitors.
PART NUMBER KEY
ETMA120750U C - P51 - WP - CXX
Base Number
example of 12 Vdc, 7.5 A
Reserved for Custom
Configurations
DC Plug Type
Input Cable:
C = included
blank = no cable
Factory
Designation
cui.com
For more information, please visit the product page.
CUI Inc │ SERIES: ETMA 110W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/04/2013 │ page 2 of 5
INPUT
parameter
conditions/description
min
voltage
frequency
current
at 90 Vac, 50 Hz
inrush current
at 230 Vac, cold start
max
units
90
typ
264
Vac
47
63
Hz
2
A
no load power consumption
70
A
0.3
W
max
units
OUTPUT
parameter
conditions/description
total regulation
19, 20, 24 V outputs
all other outputs
hold-up time
at 115 / 230 Vac, full load
min
typ
±3
±5
%
%
10
ms
temperature coefficient
±0.05
%/°C
150
150
150
%
%
%
160
%
max
units
5,656
2,121
Vdc
Vdc
PROTECTIONS
parameter
conditions/description
over voltage protection
13.5 V output
24 V output
all other outputs
118
113
120
overload protection
shutdown and auto restart
short circuit protection
output shutdown and auto restart
SAFETY & COMPLIANCE
parameter
conditions/description
isolation voltage
input to output for 2 seconds
input to frame ground for 2 seconds
min
insulation resistance
input to output, at 500 Vdc
input to frame ground, at 500 Vdc
safety approvals
UL/cUL UL 60601-1, EN 60601-1, CB IEC 60601-1
EMI/EMC
FCC Part 18 Class B, EN 60601-1 Class B, EN 55011 Class B
typ
20
20
MΩ
MΩ
leakage current
0.1
MTBF
at 25°C, max. load
RoHS compliant
yes
100,000
mA
hours
ENVIRONMENTAL
parameter
conditions/description
min
operating temperature
storage temperature
humidity
non-condensing
cui.com
max
units
0
typ
40
°C
-10
70
°C
10
90
%
For more information, please visit the product page.
CUI Inc │ SERIES: ETMA 110W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/04/2013 │ page 3 of 5
MECHANICAL
parameter
conditions/description
dimensions
6.69 x 2.55 x 1.52 (170 x 64.8 x 38.5 mm)
min
input plug
IEC320 / C14
typ
max
inch
weight
0.61
kg
MECHANICAL DRAWING
units: mm
64.8 0.50
38.5 0.50
170.0 0.50
LED
cui.com
units
For more information, please visit the product page.
CUI Inc │ SERIES: ETMA 110W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/04/2013 │ page 4 of 5
DC CORD
Aluminum Shield
PVC Jacket
PIN 2
Red (+Vo)
PIN 1
Blue (GND)
Drain Wire
PVC Insulation
PIN 4
Yellow (+Vo)
[2464#18AWG 4C+1]
SECTIONAL DRAWING
PIN 3
Black (GND)
D.W. (GND)
4P POWER DIN
FRONT VIEW
TERMINAL
ø2.4
UL Heat Shrinkable Tube
Core
6.7*14.3*28.5mm
Black+Blue+D.W.
Red+Yellow
4PIN POWER DIN
Snap and Lock Plug Type
COMPONENT LIST
1. Cable: color black, UL/CSA 2464 80/90°C 300V
18 AWG X 4C+Aluminum Shield
2. PVC Wire: U/C color: black/blue/red/yellow
3. Connector: 4PIN power DIN or EQ
4. ø2.4
5. 4PIN power DIN (female) MDP-002-4PIN (P530600004F0001)
AC CORD
(IEC 320 C13)
(NEMA 5-15P Hospital Grade)
183±5cm
SPECIFICATION:
1)
Length: 1.83m
2)
Color: black
3)
Plug: SL-22 (A)10 (V)125 Approval: UL/cUL
4)
Connector: SL-3(A)10 (V)125 Approval: UL/cUL
5)
Cable: SJT 18AWG/3C Approval: UL/CSA
20±2cm
cui.com
For more information, please visit the product page.
CUI Inc │ SERIES: ETMA 110W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/04/2013 │ page 5 of 5
REVISION HISTORY
rev.
description
date
1.0
initial release
06/05/2012
1.01
description and warranty updated
08/21/2012
1.02
replaced TUV approval with Nemko
09/04/2013
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
[email protected]
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
affect its safety or effectiveness.