Download Data Sheet

DATASHEET
Rev. A
WMHPU12 SERIES
90~264VAC Input Voltage Range
12 Watt, Single Outputs
Class II Limited Power Source
US, UK, and EU Plugs
Medical AC/DC Wall Mount Power Supplies
US Type
UK Type
EU Type
MEDICAL
FEATURES
• Class II Limited Power Source
• US, UK, and EU Types Available
• RoHS Compliant
• US Plug Medical Approvals: UL/cUL 60601-1: 2nd Edition
• Energy Star 2.0, Efficiency Level V Compliant
• Single Output
and TUV/T-mark EN60601-1:2nd Edition
• UK and EU Plug Medical Approvals: TUV/T-mark
• Wide Input Voltage Range: 90~264VAC, 47~63Hz
• Optional Output Connectors
EN60601-1: 2nd Edition
• Meets FCC Part-18 Class B and CISPR-11 EN55011 Class
• Over Load and Over Temperature Protection
B Emission Limits
DESCRIPTION
The WMHPU12 series of Class II medical AC/DC wall mount power supplies provides up to 12 Watts of continuous output
power. This series consists of single output models from 5VDC to 27VDC with a 90~264VAC input voltage range. These
supplies also have over load and over temperature protection. This series has UL/cUL 60601-1: 2nd Edition (US plug type
only) and TUV/T-mark EN60601-1: 2nd Edition (US, UK, and EU types) medical approvals. All supplies are RoHS and
Energy Star 2.0, Efficiency Level V compliant. These supplies also meet new CE requirements and are well suited for use in
health care applications. The WMHPU12 series has three types of plugs available: US (A Type), UK (B Type), and EU (C
Type). Optional output connectors are also available. Please call factory for ordering details.
07/18/2011
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 1 of 5
Rev. A
WMHPU12 Series
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II Limited Power Source
Medical AC/DC Wall Mount Power Supplies
SPECIFICATIONS: WMHPU12 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Operating Voltage Range
TEST CONDITIONS
Operating Input Voltage Range
Safety Approvals Input Voltage Range
Input Frequency
Input Current
Inrush Current
Low Line
High Line
Low Line
High Line
Min
Nom
Max
0.25
0.17
14
28
264
240
63
0.35
0.22
25
50
90
100
47
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Unit
VAC
Hz
A
A
A
A
OUTPUT SPECIFICATIONS
Output Voltage
Load Regulation
Line Regulation
Output Power
Output Current Range
Ripple & Noise (peak to peak)
Transient Response Time
Hold-Up Time
Start-Up Time
Temperature Coefficient
PROTECTION
Over Current Protection
Over Temperature Protection (See Note 2)
GENERAL SPECIFICATIONS
Efficiency
Dielectric Withstanding Voltage
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
Io = Full Load, Vin = 100VAC
0
10
0.3
-0.04
Verified by design/characteristics of integrated controller
110
-25
Io = Full Load, Vin = 230VAC
Primary to Secondary
77.7
5656
See Table
5
1
12
See Table
1
4
1
%
%
W
2
+0.04
%
ms
ms
s
%/°C
150
130
%
°C
%
VDC
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
MTBF
PHYSICAL SPECIFICATIONS
Weight
Derates linearly from 100% Load at 40°C to 50% load at 70°C
Operating Temp. at 25°C, calculated per MIL-HDBK-217F
US Plug (A Type)
UK Plug (B Type)
EU Plug (C Type)
US Plug (A Type)
Dimensions (L x W x H)
UK Plug (B Type)
EU Plug (C Type)
0
-40
0
0
40
70
85
95
75
100,000 hours
°C
°C
%
%
4.6oz (130g)
5.3oz (150g)
4.9oz (140g)
2.52 x 2.17 x 2.10 inches
(64.0 x 55.0 x 53.3 mm)
2.52 x 2.17 x 2.37 inches
(64.0 x 55.0 x 60.2 mm)
2.52 x 2.17 x 2.70 inches
(64.0 x 55.0 x 68.6 mm)
SAFETY & EMI
Safety Approvals
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
07/18/2011
US Plug (A Type)
UK & EU Plugs (B & C Types)
Vin = 220VAC
Vin = 110VAC
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
UL/cUL 60601-1: 2nd Edition
TUV/T-mark EN60601-1: 2nd Edition
TUV/T-mark EN60601-1: 2nd Edition
B
Class
B
Class
Page 2 of 5
WMHPU12 Series
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II Limited Power Source
Medical AC/DC Wall Mount Power Supplies
Rev. A
MODEL SELECTION TABLE
Model Number
Input Voltage Range
Output Voltage
Output Current
Total Regulation (3)
Output Power
WMHPU12A-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMHPU12A-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
*WMHPU12A-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
*WMHPU12A-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
*WMHPU12A-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
*WMHPU12A-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
*WMHPU12A-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
WMHPU12B-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMHPU12B-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
*WMHPU12B-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
*WMHPU12B-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
*WMHPU12B-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
*WMHPU12B-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
*WMHPU12B-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
WMHPU12C-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMHPU12C-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
*WMHPU12C-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
*WMHPU12C-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
*WMHPU12C-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
*WMHPU12C-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
*WMHPU12C-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
Plug Type
US
UK
EU
NOTES
1. Models with a “*” next to the model number are approved by CEC Level V.
2. Over Temperature Protection: Thermal shutdown by junction temperature controller; the adapter will automatically recover when the
fault condition is removed.
3. Models WMHPU12-102~104 needs to use AWG#18/6FT output cable in order to meet the total regulation specified.
Models WMHPU12-105~108 needs to use AWG#20/6FT output cable in order to meet the total regulation specified.
The regulation and efficiency will change if a different output cable is used.
4. Optional output connectors are available. Please call factory for ordering details.
Output Power (%)
DERATING CURVE
Ambient Temperature (°C)
07/18/2011
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 3 of 5
WMHPU12 Series
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II Limited Power Source
Medical AC/DC Wall Mount Power Supplies
Rev. A
MECHANICAL DRAWINGS
US TYPE (SUFFIX “A”)
Unit: inches (mm)
2.10±0.08 (53.3±2.0)
1.77±0.04 (45.0±1.0)
0.25 (6.3)
0.31 (8.0)
1.77±0.04 (45.0±1.0)
2.52±0.04 (64.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
1.22±0.04
(31.1±1.0)
LED (OPTIONAL)
0.19±0.04
(4.7±1.0)
1.81±0.04 (46.0±1.0)
UK TYPE (SUFFIX “B”)
Unit: inches (mm)
2.37±0.08 (60.2±2.0)
1.89±0.04 (48.0±1.0)
2.52±0.04 (64.0±1.0)
1.85±0.04 (47.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
1.22±0.04
(31.1±1.0)
LED (OPTIONAL)
0.25±0.04
(6.4±1.0)
1.81±0.04 (46.0±1.0)
EU TYPE (SUFFIX “C”)
Unit: inches (mm)
2.70±0.08 (68.6±2.0)
2.52±0.04 (64.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
LED (OPTIONAL)
07/18/2011
1.22±0.04
(31.1±1.0)
0.73±0.04
(18.5±1.0)
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
1.81±0.04 (46.0±1.0)
Page 4 of 5
Rev. A
WMHPU12 Series
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II Limited Power Source
Medical AC/DC Wall Mount Power Supplies
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working
closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to
provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss
your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
07/18/2011
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
5 Watson Brook Rd.
Exeter, NH 03833
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 5 of 5