Download Data Sheet

DATASHEET
Rev. A
DTMPU16 SERIES
90~264VAC Input Voltage Range
15 Watts, Single Outputs
Class I for A & C Types; Class II for B Type
3rd Edition Medical Approvals
AC/DC Desktop Power Supplies
IEC-320-C14
IEC-320-C8
IEC-320-C6
MEDICAL
FEATURES
• Class I for A & C Types; Class II for B Type
• Efficiency up to 85%
• RoHS Compliant
• IEC-320-C14, IEC-320-C8, and IEC-320-C6 Input Inlets
• Energy Star 2.0, Efficiency Level V Compliant
Available
• Meets FCC Part-18 Class B and EN55022 Class B
• Single Outputs
• Output Voltages Available from 5VDC to 36VDC
• Wide Input Voltage Range: 90~264VAC, 47~63Hz
Emission Limits
• ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd Edition) and
• Over Voltage and Over Load Protection
EN 60601-1:2006 (TUV/T-mark 3rd Edition) Medical
• Over Temperature Detection
Approvals
DESCRIPTION
The DTMPU16 series of medical AC/DC desktop power supplies provides up to 15 Watts of continuous output power. This
series consists of single output models with a 90~264VAC input voltage range. All units are UL94V-1, RoHS, and Energy
Star 2.0 Level V compliant. All models meet FCC Part-18 class B and EN55022 class B emission limits and have
ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd Edition) and EN 60601-1:2006 (TUV/T-mark: 3rd Edition) medical approvals.
These units also meet new CE requirements and have been 100% burn-in tested. The DTMPU16 series has three types of
input inlets available: IEC-320-C14 (A Type), IEC-320-C8 (B Type), and IEC-320-C6 (C Type).
02/22/2012
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 1 of 5
Rev. A
DTMPU16 Series
Class I & II, Single Outputs
3rd Edition Medical Approvals
15W AC/DC Desktop Power Supplies
SPECIFICATIONS: DTMPU16 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Nom
Max
0.25
0.17
18
38
0.25
264
240
63
0.33
0.18
23
45
0.3
Unit
INPUT SPECIFICATIONS
Operating Input Voltage Range
Safety Approvals Input Voltage Range
Operating Voltage Range
Input Frequency
Low Line
Input Current
High Line
Low Line
Inrush Current
High Line
No Load Power Consumption
OUTPUT SPECIFICATIONS
Output Voltage
Load Regulation
Line Regulation
Output Power
Output Current Range
Ripple & Noise (peak to peak)
Transient Response Time
Hold-Up Time
Start-Up Time
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Current Protection
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
No Load, Vin = 230VAC
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
Io = Full Load, Vin = 100VAC
The parameter is not subject to production test-verified by
design/characterization of integrated controller
GENERAL SPECIFICATIONS
Efficiency
Dielectric Withstanding Voltage
Isolation Resistance
Safety Ground Leakage Current
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
MTBF
PHYSICAL SPECIFICATIONS
Weight
Io = Full Load, Vin = 230VAC
Primary to Secondary
A, B, and C types
Primary to Ground
A and C types
Test Voltage = 500VDC
A and C types
Io = Full Load, Vin = 240VAC
Derating linearly from 100% Load at 50°C to 50% load at 70°C
Operating Temp. at 25°C; calculated per MIL-HDBK-217F
A, B, and C types
A type
Dimensions (L x W x H)
B and C types
SAFETY
Safety Approvals
EMI Requirements for EN55022
EMI Requirements for FCC PART-18
02/22/2012
0
1
0
10
-0.04
Over Temperature Detection (See Note 1)
AC Inlets
90
100
47
A Type
B Type
C Type
See Table
3
5
0.5
1
15
See Table
0.5
1
4
16
0.5
1
+0.04
VAC
Hz
A
A
A
A
W
%
%
W
%
ms
ms
s
%/°C
112
110
132
150
%
%
-20
125
°C
73
5656
2828
50
85
%
VDC
MΩ
mA
0.3
0
-40
0
0
50
70
85
95
95
100,000 hours
°C
°C
%
%
Approx. 6oz (170g)
4.11 x 1.65 x 1.22 inches
(104.4 x 42.0 x 31.0 mm)
3.90 x 1.65 x 1.22 inches
(99.0 x 42.0 x 31.0 mm)
IEC-320-C14
IEC-320-C8
IEC-320-C6
ANSI/AAMI ES 60601-1:2005 (UL/cUL: 3rd Ed.), EN 60601-1:2006 (TUV/T-mark: 3rd Ed.), CE
Vin = 230VAC, 50Hz
B
Class
Vin = 120VAC, 60Hz
B
Class
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 2 of 5
DTMPU16 Series
Class I & II, Single Outputs
3rd Edition Medical Approvals
15W AC/DC Desktop Power Supplies
Rev. A
MODEL SELECTION TABLE
Model Number
Input Voltage Range
DTMPU16A-102
DTMPU16A-103
DTMPU16A-104
DTMPU16A-105
DTMPU16A-106
DTMPU16A-107
DTMPU16A-108
DTMPU16A-109
DTMPU16A-110
DTMPU16B-102
DTMPU16B-103
DTMPU16B-104
DTMPU16B-105
DTMPU16B-106
DTMPU16B-107
DTMPU16B-108
DTMPU16B-109
DTMPU16B-110
DTMPU16C-102
DTMPU16C-103
DTMPU16C-104
DTMPU16C-105
DTMPU16C-106
DTMPU16C-107
DTMPU16C-108
DTMPU16C-109
DTMPU16C-110
90 ~ 264 VAC
90 ~ 264 VAC
90 ~ 264 VAC
Output Voltage
Output Current
5 ~ 6 VDC
6 ~ 8 VDC
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 36 VDC
5 ~ 6 VDC
6 ~ 8 VDC
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 36 VDC
5 ~ 6 VDC
6 ~ 8 VDC
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 36 VDC
2.60 ~ 2.16 A
2.16 ~ 1.62 A
1.87 ~ 1.36 A
1.36 ~ 1.15 A
1.15 ~ 0.93 A
0.93 ~ 0.71 A
0.71 ~ 0.55 A
0.55 ~ 0.45 A
0.45 ~ 0.41 A
2.60 ~ 2.16 A
2.16 ~ 1.62 A
1.87 ~ 1.36 A
1.36 ~ 1.15 A
1.15 ~ 0.93 A
0.93 ~ 0.71 A
0.71 ~ 0.55 A
0.55 ~ 0.45 A
0.45 ~ 0.41 A
2.60 ~ 2.16 A
2.16 ~ 1.62 A
1.87 ~ 1.36 A
1.36 ~ 1.15 A
1.15 ~ 0.93 A
0.93 ~ 0.71 A
0.71 ~ 0.55 A
0.55 ~ 0.45 A
0.45 ~ 0.41 A
Total
Regulation (2)
5%
5%
5%
5%
5%
5%
3%
3%
3%
5%
5%
5%
5%
5%
5%
3%
3%
3%
5%
5%
5%
5%
5%
5%
3%
3%
3%
Output Power
13W
13W
15W
15W
15W
15W
15W
15W
15W
13W
13W
15W
15W
15W
15W
15W
15W
15W
13W
13W
15W
15W
15W
15W
15W
15W
15W
Class
AC Inlet
Class I
IEC-320-C14
Class II
IEC-320-C8
Class I
IEC-320-C6
NOTES
1. Over Temperature Detection: Thermal shutdown by junction temperature controller. When the power system interruption is isolated, the
product will re-start after recovering by hand.
2. Models DTMPU16-102~103 need to use AWG#16/4FT output cable in order to meet the total regulation specified.
Models DTMPU16-105~110 need to use AWG#18/4FT output cable in order to meet the total regulation specified.
The regulation and efficiency will change if a different output cable is used.
DERATING CURVE
Output Power (%)
120
100
80
60
40
20
0
0
10
20
30
40
50
60
70
80
Ambient Temperature (°C)
02/22/2012
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 3 of 5
Rev. A
DTMPU16 Series
Class I & II, Single Outputs
3rd Edition Medical Approvals
15W AC/DC Desktop Power Supplies
MECHANICAL DRAWING
TYPE A
Unit: inches (mm)
0.02±0.04 (0.5±1.0)
0.04±0.04 (1.0±1.0)
1.22±0.04 (31.0±1.0)
1.65±0.04 (42.0±1.0)
4.11±0.04 (104.4±1.0)
“A” TYPE
AC INLET
IEC-320-C14
TYPES B & C
Unit: inches (mm)
3.90±0.04 (99.0±1.0)
0.02±0.04 (0.5±1.0)
0.04±0.04 (1.0±1.0)
AC INLET
IEC-320-C8
1.22±0.04 (31.0±1.0)
1.65±0.04 (42.0±1.0)
“B” TYPE
“C” TYPE
AC INLET
IEC-320-C6
02/22/2012
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 4 of 5
Rev. A
DTMPU16 Series
Class I & II, Single Outputs
3rd Edition Medical Approvals
15W AC/DC Desktop Power Supplies
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working
closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to
provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss
your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
02/22/2012
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
5 Watson Brook Rd.
Exeter, NH 03833
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 5 of 5