Echotel ® Model 961 Ultrasonic Level Switches for Hygienic Applications D E S C R I P T I O N Echotel® Model 961 Ultrasonic Level Switches use pulsed signal technology for superior performance in difficult process conditions, and to provide excellent immunity from sources of electrical noise interference. Extensive self-testing of the electronics and transducer make this advanced switch suitable for a wide variety of hygienic level switch applications. Model 961 is offered with either a 5-amp relay or a current shift electronics. The relay output electronics has a DPDT relay for level detection, and a SPDT malfunction relay. The current shift version indicates 8 mA during normal operation, 16 mA as a level alarm condition and a user-selectable 3.6 or 22 mA malfunction indication. The Model 961 switches are designed for the stringent requirements of hygienic applications. These switches are offered with a deep-drawn 304 stainless steel housing that is favorable to CIP and SIP procedures. A 20 Ra surface finish is featured with hygienic Model 961 transducers. These transducers are available with 11⁄2" and 2" Tri-Clamp® fittings, and DN65 Varivent® Type N flanges. Model 961 (with Varivent® fitting) F E A T U R E S • Self-test technology provides unsurpassed reliability and testing of electronics, transducer, piezoelectric crystals, and electromagnetic noise • 20 Ra surface finish with Tri-Clamp and Varivent fittings • Adjustable time delay for turbulent aerated liquids • Tip-sensitive transducer measures level within 1⁄4" of the vessel bottom • Pulsed signal technology provides superior performance in difficult process conditions Model 961 (with Tri-Clamp® fitting) P U L S E D S I G N A L T E C H N O L O G Y Model 961 switches use pulsed signal technology to detect the presence or absence of liquid in an ultrasonic transducer gap. The transducer uses a pair of piezoelectric crystals that vibrate at a given frequency when subjected to an applied voltage. The transmit crystal converts the applied voltage from the electronics into an ultrasonic signal. When liquid is present in the gap, the receive crystal senses the ultrasonic signal and converts it back into an electrical signal. This signal is sent to the electronics to indicate a wet gap condition. When no liquid is present, the ultrasonic signal is attenuated and is not detected by the receive crystal. Ultrasonic transducer Transmit crystal Challenging process conditions like aeration, suspended solids, and high viscosities are overcome by using pulsed signal circuitry in 961 switches. Unlike many tuning forks, pulsed signal ultrasonic switches do not need to be configured for different media densities, making these units the most universally applied level switches on the markey today. ADVANCED SELF-TEST Ultrasonic switches are often used as the last means of detecting whether a cer Transdu ls ta process vessel will overflow and cause a s & cry s ic spill of potentially hazardous liquids, or Electron e empty out and possibly cavitate the Nois pumps. In these critical applications it is desirable to have a method of periodically testing the ultrasonic switch to ensure that it is functioning properly. Receive crystal Ultrasonic signal transmission across transducer gap AND DIAGNOSTICS A microprocessor in the 961 electronics continuously monitors all self-test data. Should a fault occur, the microprocessor can determine whether the malfunction is due to the electronics, transducer, piezoelectric crystals, or the presence of environmental noise. A pushbutton and Fault LED is used to assist in troubleshooting the switch: 5 One flash of the Fault LED indicates a problem with the transducer or piezoelectric crystals Model 961 switches feature an advanced self-test technology that not only tests the electronics, transducer, and piezoelectric crystals, but also tests for the presence of industrial sources of environmental noise. Should the switch detect any problems a malfunction output is generated to alarm the user, and a red LED is lit to indicate an alarm condition. 2 55 Two flashes of the Fault LED indicates a problem with one of the electronics boards 555 Three flashes of the Fault LED indicates excessive levels of environmental noise ADJUSTABLE TIME DELAY Turbulence and splashing can cause some fixed time response switches to produce false level alarms. Model 961 switches overcome this difficulty with an adjustable time delay feature. A potentiometer allows a 1⁄2 – 45-second delay to be set to disregard waves or splashes, and reliably detect the true liquid level. A P P L I C A T I O N S Model 961 Switches may be used for high or low level alarm, empty pipe detection and pump protection in a wide variety of hygienic applications. Available with a 304 stainless steel electronics housing and a 20 Ra surface finish transducer, typical applications include: Fill Line Spray Ball High Alarm • CIP/SIP day tank point level • Water for Injection (WFI) storage • Liquid chromatography skids • Brewery fill lines Pump Protection • Condensate receiver tanks Low Alarm 3 E L E C T R O N I C S M O D E L 9 6 1 S P E C I F I C A T I O N S W I T H Supply Voltage Relay Outputs Relay Ratings R E L A Y DPDT: SPDT: Fail-safe Power Consumption Ambient Temperature M O D E L 9 6 1 W I T H 18 to 32 VDC, or 102 to 265 VAC, 50/60 Hz One DPDT level relay and one SPDT malfunction relay 5 amps @ 120 VAC, 250 VAC, and 30 VDC, 0.4 amp @ 110 VDC 5 amps @ 120 VAC, 250 VAC, and 30 VDC, 0.15 amp @ 125 VDC Selectable for high or low level Less than 3 watts -40° to +160° F (-40° to +70° C) C U R R E N T Supply Voltage Current Shift Output S H I F T O U T P U T 12 to 35 VDC 8 mA normal operation, 16 mA level alarm (±1 mA) 3.6 mA or 22 mA selectable fault signal (±1 mA) 104 ohms with 12 VDC input, 1100 ohms with 35 VDC input Selectable for high or low level Less than 1 watt -40° to +160° F (-40° to +70° C) Loop Resistance Fail-safe Power Consumption Ambient Temperature T R A N S D U C E R Operating Temperature Maximum Pressure O U T P U T S P E C I F I C A T I O N S Operating Frequency Surface Finish -40° to +325° F (-40° to +163° C) 2000 psi (138 bar) 1500 psi (103 bar) 2 MHz 20 Ra (when ordered with 4th digit code S) P E R F O R M A N C E S P E C I F I C A T I O N S Repeatability Response Time Time Delay Self-Test 1" and 2": 3" to 130": Automatic: Manual: Shock Class Vibration Class Humidity Electromagnetic Compatibility 4 ±0.078" (± 2 mm) ⁄2 second typical Variable 0.5 – 45 seconds on rising and falling levels Continuously verifies operation of electronics, transducer, piezoelectric crystals, and electrical noise Push button verifies operation of electronics, transducer, and piezoelectric crystals ANSI/ISA-S71.03 Class SA1 ANSI/ISA-S71.03 Class VC2 0-99%, non-condensing Meets CE requirements EN 61326 1 D I M E N S I O N A L I N C H E S S P E C I F I C A T I O N S ( m m ) 5.24 (133) 5.24 (133) 5.75 (146) 5.75 (146) 1.05 (26) Typical 2.70 (69) Typical Actuation length with Tri-Clamp fitting Actuation length with Varivent fitting 0.25 (6) ∅ .875 (22) 0.25 (6) ∅ .875 (22) Model 961 with Tri-Clamp® fitting Model 961 with DN65 Varivent® Flange 5 M O D E L 9 6 1 N U M B E R E L E C T R O N I C S BASIC MODEL NUMBER 961 Relay or current shift electronics INPUT POWER 2 5 7 18–32 VDC line-powered 12–35 VDC loop-powered 102 to 265 VAC line-powered OUTPUT SIGNAL 0 D mA current shift (available with Input Power code 5) 5-amp gold flash relays (available with Input Power codes 2 or 7) COVER 0 1 Standard cover Glass window cover MOUNTING 0 Integral AGENCY APPROVAL 7 FM/CSA Non-incendive (available with Output Signal codes 0, 8 & D and Housing codes 4 & 5) HOUSING & CONDUIT CONNECTION 4 5 9 6 6 1 A 0 7 Deep drawn 304 stainless steel w/ 1⁄2" NPT dual conduit entries Deep drawn 304 stainless steel w/M20 dual conduit entries M O D E L 9 6 1 N U M B E R S I N G L E P O I N T T R A N S D U C E R TRANSDUCER UNIT OF LENGTH A M English (length in inches) Metric (length in centimeters) MATERIALS OF CONSTRUCTION S 316/316L with 20 Ra hygienic finish (use only with Process Connection codes 3T, 4T, or VV) PROCESS CONNECTIONS 3T 4T VV 11⁄2" Tri-Clamp® 16 AMP fitting 2" Tri-Clamp® 16 AMP fitting DN65 – Varivent® Type N flange ACTUATION LENGTH (unit of length specified in second digit) 2" to 130" in 1" increments Example: 4 inches = 004 5 cm to 330 cm in 1 cm increments Example: 6 centimeters = 006 9 1 A 7 Q U A L I T Y The quality assurance system in place at Magnetrol® guarantees the highest level of quality throughout the company. MAGNETROL is committed to providing full customer satisfaction both in quality products and The MAGNETROL quality assurance system is registered to ISO 9001 affirming its commitment to known international quality standards providing the strongest assurance of product and service quality available. quality service. E S P Expedite Ship Plan Several ECHOTEL Model 961 units are available for quick shipment, usually within one week after factory receipt of a complete purchase order, through the Expedite Ship Plan (ESP). ESP service may not apply to orders of ten units or more. Contact your local representative for lead times on larger volume orders, as well as other products and options. To take advantage of ESP, simply match the color coded model number codes (standard dimensions apply). W A R R A N T Y All MAGNETROL electronic level and flow controls are warranted free of defects in materials or workmanship for one full year from the date of original factory shipment. If returned within the warranty period; and, upon factory inspection of the control, the cause of the claim is determined to be covered under the warranty; then, MAGNETROL will repair or replace the control at no cost to the purchaser (or owner) other than transportation. MAGNETROL shall not be liable for misapplication, labor claims, direct or consequential damage or expense arising from the installation or use of equipment. There are no other warranties expressed or implied, except special written warranties covering some MAGNETROL products. For additional information, see Instruction Manual 51-646. 705 Enterprise Street • Aurora, Illinois 60504-8149 • 630-969-4000 • Fax 630-969-9489 [email protected] • www.magnetrol.com Copyright © 2013 Magnetrol International, Incorporated. All rights reserved. Printed in the USA. Performance specifications are effective with date of issue and are subject to change without notice. Magnetrol, Magnetrol logotype and Echotel are registered trademarks of Magnetrol International, Incorporated. AS-Interface® is a registered trademark of the AS-International Association Kynar® is a registered trademark of Elf Atochem North America, Inc. Hastelloy® is a registered trademark of HAYNES INTERNATIONAL, INC. (DELAWARE CORPORATION) Monel® is a registered trademark of the INCO family of companies Tri-Clamp® is a registered trademark of Ladish Co. Varivent® is a registered trademark of Tuchenhagen GmbH LTD BULLETIN: 51-160.3 EFFECTIVE: October 2013 SUPERSEDES: August 2011