English User's Manual

User’s Guide
400A AC Clamp Meter
Model MA200
Introduction
Thank you for selecting the Extech MA200 AC Clamp Meter. This professional meter, with proper care,
will provide years of safe reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
SAFETY NOTES










Do not exceed the maximum allowable input range of any function
Do not apply voltage to meter when resistance function is selected.
Set the function switch OFF when the meter is not in use.
Remove the battery if meter is to be stored for longer than 60 days.
If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
This device is not a toy and must not reach children’s hands. It contains hazardous objects as
well as small parts that the children could swallow. In case a child swallows any of them,
please contact a physician immediately
Do not leave batteries and packing material lying around unattended; they can be dangerous
for children if they use them as toys
In case the device is going to be unused for an extended period of time, remove the batteries
to prevent them from training
Expired or damaged batteries can cause cauterization on contact with the skin. Always,
therefore, use suitable hand gloves in such cases
See that the batteries are not short-circuited. Do not throw batteries into the fire.
WARNINGS





Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
Do not measure current on a circuit whose voltage exceeds 240V.
When changing ranges using the selector switch always disconnect the test leads from the
circuit under test.
Do not exceed the maximum rated input limits.
OVERVOLTAGE CATEGORY III 600V
This meter meets the IEC 610-1-95 standard for OVERVOLTAGE CATEGORY III 600V. Cat III
600V meters are protected against overvoltage transients in fixed installation at the distribution
level. Examples include switches in the fixed installation and some equipment for industrial use with
permanent connection to the fixed installation.
2
MA200-en-EU_V2.4 8/14
Meter Description
1.
Transformer jaws
2.
MAX Hold button
3.
Hold button
4.
LCD Display
5.
COM Input Terminal
6.
Jaw opening trigger
7.
Function select switch
8.
Voltage, Resistance Input Terminal
9.
Mode Select Button
10.
Backlight Button
11.
Range Select Button
AC
DC
AUTO
MAX
•)))
HOLD
Alternating current
Direct currrent)
AutoRange mode
MAX Hold mode
Diode test mode
Audible Continuity
Data Hold mode
Low Battery icon
3
MA200-en-EU_V2.4 8/14
Operation
NOTICES: Read and understand all warning and precaution statements listed in the safety
section of this operation manual prior to using this meter. Set the function select switch to the OFF
position when the meter is not in use.
AC Voltage Measurements
1.
Set the rotary function switch to the VAC position.
2.
Insert the black test lead banana plug into the negative (COM) jack
and the red test lead banana plug into the positive (V/) jack
3.
Touch the test probe tips to the circuit under test
4.
Read the voltage in the display. The display will indicate the proper
decimal point and value
DC Voltage Measurements
1.
Set the rotary function switch to the VDC position.
2.
Insert the black test lead banana plug into the negative (COM) jack and
the red test lead banana plug into the positive (V/) jack
3.
Touch the test probe tips to the circuit under test. Be sure to observe the
correct polarity (red lead to positive, black lead to negative)
4.
Read the voltage in the display. The display will indicate the proper
decimal point and value. If the polarity is reversed, the display will show () minus before the value
4
MA200-en-EU_V2.4 8/14
AC Current Measurements
WARNING: Ensure that the test leads are disconnected from the meter
before making current clamp measurements.
1. Set the Function switch to the 400 or 200A range. If the range of the
measured is not known, select the higher range first then move to the
lower range if necessary.
2. Press the trigger to open jaw. Fully enclose one conductor to be measured.
3. The clamp meter LCD will display the reading.
Resistance Measurements
•))) position.
1. Set the function switch to the 
2. Insert the black test lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (V jack.
3. Touch the test probe tips across the circuit or part under test. It is best to
disconnect one side of the part under test so the rest of the circuit will not
interfere with the resistance reading.
4. Read the resistance in the display. The display will indicate the proper
decimal point and value.
Continuity Check
1. Set the function switch to the 
•))) position.
2. Push the mode button to indicate •))) on the display.
3. Insert the black lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (V) jack.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 30, the audible signal
will sound. If the circuit is open, the display will indicate “OL.”.
5
MA200-en-EU_V2.4 8/14
Diode Test
1. Turn the rotary switch to the 
•))) position.
2. Insert the black test lead banana plug into the negative (COM) jack and the
red test lead banana plug into the positive (V jack.
3. Push the mode button to indicate
on the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate
400 to 700mV. Reverse voltage will indicate “OL”. Shorted devices will
indicate near 0mV. Shorted devices will indicate near 0mV and an open
device will indicate “OL” in both polarities.
Data Hold
To freeze the LCD meter reading, press the data hold button. The data hold button is located on the
left side of the meter (bottom button). While data hold is active, the HOLD display icon appears on
the LCD. Press the data hold button again to return to normal operation.
MAX Hold
To hold the highest reading on the LCD, press the MAX button. The max hold button is located on
the left side of the meter (top button). While data hold is active, the MAX display icon appears on
the LCD. The meter reading will not change as readings change, rather it will only display the
highest reading encountered since the max hold button was pressed. Press the max hold button
again to return to normal operation.
6
MA200-en-EU_V2.4 8/14
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input terminals and turn OFF the meter before opening the case. Do not operate with open
case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If
the meter is not to be used for periods of longer than 60 days, remove the batteries and store them
separately.
Battery Replacement
1. Remove the two rear Phillips head screws
2.
Open the battery compartment
3.
Replace the two 1.5V AAA batteries.
4.
Re-assemble the meter
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever batteries
/ accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
7
MA200-en-EU_V2.4 8/14
Specifications
Function
AC Current
AC Voltage
DC Voltage
Resistance 
Jaw size
Display
Continuity
Diode Test
AC V bandwidth
AC A bandwidth
Low battery indication
Overrange indication
Auto Power OFF
Measurement rate
Input Impedance
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Batteries
Weight
Size
Safety Approval
Safety information
Range
Accuracy (of reading)
2.000 ACA
± (2.5% + 10 digits)
20.00 ACA
± (2.5% + 4 digits)
200.0 ACA
± (2.5% + 4 digits)
400 ACA
± (3.0% + 5 digits)
200.0mV,
± (1.5% + 30 digits)
2.000V
± (1.5% + 3 digits)
20.00V
200.0V
600V
± (2.0% + 4 digits)
200.0mV
± (0.5% + 5 digits)
2.000V
± (1.2% + 3 digits)
20.00V
200.0V
600V
± (1.5% + 3 digits)
200.0
± (1.0% + 4 digits)
2.000k
± (1.5% + 2 digits)
20.00k
200.0k
2.000M
± (2.0% + 3 digits)
20.00M
± (3.0% + 5 digits)
23mm (0.9") approx.
3-1/2 digits (2000 counts) LCD
Audible tone < 120 approx.
Open circuit voltage < 1.5VDC;
Test current 0.3mA (typical)
50Hz to 400Hz
50/60Hz
is displayed
“OL” is displayed
After 15 minutes
2 per second, nominal
7.8MΩ (V DC and V AC)
5ºC to 40ºC (41ºF to 104ºF)
o
o
o
o
-20 C to 60 C (-4 F to 140 F)
Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF)
<80%
3000m (9800ft)
(2) 1.5V AAA batteries
200g (0.44lb)
200 x 50 x 35mm (7.87” x 1.97” x 1.38”)
CE
For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (1995): EN61010-1 (1995) Overvoltage
Category III 600V, Pollution Degree 2.
Copyright © 2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com 8
MA200-en-EU_V2.4 8/14