English User's Manual

 Heat Stress WBGT Meter Model HT200 USER GUIDE EXTECH INSTRUMENTS Introduction Thank you for selecting the Extech Instruments Model HT200 Heat Stress WBGT Meter. The HT200 accurately determines the Heat Stress by factoring a combination of parameters: Humidity, Temperature, and Direct Solar Radiation. These factors affect how high body temperatures rise, as well as the ability to cool down. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.extech.com) to check for the latest version of this User Guide, Product Updates, Product Registration, and Customer Support. Features  Fast responding capacitance sensor  Accurate measurements for: Wet bulb globe temperature (WBGT), Black globe temperature (TG), Humidity (%RH), Air temperature (TA), Wet bulb (WET), and Dew point (DEW)  Maximum, minimum, and data hold Recording  Low battery indication  LCD display with LED backlighting  Manual data recording (up to 50 reading sets)  View data records  Selectable temperature units C or F  Brass black ball 50mm (2”) diameter  WBGT high and low audible and visual alarm functions  Auto Power OFF with disable function Safety WARNING: Avoid electromagnetic interference (EMI) to avoid erratic meter readings. WARNING: The measurements will be inaccurate if the black ball is touched during a test. WARNING: If the black ball is damage or deformed in shape, inaccurate measurement values will result. CAUTION: Please place the meter in a warm temperature and low humidity environment for 24 hours after the meter has been used in an overly humid environment. CAUTION: Please have the meter calibrated yearly for the best results. WARNING: Do not store this device in direct sunlight or in overly hot and/or damp areas. 2 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Meter Description 1.
Up Arrow/MAX‐MIN button 2.
MODE/SET button 3.
Power ON‐OFF button 4.
UNIT/DATA HOLD button 5.
LCD Display 6.
Temperature & Humidity sensor 7.
Black Globe sensor 8.
Backlight/MEM button 9.
SEL/Alarm button 10.
Down arrow/REC button 11.
AC Adaptor power jack Note: Battery compartment on back 7
6
5
4
8
3
2
1
9
10
11
3 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS LCD Display Description 1.
Alarm symbol 2.
DATA HOLD symbol 3.
WBGT Mode 4.
Max reading symbol 5.
Air temperature mode 6.
Black globe temperature mode 7.
MIN (minimum) reading symbol 8.
Wet bulb temperature mode 9.
USB symbol (unused in HT200) 10. Dew point temperature mode 11. Primary display area 12. Temperature units 13. Relative humidity mode 14. Secondary display area 15. Stored reading ‘R’ alert & Memory Recall ‘M’ icon 16. Time symbol (unused in HT200) 17. Date symbol (unused in HT200) 18. Data Record mode symbol 19. Set Mode symbol 20. Indoor symbol (WBGT) 21. Outdoor symbol (WBGT) 22. Low battery symbol 23. Auto power off symbol 4 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Operation Powering the meter Press the Power
button to power the meter ON or OFF. To show all of the display symbols button. The firmware version will at once: with the meter OFF, press and hold the Power
display for one second (after the button is released). Auto Power Off The HT200 automatically powers off after approximately 15 minutes of inactivity. Enable‐Disable Auto Power Off With power ON, press and hold the Power
button for at least 2 seconds to disable or enable auto power‐off. The clock symbol appears on the LCD screen when the Auto Power OFF feature is enabled. The symbol disappears when the Auto Power OFF feature is disabled. Note: The auto power‐off will be disabled if the SET function or the alarm is enabled. Alarm ON‐OFF Press and hold the SEL/Alarm button for at least 2 seconds to turn the alarm function ON (armed) or OFF (disarmed). When armed the ALARM display icon will appear. When disarmed, the ALARM display icon disappears. Note: The alarm function is disabled when the HOLD, SET, or VIEW DATA RECORDSD function is enabled. Display Backlight Press the Backlight/Mem
button to turn on the LCD backlight. The LCD backlight will automatically turn off after 15 seconds. Selecting Temperature C/F units of measure Press the Unit/DH
button to select the desired temperature unit of measure. Mode Selection Press the Mode/Set button to change the mode. The available modes are WBGT, TA, %RH, TG, WET, DEW, and back to WBGT (see definitions below): o Wet bulb globe temperature (WBGT) o Air temperature (TA) o Humidity (%RH) o Black globe temperature (TG): monitors the effects of direct solar radiation o Wet bulb temperature (WET) o Dew point temperature (DEW) button to toggle IN (indoor: without direct sun In the WBGT mode, press the Sel/Alarm exposure) and OUT (outdoor: with direct sun exposure) modes. Note: The mode function is unavailable in the HOLD function and SET function modes. 5 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Data Hold Press and hold the Unit/DH button for 2 seconds to enable the Data Hold function. The HOLD display icon will appear the displayed reading will be frozen until the button is pressed and held again. Note: The HOLD function is unavailable when the SET function is enabled. Manual Data Recording Press the REC button to manually record (store) the current reading. The REC and R symbol will briefly appear on the LCD while the reading is stored. Note: The Record Reading function is unavailable in the HOLD, SET, and VIEW DATA RECORDS modes. View Data Records Press and hold the Backlight/Mem button to turn access (or exit) the View Data Records mode. When the View Data Records mode is accessed, the record number is shown on the bottom of the display, the associated reading is shown at the center or the LCD, and the ‘M’ memory and ‘REC’ icons will be visible. Use the arrow buttons to scroll through the stored readings. Press the Mode/Set button to change the measurement type (i.e. WBGT, TA, %RH) for the currently selected record. Press the Unit/DH
button to switch the temperature unit. button to exit the viewing data records mode. Press and hold the Backlight/Mem Note: The View Data Records mode is unavailable when the HOLD or SET function is enabled. Maximum‐Minimum (MAX‐MIN) Reading Recording Press and hold the Max/Min button to enter the MAX‐MIN recording mode. The LCD will show the MAX icon along with the maximum temperature measurement. button to toggle between Maximum (MAX) and Minimum (MIN) Use the Max/Min button for more than 2 seconds to exit the readings. Press and hold the Max/Min maximum/minimum mode. Note: The MAX‐MIN mode is unavailable when the HOLD, SET, or VIEW DATA RECORDS function is enabled. 6 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Setting Mode Press and hold the button to access the Setting mode where a series of parameters can be customized. The available parameters are explained individually below. To scroll through the parameters use the button. Press the button to start editing a particular parameter as explained below. WBGT HIGH ALARM THRESHOLD (HI)
1.
Step to the WBGT High Alarm Threshold screen using the button. button to begin editing; the display 2.
At the High Alarm Threshold screen, press the will begin flashing. 3.
Press the
4.
Use the up and down arrow buttons to set the WBGT High Alarm Threshold value. 5.
Use the 6.
7.
button again to save the setting. Press the The meter will now aurally and visually alarm when the High Alarm Threshold has been exceeded. To arm/disarm the alarm function, refer to the ‘ALARM ON‐OFF’ section. button to select the desired temperature units. button to select the digit to edit. WBGT LOW ALARM THRESHOLD (LO) 1.
Step to the WBGT Low Alarm Threshold screen using the button. 2.
At the Low Alarm Threshold screen, press the will begin flashing. 3.
Press the
4.
Use the up and down arrow buttons to set the WBGT Low Alarm Threshold value. 5.
Use the button to begin editing; the display button to select the desired temperature units. button to select the digit to edit. button again to save the setting. 6. Press the The meter will now aurally and visually alarm when the Low Alarm Threshold is reached. To arm/disarm the alarm function, refer to the ‘ALARM ON‐OFF’ section. 7 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS CLEAR STORED READINGS 1.
Step to the ‘Clear Stored Reading’ screen (example shown above) using the button. The number at the bottom of the display indicates the number of readings stored (50 maximum). 2.
3.
Press the button and the display will begin flashing. Use the up or down arrow button to select YES or NO. Select YES to erase all stored readings. Press NO to retain the readings in memory. 4.
Press the button to perform the selected action (YES or NO). AIR TEMPERATURE OFFSET 1.
Step to the Air Temperature Offset screen using the above). 2.
3.
button to begin editing; the display will begin flashing. Press the Use the up and down arrow buttons to set the offset value (from ‐9.9 to +9.9) 4.
Press the
button to select the desired temperature units. 5.
Press the button again to save the setting. button (example screen shown RELATIVE HUMIDITY % OFFSET 1.
Step to the RH% Offset screen using the 2.
3.
button to begin editing; the display will begin flashing. Press the Use the up and down arrow buttons to set the offset value (from ‐9.9 to +9.9). 4.
Press the button (example screen shown above). button again to save the setting. 8 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS BLACK GLOBE TEMPERATURE OFFSET 1.
Step to the Black Globe Temperature Offset screen using the screen shown above). 2.
3.
Press the button to begin editing; the display will begin flashing. Use the up and down arrow buttons to set the offset value (from ‐9.9 to +9.9). 4.
Press the
button to select the desired temperature units. 5.
Press the button again to save the setting. button (example Battery Replacement When the battery icon ( ) appears on the LCD, the 9V battery must be replaced. Turn off the power and unplug all cables. Remove the battery from the rear battery compartment and replace it with a new 9V battery. Please observe correct battery polarity and do not switch the unit ON until the battery door is closed and secured. When the meter is not in use, please remove the battery. Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold. Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end‐of‐life devices to a designated collection point for the disposal of electrical and electronic equipment. AC Adaptor Power The AC power adaptor jack is located on the bottom of the meter. See specifications below: External AC to DC adapter: Voltage 9 VDC (8~14 VDC Max) / 500mW. Socket: Pin positive, Ground casing negative. External diameter 5.5mm internal diameter 2.1mm 9 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Specifications Wet bulb globe temperature (WBGT) Range Resolution Accuracy @15~40 oC °C 0~59.0 0.1 ±1.0 °F 32.0~ 138.0 0.1 ±1.8 °C 0~56.0 0.1 ±1.5 °F 32.0~ 132.0 0.1 ±2.7 Unit Indoor & Outdoor Without sunlight Outdoor With sunlight Indoor & Outdoor without sunlight: WBGT = (0.7×WET)+(0.3×TG) Outdoor with sunlight: WBGT=(0.7×WET)+(0.2×TG)+(0.1×TA)
Air Temperature (TA) Unit Range Resolution °C °F 0~50.0 32.0~122.0 0.1 0.1 o
Accuracy @15~40 C ±0.8 ±1.5 Black globe temperature (TG) Unit °C °F Range 0~80.0 32.0~176.0 Resolution 0.1 0.1 Accuracy @15~40 oC ±0.6 ±1.1 Relative Humidity (%RH) Measurement Range Accuracy Resolution 1%~99% ±3.0%RH (20~80%) ±5.0%RH (<20% or >80%) 0.1% 10 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Dew point temperature (DEW) Unit Range Resolution °C ‐35.3~48.9 0.1 °F ‐31.5~120.1 0.1 The value is calculated from the RH and Air temperature Wet bulb temperature (WET) Unit Range Resolution °C ‐21.6~50.0 0.1 °F ‐6.9~122.0 0.1 The value is calculated from the RH and air temperature Meter Dimensions300 x 70 x 50mm (11.81 x 2.76 x 1.97in.) [L x W x H] LCD 52 mm (W) x 36mm (L) [2.05 in (W) x 1.42 in (L)] Ball Dimensions 50mm diameter, 19mm height (2” diameter and 0.75” height) Weight 220g (7.76oz.) without batteries Operating Altitude 2000m (6562ft) max. Sampling Rate One per second (1 Hz) Operating Temperature & Humidity 0°C to +50°C (32°F to 122°F), <95%RH. (Non‐condensing) Storage Temperature & Humidity ‐10°C to +50°C (14°F to 122°F), <70% RH. (Non‐condensing) Power supply 9V battery or AC Adaptor 100~240V DC 9V/ 0.5A (9mm) Battery life 200 hours Accessories 9V battery, Carrying Case, and AC100~240V to DC 9V/0.5A (9mm) Adaptor 11 HT200‐en‐GB_V1.3 1/15 EXTECH INSTRUMENTS Thermal Hazard Prevention Standards Screening Criteria for Heat Stress Exposure (WGBT values in °C); for reference purposes only.
Example of Activities within Metabolic Rate Categories* Categories Resting Light Moderate Heavy Very heavy Examples/Activities Sitting quietly Sitting with moderate arm movements Sitting with moderate arm and leg movements Standing with light work at machine or bench while using mostly arms Using a table saw Standing with light or moderate work at a machine or bench and some walking about Scrubbing in a standing position Walking about with moderate lifting or pushing Walking level at 3.7 mph (6Km/hr) while carrying 6.6 lbs. (3 Kg) weight Carpenter sawing by hand Shoveling dry sand Heavy assembly work on a non‐continuous basis Intermittent heavy lifting with pushing/pulling (pick‐and‐shovel work) Shoveling wet sand *According to the American Conference of Governmental Industrial Hygienists [ACGIH] (2005) Copyright © 2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com 12 HT200‐en‐GB_V1.3 1/15