File E116994 Vol. 1 Sec. 27 and Report Page 1 Issued: Revised: 1992-05-19 2003-10-23 DESCRIPTION PRODUCT COVERED: Component - Switching Power Supplies for use in Medical and Dental Equipment; Models GPM140-5, -12, -13, -15, -24, -28, -30; may be followed by C, -T, -T1, or -T2; MSP1363, MSP1402. ELECTRICAL RATING: Input: Models GPM140 Series, MSP1363, MSP1402 100-240 V ac, 50/60 Hz, 4.0 A Outputs: Model GPM140-5 GPM140-12 GPM140-13 GPM140-15 MSP1363 GPM140-24 GPM140-28 GPM140-30 MSP 1402 Volts 5 12 13 15 12 24 28 30 17 MAXIMUM OUTPUT AMPS AND WATTS CONVECTION COOLING FORCED AIR COOLING (1) Without Cover With Cover With & Without Cover 26.0 11.7 10.8 9.3 11.7 5.8 5.0 4.7 8.2 A A A A A A A A A 130 140 140 140 140 149 140 140 140 W W W W W W W W W 24.0 10.0 9.2 8.0 10.0 5.8 5.0 4.7 - A A A A A A A A 120 120 120 120 120 140 140 140 - W W W W W W W W 32.0 13.4 12.3 10.7 13.4 6.7 5.8 5.3 9.4 A A A A A A A A A 160 160 160 160 160 160 160 160 160 W W W W W W W W W (1) Reference the Installation Instructions, ILL. 1. ENGINEERING CONSIDERATIONS (NOT FOR FIELD REPRESENTATIVE'S USE): For use in products where the acceptability of the combination is determined by Underwriters Laboratories Inc. The products, models GPM140-5, -12, -13, -15, -24, -28, and -30, were evaluated to the Second Edition of the Standard for Medical and Dental Equipment, UL 544; and the First Edition of the Standard For Medical Electrical Equipment, Part 1: General Requirements for Safety, UL 2601-1. An insulation diagram is provided as Ill. 3 and the manufacturer's installation instructions are provided as Ill. 1. The products, models MSP1363 and MSP1402 were evaluated to the Second Edition of the Standard for Medical and Dental Equipment, UL 544 only. File E11699 Vol. 1 Sec. 27 and Report Page 2 Issued: Revised: 1992-05-19 2003-10-23 Conditions of Acceptability - When installed in the end-use equipment, the following are among the considerations to be made: 1. The power supplies, models GPM140-5, -12, -13, -15, -24, -28, and -30, have been judged on the basis of the required spacings in the First Edition of the Standards for Medical Electrical Equipment, Part 1: General Requirements for Safety, UL 2601-1, and the second edition of the Standard for Medical and Dental Equipment, UL 544, which covers the end use product for which the component is designed. The power supplies, models MSP1363 and MSP1402 have been judged on the basis of the required spacings in the Second Edition of the Standard for Medical and Dental Equipment, Professional, UL 544, which covers the end-use product for which the component designed. 2. The device shall be installed in compliance with the enclosure, mounting, spacing, casualty markings and segregation requirements of the end-use application. 3. Consideration should be given to measuring the temperature on power electronic components and transformer windings when the power supply is installed in the end-use equipment. All transformers comply with Class 155 limits. 4. Input and output connectors, when provided, are not acceptable for field connections, they are only intended for connection to mating connectors of internal wiring inside the end-use machine. The acceptability of the mating connectors relative to secureness, insulating materials, and temperature shall be considered. 5. The end-use product shall ensure that a fuse warning marking is provided adjacent to the primary fuse (F1). The marking shall include the following wording: "WARNING - For continued protection against risk of fire, replace only with the same type and ratings of fuse" and the fuse ratings. The minimum letter height 7/64 in. 6. The power supply should be properly bonded to ground in the end-use product. 7. The power supply has been evaluated for patient care equipment, but not patient connected. 8. The output circuits have not been evaluated for direct patient connection (Type B, BF or CF). File E116994 Vol. 1 Sec. 27 and Report Page 3 Issued: Revised: 1992-05-19 1997-09-03 9. This power supply has been evaluated for use in 25°C ambient. The end use product shall ensure that the airflow and power ratings listed in the electrical ratings output table are not exceeded. 10. Leakage current testing should be repeated in the end product application. 11. The UL 2601 power supplies were evaluated as Reinforced insulation between primary and secondary; basic insulation between primary to ground; and operational insulation only between secondary to ground. 12. The UL 2601 power supplies have been evaluated as Class I, continuous operation, ordinary equipment and has not been evaluated for use in the presence of a flammable anaesthetic mixture with air, oxygen, or nitrous oxide. 13. Fusing in the end product shall be considered since only one fuse rated 5 A, 250 V is provided in the hot side of the input supply circuit. 14. For the UL 2601-1 power supplies, under normal and single fault conditions, the outputs do not exceed 25 V ac or 60 V dc. DC Power Supplies, Inc. STATEMENT OF COMPLIANCE File Number: LR 46516 Date Issued: October 16, 2003 In accordance with the conditions established by the CSA Category Certification Program, the models listed below are eligible to bear the CSA mark. Issued By: Ross Sacolles Product Safety Engineer Signature: CLASS 5311-03 - POWER SUPPLIES - Component Type PRODUCTS Component power supplies for use with Information Processing and Business Equipment, where the suitability of the combination is to be determined by the Canadian Standards Association. - Model GPC140-5, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 5 V/32 A; 160 W with 26 cfm airflow; 130 W with convection cooling, 120 W with cover. - Model GPC140-12, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 12 V/13.4 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 120 W with cover. - Model GPC140-15, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 15 V/10.7 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 120 W with cover. - Model GPC140-24 and GPC140-24P0, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 24 V/6.7 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 140 W with cover. - Model GPC140-28 and SP1382, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 28 V/5.8 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 140 W with cover. CLASS 5311-20 - POWER SUPPLIES - Component Type PRODUCTS Component power supplies for use with Medical Equipment, where the suitability of the combination is to be determined by the Canadian Standards Association. - Model GPM140-5, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 5 V/32 A; 160 W with 26 cfm airflow; 130 W with convection cooling, 120 W with cover. - Model GPM140-12, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 12 V/13.4 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 120 W with cover. - Model GPM140-13, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 13 V/12.3 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 120 W with cover. - Model GPM140-15, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 15 V/10.7 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 120 W with cover. - Model GPM140-24, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 24 V/6.7 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 140 W with cover. - Model GPM140-28, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 28 V/5.8 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 140 W with cover. - Model GPM140-30, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 30 V/5.3 A; 160 W with 26 cfm airflow; 140 W with convection cooling, 140 W with cover. - Model MSP1363, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 12 V/13.4 A; 160 W with 26 cfm airflow. - Model MSP1402, (Level 3), input rated 100-240 V, 50/60 Hz, 4 A; dc output rated 17 V/9.2 A; 160 W with 26 cfm airflow; 140 W with convection cooling. Notes 1. Maximum ambient temperature for continuous output power of 160 Watts is 50 °C. 2. Model numbers may include the following suffixes: -C for cover. -T for input and output terminal blocks, -T1 for input terminal block or -T2 for output terminal block. APPLICABLE REQUIREMENTS CSA Standard C22.2 No. 0-1991: 0.4-M1982: 234-M90: 601.1-M90: TIL CA-08: General Requirements – Canadian Electrical Code, Part II Bonding and Grounding of Electrical Equipment (Protective Grounding) Safety of Component Power Supplies Medical Electrical Equipment Power Supplies for use in Medical Electrical Equipment