HS-201HSRH Neutron Test Report

Neutron testing of the HS-201HSRH quad analog switch
Nick van Vonno
Intersil Corporation
October 2013
Revision 0
Table of Contents
1. Introduction
2. Part Description
3. Test Description
3.1 Irradiation facility
3.2 Characterization equipment
3.3 Experimental Matrix
4 Results
4.1 Test results
4.2 Variables data
5 Discussion and conclusion
6 Appendices
7 Document revision history
1. Introduction
This report summarizes results of 1 MeV equivalent neutron testing of the HS-201HSRH quad analog
switch. The test was conducted in order to determine the sensitivity of the part to the displacement damage
caused by the neutron environment. Neutron fluences ranged from 5 x 1011 n/cm2 to 1 x 1014 n/cm2 in an
approximately logarithmic sequence. This project was carried out in collaboration with Honeywell Aerospace
(Clearwater, FL), and their support is gratefully acknowledged.
2: Part Description
The HS-201HSRH, HS-201HSEH are monolithic CMOS analog switches featuring power-off high
input impedance, very fast switching speeds and low ON-resistance. Fabrication on the Intersil DI RSG
process assures SEL immunity and good low dose rate performance. These Class V/Q devices are acceptance
tested to 300krad (Si) at high dose rate and to 50krad(Si) at low dose rate. Power-off high input impedance
enables the use of this device in redundant circuits without causing data bus signal degradation. ESD
protection, overvoltage protection, fast switching times, low ON resistance and guaranteed radiation hardness
make the HS-201HSRH ideal for any space application where improved switching performance is required.
Specifications for Rad Hard QML devices are controlled by the Defense Logistics Agency Land and
Maritime (DLA). The SMD numbers listed below must be used when ordering. Detailed Electrical
Specifications for these devices are contained in SMD 5962-99618. A “hot-link” is provided on the Intersil
homepage for downloading this and other documents.
1
• QML qualified per MIL-PRF-38535
• Radiation performance:
- High dose rate (50-300rad(Si)/s)
- Low dose rate (0.01rad(SI)/s)
- SEL immune
• Overvoltage protection (power on, switch off)
• Power off high impedance
• Fast switching times
- tON
- tOFF
• Low “ON” resistance
• Pin compatible with industry standard 201 switches
• Operating supply range
• Wide analog voltage range (±15V supplies)
300krad(Si)
50krad(Si)
DI RSG process
±30V
±17V
110ns (max)
80ns (max)
50Ω (max)
±10V to ±15V
±15V
3: Test Description
3.1 Irradiation Facilities
Neutron irradiation was performed by the Honeywell (Clearwater, FL) team at the Fast Burst Reactor
facility at White Sands Missile Range (White Sands, NM), which provides a controlled 1MeV equivalent
neutron flux. Parts were tested in an unbiased configuration with all leads open. As neutron irradiation
activates many of the elements found in a packaged integrated circuit, the parts exposed at the higher neutron
levels required (as expected) significant „cooldown time‟ before being shipped back to the Intersil facility
(Palm Bay, FL) for electrical testing.
3.2 Characterization equipment and procedures
Electrical testing was performed before and after irradiation using the Intersil production automated test
equipment (ATE). All electrical testing was performed at room temperature.
3.3 Experimental matrix
Testing proceeded in general accordance with the guidelines of MIL-STD-883 Test Method 1017. The
experimental matrix consisted of five samples irradiated at 5 x 1011 n/cm2, five samples irradiated at 2 x 1012
n/cm2, five samples irradiated at 1 x 1013 n/cm2 and five samples irradiated at 1 x 1014 n/cm2. Two control
units were used.
4: Results
4.1 Attributes data
Neutron testing of the HS-201HSRH is complete and the results are reported in the balance of this
report. It should be realized when reviewing the data that each neutron irradiation was made on a different 5unit sample; this is not total dose testing, where the damage is cumulative. Table 1 summarizes the results.
2
Table 1: Neutron irradiation test results.
Neutron fluence Sample size
Bin 1
Rejects
5 x 1011 n/cm2
2 x 1012 n/cm2
1 x 1013 n/cm2
1 x 1014 n/cm2
5
5
5
5
0
0
0
5
5
5
5
5
4.2 Variables data
The plots in Figs. 1 through 15 show data plots for key parameters before and after irradiation to each
level. The plots show the average, minimum and maximum of each parameter for each of the four channels as
a function of neutron irradiation. The exception is Fig. 1 which shows the total power supply current, positive
and negative, for the four channels.
Power supply current, mA
15.0
10.0
5.0
0.0
-5.0
ICCP24 Avg
ICCP24 Max
ICCP08 Min
IEEP24 Avg
IEEP24 Max
IEEP08 min
Spec limit
ICCP24 Min
ICCP08 Avg
ICCP08 Max
IEEP24 Min
IEEP08 Avg
IEEP08 max
Spec limit
-10.0
-15.0
Pre-rad
1.00E+11
1.00E+12
1.00E+13
Neutron fluence,
1.00E+14
n/cm2
Fig. 1: HS-201HSRH positive and negative power supply current at ±8.0V and ±24V as a function of
neutron irradiation, showing the mean, minimum and maximum of the populations at each level.
Sample size was 5 for each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2),
with two control units. The post-irradiation SMD limits are +12mA maximum (ICC) and -12V
maximum (IEE).
3
LOW input threshold voltage
15.0
10.0
5.0
VALA1 Avg
VALA1 Max
VALA2 Min
VALA3 Avg
VALA3 Max
VALA4 Min
Spec limit
VALA1 Min
VALA2 Avg
VALA2 Max
VALA3 Min
VALA4 Avg
VALA4 Max
0.0
-5.0
-10.0
-15.0
1.00E+11
Pre-rad
1.00E+12
1.00E+13
Neutron fluence,
1.00E+14
n/cm2
Fig. 2: HS-201HSRH LOW input threshold voltage as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 0.8V maximum.
HIGH input threshold voltage
15.0
10.0
5.0
0.0
-5.0
-10.0
-15.0
1.00E+11
Pre-rad
VAHA1 Avg
VAHA1 Max
VAHA2 Min
VAHA3 Avg
VAHA3 Max
VAHA4 Min
Spec limit
1.00E+12
VAHA1 Min
VAHA2 Avg
VAHA2 Max
VAHA3 Min
VAHA4 Avg
VAHA4 Max
1.00E+13
Neutron fluence,
1.00E+14
n/cm2
Fig. 3: HS-201HSRH HIGH input threshold voltage as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 2.4V minimum.
4
120.0
ON resistance, +10V
100.0
80.0
RON1P10 Avg
RON1P10 Max
RON2P10 Min
RON3P10 Avg
RON3P10 Max
RON4P10 Min
Spec limit
RON1P10 Min
RON2P10 Avg
RON2P10 Max
RON3P10 Min
RON4P10 Avg
RON4P10 Max
60.0
40.0
20.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 4: HS-201HSRH ON resistance at +10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 50.0 ohms maximum.
120.0
ON resistance, -10V
100.0
80.0
RON1N10 Avg
RON1N10 Max
RON2N10 Min
RON3N10 Avg
RON3N10 Max
RON4N10 Min
Spec limit
RON1N10 Min
RON2N10 Avg
RON2N10 Max
RON3N10 Min
RON4N10 Avg
RON4N10 Max
60.0
40.0
20.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 5: HS-201HSRH ON resistance at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 50.0 ohms maximum.
5
Source OFF leakage, +10V, nA
12.0
10.0
8.0
ISOFF1P14 Avg
ISOFF1P14 Max
ISOFF2P14 Min
ISOFF3P14 Avg
ISOFF3P14 Max
ISOFF4P14 Min
Spec limit
ISOFF1P14 Min
ISOFF2P14 Avg
ISOFF2P14 Max
ISOFF3P14 Min
ISOFF4P14 Avg
ISOFF4P14 Max
6.0
4.0
2.0
0.0
-2.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 6: HS-201HSRH source OFF leakage at +10.0V as a function of neutron irradiation, each
channel, showing the mean, minimum and maximum of the populations at each level. Sample size was
5 for each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control
units. The post-irradiation SMD limit is +10.0nA maximum.
Source OFF leakage, -10V, nA
0.0
-2.0
-4.0
-6.0
-8.0
-10.0
-12.0
Pre-rad
1E+11
ISOFF1N14 Avg
ISOFF1N14 Max
ISOFF2N14 Min
ISOFF3N14 Avg
ISOFF3N14 Max
ISOFF4N14 Min
Spec limit
1E+12
ISOFF1N14 Min
ISOFF2N14 Avg
ISOFF2N14 Max
ISOFF3N14 Min
ISOFF4N14 Avg
ISOFF4N14 Max
1E+13
1E+14
Neutron fluence, n/cm2
Fig. 7: HS-201HSRH source OFF leakage at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is +10.0nA maximum
6
Drain OFF leakage, +10V, nA
12.0
10.0
8.0
6.0
IDOFF1P14 Avg
IDOFF1P14 Max
IDOFF2P14Min
IDOFF3P14 Avg
IDOFF3P14 Max
IDOFF4P14 Min
Spec limit
IDOFF1P14 Min
IDOFF2P14 Avg
IDOFF2P14 Max
IDOFF3P14 Min
IDOFF4P14 Avg
IDOFF4P14 Max
4.0
2.0
0.0
-2.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 8: HS-201HSRH drain OFF leakage at +10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is +10.0nA maximum.
Drain OFF leakage, -10V, nA
0.0
-2.0
-4.0
-6.0
-8.0
-10.0
-12.0
Pre-rad
1E+11
IDOFF1N14 Avg
IDOFF1N14 Max
IDOFF2N14 Min
IDOFF3N14 Avg
IDOFF3N14 Max
IDOFF4N14 Min
Spec limit
1E+12
IDOFF1N14 Min
IDOFF2N14 Avg
IDOFF2N14 Max
IDOFF3N14 Min
IDOFF4N14 Avg
IDOFF4N14 Max
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 9: HS-201HSRH drain OFF leakage at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is -10.0nA maximum.
7
Drain ON leakage, +10V, nA
12.0
10.0
8.0
IDON1P14 Avg
IDON1P14 Max
IDON2P14 Min
IDON3P14 Avg
IDON3P14 Max
IDON4P14 Min
Spec limit
IDON1P14 Min
IDON2P14 Avg
IDON2P14 Max
IDON3P14 Min
IDON4P14 Avg
IDON4P14 Max
6.0
4.0
2.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 10: HS-201HSRH drain ON leakage at +10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is +10.0nA maximum.
2.0
Drain ON leakage, -10V, nA
0.0
-2.0
-4.0
-6.0
-8.0
-10.0
IDON1N14
IDON1N14
IDON1N14
IDON2N14
IDON2N14
IDON2N14
-12.0
IDON3N14
IDON3N14
IDON3N14
IDON4N14
IDON4N14
IDON4N14
-14.0
-16.0
Pre-rad
1E+11
Spec limit
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 11: HS-201HSRH drain ON leakage at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is -10.0nA maximum.
8
Switch turnon time, +10V, ns
250.0
200.0
150.0
TON1P10 Avg
TON1P10 Max
TON2P10 Min
TON3P10 Avg
TON3P10 Max
TON4P10 Min
Spec limit
TON1P10 Min
TON2P10 Avg
TON2P10 Max
TON3P10 Min
TON4P10 Avg
TON4P10 Max
100.0
50.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 12: HS-201HSRH switch turnon time at +10.0V as a function of neutron irradiation, each
channel, showing the mean, minimum and maximum of the populations at each level. Sample size was
5 for each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control
units. The post-irradiation SMD limit is 130.0ns maximum.
Switch turnon time, -10V, ns
250.0
200.0
150.0
TON1N10 Avg
TON1N10 Max
TON2N10 Min
TON3N10 Avg
TON3N10 Max
TON4N10 Min
Spec limit
TON1N10 Min
TON2N10 Avg
TON2N10 Max
TON3N10 Min
TON4N10 Avg
TON4N10 Max
100.0
50.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 13: HS-201HSRH switch turnon time at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 130.0ns maximum.
9
Switch turnoff time, +10V, ns
160.0
140.0
120.0
100.0
TOFF1P10 Avg
TOFF1P10 Max
TOFF2P10 Min
TOFF3P10 Avg
TOFF3P10 Max
TOFF4P10 Min
Spec limit
TOFF1P10 Min
TOFF2P10 Avg
TOFF2P10 Max
TOFF3P10 Min
TOFF4P10 Avg
TOFF4P10 Max
80.0
60.0
40.0
20.0
0.0
Pre-rad
1E+11
1E+12
1E+13
Neutron fluence,
1E+14
n/cm2
Fig. 14: HS-201HSRH switch turnoff time at +10.0V as a function of neutron irradiation, each
channel, showing the mean, minimum and maximum of the populations at each level. Sample size was
5 for each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control
units. The post-irradiation SMD limit is 80.0ns maximum.
Switch turnoff time, -10V, ns
160.0
140.0
120.0
100.0
TOFF1N10 Avg
TOFF1N10 Max
TOFF2N10 Min
TOFF3N10 Avg
TOFF3N10 Max
TOFF4N10 Min
Spec limit
TOFF1N10 Min
TOFF2N10 Avg
TOFF2N10 Max
TOFF3N10 Min
TOFF4N10 Avg
TOFF4N10 Max
80.0
60.0
40.0
20.0
0.0
Pre-rad
1E+11
1E+12
1E+13
1E+14
Neutron fluence, n/cm2
Fig. 15: HS-201HSRH switch turnoff time at -10.0V as a function of neutron irradiation, each channel,
showing the mean, minimum and maximum of the populations at each level. Sample size was 5 for
each cell (5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2), with two control units.
The post-irradiation SMD limit is 80.0ns maximum.
10
5: Discussion and conclusion
This document reports the results of neutron testing of the HS-201HSRH quad analog switch. Samples
were irradiated to levels of 5 x 1011 n/cm2, 2 x 1012 n/cm2, 1 x 1013 n/cm2 and 1 x 1014 n/cm2 with a sample
size of five parts per cell. It should again be realized when reviewing the data that each neutron irradiation
was made on a different 5-unit sample; this is not total dose testing, where the damage is cumulative. ATE
characterization testing was performed before and after the irradiations, and three control units were used to
insure repeatable data. Variables data for monitored parameters is presented in Figs. 1 through 15. The 2 x
1012 n/cm2 level is of some interest in the context of recent developments in the JEDEC community, where
the discrete component vendor community have signed up for characterization testing (but not for acceptance
testing) at this level.
The HS-201HSRH is not formally designed for neutron hardness. The part is built in a DI BiCMOS
process. There are some bipolar transistors in the design; these are minority carrier devices may be expected
to be sensitive to displacement damage (DD) at the higher levels. This expectation turned out to be correct.
We will discuss the results on a parameter by parameter basis and then draw some conclusions.
The positive and negative power supply current (Fig. 1) showed good stability after 5 x 1011 n/cm2 and 2
x 1012 n/cm2, a slight decrease after 1 x 1013 n/cm2 irradiation and a further decrease after 1 x 1014 n/cm2
irradiation.
The switch turnon time at +10.0V and HIGH input threshold voltage (Figs. 2 and 3) showed good
stability after 5 x 1011 n/cm2, 2 x 1012 n/cm2 and 1 x 1013 n/cm2 irradiation but was basically nonfunctional
after 1 x 1014 n/cm2 irradiation.
The switch ON resistance at +10.0V and -10.0V (Figs. 4 and 5) showed good stability after 5 x 1011
n/cm2, 2 x 1012 n/cm2 and 1 x 1013 n/cm2 irradiation but was also nonfunctional after 1 x 1014 n/cm2
irradiation. At this point the reader will note that the samples were nonfunctional after irradiation to 1 x 1014
n/cm2, and the data taken at this level do not carry a great deal of meaning.
The source OFF leakage at +10.0V and -10.0V (Figs. 6 and 7), the drain OFF leakage at +10.0V and 10.0V (Figs. 8 and 9) and the drain ON leakage at +10.0V and -10.0V (Figs. 10 and 11) showed good stability
after 5 x 1011 n/cm2, 2 x 1012 n/cm2 and 1 x 1013 n/cm2 irradiation but were nonfunctional after 1 x 1014 n/cm2
irradiation.
The switch turnon time at +10.0V and -10.0V (Figs. 12 and 13) and the switch turnoff time at +10.0V and
-10.0V (Figs. 14 and 15) showed good stability after 5 x 1011 n/cm2, 2 x 1012 n/cm2 and 1 x 1013 n/cm2
irradiation but were nonfunctional after 1 x 1014 n/cm2 irradiation.
We conclude that the HS-201HSRH is capable of post 1 x 1013 n/cm2 operation within the SMD post-total
dose parameters. The part is not capable of post 1 x 1014 n/cm2 operation as it was found to be nonfunctional
after irradiation to this level.
11
6: Appendices
6.1: Reported parameters.
Fig.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Parameter
Limit,
low
Limit,
high
Units
2.4
-
±12.0
0.8
50
50
+10.0
-10.0
+10.0
-10.0
+10.0
-10.0
130.0
130.0
80.0
80.0
mA
V
V
ohms
ohms
nA
nA
nA
nA
nA
nA
ns
ns
ns
ns
Positive and negative supply current
ON input threshold
OFF input threshold
ON resistance, +10.0V
ON resistance, +10.0V
Source OFF leakage, +10.0V
Source OFF leakage, -10.0V
Drain OFF leakage, +10.0V
Drain OFF leakage, -10.0V
Drain ON leakage, +10.0V
Drain ON leakage, -10.0V
Switch turnon time, +10.0V
Switch turnon time, -10.0V
Switch turnoff time, +10.0V
Switch turnoff time, -10.0V
7: Document revision history
Revision
0
Date
11 October 2013
Pages
All
Comments
Original issue
12
Notes