DATA LOGGER LR5000 Series Data Loggers Complete Line of Easy-to-Use Compact Loggers with Expanded Memory The new HIOKI compact data logger series easily records temperature, voltage, current, and instrumentation signals over long periods. Carried over from its highly reputed predecessor, this series includes features and functions such as 7 times the recording capacity of former models, data import during recording, continuous measurement even during battery replacement, and intuitive PC software. Flexible and easy-to-use at single and multiple locations, the new HIOKI compact data logger series is ideal for any application that requires simple set-up but long-term, reliable recording capabilities. 2 Meet a Wide Variety of Data Logging Applications Temperature Logger /Humidity Logger Manage the temperature and humidity in offices and factories. Visually monitor the data to save on airconditioning and heating costs. Instrumentation Logger / Voltage Logger Record fluid flow such as for water, gas and oil. Measure flow meter output signals to monitor flow trends. Clamp Logger Manage the current consumption of plant and building equipment. Visually monitor power costs to efficiently conduct energyand cost-saving activities. Use as a Voltage Logger to record pyranometer output for evaluating insulation. Use as a Clamp Logger and leakage sensor to record and monitor leakage trends. pyranometer Voltage logger has a Preheat function Use as a Temperature Logger to record warehouse temperatures for visually monitoring temperature changes of products and goods. Use as an Instrumentation Logger to record pressure sensor output and monitor fluctuations in air or oil pressure. 3 Easy operation in just steps ! Install a Data Logger, set an interval, and start measuring. Easy to start recording Unlimited installation capabilities STEP Set up &Record Wall-mounted Holder (sold separately) Not usable with LR5051 Set your recording interval. (from one second to 60 minutes) Download data using infrared communication. STEP Transfer data from Data Logger to PC ! Kickstand (included, except for Model LR5051) Grab the Data Logger from the worksite and connect to a PC. Communication Adapter LR5091 Retrieving the Data Logger Transfer data from Data Logger to PC using a USB connection Using the Data Collector's internal memory, import data from up to 16 Data Loggers installed on site.* Download data u s i n g i n f r a re d communication. Requires optional Communication Adapter Importing to a Data Collector or Data Collector Magnetic Strap (sold separately) Hold the REC button for two seconds to start recording. Front Back Data Collector LR5092-20 Retrieving the Data Collector Transfer data from Data Logger to PC using a USB connection * Data for up to 16 channels can be stored. Combine up to 16 single-channel Data Loggers (Models LR5011, LR5031, LR5041, LR5042, and LR5043), or up to eight 2-channel Data Loggers (Models LR5001, and LR5051). Using an optional SD Memory Card, the amount of data that can be imported is practically limitless. Top surface of LR5092-20 SD Card slot or USB Connector Importing to an SD Memory Card Transfer data from SD Memory Card within Data Collector to PC using USB View data graphically and easily print using the bundled software. STEP View graphs and manage data Transfer data directly from SD Memory Card 4 Advanced Features and Functions Install Almost Anywhere Recording capacity up to 7 times previous models Easily mount the light-weight, pocket-sized loggers in tight spaces. Large internal memory stores 60,000 data points per channel. Long-term recording capability exceeds that of previous models. Interval times 1s 2s 5s 10s 15s 20s 30s 1m 2m 5m 10m 15m 20m 30m 60m Actual size Easy-to-see dual display Instantaneous value Statistical value 16h 9h 11h 22h 10h 21h 20h 16h 8h 8h 16h 8h 20h 17h 14h 11h 5h 10h 20h 2h 4h 6h 8h 12h 1d 3d 6d 10d 13d 20d 41d 83d 208d 416d 625d 833d 1250d 2500d 40m 20m 20m 40m 20m 8h 1d 2d 3d 5d 10d 20d 52d 104d 156d 208d 312d 625d 20m 50m 40m 30m 20m The maximum recording time depends on battery life. The battery may need to be replaced during long-term recording. Temperature and humidity or current channels can be displayed. View maximum and minimum values while measuring. Moist environments Record without missing fluctuations IP54 splash-proof rating withstands operation in extremely humid environments like kitchens and pipe rooms. (Except Model LR5051) With usual (instantaneous value) recording at long intervals, detailed fluctuations occurring within the intervals are missed. However, with the statistical value recording mode, detailed fluctuations are captured even when they occur during long recording intervals. In STAT mode, measurement is taken every second, and the maximum, minimum, average, and instantaneous values within the specified interval are recorded. Data interval(2sec to 60min) Transfer data even during recording 1sec Continue to record even when transferring data. Data interval(2sec to 60min) Maximum data Maximum data Transfer data from Data Logger to PC! Average data Instant data In the middle of Recording ... Input signal (Temperature, Voltage, etc.) Batteries last up to 2 years Energy-efficient design provides up to two years of battery life (For the LR5011 only. Actual battery life depends on model type and settings). Average data Instant data Minimum data Minimum data Record 4 data for interval Time Record 4 data for interval Never worry about a dead battery The worry-free backup function preserves measurement data even after the battery dies. Replace batteries while recording Never worry about operating errors Recording continues for about 30 seconds even with the battery removed. Worry-free backup preserves recorded data even if a new measurement is started by mistake. Note. With the LR5001, re c o rd i n g i s i n t e r r u p t e d during battery replacement if the battery is very weak. After batteries are replaced, recording resumes automatically. Previously recorded data is not lost during battery replacement. ! 5 Bundled Software Ensures Smooth and Easy Data Analysis Import data to a PC and create graphs Use the LR5000 Utility program to import Data Logger data to a PC to make graphs and analyze data further. Easily print results using your PC. Show specific values using the cursor function Use the A/B cursors to select any location on a graph and display its value. The PC software can also calculate maximum, minimum, and average values between A and B cursors. Simple file aggregation and management Transferred data can be combined with data previously transferred (from the same Data Logger unit) into one file on the PC. Past data New data Saving 1 file Display data from former Data Logger models The PC application also supports data collected from the HIOKI 36XX Series Data Loggers. Print function Configurating Data Logger Graph display LR5000 Utility Specifications •Import/export Data Logger settings (LR5091 or LR5092-20 required) •Settings sent to each LR5000 logger are also saved to the PC. •Graphically display data for up to 16 channels •Select colors and display/hide any channel and graph •Copy graph images to clipboard •Display statistical data (maximum, minimum and average) •Scaling function •Print graphs •Print statistical data. •Scaling •Power calculation •Energy cost calculation Data processing •Operating ratio calculation •Integration •Dew point temperature •Calculate between channels •OS:Windows XP (SP2 or later) Windows Vista (SP1 or later) / Windows 7 •CPU : 1GHz or more Operating •Memory : 512MB or more environment •Interface : USB •Free space in hard disk:30MB or more 6 Communication Adapter and Data Collector Specifications (Product guaranteed for one year) Physical appearance Model Communication Adapter LR5091 Features Data Collector LR5092-20 •Collect recorded data from the Data Logger to internal memory or SD card •View collected data in a graph •Transfer data from a Data logger to a PC •Transfer Data Logger configurations or clock settings from internal •Transfer Data Logger configurations or clock settings from a PC memory or SD card to the Data Logger •Transfer data from a Data Logger to a PC to the Data Logger •Transfer Data Logger configurations or clock settings from a PC to the Data Logger Interface with Data Logger Interface with PC Clock functions Display - Display items - Infrared optical communications Internal memory capacity of data Removable storage media Operating environment Power supply USB2.0, Full Speed, Series Mini B Receptacle Auto calender, auto leap year Dot-matrix LCD (128 × 64 dots) Data Logger configurations (Interval, Start/Stop method, Recording mode, Scaling, Alarm, Power-saving mode, Clock, Range) Collected data (Record list , Maximum data, Minimum data, Average, Graph, Value) 60,000 data elements ×16ch (instantaneous value mode) 15,000 data elements ×16ch (statistical value mode) SD Card (SDHC, Max 32GB) Save data and configurations Indoors DC3V (LR6 (AA) Alkaline battery 1.5V×2) or DC5V (USB bus power) Maximum rated power 1VA Approx. 12 hours or 500 times of data collection - DC5V (USB bus power) Maximum rated power 0.5VA - Battery life Operating temperature and humidity 0˚C(32˚F) to 40˚C(104˚F) , 80%rh or less (non-condensating) Dimensions & Mass Approx. 83mm(3.27 in)W×61mm(2.40in)H×19mm(0.75in)D 43g(1.5oz) 91mm(3.58in)W×141mm(5.55in)H×31mm(1.22in)D 215g(7.6oz) (excluding batteries) Accessories USB cable (1m)×1, CD (Application software "LR5000 Utility") × 1 Instruction manual ×1, Operation manual×1, LR6 (AA) Alkaline battery 1.5V×2, USB cable(1m)×1, CD (Application software "LR5000 Utility") × 1 LR5092-20 Option SD Memory Card (2GB) Z4001 LR5000 Series Common specifications (Product and accuracy guaranteed for one year) Recording interval 1/ 2/ 5/ 10/ 15/ 20/ 30 seconds 1/ 2/ 5/ 10/ 15/ 20/ 30/ 60 minutes Display items Recording methods One time recording Stop recording when the memory capacity is full. Endless recording Continue recording even when the memory capacity is full. (old data is overwritten.) Recording start / stop Instantaneous recording Instantaneous values are recorded at every Recording modes recording interval. Statistical value recording (instantaneous value Measure at one second intervals, and record the mode/statistical instantaneous, maximum, value mode) minimum, and average values within every recording interval. Storage capacity Instantaneous value mode 60,000 data sets per channel Statistical value mode 15,000 data sets per channel LR5000 Series common options Magnetic Strap Z5004 Wall-mounted Holder LR9901 Not compatible with Model LR5051 Data backup Measured value, Interval configration, Date, Time, Alarm, Remaining battery power, Number of data, Maximum data, Minimum data Recording start Manual start Timer start Recording stop Manual stop Timer stop When the memory capacity is full (One time recording) Data from the last recording session is always backed up. Back up recorded data and configuration when battery is dead. Interface Infrared optical communications with LR5091, LR5092-20 Power supply During battery replacement, recording and clock operations are preserved for about 30 seconds. (Recording operation continues if the battery is replaced within about 30 seconds.) Note. With the LR5001, recording is interrupted during battery replacement if the battery is very weak. After batteries are replaced, recording resumes automatically. Previously recorded data is not lost during battery replacement. 7 Product Specifications See page 6 for Common specifications (Product and accuracy guaranteed for one year) Physical appearance *Sensor is sold separately. Model Humidity Logger LR5001 Temperature measurement with external temperature sensor. Select the sensor according to the measurement object Temperature 1ch and Humidity 1ch Temperature 1ch ±15%rh Accuracy -40.0˚C to 180˚C *Depends on measurement range of sensor. (LR5011+Sensor) 180 ±5.0˚C 120 ±2.0˚C 70 ±1.0˚C 35 ±0.5˚C 0 ±1.0˚C -40 Temperature : ˚C Temperature : -40˚C to 85˚C Humidity : 0% to 100%rh Temperature Humidity (LR5001+Sensor) 100 (LR5001 + Sensor) 90 *±10%rh ±10%rh ±8%rh ±10%rh 85 70 ±2.0˚C *±8%rh ±8%rh ±6%rh ±8%rh 70 50 ±1.0˚C *±6%rh ±6%rh ±5%rh ±6%rh 35 * ±0.5˚C 10 0 ±1.0˚C 0 -40 0 10 20 30 50 85 -40 Temperature : ˚C Measurement items Measurement range Temperature and humidity are recorded simultaneously using supplied or optional temperature and humidity sensors. Relative humidity : %rh Features Temperature Logger LR5011 Temperature : ˚C * Values provided for reference only. Waterproof and dustproof performance Operating temperature and humidity Dimensions & Mass Power supply IP54 (splash-proof construction) -20˚C(-4˚F) to 70˚C(158˚F) , 80%rh or less (non-condensating) Approx. 79mm(3.11in)W×57mm(2.24in)H×28mm(1.10in)D 105g(3.7oz) LR6 (AA) Alkaline battery 1.5V×1 Kickstand Humidity sensor LR9504×1, Kickstand Accessories LR6 (AA) Alkaline battery 1.5V×1, Instruction manual ×1, Operation manual×1 Case 1 : Approx. 3 months Case 2: Approx. 20 days Case 1 : Approx. 2 years Case 2: Approx. 2 months Battery life Case 1 : 1min. recording interval, power-saving mode, Instantaneous recording, environmental temp.20˚C Case 2 :1sec. recording interval, power-saving mode, Instantaneous recording, environmental temp.20˚C Analysis of measurement data on a PC requires the optional LR5091 Communication Adapter or LR5092-20 Data Collector. See page 6 for details. (Reference) When the recording interval is set to 10 minutes, the LR5001 Temperature and Humidity Logger can measure for about one year between battery replacements. LR5001 Options Humidity Sensor ■ Humidity Sensor 30mm 1m (LR9501) 5m (LR9502) 10m (LR9503) 40mm (LR9504) LR5011 Options ■ Molded plastic type diameter:6mm : LR9501 (1m) LR9502 (5m) LR9503 (10m) LR9504 (40mm, bundled accessory) ●Temperature range : -40.0˚C to 85.0˚C ●Humidity range : 0.0%rh to 100.0%rh ●Response time : Approximately 300 seconds (Temperature and humidity; 90% response time) ●Models (length) (Thickness : 8mm) 13mm Te m p e r a t u r e S e n s o r 28mm ■ Lug type Outer diameter:7mm Inner diameter:3.2mm 1m (LR9601) 5m (LR9602) 10m (LR9603) 1m (LR9611) 5m (LR9612) 10m (LR9613) 45mm (LR9604) ●Models (length) ●Temperature range ●Response time ●Material : LR9601 (1m) LR9602 (5m) LR9603 (10m) LR9604 (45mm) : -40˚C to 180˚C : Approx. 100 seconds (90% response time) : Cable : Silicon Sensor head : Silicon ■ Sheathed type ■ Needle type 25mm 180mm diameter 4mm ●Models(length) ●Temperature range ●Response time ●Material : LR9611 (1m) LR9612 (5m) LR9613 (10m) ●Temperature range : -30˚C to 180˚C ●Response time : Approx. 45 seconds (90% response time) ●Material: Cable : Silicon Sensor head : Nickel-plated brass ●Models(length) 1m : : : : LR9621(1m) -40˚C to 120˚C Approx. 90 seconds (90% response time) Cable : Silicon Sensor head : SUS304 diameter 1.3mm ●Models(length) ●Temperature range ●Response time ●Material 1m : : : : LR9631(1m) -40˚C to 120˚C Approx. 20 seconds (90% response time) Cable : Silicon Sensor head : SUS304 Product Specifications (Product and accuracy guaranteed for one year) See page 6 for Common specifications Physical appearance LR5042(5V) Model Instrumentation Logger LR5031 Features For recording 4-20 mA instrumentation signals, etc. Measurement items For Instrumentation / 0 to 20mA DC, 1ch Measurement range DC -30.00 to 30.00mA Voltage Logger LR5041, LR5042, LR5043 For recording instrumentation signals and measuring analog outputs from sensors and other devices DC voltage 1ch LR5041: -50.00mV to 50.00mV LR5042: -5.000V to 5.000V LR5043: -50.00V to 50.00V ±0.5%rdg. ±5dgt. (@23˚C±5˚C) ±0.5%rdg. ±5dgt. (@23˚C±5˚C) Accuracy Waterproof and IP54 (splash-proof construction) dustproof performance Operating temperature -20˚C(-4˚F) to 70˚C(158˚F) , 80%rh or less (non-condensating) and humidity Approx. 79mm(3.11in)W×57mm(2.24in)H×28mm(1.10in)D, 105g(3.7oz) Dimensions & Mass LR6 (AA) Alkaline battery 1.5V×1 Power supply Connection Cable LR9801×1, Kickstand Connection Cable LR9802×1, Kickstand Accessories LR6 (AA) Alkaline battery 1.5V×1, Instruction manual ×1, Operation manual×1 Case 1 : Approx. 2 years Case 2: Approx. 2 months Battery life Case 1 : 1min. recording interval, power-saving mode, Instantaneous recording, environmental temp.20˚C Case 2 : 1 sec. recording interval, power-saving mode, Instantaneous recording, environmental temp.20˚C Preheat function (When using preheat function, a separate external Other power supply is required.) Analysis of measurement data on a PC requires the optional LR5091 Communication Adapter or LR5092-20 Data Collector. See page 6 for details. LR5031 Option LR5041,LR5042,LR5043 Option 4 wires 2 wires 1m 1m Connection Cable LR9802 (Bundled accessory) Connection Cable LR9801(Bundled accessory) Product Specifications (Product and accuracy guaranteed for one year) See page 6 for Common specifications *Sensor is sold separately.. Features Measurement items Measurement range Cord length : Approx. 3m Cord length : Approx. 3m CAT III II 600V CAT III II 600V Connection cord 9219 is required (sold separately) Physical appearance Physical appearance Model LR5051 Options Clamp Logger LR5051 Recording load current of 50Hz/60Hz Recording leak current AC Current (2 channels) When Using 9669 When Using CT6500 When Using 9695-02 When Using 9675 When Using 9657-10 : 1000A range : 50.00A / 500.0A range : 5.000A / 50.00A range : 500.0mA / 5.000A range : 500.0mA / 5.000A range ±0.5%rdg. ±5dgt. +Clamp sensor accuracy Accuracy Waterproof and Not waterproof dustproof performance Operating temperature -0˚C(32˚F) to 50˚C(122˚F) , 80%rh or less (non-condensating) and humidity Dimensions & Mass Approx. 79mm(3.11in)W×70mm(2.76in)H×37mm(1.46in)D, 165g(5.8oz) LR6 (AA) Alkaline battery 1.5V × 2 Power supply LR6 (AA) Alkaline battery 1.5V × 2 Accessories Instruction manual ×1, Operation manual×1 Case 1 : Approx. 1 years Case 2: Approx. 1 months Case 1 : 1min. recording interval, power-saving mode, Instantaneous Battery life recording, environmental temp.20˚C Case 2 : 1 sec. recording interval, power-saving mode, Instantaneous recording, environmental temp.20˚C Analysis of measurement data on a PC requires the optional LR5091 Communication Adapter or LR5092-20 Data Collector. See page 6 for details. Model Measurable conductor diameter Primary current rating Accuracy (45Hz to 66Hz) Maximum rated voltage to earth Maximum allowable input (45 to 66 Hz) Dimensions & Mass CAT III 300V Clamp on Sensor 9669 Clamp on Sensor CT6500 Clamp on Sensor 9695-02 φ55 mm (2.17") or less, φ46 mm (1.81") φ15 mm (0.59") or or less less 80 (3.15") × 20 (0.79") mm busbar 1000A AC 500A AC 50A AC ±1.0%rdg. ±0.01%f.s. ±1.5%rdg. ±0.03%f.s. ±0.3%rdg. ±0.02%f.s. 600Vrms (insulated conductor) 600Vrms (insulated conductor) 300Vrms (insulated conductor) 1000A continuous 600A continuous 60A continuous 99.5 (3.92")W × 188 (7.40")H × 77 (3.03")W × 151 (5.94")H × 51 (2.01")W × 58 (2.28")H × 42 (1.65")D mm, 590g (20.8 oz.) 42 (1.65")D mm, 360g (12.7 oz.) 19 (0.75")D mm, 50g (1.8 oz.) length : 3m(9.84ft) Connection Cord 9219(For 9695-02 connection) Cord length : Approx. 3m Cord length : Approx. 3m Physical appearance CAT III 300V CAT III 300V Clamp on Leak Sensor 9675 Clamp on Leak Sensor 9657-10 Model φ30mm φ40mm Measurable conductor diameter 10A AC 10A AC Primary current rating ±1.0%rdg.±0.005%f.s. ±1.0%rdg.±0.05%f.s. Accuracy (45Hz to 66Hz) Lag current 1mA(When 10A AC is input) Maximum rated voltage to earth 300Vrms (insulated conductor) 10A continuous Maximum allowable input (45 to 66 Hz) 60 (2.36")W × 113 (4.45")H × Dimensions & Mass 24 (0.94")D mm, 160g (5.6 oz.) 5mA(When 100A AC is input) 300Vrms (insulated conductor) 30A continuous 74 (2.91")W × 145 (5.71")H × 42 (1.65")D mm, 380g (13.4 oz.) Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies. HIOKI (Shanghai) Sales & Trading Co., Ltd. : HEAD OFFICE : 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 http://www.hioki.co.jp / E-mail: [email protected] HIOKI USA CORPORATION : 6 Corporate Drive, Cranbury, NJ 08512 USA TEL +1-609-409-9109 / FAX +1-609-409-9108 http://www.hiokiusa.com / E-mail: [email protected] 1608-1610 Shanghai Times Square Office, 93 Huai Hai Zhong Road, Shanghai, P.R.China POSTCODE: 200021 TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360 http://www.hioki.cn / E-mail: [email protected] Beijing Office : TEL +86-10-5867-4080/4081 Guangzhou Office : TEL +86-20-38392673/2676 DISTRIBUTED BY HIOKI SINGAPORE PTE.LTD. : 33 UBI AVENUE 3, #03-02 VERTEX, SINGAPORE 408868 TEL +65-6634-7677 FAX +65-6634-7477 E-mail: [email protected] HIOKI INDIA PRIVATE LIMITED : Khandela House, 24 Gulmohar Colony Indore 452 018 (M.P.), India TEL +91-731-4223901, 4223902 FAX +91-731-4223903 http://www.hioki.in / E-mail: [email protected] All information correct as of Feb. 3, 2011. All specifications are subject to change without notice. LR5000sE2-12B