150 kV MICROFOCUS X-RAY X-RAY SOURCE SOURCE L12161-07 FEATURES APPLICATIONS ●Focal spot size: 5 µm (at 4 W) ●Non-destructive inspection ●X-ray CT ●High power: Maximum output 75 W ●External control via RS-232C interface ●High speed ramping-up [Applicable objects] The focal spot of 5 µm of the sealed type X-ray tube offers sharp and clear X-ray images even at a high magnification. Ramping-up speeded up to 3 times faster than conventional 150 kV microfocus X-ray source. ●No high voltage cable connection required High voltage power supply is integrated. •Semiconductor device •Electronic component •Printed circuit board •Ceramic •Plastic component •Die casting •Metal component SPECIFICATIONS GENERAL Parameter X-ray tube voltage setting range X-ray tube current setting range X-ray tube voltage operational range 1 X-ray tube current operational range 1 Small Focus Mode Middle Focus Mode Maximum output Large Focus Mode Small Focus Mode X-ray focal spot size Middle Focus Mode (Nominal value) Large Focus Mode X-ray beam angle 2 Focus to object distance (FOD) Rated output Communication method Description / Value 0 to 150 0 to 500 40 to 150 10 to 500 10 30 75 7 (5 µm at 4 W) 20 50 Approx. 43 Approx. 17 Continuous rating Interface: RS-232C (9-pin D-sub connector) Unit kV µA kV µA degree mm — — Description / Value Beryllium / 0.2 Tungsten +10 to +40 0 to +50 20 to 85 (No condensation) Approx. 13.5 Unit µm — °C °C % kg Description / Value Single phase 100 to 240 (50 Hz / 60 Hz) Less than 220 +10 to +40 0 to +50 20 to 85 (No condensation) Approx. 6 Unit V W °C °C % kg Description EN 50581 Category 9 Emission limits: CISPR 11 Group 1 Class A IEC/EN 61326-1 Immunity requirements: Table 2 IEC/EN 61010-1 Unit — Description PC / AT compatible Windows® XP, 7 RS-232C interface Unit — — — W µm X-RAY TUBE UNIT Parameter X-ray output window material / Thickness Target naterial Operating ambient temperature Storage ambient temperature Operating and storage humidity Weight X-RAY CONTROL UNIT Parameter Input voltage (AC) Power consumption Operating ambient temperature Storage ambient temperature Operating and storage humidity Weight 3 REGULATION AND STANDARDS Parameter RoHS directive EMC Safety — — CONTROL SOFTWARE 4 Parameter Applicable PC Applicable OS Interface NOTE: 1See the graph of the X-ray tube voltage / current operation range. 2Reference value: With 50 % of maximum X-ray emission. 3This weight includes the accessories of approx. 1.5 kg. 4The control software is provided as a sample software for the purpose of MFX operation. The performance of the software is not guaranteed. ! PRECAUTIONS TO USE · This microfocus X-ray source generates X-rays harmful to the human body. Use sufficient caution when handling the equipment to avoid direct or inadvertent exposure to X-rays. Install the X-ray source or the X-ray tube unit in an X-ray shielded cabinet or room equipped with safety interlock functions to prevent accidental exposure to X-rays. OPERATIONAL CAUTION · This microfocus X-ray source generates X-rays and must therefore be used only under the supervision of qualified personnel. · This microfocus X-ray source shall be used in compliance with health and safety regulations enforced in order to prevent health hazards problems due to ionizing radiation. X-RAY TUBE VOLTAGE / CURRENT OPERATION RANGE 600 TXPRB0021EB LARGE FOCUS MODE TUBE CURRENT (µA) 500 400 MIDDLE FOCUS MODE 300 200 SMALL FOCUS MODE 100 10 40 50 60 70 80 90 100 110 120 130 140 150 * The X-ray tube voltage guaranteed range is 40 kV to 150 kV. * Operation is not guaranteed when the tube current is below 10 µA. TUBE VOLTAGE (kV) DIMENSIONAL OUTLINES (Unit: mm) 70 ± 2 X-RAY TUBE UNIT L8122-01 43° 19.5 228 ± 2 X-RAY FOCAL SPOT X-RAY BEAM ANGLE 43° FOD 17 3 × M3 DEPTH 5 COOLING FAN 200 ± 0.5 COOLING FAN 32 PROTECTIVE EARTH TERMINAL 300 ± 2 4 × M6 DEPTH 8 (MOUNTING SCREW HOLES) CONTROL CONNECTOR 140 ± 2 23 100 ± 0.5 200 ± 0.5 298 ± 2 100 ± 0.5 11 100 MIN. 4 × M6 DEPTH 8 (MOUNTING SCREW HOLES) TXPRA0014EB X-RAY CONTROL UNIT C12163 100 MIN. X-RAY150 CONTROL UNIT X-RAY ON ERROR 12 426 ± 2 482 ± 2 99 ± 2 INTER LOCK 50 POWER 33 Max. (CONNECTOR) 40 300 ± 2 30 COOLING FUN POWER RS-232C MONITOR RS-232C CONTROL INTER LOCK 1 INTER LOCK 2 CONTROL OUTPUT AC INPUT FUSL 14 .0/250V CONNECTION CABLE LENGTH: 5 m AC POWER INPUT CONNECTOR TXPRA0015EA RELATED PRODUCTS X-RAY IMAGE INTENSIFIER DIGITAL CAMERA UNIT C7336-05/-52 The C7336 series consist of a high resolution, high contrast 4-inch X-ray image intensifier (X-ray I.I.) and a 2.8 megapixel CMOS image sensor. The X-ray I.I. used has a fixed field-of-view of 100 mm diameter or a 4 inches/2 inches adjustable field-of-view and an input window made of thin aluminum which is excellent in X-ray transmission and causes less scattering of X-rays. These features allow real-time detection at X-ray energy levels from about 20 keV. The captured images can be transferred to PC directly by interface of IEEE1394b. X-CUBE™ (COMPACT X-RAY CCD CAMERA) H8480, H8481, H8953 TXPRF0020 X-CUBEs are compact X-ray CCD camera designed for non-destructive inspection, which make X-ray imaging as easy as an ordinary CCD camera in handling. The H8480 and H8953 use a 2/3 type CCD coupled to large-diameter tapered FOPs which are coated with CsI. The H8481 uses a straight type FOP instead of the large FOP, achieving a high resolution of 20 Lp/mm. Left: H8480 Center: H8953 Right: H8481 FOS (Fiber optic plate coated with X-ray scintillator) The FOS is an optical device for X-ray imaging, fabricated by coating an X-ray scintillator material over a fiber optic plate consisting of more than tens of million glass fibers each a few micrometers in diameter. The FOS provides higher sensitivity and resolution than currently used sensitized paper films and also allows real-time digital radiography when directly coupled to a commercially available CCD. The fiber optic plate used in the FOS has excellent X-ray absorption characteristics, so that X-rays penetrating the X-ray scintillator and directly entering the CCD are minimized to less than 1 %. This protects the CCD from the deterioration and increased noise caused by X-ray irradiation, assuring a long service life and maintaining high image quality. Various sizes and shapes of FOS are available to meet your particular needs, including tapered FOP types. ■Structure TMCPF0072 Protection Film Scintillator : CsI(Tl) FOP (Fiber optic plate) 2D Image-Device Windows ® is a registered trademark of Microsoft Corporation in the United States and/or other countries. Other product and software names mentioned herein may be either registered trademarks or trademarks of their respective owners. Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2015 Hamamatsu Photonics K.K. HAMAMATSU PHOTONICS K.K. www.hamamatsu.com HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected] Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected] France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected] United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44)1707-294888, Fax: (44)1707-325777 E-mail: [email protected] North Europe: Hamamatsu Photonics Norden AB: Torshamnsgatan 35 SE-164 40 Kista, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected] TXPR1023E02 Italy: Hamamatsu Photonics Italia S.r.l.: Strada della Moia, 1 int. 6, 20020 Arese (Milano), Italy, Telephone: (39)02-93581733, Fax: (39)02-93581741 E-mail: [email protected] NOV. 2015 IP China: Hamamatsu Photonics (China) Co., Ltd.: B1201 Jiaming Center, No.27 Dongsanhuan Beilu, Chaoyang District, Beijing 100020, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: [email protected]