100 kV X-RAY MICROFOCUS X-RAY SOURCE SOURCE L10101 Short FOD: 6.8 mm TLSOF0185 FEATURES ●Short FOD (Focus to Object Distance): 6.8 mm X-ray image with Higher Magnification ratio ●High Power: Maximum Output 20 W ●Focal Spot Size: 5 µm (at 4 W) The focal spot of 5 µm of the sealed type X-ray tube enables sharp and clear X-ray images even at a high magnification. APPLICATIONS ●Non-destructive Inspection ●X-ray CT ●In-line X-ray Inspection [Applicable Objects] •Electronic component •Printed circuit board •Plastic component •Metal component SPECIFICATIONS GENERAL Parameter Input Voltage (DC) Power Consumption (Max.) Operating Ambient Temperature Storage Temperature Operating and Storage Humidity Weight Conformance Standards Operation High Voltage Power Supply Description / Value +24 100 +10 to +40 0 to +50 Below 85 (No Condensation) Approx. 10 CE (EMC: IEC 61326-1, Group1, Class A) Continuous Built-in Unit V W °C °C % kg — — — Description / Value Sealed Type Convection Cooling Beryllium / 150 Tungsten 20 to 100 10 to 200 (20 W Max.) 20 5 to 30 42 6.8 Unit — — µm — kV µA W µm degrees mm Description Tube Voltage and Tube Current Preset / Auto Warm-up Interlock RS-232C Windows® 2000 Professional, XP Professional CPU: Intel Pentium or Higher, Memory: 64 MB or More Unit — — — — — X-RAY TUBE Parameter X-ray Tube X-ray Tube Cooling Method X-ray Tube Window Material / Thickness Target Material Tube Voltage Operational Range Tube Current Operational Range 1 Maximum Output X-ray Focal Spot Size X-ray Beam Angle (Max.) Focus to Object Distance (FOD) X-RAY CONTROL PART Parameter Function Protection External Control Applicable OS Computer Operating Conditions NOTE: 1See the graph of the tube current operational range. ! PRE-CAUTION TO USE 1. X-ray emitted from this device is harmful for human body. And it should be necessary for the operator to protect himself/herself from it. 2. During an operation, the X-ray tube unit should be installed in the X-ray shielded facility or area in order to avoid any X-ray leakage. Also the interlock system in X-ray control unit should be always used in order to avoid any misoperation. OPERATIONAL CAUTION The product may be subject to governmental occupational radiation hazardous regulation therefore the necessary application must be field according to the local regulation. X-RAY FOCAL SPOT SIZE vs. OUTPUT TLSOB0102EA 250 20 kV to 40 kV Range: Reference Data TLSOB0103EA 30 TUBE CURRENT (µA) 200 25 20 15 10 150 100 50 5 0 20 0 0 5 10 15 20 30 40 OUTPUT (W) 50 60 70 80 TUBE VOLTAGE (kV) DIMENSIONAL OUTLINE (Unit: mm) 85.5 ± 1 19 X-RAY FOCAL SPOT 78 ± 1 100 33 FOD 6.8 X-RAY BEAM ANGLE 42° 42° 4-M2 DEPTH 3 COOLING FAN INTERLOCK-1 INTERLOCK-2 288.5 ± 1 195 ± 1 X-RAY FOCAL SPOT POWER DC 24 V 44 GND 167 ± 1 85 MIN. (REFERENCE)* 9.5 147 ± 0.5 4-M6 DEPTH 8 (MOUNTING SCREW HOLES) 171 ± 1 4-M6 DEPTH 8 (MOUNTING SCREW HOLES) 12 23.5 RS-232C 147 ± 0.5 X-RAY FOCAL SPOT SIZE (µm) 35 TUBE CURRENT OPERATIONAL RANGE , * Connection cables aren t included. TLSOA0100ED 90 100 RELATED PRODUCTS X-RAY IMAGE INTENSIFIER CAMERA UNIT (4-inch Beryllium Window) C7876, C7876-10 TXPRF0005 The C7876 is an X-ray image intensifier camera unit ideal for non-destructive inspection of light-element materials and radiation imaging at low energy X-ray levels. The C7876 remarkably improves X-ray transmittance at low energy X-ray levels by using a beryllium window instead of an aluminum window currently used for most X-ray image intensifiers. The results are sharp and clear, high contrast images taken in real time even at low energy X-ray levels down to a few keV. An Aluminum window type is also available. X-RAY IMAGE INTENSIFIER DIGITAL CAMERA UNIT C7336-03 TIIF0170 The C7336-03 consist of a high resolution, high contrast 4-inch X-ray image intensifier (X-ray I.I.) and a 1.45 megapixel digital CCD camera. The X-ray I.I. used has a fixed field-of-view of 100 mm diameter and an input window made of thin aluminum which is excellent in X-ray transmission and causes less scattering of X-rays. These features allow real-time detection at X-ray energy levels from about 20 keV. The 1.45 megapixel digital CCD camera captures high-quality images which are clearer than those taken with conventional analog cameras. X-CUBE™ (COMPACT X-RAY CCD CAMERA) H8480, H8953, H8481 TAPPF0174 X-CUBE™ is a compact X-ray CCD camera designed for non-destructive inspection. Using a general-purpose CCD chip mounted in a rugged but lightweight camera head, X-CUBE™ makes X-ray imaging as easy as an ordinary CCD camera in handling. FOS (Fiber optic plate coated with X-ray scintillator) Left: H8480 Center: H8953 Right: H8481 The FOS is an optical device for X-ray imaging, fabricated by coating an X-ray scintillator material over a fiber optic plate consisting of more than tens of million glass fibers each a few micrometers in diameter. The FOS provides higher sensitivity and resolution than currently used sensitized paper films and also allows real-time digital radiography when directly coupled to a commercially available CCD. The fiber optic plate used in the FOS has excellent X-ray absorption characteristics, so that X-rays penetrating the X-ray scintillator and directly entering the CCD are minimized to less than 1 %. This protects the CCD from the deterioration and increased noise caused by X-ray irradiation, assuring a long service life and maintaining high image quality. Various sizes and shapes of FOS are available to meet your particular needs, including tapered FOP types. ■Structure TMCPF0072 Protection Film Scintillator : CsI(Tl) FOP (Fiber optic plate) 2D Image-Device Windows ® is a registered trademark of Microsoft Corporation in the United States and/or other countries. Other product and software names mentioned herein may be either registered trademarks or trademarks of their respective owners. Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2012 Hamamatsu Photonics K.K. HAMAMATSU PHOTONICS K.K. www.hamamatsu.com HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected] Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected] France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected] United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected] North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected] TXPR1012E01 Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected] China: HAMAMATSU PHOTONICS (CHINA) Co., Ltd.: 1201 Tower B, Jiaming Center, No.27 Dongsanhuan Beilu, Chaoyang District, Beijing 100020, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: [email protected] MAR. 2012 IP