Increased Sample Throughput for ICP-OES on Drinking Water and Waste Water

a p p l i c at i o n N o t e
Inductively Coupled Plasma –
Optical Emission Spectroscopy
Authors
Laura Thompson
Zoe Grosser, Ph.D.
Paul Krampitz
PerkinElmer, Inc.
Shelton, CT 06484 USA
Increased Sample
Throughput for
ICP-OES on Drinking
Water and Waste Water
Abstract
The application of an SC-FAST sample introduction system to the analysis of
natural and certified water samples is described. The SC-FAST system consists
of an autosampler, sample loop, switching valve, high efficiency nebulizer and
a glass cyclonic spray chamber to perform analysis by direct nebulization.
The potential benefits of this introduction system are numerous and include:
increased throughput, reduced memory effects, increased stability, lower
reagent consumption and reduced instrument maintenance. Results indicate that
sample throughput can be nearly tripled while still meeting the requirements.
Sample-to-sample analysis is accomplished in 77 s with significantly improved
washout compared to ICP-OES analysis by conventional introduction.
Introduction
Experimental
The analysis of drinking water and wastewater for trace
metal contamination is an important step in ensuring human
and environmental health. More productive analyses make
better use of public dollars and provide laboratories with a
better cost of ownership for instrumentation. One way to
improve productivity for metals analysis is by using a more
sophisticated and automated sample introduction system to
maximize the time spent on measurements and minimize
the time spent on wash-in and wash-out of the sample.
This work describes the coupling of the Optima™ 7300 DV
with the ESI SC-FAST sample introduction system.
Instrument
The SC-FAST sample introduction system consists of an
autosampler, sample loop, vacuum pump, 6-port switching
valve, a sea spray nebulizer and utilizes flow injection to
perform analysis by direct nebulization. The FAST system
provides a number of advantages over conventional ICP-OES
introduction systems, the most significant of which is higher
sample throughput and reduced memory effects. The sample
loop is in close proximity to the nebulizer which reduces the
time required for sample uptake. Furthermore, the sample
loop prevents samples from contacting the peristaltic pump
tubing, which greatly improves sample washout. In addition
to higher throughput and reduced memory effects, the FAST
system allows for the online addition of internal standards
which simplifies sample preparation and helps minimize
errors and contamination.
An Optima 7300 DV (PerkinElmer®, Shelton, CT) was used
for the analysis of all samples described in this work. An
SC-FAST (Elemental Scientific Inc., Omaha, NE) was coupled
to the ICP-OES for sample introduction. The FAST system is
controlled through the Optima WinLab32™ software and is
shown schematically in Figure 1. The wavelengths monitored
and the viewing modes used for the analytes are listed in
Table 2. Instrument conditions for the Optima ICP-OES and
FAST, as well as experimental parameters used throughout
this work, are listed in Table 3.
Standards
All solutions were prepared using ASTM Type I (>18 MΩ-cm)
water and double-distilled nitric acid. All acid concentrations
reported in this document are described as a relative (v/v)
percentage. Reference materials for this work were obtained
from High Purity Standards (Charleston, SC) and from NIST®,
(Gaithersburg, MD).
Summary of Method
The procedure followed in this work is summarized in Table 1.
Table 1. Summary of Method.
Establish Initial Performance Data
1. Linear Range
2. Perform IDLs and MDLs
3. Analyze Quality Control Samples with acceptable performance
Daily Analysis
1. Light plasma, allow 15 minutes for warm-up
2. Record Instrument Sensitivity
a. Use 1 ppm Mn for axial view, 10 ppm Mn for radial view
b. Record the counts for each and watch for significant changes as compared to signals obtained in previous days/weeks
3. Calibrate using blank and standards
4. Examine data and adjust background and measured wavelengths as needed
5. Screen new samples for relative levels and natural presence of internal std elements
6. Run instrument performance QCS
7. Run analytical QCS
8. Run samples
9. Review results of quality control samples for PASS/FAIL criteria
2
Figure 1. Schematic of a FAST introduction system.
Table 3. FAST-Optima 7300 DV Instrumental Conditions and
Experimental Parameters.
Table 2. Wavelengths Monitored and Viewing Modes.
Analyte
Symbol
Wavelength
Monitored (nm)
View
Optima 7300 DV Parameters
RF Power
1450 watts
Axial
Plasma Gas Flow
15 L min-1
188.979
Axial
Auxiliary Gas Flow
0.2 L min-1
Ba
233.527
Axial
Nebulizer Gas Flow
0.6 L min-1
Beryllium
Be
313.042
Radial
Peristaltic Pump Speed
0.85 mL min-1
Boron
B
249.677
Radial
Nebulizer/Spray Chamber
Sea Spray
Cadmium
Cd
226.502
Axial
Torch Cassette Position
-3
Calcium
Ca
315.887
Radial
Purge
Normal
Cerium
Ce
413.765
Radial
Chromium
Cr
267.716
Axial
Resolution
Normal
Cobalt
Co
228.616
Axial
Integration Time
2 s min/5 s max
Copper
Cu
327.393
Axial
Read Delay
14 s
Iron
Fe
238.204
Radial
Wash Time
1s
Lead
Pb
220.353
Axial
Number of Replicates
3
Lithium
Li
670.784
Radial
Magnesium
Mg
285.213
Radial
Manganese
Mn
257.610
Axial
Mercury
Hg
194.168
Axial
Molybdenum
Mo
202.035
Nickel
Ni
Phosphorus
P
Aluminum
Al
308.215
Radial
Antimony
Sb
206.836
Arsenic
As
Barium
FAST Parameters
Sample Loop Volume
2 mL
Sample Loop Fill Rate
27 mL min-1
Carrier Pump Tubing
Black/Black (0.76 mm i.d.)
Axial
Sample Load Time
7s
231.604
Axial
Rinse
5s
213.617
Axial
Analysis Time (total)
77 s (sample-to-sample)
Experimental Parameters
Potassium
K
766.490
Radial
Selenium
Se
196.026
Axial
Carrier Solution
1% HNO3
Silicon
Si
251.611
Axial
Rinse Solution
1% HNO3
Silver
Ag
328.068
Axial
Acidity of Stds/Samples
1% HNO3
Sodium
Na
589.592
Radial
Strontium
Sr
407.771
Radial
Thallium
Tl
190.801
Axial
Tin
Sn
189.927
Axial
Titanium
Ti
334.940
Axial
Vanadium
V
292.402
Axial
Zinc
Zn
206.200
Axial
Internal Standards
Yttrium
Y
371.029
Radial/Axial
Tellurium
Te
214.281
Radial/Axial
Ce
413.764
Radial
Interference Check
Cerium
Sensitivity Check
Solutions containing 1 ppm Mn and 10 ppm Mn were
analyzed periodically to monitor the sensitivity of the
instrument. The Mn solutions were analyzed weekly and
after the initial installation of the introduction system.
The 1 ppm and 10 ppm Mn solutions were prepared by
diluting 50 μL and 500 μL into 50 mL of 1% HNO3,
respectively.
Internal Standards
All solutions were spiked with 1.5 ppm Y and 2.5 ppm
of Te. The spiking solution was made from single element
stock solutions.
Calibration
Since this method can cover drinking water, waste water
and soils, each element was calibrated to levels typically
encountered in those samples. The concentrations used in
the calibration standards are listed in Table 4. Each standard
contained all elements listed in Table 4.
3
Table 4. Calibration Standard Concentrations.
Analytes
Standard Concentration μg L-1
Al, As, Ag, B, Ba, Be, Ca, Cd,
Ce, Co, Cr, Cu, Fe, Hg, K, Li,
Mg, Mn, Mo, Na, Ni, P, Pb, Sb,
Si, Se, Sn, Sr, Ti, Tl, V, Zn
1000
The linear range results should be viewed with the understanding that a combination of elements in the presence
of a complicated matrix can cause interference effects
and reduce the linear range for a number of elements. For
results that more accurately reflect an individual experiment,
the linear range should be established using standards in a
matrix that replicates the sample matrix as closely as possible.
Monitored Wavelengths
Table 5. Optima 7300 DV IDLs, MDLs and Linear Ranges.
As mentioned earlier, the wavelengths monitored, along
with the viewing mode used for each analyte are listed in
Table 2. All analyses were performed using the instrument’s
auto-integration feature with a minimum of 2 seconds and
a maximum of 5 seconds.
IDL
MDL
Analyte Wavelength (ppb)
(ppb)
MDL
Spike Linear
Level Range
Ag 328.068
0.5
1.1
5
100
Al 308.215
1.5
6.6
20
2000
Initial Performance Demonstration
As 188.979
1.8
1.2
20
100
IDLs
B
249.677
2.0 (0.4)
1.9 (0.2)
5
2000
Ba 233.527
0.2
0.2
2
25
Be 313.107
0.2
0.5
2
50
Ca 317.933
1.3 (0.3)
1.0 (0.1)
5
900
Cd 226.502
0.4
0.4
2
100
Ce 413.764
7.6 (0.9)
12.8 (1.4)
20
100
Co 228.616
0.3
0.5
2
250
MDLs
Cr 267.716
0.4
0.6
2
50
Method detection limits (MDLs) were based upon seven replicate measurements of a series of spiked calibration blanks.
Each blank solution was spiked with analytes at concentrations between 2 and 10 times the calculated IDLs. The MDL
was calculated by multiplying the standard deviation of the
seven replicate measurements by the appropriate Student’s t
test value according to:
Cu 327.393
0.4
0.3
2
300
Fe 259.939
1.0 (0.5)
0.7 (0.2)
5
400
K
766.490
41.1 (2.1)
28.1 (3.6)
100
2000
Mg 279.077
2.5 (0.4)
4.6 (1.0)
20
700
Mn 257.610
0.1
0.5
2
40
Mo 202.031
0.4
0.5
5
125
Na 589.592
7.6 (0.2)
4.7 (0.5)
20
900
Ni 231.604
0.7
0.5
2
125
Pb 220.353
0.6
1.1
20
100
P
213.617
2.9
5.6
20
3000
Li 670.784
0.5 (0.01)
0.5 (0.1)
5
200
Hg 253.652
1.6
9.6
20
100
Sb 206.836
2.6
2.1
20
100
Se 196.026
1.1
1.8
20
100
Si 251.611
2.6 (1.3)
13.3 (4.7)
5
2500
Sr 421.552
1.6 (0.1)
1.6 (0.5)
2
50
Sn 189.927
5.6 (0.5)
6.7 (1.9)
20
2000
Ti 334.940
0.1
0.5
2
50
Tl 190.801
1.1
1.8
20
100
V
292.402
0.2
0.6
2
50
Zn 206.200
0.2
0.4
2
100
Instrument detection limits (IDLs) were estimated using
multiple replicate measurements of the calibration blank
(1% nitric acid). The IDL was calculated to be the concentration
equal to three times the standard deviation of those replicate
measurements; results are shown in Table 5. The IDL calculation was followed according to the procedure outlined.
MDL = (S) x (t)
Note that the Student's t-value is based on a 99% confidence level. Both the Student’s t-value and the standard
deviation are based on n-1 degrees of freedom (t = 3.14
for six degrees of freedom).
Linear Range
A linear calibration range was established for each element
listed. No special detector optimization was done prior to
conducting this procedure. The linear dynamic range for
each analyte was calculated to be the highest concentration
for which the recovery was within ±10% of the true (i.e.,
known) value of the standard. The results from this study
are based upon multi-element standards in a 1% nitric acid
matrix and are given in Table 5.
( ) = axial
4
Table 6. Precision and Recovery Data for High Purity “Trace Metals in Drinking Water” (CRM).
Analyte
Avg Meas.
Conc.
μg L-1
Std.
Dev.
μg L-1
%
RSD
μg L-1
Certified
Value
μg L-1
Recovery of
Certified
Value (%)
Spike
Level
μg L-1
Avg. Spike
Recovery
(%)
Std. Dev.
of Spike
Rec
%
RSD
Ag
---
---
---
2
---
100
99.4
0.8
0.6
As
69.9
0.7
1.0
80
87.4
100
102
3.0
2.9
B
---
---
---
NA
---
100
99.7
1.1 1.1
Be
19.8
0.1
0.5
20
99.0
100
102
0.3
0.3
Ca
33400
18
0.1
35000
95.4
100
---
---
---
Cd
9.66
0.2
1.7
10
96.6
100
101
1.0
1.0
Co
23.3
0.2
0.8
25
93.2
100
100
0.2
0.2
Cu
19.8
0.7
3.4
20
99.0
100
102
1.0
1.0
Cr
19.4
0.3
1.5
20
97.0
100
102
0.7
0.7
Fe
93.2
1.9
2.0
100
93.2
100
102
5.0
4.9
K
2380
31
1.3
2500
95.2
100
---
---
---
Li
19.3
0.8
4.3
20
96.5
100
102
1.4
1.4
Mg
8530
125
1.5
9000
94.8
100
---
---
---
Mn
38.6
0.5
1.3
40
96.5
100
103
1.2
1.1
Mo
101
0.7
0.6
100
101
100
101
1.6
1.6
Na
5580
6.9
0.1
6000
93.0
100
---
---
---
Ni
57.8
0.5
0.9
60
96.3
100
98.6
2.5
2.6
Pb
39.7
3.1
7.7
40
99.2
100
103
2.9
2.8
Sb
---
---
---
10
---
100
107
2.8
2.7
Se
---
---
---
10
---
100
110
6.5
5.9
Sn
---
---
---
NA
---
100
100
2.2
2.2
Sr
246
0.4
0.2
250
98.4
100
94.3
3.0
3.2
Ti
---
---
---
NA
---
100
105
0.4
0.4
Tl
---
---
---
10
---
100
99.5
4.7
4.7
V
29.8
0.4
1.4
30
99.3
100
103
0.5
0.5
Zn
69.5
0.6
0.9
70
99.3
100
98.8
1.6
1.7
NA = Not Applicable
Data Handling
All data from the Optima 7300 DV was collected using a
desktop computer attached to the instrument. The analytical
results presented here were computed using the WinLab32
software and exported as report files. The text and supporting data tables were generated using Microsoft® Word and
Excel®.
Sample Analysis/Quality Control
The accuracy and precision of the above-described method
were verified using certified reference materials and a local
drinking water sample. Certified reference materials were
analyzed without modification to determine the accuracy as
compared to the certified values. Recoveries of spiked reference
materials and the local drinking water sample were calculated.
Results from High Purity Standards “Trace Metals in Drinking
Water”, NIST SRM 1643e “Trace Metals in Water” and a
local drinking water sample are listed in Tables 6, 7 and
8, respectively. The results for the analysis of High Purity
Standards interference check standards “INFCS I + INFCS IV”
are listed in Table 9. Note that INFCS I was diluted 10,000
fold and INFCS IV was diluted 1,000 fold before the standards were combined into one sample vial.
5
Table 7. Precision and Recovery Data for NIST SRM 1643e “Trace Elements in Water”.
Avg. Meas.
Certified
Analyte Conc. μg L-1 Std. Dev. μg L-1 % RSD μg L-1
Value μg L-1
Rec. of
Certified
Spike
Avg. Spike
Std. Dev.
Value (%) Level μg L-1 Recovery (%) of Spike Rec % RSD
Ag
---
---
---
1.062
---
100
90.3
1.7
1.9
Al
133
0.3
0.3
141.8
94.1
100
94.0
1.1
1.2
As
60.2
1.9
3.7
60.45
99.6
100
109
4.3
4.0
B
150
0.3
0.2
157.9
95.0
100
98.5
2.4
2.4
Ba
504
11
2.1
544.2
92.6
100
---
---
---
Be
13.6
0.1
0.5
13.98
97.3
100
104
0.5
0.4
Ca
29600
95
0.3
32300
91.6
100
---
---
---
Cd
6.00
0.2
4.1
6.568
91.4
100
102
1.6
1.6
Co
25.2
0.2
0.6
27.06
93.1
100
103
0.9
0.9
Cr
20.4
0.2
0.8
20.40
100
100
105
1.0
0.9
Cu
23.6
0.3
1.2
22.76
101
100
108
0.2
0.1
Fe
97.6
7.0
7.2
98.1
99.5
100
99.7
4.8
4.9
K
1920
46
2.4
2034
94.4
100
---
---
---
Li
18.3
0.9
5.0
17.4
105
100
104
0.8
0.8
Mg
7600
70
0.9
8037
94.5
100
---
---
---
Mn
38.7
1.6
4.0
38.97
99.4
100
102
1.1
1.1
Mo
125
0.3
0.2
121.4
103
100
105
1.4
1.3
Na
19100
100
0.5
20740
92.1
100
---
---
---
Ni
59.5
0.9
1.4
62.41
95.3
100
102
1.1
1.1
Pb
19.4
0.5
2.6
19.63
98.8
100
107
2.1
2.0
Sb
59.9
3.7
6.1
58.30
103
100
102
0.8
0.7
Se
---
---
---
11.97
---
100
105
5.3
5.0
Sr
319
0.9
0.3
323.1
98.7
100
101
9.5
9.4
Sn
---
---
---
n/a
---
100
104
0.03
0.03
Ti
---
---
---
n/a
---
100
108
0.3
0.3
Tl
---
---
---
7.445
---
100
106
6.4
6.1
V
37.0
0.2
0.5
37.86
97.7
100
107
0.6
0.6
Zn
76.6
0.7
0.9
78.5
97.6
100
101
0.7
0.6
NA = Not Applicable
6
Table 8. Results and Spike Recoveries for Local Drinking Water. Analyte Wavelength (nm) LDW Conc. μg L-1
Low Spike
Level μg L-1
Low Spike
Low Spike
Results μg L-1 % Recovery
High Spike
Level μgL-1
High Spike
Results μg L-1
High Spike
% Recovery
Al
100
100.3
500
475
95.0
308.215
7.50
100.3
As
188.979
< MDL
100
102.9
102.9
500
487
97.4
B
249.677
11.2
100
100.1
100.1
500
472
94.4
Ba
233.527
69.0
200
192.9
96.5
1000
946
94.6
Be
313.042
0.120
100
103.8
103.8
500
490
98.0
Ca
315.887
7120
100
---
---
500
---
---
Cd
226.502
< MDL
100
98.8
98.8
500
484
96.8
Co
228.616
< MDL
100
102.1
102.1
500
486
97.2
Cr
267.716
< MDL
100
101.6
101.6
500
484
96.8
Cu
327.393
97.6
100
106
106
500
478
95.6
Fe
238.204
83.5
100
94.5
94.5
500
491
98.2
K
766.490
877.7
100
---
---
500
---
---
Li
670.784
< MDL
100
102
102
500
492
98.4
Mg
285.213
1710
100
---
---
500
---
---
Mn
257.610
16.4
100
102
102
500
503
101
Mo
202.035
< MDL
100
101
101
500
480
96.0
Na
589.592
12800
100
---
---
500
---
---
Ni
231.604
0.730
100
99.4
99.4
500
482
96.4
P
213.617
< MDL
100
99.0
99.0
500
476
95.2
Pb
220.353
< MDL
100
99.7
99.7
500
487
97.4
Sb
206.836
< MDL
100
102
102
500
492
98.4
Se
196.026
< MDL
100
105
105
500
525
105
Sn
189.927
2.94
100
97.4
97.4
500
475
95.0
Sr
407.771
52.4
100
101
101
500
488
97.6
Ti
334.940
< MDL
100
103
103
500
503
101
Tl
190.801
< MDL
100
104
104
500
475
95.0
V
292.402
0.220
100
103
103
500
487
97.4
Zn
206.200
47.2
100
100
100
500
478
95.6
7
Table 9. Results for High Purity INFCS I + INFCS IV.
Analyte
Wavelength (nm)
INFCS Run #1
μg L-1
INFCS Run #2
μg L-1
INFCS Avg. Result μg L-1
Duplicate
RPD
INFCS True
Value μg L-1 % Recovery
As
188.979
110
111
111
0.8
100
111
Ba
233.527
30.0
28.5
29.3
5.2
30
97.7
Be
313.042
10.9
10.9
10.9
0.1
10
109
Ca
315.887
4950
4910
4930
0.8
5000
98.6
Cd
226.502
32.2
33.2
32.7
2.9
30
109
Co
228.616
31.0
31.5
31.3
1.6
30
104
Cr
267.716
31.5
31.4
31.5
0.4
30
105
Cu
327.393
31.3
31.1
31.2
0.5
30
104
Fe
238.204
5230
4910
5070
6.2
5000
101
Hg
194.168
4.97
5.54
5.30
10.8
5
105
K
766.490
2130
2150
2140
0.7
2000
107
Mg
285.213
5070
5050
5060
0.3
5000
101
Mn
257.610
20.8
19.5
20.1
6.4
20
100
Ni
231.604
30.9
31.5
31.2
2.1
30
104
Pb
220.353
105
108
106
2.9
100
106
Se
196.026
53.9
48.8
51.4
10.1
50
103
Tl
190.801
102
103
103
1.1
100
103
V
292.402
31.1
31.1
31.1
0.2
30
104
Zn
206.200
34.8
32.6
33.7
6.7
30
112
Conclusions
Acknowledgement
The FAST system coupled with the Optima 7300 DV has been
shown to produce results while increasing sample productivity
over 200% when compared to analyses with conventional
introduction systems. Since the FAST system eliminates virtually
all of the rinse and read delay times, most of the time is now
spent running samples, therefore increasing productivity.
Also, since the FAST reaches a steady state signal much more
quickly than conventional sample introduction, instrument and
method detection limits are improved almost 2-fold for many
analytes. When used in conjunction with the SC autosampler
and Optima 7300 DV, the FAST system provides a rugged,
automated sample introduction system that can significantly
reduce labor costs and improve laboratory productivity.
We would like to thank Maura Mahar, a graduate student at
University of Massachusetts and summer intern for her excellent
work contributing to this application. Special thanks also to
Stan Smith, PerkinElmer Product Specialist and to Matthew
Knopp of ESI for their time and input.
References
1.EPA Method 200.7, “Determination of Metals and Trace
Metals in Water and Wastes by Inductively Coupled
Plasma-Atomic Emission Spectrometry,” Revision 4.4,
1994, Environmental Monitoring Systems Laboratory,
Office of Research and Development, United States
Environmental Protection Agency, Cincinnati, OH 45268.
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2009, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
008710_02