5 © Siemens AG 2015 General information 5/2 Communication 5/2 5/7 5/8 Continuous gas analysis Gas analysis library for SIMATIC PCS 7 Process gas chromatography 5/11 Operator functions of Series 6 5/12 FAT & factory certificates 5/13 5/13 5/21 5/33 Ex versions Continuous gas analysis, extractive Continuous gas analysis, extractive ATEX II 2G control unit Continuous gas analysis, extractive ATEX II 2G control unit, leakage compensation Continuous gas analysis, extractive ATEX II 2G/3G control unit, continuous purging Continuous gas analysis, extractive Purging unit FM (Class I Div 2) Continuous gas analysis, extractive Additional units Continuous gas analysis, in-situ LDS 6 Continuous gas analysis, in-situ LDS 6, EEx barrier Continuous gas analysis, in-situ SITRANS SL Process gas chromatography 5/34 5/34 5/35 5/37 Tables Conversion tables Dew point/saturation table International standards 5/39 Definitions 5/22 5/24 5/26 5/27 5/29 5/31 5/32 Siemens AP 01 · 2015 © Siemens AG 2015 General information Communication Continuous gas analysis ■ Overview Reliable functioning of analyzers is of decisive importance for process control. It is necessary to record, correct and transmit measured values, to set and modify parameters, to check functions, to update calibrations, and to scan status signals e.g. for preventive maintenance. Communication between the operator and device is therefore an important part of process analysis, and the offered facilities have become a decisive performance feature of analyzers. Networking over ELAN ELAN communication is used e.g. for the correction of cross-interference. Direct connection is only possible between Siemens gas analyzers. Specification for the interface cable Surge impedance 100 ... 300 Ω, with a measuring frequency of > 100 kHz Extractive Cable capacitance Typ. < 60 pF/m The gas analyzers of Series 6 (ULTRAMAT 6, ULTRAMAT/OXYMAT 6, OXYMAT 6, OXYMAT 61, FIDAMAT 6 and CALOMAT 6) as well as the ULTRAMAT 23 offer the following communications facilities in addition to data transmission over analog and binary outputs: • RS 485 interface • SIPROM GA • PROFIBUS DP/PA • Generic communications interface (only OXYMAT 6, ULTRAMAT 6 and ULTRAMAT/OXYMAT 6). Core cross-section > 0.22 mm2, corresponds to AWG 23 Cable type Twisted pair, 1 x 2 conductors of cable section Signal attenuation Max. 9 dB over the whole length Shielding Copper braided shield or braided shield and foil shield Connection Pin 3 and pin 8 RS 485 interface The serial interface integrated as standard permits communication between several analyzers over the internal bus (ELAN). Parameterization is carried out using the analyzer’s menu. *1' 0 5 9 0 *1' 0 9 0 *1' 0 9 0 Bus cable with plug connections, ELAN networking 5/2 Siemens AP 01 · 2015 9-pin connector (RS 485) (device 1) 9-pin connector (RS 485) (device 2) 9-pin connector (RS 485) (device 3) © Siemens AG 2015 General information Communication Continuous gas analysis Bus terminating resistors Hardware requirements Pins 3-7 and 8-9 of the first connector of a bus cable must be bridged (ELAN networking). The following hardware and system requirements must be provided for the PC/laptop configuration in order to use SIPROM GA: • Windows computer with Pentium 133 MHz and 32 MB RAM: Recommendation: Pentium II 266 MHz and 64 MB RAM • CD-ROM drive (for installation) • Vacant hard disk capacity of at least 10 MB • VGA graphics card (Windows-supported); resolution: 1024 x 768 • Printer (Windows-supported) • MS-Windows 95, ME; NT 4, Windows 98, Windows 2000 or Windows XP operating system • Vacant COM port (COM 1, 2 ...) - The RS 485 / RS 232 interface converter is required for coupling to the RS 485 ELAN network - A standard 10-Mbit or 100-Mbit network (RJ45 connection) with TCP/ IP is required for connecting the Ethernet / RS 485 interface converter Note It is advisable to install a repeater on the device side in the case of a cable length of more than 500 m or with high interferences. Networking with SIPROM GA When used externally, the RS 485 interface requires software matched to the analyzers, e.g. SIPROM GA. SIPROM GA is a software program for communication between PC/laptop and analyzers. A maximum of 12 devices (electronics modules) with up to four channels/measured components of the following type can be connected, displayed and remote-controlled per COM interface: • OXYMAT 6/61 • OXYMAT 64 • ULTRAMAT 6 • CALOMAT 6 • CALOMAT 62 • FIDAMAT 6 • ULTRAMAT 23 SIPROM GA allows access to device parameters, right up to the configuration of devices. All analyzer functions (except factory default functions) can be remote-controlled and monitored in this manner. SIPROM GA is therefore an ideal servicing and maintenance tool for Siemens gas analyzers. In addition to remote control of all operator functions, SIPROM GA offers complete access to all diagnostics data. SIPROM GA therefore permits preventive maintenance as well as fast responses when maintenance becomes necessary or when the production sequence is changed. Accessories for the network For cables, connectors, repeaters etc., see Catalog IK PI or CA 01 on the Mall under SIMATIC NET communications systems/PROFIBUS/network components. Networking with SIPROM GA via converter Up to 12 analyzers with max. four components each can be networked. The functional principle is shown in the following illustration. up to 12 analyzers 5 SIPROM GA ensures: • High operational reliability • High availability • Central, comprehensive information • Fast response time • Flexibility • Economical system integration In addition to output of TAG No., components, current measured values, comprehensive diagnostics information (status) and parameter settings on the analyzer display, SIPROM GA also offers the following features: • Bargraph display • Recorder display of one or more measured values with printer output • Calibration functions (adjustment of all setpoints for calibration, remote calibration) • Saving of all device data • Remote control of all device functions • Remote calibration • Online help • Downloading of new device firmware • Cyclic saving of measured values on hard disk • Writing user data to the device’s EEPROM, or downloading data from it. RS 485/RS 232 converter Typical design of an RS 485 network with SIPROM GA The gas analyzers can be installed at distances up to 500 m. One network can be connected to each COM port. Access to the analyzers using SIPROM GA is carried out either: • Directly from the PC over an RS 485 interface or • Over an Ethernet gateway Siemens AP 01 · 2015 5/3 © Siemens AG 2015 General information Communication Continuous gas analysis Networking with SIPROM GA via Ethernet When networking with Ethernet, there are no limitations for the distance between PC and gateway. In addition, communication over Ethernet permits connection of several gateways to one COM port, and thus the possibility for monitoring and operating several widely distributed or separately installed analyzers/systems from one station. up to 12 analyzers Ethernet Gateway Ethernet Gateway Typical design of an RS 485 Ethernet network with SIPROM GA 5 5/4 Siemens AP 01 · 2015 © Siemens AG 2015 General information Communication Continuous gas analysis PROFIBUS The usual transmission of measured values and fault messages via analog and binary outputs requires complex cabling. On the other hand, when using PROFIBUS DP and PROFIBUS PA, one single two-wire conductor is sufficient for digital transmission e.g. of all measured values (also from several channels), status information or diagnostics functions for preventive maintenance. The PROFIBUS DP version with its high transmission rate for relatively small data quantities per device is widely used in production automation, whereas PROFIBUS PA takes into account the features important for process engineering, e.g. large data quantities and use in hazardous areas. The limited dynamic performance of 4 to 20 mA mA signals can be replaced, the laborious configuring of measuring ranges can be omitted. By using simulated measured values without media, increased safety can be provided for the plant configuration, and configuration errors can be avoided. Parameter sets can be generated offline (from your desk) and subsequently downloaded and saved in the device. Local operations can thus be reduced to a minimum. The Siemens gas analyzers • OXYMAT 6/61 • OXYMAT 64 • ULTRAMAT 23 • ULTRAMAT 6 • CALOMAT 6 • CALOMAT 62 • FIDAMAT 6 are PROFIBUS-compatible when using an optional plug-in card (retrofitting also possible) and therefore comply with the "Device profile for analyzers" defined as binding by PI (PROFIBUS International). Customer benefits include an enormous savings potential in all plant areas, covering configuration and commissioning, operation and maintenance, up to subsequent plant expansions. Operation of the gas analyzers from a control system or a separate PC is possible using the SIMATIC PDM (Process Device Manager); this is software which executes under Windows and which can also be integrated into the SIMATIC PCS 7 process control system. This permits a clear presentation for integration of the analyzers in the system as well as for the complex parameter structure of the analyzers. Direct connection of the analyzers to a control system is also possible without PDM, e.g. using STEP7, but this necessitates additional programming and offers less user friendliness. In most cases, this direct connection is therefore only applicable if acyclic (device operation) data are not used. A differentiation is made between cyclic and acyclic services. Cyclic services are used to transmit time-critical data such as measured values and statuses. Acyclic services permit scanning or modification of device parameters during operation. Both graphic displays and values can be output on a PC. Signaling of maintenance, fault and diagnostics information is also cyclic. These data are displayed in plain text when using SIMATIC PDM. The binary outputs can also be switched using cyclic services, thus also permitting triggering of relays over PROFIBUS (e.g. for measuring point switchover, calibration etc.). 5 3URFHVVFRQWUROV\VWHP 6 '33$FRXSOHU '33$OLQN *DVDQDO\]HUV *DVDQDO\]HU *DVDQDO\]HU Schematic structure of a PROFIBUS system Siemens AP 01 · 2015 5/5 © Siemens AG 2015 General information Communication Continuous gas analysis The following acyclic device parameters and configurations can be used in PROFIBUS DP and PROFIBUS PA by means of SIMATIC PDM: • Factory data • Diagnostics values • Logbook • Display measuring ranges • Zero calibration • Sensitivity calibration • Setpoints for zero/sensitivity • Total/single calibration and AUTOCAL • Select measuring ranges • Define measuring ranges • Electric time constants • On/off functions • Chopper frequency • Magnetic field frequency • Date/time • Measuring point switchover • Logbook settings • Relay assignment • Binary inputs • Reset • Save/load data • Suppression of short noise signals • Calibration tolerances • Switch valves • PROFIBUS configuration 5 Use of PROFIBUS offers the following customer benefits: • Cost reductions for planning, installation and operation • Use of (distributed) device intelligence • Replaceability of devices • Only one cable for everything, no complex cabling • No limited 4 to 20 mA resolution • No laborious parameterization of measuring ranges • Simulation of measured values • Simplification of commissioning • Testing of network/AS • Avoidance of errors during startup • Online diagnostics • Offline parameterization 5/6 Siemens AP 01 · 2015 Generic communications interface (only OXYMAT 6, ULTRAMAT 6 and ULTRAMAT/OXYMAT 6) User benefits are offered thanks to numerous functions which are mainly required in the automotive industry, for example to carry out repeated linearization. In contrast to PROFIBUS and ELAN, communication is only possible between one device and one PC, and takes place according to the master/slave principle. The device only transmits data when requested by a command telegram, where only one command can be processed and replied to at a time. The generic communications menu can be called using Function88, and the parameters adjusted. Continuous gas analysis/in-situ LDS 6 can send and receive data over an Ethernet connection together with the LDScom software. This installation and service tool is able to check and adapt device status and calibration parameters from a remote location. If necessary, even a complete system check can be carried out over the remote connection. If servicing is necessary, the required information can be sent to the Siemens service engineer by modem, and he can then carry out the appropriate measures from the remote location. This facility for remote maintenance and diagnostics is implemented using a standard LAN modem. Public telephone network Modem Ethernet LDS 6 Preset IP address in network 192.168.XXX.XXX LDS 6 additionally possible LDS 6 additionally possible LDS 6 additionally possible External connection of LDS 6 via a modem for implementing remote maintenance measures © Siemens AG 2015 General information Communication Gas analysis library for SIMATIC PCS 7 ■ Overview PCS 7 Add-on ■ Function fit for SIMATIC PCS 7 V7 Driver blocks The gas analyzers are integrated into the hardware configuration of the SIMATIC PCS 7 process control system using their GSD files. Parameterization of the driver blocks is subsequently carried out corresponding to the device configuration. The driver blocks provide the following functions: • Reading of analyzer values • Starting of autocalibration • Evaluation of device-specific diagnostics • Standard diagnostics • Alarms for analyzer values (alarm limits adjustable on the block) • Simulation Symbols and faceplates The symbols are automatically created and connected using the wizard "Create block symbols". The faceplates can be displayed in various views: • Standard • Maintenance • Configuration • Limits • Trend • Alarm ■ More information The driver blocks from the gas analysis library permit integration of the following gas analyzers into the SIMATIC PCS 7 process control system over PROFIBUS DP: • ULTRAMAT 6 and ULTRAMAT 23 • CALOMAT • OXYMAT Please contact your Siemens sales partner for further information and for ordering. 5 The driver blocks permit access to the measured values and to the calibration functions of these devices. They can also be used to evaluate and display diagnostics information, and to trigger alarms if necessary. Note: The gas analysis library can be used together with SIMATIC PCS 7 V6 and V7. Siemens AP 01 · 2015 5/7 © Siemens AG 2015 General information Communication Process gas chromatography ■ Overview The MAXUM edition II and MicroSAM gas chromatographs can transfer measured results and status information to process control systems, operator panels or printers during operation. Interfaces Chromatograph, operator panel, printer and control system use special electronic interfaces: • Electrical connection The device interfaces are connected by electric cables. The electrical properties of the interfaces are standardized. • Control of communication and language Rules must be observed to control the communication. It must be clearly defined in networks who is the "sender" and who is the "receiver" of the data. Both communication partners must use the same protocol. MODBUS MODBUS is a rule for controlling data transfer between two computer systems - a transmission protocol. MODBUS is the "de facto" industry standard for connecting measuring and control devices to process control systems (PCS). Most process control systems can be equipped with serial interfaces and MODBUS. 5 Memory division To ensure that the meaning of the registers is known to each network station, this must be defined in the configuration. The results of each component in each sample must be written into defined positions in the PCS memory. These address declarations depend on the number of chromatographs, samples and components. The same applies to status, sample sequence and sample release. Standard addresses are also defined in these cases. OPC server (OLE for Process Control) OPC is a vendor-independent software interface. It allows standardized access from Windows applications to chromatograph data. OPC corresponds to a typical client/server architecture. OPC allows a universal connection between any Windows application which supports an OPC client interface and the MAXUM edition II / MicroSam. The OPC server is usually installed on a separate PC. OPC is a modern alternative to MODBUS. MAXUM edition II and MicroSAM do not require an additional interface, they use the existing Ethernet connection (TCP/IP). Using the MODBUS coupling, information can be sent interference-proof via just one data line. Information can be read from process gas chromatographs (PGC), and certain functions of the PGC can also be parameterized. OPC standardizes the access to measured values, status functions, control functions and analytical data in a manner similar to MODBUS. Advantages: • Information on the PGC status during operation • Supply of protected data in numerical form without falsification by interfering pulses • Reduced cabling overhead Advantages with OPC applications: • Reduced maintenance costs • Simple GUI for configuration • Reduced system integration costs • Reduced test costs • Reduced maintenance costs The MODBUS coupling can: • Transmit measured values • Transmit status information • Output information on the current analysis • Trigger control functions The MODBUS uses a master/slave transmission procedure. The control system is always the master, the process gas chromatograph is the slave. The representation of data in the message frames is based on the compact RTU format. 23&FOLHQW 23&FOLHQW 3&66LHPHQV3&6 OPC server 5/8 Siemens AP 01 · 2015 23&VHUYHU © Siemens AG 2015 General information Communication Process gas chromatography Hardware components Software NAU - Network Access Unit Modern chromatographs are controlled by microprocessors. We differentiate between software in the device and software on a PC operator panel. An NAU expands and supplements a GC network, and has three fundamental functions: • Enclosure for 7 additional I/O plug-in cards • Connection of serial ASCII printers and external host PCs (control system) • Central operation of a GC network from one point The Network Access Unit (NAU) is an input/output station for the Siemens process chromatographs. It can be used to centrally call, process and pass on data. It is used if it is not possible to connect the electronics close to the analyzer, if the number of MicroSam inputs/outputs is insufficient, or if installation in a central control room is required. This significantly reduces the required wiring to the control room. The NAU can also be used as a central control unit for function testing, data output and parameterization of the MicroSam. It permits access to the MicroSam to which it is connected, as well as to all other Siemens gas chromatographs and further NAUs of the system which are networked over the system bus. The NAU is connected to the Ethernet or DataNet and has a total of 7 slots to accommodate a wide range of electronics cards. These comprise cards for analog and digital signal processing as well as interfaces for host computers and process control systems. An NAU can be expanded by a further 10 slots using a CAN Extension Unit. A total of 7 different electronics cards are available: Software in the chromatograph The chromatograph can carry out analyses independently, without an operator panel being connected. It then requires its own control software and local operating software (HMI). Software in the operator panel (PC) Siemens gas chromatographs can be operated over Ethernet and a PC, by using the built-in control panel (HMI), or with a Network Extension Unit (NAU). Workstation software The most important programs this contains are the MAXUM System Manager, the MAXUM EZChrom and the HMI-Emulation. In addition, it contains useful MAXUM utilities and loadable extensions such as: MaxBasic For modification of MaxBasic programs in the gas chromatographs or the NAU. MAXUM OPC Server For coupling of the MAXUM e.g. to control systems. Simulated Distillation For import/export of methods for simulated distillation. MAXUM System Tools • Input/output module 8 analog outputs • Input/output module 4 digital outputs, 4 digital inputs • Input/output module 2 digital outputs, 2 digital inputs, 2 analog outputs, 2 analog inputs • Communication module 10 Base FO Ethernet (fiber-optic coupling) • Communication module DataNET Copper (redundant system bus) • Communication module DataNET Fiber Optic • Communication module Advanced Data Highway (OptiCHROMe Advance coupling) CEU - CAN Extension Unit (only MAXUM edition II) • Increases the I/O capacity of the GC or NAU • 10 additional I/O plug-in cards • Has its own power supply • Also for zone 1 ANG - Advance Network Gateway Connects OptiCHROMs to the Ethernet network. DNH - DataNET Hub Communications router completely redundant in the network: has 2 own TCP/IP addresses, dual electronics and power supply. ANCB - Advance Network Communication Board Converts the communications protocols in the device For data logging, for firmware updating, or for operation of discontinued OptiCHROMs (APC 8.0). Different levels of operation Operation of the PC is in three levels: • System Manager - network monitoring and configuration of the chromatograph • EZChrom - method development and control of analysis • MMI emulation - operator control and monitoring System Manager: Configuration of database and applications The System Manager provides the connection to the chromatograph as well as an overview of the network. It is additionally used for configuration and high-level monitoring of the chromatograph, and additionally branches to the MMI emulation, EZ Chrom and Basic editor. The System Manager almost exclusively outputs static displays, i.e. it fetches a table from the chromatograph’s database, and saves it again there later. Only results and alarms are displayed dynamically. The System Manager presents the hardware in tables, e.g. the system table contains all hardware components of the chromatograph: detectors, valves, ovens, digital outputs, etc. In addition, the System Manager can: • Save databases of a chromatograph as a file on the PC, or load them from the PC into the chromatograph. • Upgrade the chromatograph software. • Call EZChrom, MMI emulation, Datalogger or the MaxBasic editor. Two versions: • MAXUM directly to DataNET • MAXUM directly to OptiCHROM data highway Siemens AP 01 · 2015 5/9 5 © Siemens AG 2015 General information Communication Process gas chromatography EZChrom: Generate methods and sample sequences The EZChrom software is installed on the PC and is also included in the operator software in the chromatograph, e.g. it integrates the detector signals there, calculates the results, or switches time-controlled events. EZChrom on the PC is responsible for the following tasks: • Generate or modify methods • Carry out re-integration • Calibrate a method • Display and print saved and real-time chromatograms • Archive chromatograms • Generate and modify sample flow sequences • View analysis clock • Switch the chromatograph to Run or Hold. MMI emulation: Operator control and monitoring This is identical to operation on the built-in control panel of a MAXUM or an NAU. It is used for operator control and monitoring. For example, it is possible to display results, switch valves or modify temperatures. However, there are only minimum possibilities for editing the configuration and tables. The MMI is always a dynamic display. 5 5/10 Siemens AP 01 · 2015 APC 8.0 This is an interface to the OptiCHROM Advance gas chromatographs. It is started from the MAXUM System Manager. It permits operation of older types of chromatograph via a PCI card. The following can be executed with APC 8.0: • Service panel emulation on the PC • Data logging • Viewing of chromatograms • Editing of tables in OptiCHROM. GC-Tools operating software GC-Tools is the Windows-based operating software for discontinued Siemens gas chromatographs of Series 202 and 302. This software expands the BEDI operating software, which is based on MS-DOS. The Network Explorer shows all information at the device level, e.g.: • Chromatogram display • Manual control of all parameters • Drivers for the four serial interfaces of the chromatograph © Siemens AG 2015 General information Operator functions of Series 6 Main menu No. Function designation Diagnostics 1 2 Calibration Manual SIPROM GA PA/DP V1.6.0 PA/DP V2.0.0 Factory data X Diagnostic values X X X X X — 3 Log book X X X — 4 X Display measuring ranges X X — X 20 Zero calibration X X — X 21 Sensitivity calibration X X — X 22 Zero point/sensitivity setpoints X X — X 23 Total/individual calibration X X — X 24 AUTOCAL X X X X 25 Drift values X X — — 26 Calibration with air (OXYMAT 64 only) X — — — Measuring ranges (Code 1) 40 Select measuring ranges X X — X 41 Define measuring ranges X X — X Parameter (Code 1) 50 Electrical time constants X X — X 51 Limit values X X — — 52 On/off functions X — — X 53 Status messages X X — — 54 Graphical measured value representation X X — — 55 Measured-value display X X — — 56 LCD contrast X — — — 57 Chopper frequency (ULTRAMAT 6 only) Magnetic field frequency (OXYMAT 6 only) Flame ignition (FIDAMAT 6 only) X X — X 58 Date/time X X — X 59 Measuring point switchover X X — — 60 Log book settings X — — X 61 Vibration compensation (OXYMAT 6 only) Switch internal valves (FIDAMAT 6 only) X X — X 62 Switch external pressures (FIDAMAT 6 "without pump" only) X X — X 70 Analog output X X — — 71 Relay assignment X X — X 72 Binary inputs X X — X 73 ELAN configuration X X — — 74 Reset X X — X 75 Save, load data X X — X 76 Suppression of short noise signals X X — X 77 Measured value memory (analog output) X X — — 78 Calibration tolerances X X — X 79 Change codes X X — — 80 Unit test X X — — 81 Language selection X X — — 82 Pressure correction (ULTRAMAT 6, OXYMAT 6, OXYMAT 64 and CALOMAT 62 only) X X — — 83 Correction of cross-interference X X — — 84 Phase calibration (ULTRAMAT 6 and OXYMAT 6 only) X — — — 85 Switch valves X — — — 86 Linear temperature compensation X X — — 87 Fault on/off X — — — 88 AK configuration (ULTRAMAT 6 and OXYMAT 6 only) X — — — 89 Sample chamber heating (ULTRAMAT 6, OXYMAT 6 and CALOMAT 62 only) X X — — 90 PROFIBUS configuration X X X X 91 Startup state (FIDAMAT 6 only) X X — — 92 Pressure values (FIDAMAT 6 only) X X — — 93 Units (FIDAMAT 6 only) X — — — — Control of external valves — — — X — Software download — X — — Configuration (Code 2) Siemens AP 01 · 2015 5/11 5 © Siemens AG 2015 General information FAT & factory certificates Selection and ordering data Article No. FAT & factory certificates for extractive gas analyzers of Series 6 and ULTRAMAT 23 7MB8100- 7 7 7 7 7 - 7 7 7 7 not applicable for Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Factory acceptance (FAT) with customer Visual inspection and basic settings None Visual acceptance, 1 to 8 devices, incl. function test and calibration Visual acceptance, 9 devices and more, incl. function test and calibration 0 1 2 Measured signal response None Noise, drift Noise, drift, linearity, T90 time A B C Compensation, cross-interference None Pressure compensation 1 interfering gas3) 2 ... 3 interfering gases3) Pressure compensation and 1 interfering gas3) Pressure compensation and 2 or 3 interfering gases3) A B C D E F Factory acceptance, explosion protection None Pressurized enclosure for explosion-proof units (functionality) Relay test Pressurized enclosure for explosion-proof units and relay test FIDAMAT FIDAMAT FIDAMAT 0 1 2 3 Number of test channels None 1 ... 3 4 ... 6 7+ 19" rack units 19" rack units 0 2 3 4 Certificates 5 General certificates Factory certificate DIN EN 10204 2.1 (quality test certificate)1) Test report according to DIN EN 10204 2.21) Certificate of origin1) Certificate of origin1) and test report according to DIN EN 10204 2.2 Certificate of origin1) and factory certificate DIN EN 10204 2.1 Certificate of origin, factory certificate 2.1 and test report 2.2 according to DIN EN 10204 Factory certificate 2.1 and test report 2.2 according to DIN EN 10204 Factory certificate EN 10204 2.1 following repair Parameter sheets (only with suffix Y22) 0 1 2 3 4 5 6 7 8 Acceptance test certificate according to DIN EN 10204 3.1 with suffix Y22 None Measured values, noise, drift, linearity Measured values, noise, drift, linearity, pressure compensation Measured values, noise, drift, linearity, pressure and temperature compensation A B C D Acceptance test certificate according to DIN EN 10204 3.1 extended None Cross-interference of residual gases (H2O and 2 other gases) T90 time Influence of atmosphere containing CO2 Cross-interference of residual gases2) and T90 time Cross-interference of residual gases2) and influence of atmosphere containing CO2 T90 time and influence of atmosphere containing CO2 Cross-interference of residual gases2), T90 time and influence of atmosphere containing CO2 A B C D E F G H Factory insepction certificate DIN EN 10204, language German English French Required analyzer information FIDAMAT FIDAMAT 0 1 2 Order code Add "-Z" to Article No. and specify Order codes. Information on product/order with order item and contact partner (Sales Region, region or distributor) Y22 1) Can also be ordered following delivery 2) H2O and 2 other gases 3) Special gases are invoiced with a surcharge as required. Orders for factory acceptance (FAT) with customer should be ordered with quantity 1. Orders for certificates can be ordered with quantity ≥ 1. 5/12 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive ■ Overview Use of Series 6 in hazardous areas Dependent on the application, the measuring equipment can include the following parts: • Analyzer • EEx p safety equipment (purging unit) • Flame arrestors • Ex i isolation amplifier • Isolating relay Gas analyzers Suitability-tested field analyzers of Series 6 must be used to measure gases in hazardous areas. The Series 6 analyzers are approved in accordance with Ex type of protection "Pressurized enclosure EEx p" for Zone 1 and Zone 2. In addition, these analyzers must be connected to monitoring equipment which must also be suitability-tested. Exception: a pressurized enclosure is not required in zone 2 for the measurement of gases whose composition always remains below the lower explosive limit (LEL); in this case, it is sufficient for the field housing to be gas fume-proof (type of protection EEx n R). Following pre-purging of 5 minutes, the monitoring equipment ensures that no gas fumes can enter the enclosure, and accumulation of the sample gas in the enclosure is prevented. The volume flow during the pre-purging phase is > 50 l/min. The protective gas is usually fed into the analyzer enclosure from a supply network via the monitoring equipment. Category ATEX II 2G (Ex zone 1) Two versions of pressurized enclosure EEx p complying with the directive 94/9/EC are available for use in zone 1: • Pressurized enclosure with compensation of losses resulting from leaks The principle of this type of protection is based on prevention of ingress of the surrounding atmosphere or of the sample gas into the enclosure of the electrical equipment. Only that volume of protective gas is fed into the enclosure that is required to maintain an overpressure of at least 50 Pa compared to the sample gas pressure and atmospheric pressure. The maximum purging gas pressure is 165 hPa; this results in a maximum permissible sample gas pressure of 164 hPa; If the sample gas is combustible or occasionally flammable, the analyzer enclosure must be additionally purged with inert gas (e.g. nitrogen). In these cases, you must additionally ensure that the internal enclosure pressure is at least 5 mbar higher than the fail-safe-regulated sample gas pressure. If the pressure control of the sample gas is not fail-safe (= "double fault safety"), but only operationally safe (="single fault safety"), a differential pressure switch of the EEx p safety equipment must be used to signal if the sample gas pressure exceeds the purging gas pressure. This measure trips a safety shutdown. With occasionally flammable sample gas mixtures, flame arrestors must be additionally mounted externally at the sample gas inlet and outlet. Both the differential pressure switch and the flame arrestors come into contact with the sample gas and must therefore be made of corrosion-proof material, if applicable. Test certification: PTB 00 ATEX 2022 X Device identification: II 2 G Eex p [ia] ia IIC T4 • Pressurized enclosure with continuous purging The principle of this type of protection is based on having continuous purging of the EEx p enclosure after the pre-purge. It prevents ingress of the surrounding atmosphere and ensures that, for example, sample gas released through leaks is thinned to the extent that a combustible mixture cannot be created. The volume flow of the protective gas is fixed at 1 l/min and exceeds the maximum release volume by a factor of more than 100. Protective gas flows continuously through the enclosure with a volume flow of at least 1 l/min; in addition, the flow ensures that the enclosure pressure is increased to at least 50 Pa higher than the surrounding pressure. The max. permissible purging gas pressure is 25 hPa. The max. permissible sample gas pressure is equivalent to the permissible analyzer sample gas pressure. Test certification TÜV 01 ATEX 1708 X Device identification: II 2 G EEx p [ia] ia IIC T4. The fundamental safety requirements of both versions are satisfied by compliance with the European standards EN 50014:1997, EN 50016:1995, EN 50020:1994 and EN 954:1996. The purging gas is monitored using EEx p monitoring equipment: This is a stand-alone unit which is connected electrically and pneumatically to the analyzer. Explosion protection is only provided when both devices are combined (analyzer and purging unit, and possibly further measures) (see below). Category ATEX II 3G (Ex zone 2) The principle of the type of protection "Pressurized enclosure for devices of Category 3" is based on preventing the ingress of any hazardous atmosphere into the gas analyzer. Two versions complying with Directive 94/9/EC are available for use in Zone 2. In both cases, the standard devices of Series 6 (field version, not Ex) can be used. • Explosion protection due to gas-fumes-proof enclosure The enclosure of the Series 6 gas analyzers (standard, field version) is sealed sufficiently to prevent gas fumes from penetrating. With this type of protection, only sample gases may be fed in which are below the LEL. Test certificate: TÜV 01 ATEX 1686 X Device identification: II 3 G Ex nR op pr IIC T6 or T4 Gc It is not necessary to install a purging unit here. • Pressurized enclosure with continuous purging Protective gas continuously flows through the enclosure with a volume flow of at least 1 l/min; furthermore, the flow results in an overpressure in the enclosure of at least 50 Pa compared to atmospheric pressure. The max. permissible purging gas pressure is 25 hPa. The max. permissible sample gas pressure is equivalent to the permissible analyzer sample gas pressure. Test certification TÜV 01 ATEX 1697 X Device identification: II 2/3 G EEx n P II T4 The purging gas is monitored using E Ex p monitoring equipment. This is a stand-alone unit which is connected electrically and pneumatically to the analyzer. Explosion protection is only provided when these two units (analyzer and purging unit) are combined. (see below, purging unit) The fundamental safety requirements of both versions are satisfied by compliance with the European standards EN 50014:1997, EN 50016:1995, EN 50020:1994 and EN 954:1996. The EEx p monitoring equipment is a stand-alone unit which is connected electrically and pneumatically to the analyzer. Explosion protection is only provided when these two units are combined. Siemens AP 01 · 2015 5/13 5 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive Category ATEX II 3D (Ex zone 22) Ex zone 22 concerns the so-called dust protection. This is the European successor to the previous German zone 11. Zone 22 concerns the area in which during normal operation it is not expected that potentially explosive atmospheres occur in the form of a cloud of flammable dust in the air. Should such a cloud occur, however, then only briefly. Considering the more stringent conditions for zone assignment, it can be expected that there will be increased demand for dustprotected analyzers. The field versions of CALOMAT 6, OXYMAT 6 and ULTRAMAT 6 can be used in this zone according to the conformity statement TÜV 03 ATEX 2278 X. They are assigned the Ex identification II 3 D IP65 T60 °C or T65 °C or T85 °C or T135 °C. However, this only concerns the so-called external explosion protection. With respect to the measurement of flammable gases, the additional measures applicable to gas explosion protection apply in addition, such as flame inhibitors. These separate certificates apply here. FM/CSA Class I Div 2 The field versions of the standard analysis units can be used. Explosion protection is only provided when combined with the suitable equipment. Definitions Non-flammable gas Gas or gas compositions with concentrations below the lower explosion limit (LEL). Non-explosive, even in contact with air. Example: CH4 < 4.4 %; H2 < 4 %; C2H2 < 2.3 % Flammable gas 5 Gas or gas composition with concentrations above the LEL. Explosive, but additionally requires air and ignition energy. Example: CH4 > 4.4 %; H2 > 4 %; C2H2 > 2.3 % Explosive gas Mixture of flammable gas and a gas matrix containing oxygen; between the LEL and the UEL (upper explosion limit). Already contains O2 and is explosive without additional air. Example: 4,4 % ... 16.5 % CH4 in air Note: Very little data is available on the existing LEL and UEL for oxygen concentrations other than ambient air (20.95 % O2) or for sample pressures other than atmospheric pressure. Although the IEC and EN directives IEC 60079-10, EN 60079-10 (gas) and IEC 61241-10, EN 50281 (dust) do not specifically define the terms seldom, occasional, frequent, and permanent, the following interpretation is customary: • Frequent or permanent: > 1 000 hours per year → a frequent explosive atmosphere corresponds to Zone 0 or Class I, Div. 1 • Occasional: 10 to 1 000 hours per year → an occasional explosive atmosphere corresponds to Zone 1 or Class I, Div. 1 • Seldom: < 10 hours per year → a seldom explosive atmosphere corresponds to Zone 2 or Class I, Div. 2 5/14 Siemens AP 01 · 2015 The following additional safety mechanisms are recommended for continuous gas analyzers for measuring explosive gases (internal explosion protection). These requirements are based on the European ATEX approvals for analyzers, but can also be used as directives in the USA since no other specific definition exists there. Purging requirements The continuous analyzers from Siemens with approvals for Class I, Div. 2 never require purging in a hazardous area in accordance with Class I, Div. 2 / Zone 2 under the aspect of area classification. All electronic and mechanical components are classified as “non-explosive” and can be used in environments in accordance with Class I, Div. 2 / Zone 2. However, purging may be necessary for a specific application, depending on the type of sample gas and the respective analyzer model in order to comply with the NEC and NFPA standards and to guarantee maximum possible safety as well as protection of the system. NFPA 496 requirements for continuous gas analyzers and systems from Siemens The NFPA 496 "Standard for Purged and Pressurized Enclosures for Electrical Equipment" describes in great detail and clarity the requirements for purging and for the pressurized enclosure for electric systems depending on 1) the external hazardous area classification, 2) the classification/grading of the system, 3) the type of gas in the gas path, and 4) the expected discharge of gas (none/limited/unlimited). It is assumed for the internal gas path of a continuous gas analyzer that it exhibits only low losses under normal conditions and uncontrolled losses in the case of a mechanical failure (abnormal conditions). When connecting gases with flammable components (> LEL) to the gas path of an analyzer with a hermetically sealed enclosure, the flammable component can become enriched in the inside of the analyzer enclosure – even under normal conditions – beyond a limit for continuous explosiveness and change the area classification (inside the analyzer enclosure) from "General Purpose" (Universal) or Class I, Div. 2 / Zone 2 to Class I, Div. 1 / Zone 0. This can also occur under abnormal conditions in any type of analyzer enclosure (including NEMA 1). Analyzers for installation in the field – O6F, U6F and C6F – have a gas-tight enclosure (IP65 / NEMA 4 equivalent in accordance with IEC/EN 60529 and NEMA Standards Publication 250). Only a small natural exchange of air takes place with the environment. In accordance with NFPA 496, a limited discharge of gas is to be expected under normal conditions, and an unlimited discharge under abnormal conditions. Analyzers for 19" rack mounting – O6E, U6E, U/O6, C6E, U23, O61, FID5 and FID6 – have an "open" enclosure (IP20 in accordance with IEC/EN 60529, no exact NEMA equivalent to IP20 available). A high natural exchange of air takes place with the environment unless the exchange is restricted. In accordance with NFPA 496, no discharge of gas is to be expected under normal conditions, but an unlimited discharge under abnormal conditions. In the case of analyzers designed for general applications, it is assumed that they can ignite an explosive gas mixture at any time, and therefore no type of explosive atmosphere whatsoever may be present in the vicinity of these analyzers or within the enclosure at any time. © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive In the case of analyzers designed for Class I, Div. 2 / Zone 2 it is assumed that they cannot ignite an explosive gas mixture under normal conditions (single fault safety), and these analyzers can therefore be used in an occasionally explosive atmosphere in the environment or within the enclosure in accordance with the definition of Class I, Div. 2 / Zone 2. However, a frequent or permanent explosive atmosphere must be avoided since a simultaneous fault occurring on the electrical components of the analyzer could constitute an ignition source. When purging a continuous gas analyzer or when purging/venting a continuous gas analyzer system suitable for Class I, Div. 2 / Zone 2 with instrument air or ambient air, and if failure of the safety vessel is not obvious, a leak detector (measurement in % of LEL) or similar equipment should be used in order to detect the unlimited discharge under abnormal conditions and to avoid a frequent or permanent explosive atmosphere inside the analyzer or in its environment. The leak detector must be fitted at a location where the escaping sample gas can be measured before becoming too greatly diluted. The alarm limit of the leak detector must be set to a level which enables detection of a dangerous state with consideration of the fact that the discharged sample gas has most probably already been diluted before it reaches the sensor. Further important information Gas paths material It is strongly recommended that you use gas paths made of metal for applications with flammable gases since such gas paths offer the greatest safety. This particularly applies to analyzers or systems which are purged with instrument air or ambient air since an explosive atmosphere can be produced under abnormal conditions. This immediate danger does not exist in the case of analyzers or systems purged with inert gas. It should be mentioned that, with an integrated system, all parts containing flammable gas (pumps, gas coolers, filters etc.) must be assessed in the same manner. Purging of left-hand analyzer side (electronics side) of continuous gas analyzers for field installation Since the left electronics side and the right measurement side of continuous gas analyzers are separated gas-tight from each other, it is unnecessary to purge the electronics side in most cases – purging of the (right-hand) measurement side is sufficient. However, if doubt exists that flammable gas could penetrate the left-hand electronics side and become enriched there, it is advisable to purge both sides. Further reasons for purging analyzers • Corrosive sample gases: Purging with air or inert gas is necessary to prevent the enrichment of corrosive gas inside the analyzer, whereby operators or servicing personnel could be injured or the analyzer unit could be damaged. The discharged purging gas should be released at a non-critical point (collective vent etc.) • Toxic gases: Purging with air or inert gas is necessary to prevent the enrichment of toxic gas inside the analyzer, whereby operators or servicing personnel could be injured. The discharged purging gas should be released at a non-critical point (collective vent etc.). Further information can be found in the OSHA directives for handling toxic materials. Purging rate / applied pressure Purging with air: The air throughput for purging an analyzer for field installation must be sufficient such that the concentration of flammable gases is less than 25 % of the LEL (see NFPA 496, Section 8.3). An air throughput of 1 l/min is recommended. Purging with inert gas: The inert gas throughput for purging an analyzer for field installation must be sufficient such that the oxygen level is less than 5 % of the volume or, at a maximum, less than 50 % of the oxygen required to form an explosive mixture (see NFPA 496, Section 8.3). An inert gas throughput of 1 l/min is recommended. Applied pressure with inert gas: The purging and holding pressure applied to an analyzer for field installation must be sufficient such that the oxygen level is less than 5 % of the volume or, at a maximum, less than 50 % of the oxygen required to form an explosive mixture (see NFPA 496, Section 8.3). A pressure of 25 Pa (0.1 inch water column) in accordance with NFPA 496 is recommended. It should be taken into consideration when applying a pressure to an analyzer, that instead of continuous purging, flammable gas can collect within the analyzer if the sample gas pressure is higher than the purging pressure. It is recommendable to appropriately adapt the purging pressure within the permissible pressure range of the field device enclosure. Exceptions: Inert gas should not be used as the purging gas for certain applications. This particularly applies to safety-related measurements of oxygen (LEL proof) where the sample gas has a slight overpressure and the inert gas used for purging could dilute the sample under abnormal conditions. Such applications require individual assessment of the purging equipment required and the mode of operation. Integrated systems and analyzer containers: Purging or the application of pressure to continuous gas analyzer systems must be designed such that the requirements of NFPA 496 are complied with. Leak detector When purging a continuous gas analyzer or when purging/venting a continuous gas analyzer system suitable for Class I, Div. 2/ Zone 2 with instrument air or ambient air, and if failure of the safety vessel is not obvious, a leak detector (measurement in % of LEL) or similar equipment should be used in order to detect unlimited discharges under abnormal conditions and to avoid a frequent or permanent explosive atmosphere inside the analyzer or in its environment. The leak detector must be fitted at a location where the escaping sample gas can be measured before becoming too greatly diluted. The alarm limit of the leak detector must be set to a level which enables detection of a dangerous state with consideration of the fact that the discharged sample gas has most probably already been diluted before it reaches the sensor. Siemens AP 01 · 2015 5/15 5 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive ■ Application Differentiation of cases: Ex zones/danger through flammable sample gas Gas type Sample gas non-flammable Sample gas is flammable below the lower explosive and/or is rarely, and then limit (LEL) only briefly, above the LEL Sample gas is flammable and/or is occasionally above the LEL Category ATEX II 1G (zone 0) Individual acceptance test (on request) Individual acceptance test (on request) Individual acceptance test (on request) Category ATEX II 2G Analyzer (zone 1) Operating mode "Leakage compensation" Ex analyzer EEx p (certificate ATEX 2022X) Ex analyzer EEx p (certificate ATEX 2022X) Ex analyzer EEx p (certificate ATEX 2022X) Gas path Pipe gas path Pipe gas path Pipe gas path Flame arrester — — Flame arrester in sample gas inlet and outlet Monitoring EEx p control device (certificate ATEX E 082) EEx p control device sample EEx p control device sample gas pressure < 165 hPa, fail- gas pressure < 165 hPa, failsafe (certificate ATEX E 082) safe (certificate ATEX E 082) Pressure switch — Differential pressure switch Differential pressure switch (when sample gas pressure is (when sample gas pressure is not controlled fail-safely) not controlled fail-safely) Analyzer Ex analyzer EEx p (certificate ATEX 1708X) Ex analyzer EEx p (certificate ATEX 1708X) Ex analyzer EEx p (certificate ATEX 1708X) Gas path Pipe gas path Pipe gas path Pipe gas path Flame arrester — — Flame arrester in sample gas inlet and outlet Monitoring EEx p control device (certificate DMT 99 ATEX E 082) EEx p control device (certificate DMT 99 ATEX E 082) EEx p control device (certificate DMT 99 ATEX E 082) Pressure switch — — — Analyzer Standard analyzer in field housing (addition E11: certificate ATEX 1686X) Standard analyzer in field housing (addition E12: certificate ATEX 1697X) Standard analyzer in field housing (addition E12: certificate ATEX 1697X) Gas path Pipe or hose gas path Pipe gas path Pipe gas path Flame arrester — — Flame arrester in sample gas inlet and outlet Monitoring — EEx p control device (certificate DMT 99 ATEX E 082) EEx p control device (certificate DMT 99 ATEX E 082) Analyzer Analyzer as rack unit or in field housing Analyzer as rack unit or in field housing Analyzer as rack unit or in field housing Gas path Pipe or hose gas path Pipe gas path, recommended Pipe gas path, recommended Enclosure purging with inert Enclosure purging with inert gas (N2) recommended gas (N2) recommended Flame arrester — — Flame arrester in sample gas inlet and outlet Monitoring — — Simplified monitoring of purging recommended Zone Category ATEX II 2G (zone 1) Operating mode "Continuous purging" 5 Category ATEX II 3G (zone 2) Non-hazardous zone Ex configurations – principle selection criteria Signal line routing Within zone 1 From zone 1 to zone 2 From zone 1 to non-hazardous zone Ex i isolation amplifier Required Conditional use (when energy feedback cannot be excluded) Conditional use (when energy feedback cannot be excluded) Isolating relay Required Conditional use (when energy feedback cannot be excluded) Conditional use (when energy feedback cannot be excluded) Additional units, selection criteria (ATEX 2G) 5/16 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive Use of OXYMAT 6 in hazardous area and/or for measurement of flammable gases Article No. Category Operating mode ATEX II 2G Leakage (zone 1) compensation Continuous purging 7MB2011-***0*-2*** 7MB2011-***0*-3*** 7MB2011-***0*-6*** 7MB2011-***0*-7*** ATEX II 3G Flammable gases 7MB2011-***0*-0*** (zone 2) 7MB2011-***0*-1*** Non-flammable gases Non-hazardous zone CLASS 1 Div 2 Non-hazardous gas zone Flammable and non-flammable gases Certification (short codes) Additional unit Gas Dust Purging unit Flame arrestor Pressure switch Ex i isolation amplifier Ex i isolating relay Ex zone Ex zone 7MB8000- 7MB8000- 7MB8000- 7MB8000- 7MB8000- X — X — 2BB 6BA/6BB1) 5AA1) 3AB1) 4AB1) 2BA 6BA/6BB1) 5AA1) 3AB1) 4AA1) o 3AB1) 4AB1) X — 2CB 6BA/6BB1) X — 2CA 6BA/6BB1) o 3AB1) 4AA1) E12 — 2CB 6BA/6BB1) o o o 2CA 6BA/6BB1) o o o 7MB2011-***0*-0*** E42 — 6BA/6BB1) o o o 7MB2011-***0*-1*** E42 — 6BA/6BB1) o o o E12 — 7MB2011-***0*-0*** E11 — o o o o o 7MB2011-***0*-1*** E11 — o o o o o 7MB2011-***0*-0*** E41 — o o o o 7MB2011-***0*-1*** E41 — o o o o 7MB2011-***0*-0*** E40 — o o o o 7MB2011-***0*-1*** E40 — o o o o 7MB2011-***0*-0*** — — o o o o o 7MB2011-***0*-1*** — — o o o o o 7MB2021-****0-**** X — o o o o o 7MB2011-***0*-0*** E20 — 1AA 6BA/6BB o o o 7MB2011-***0*-1*** E20 — 1AA 6BA/6BB o o o 7MB2021-*****-****2) E20 — 1AA 6BA/6BB o o o — Combination not allowed X Possible combination, no additional data required o Not required Ex configurations, possible combinations 1) Conditionally required: see table of Ex configurations, selection criteria 2) Installation in additional enclosure required Siemens AP 01 · 2015 5/17 5 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive Use of ULTRAMAT 6 in hazardous area and/or for measurement of flammable gases Article No. Category Operating mode 7MB21117MB2112- ATEX II 2G (zone 1) Leakage compensation *****-2*A* ATEX II 3G (zone 2) Dust Ex zone Ex zone 7MB8000- X X *****-7*A* X Flammable gases *****-0*A* Flammable and non-flammable gases 7MB8000- Pressure switch Ex i isolation Ex i isolating amplifier relay 7MB8000- 7MB8000- 7MB8000- 6BA/6BB1) 5AA1) 3AB1) 4AB1) 2BA 6BA/6BB1) 5AA1) 3AB1) 4AA1) — 2CB 6BA/6BB1) o 3AB1) 4AB1) — 2CA 6BA/6BB1) o 3AB1) 4AA1) — 6BA/6BB1) o o o — E42 — 6BA/6BB1) o o o *****-0*A* E12 — 2CB o o o o *****-1*A* E12 — 2CA o o o o *****-1*A* E42 *****-0*A* E41 — o o o o *****-1*A* E41 — o o o o o o o o o *****-0*A* E11 — *****-1*A* E11 — o o o o o *****-0*A* X E40 — o o o o *****-1*A* X E40 — o o o o 7MB2111*****-0*A*, E20 — 1AA 6BA/6BB o o o 7MB2111*****-1*A* E20 — 1AA 6BA/6BB o o o 7MB212******-****2) E20 — 1AA 6BA/6BB o o o — Combination not allowed X Possible combination, no additional data required o Not required 1) Conditionally required; see table of Ex configurations, selection criteria. 2) Installation in additional enclosure required 5/18 Purging unit Flame arrestor 2BB — X *****-6*A* Non-hazard- Non-hazardous ous zone gas zone 5 Gas Continuous purging Non-flammable gases CLASS 1 Div 2 *****-3*A* Certification and Additional unit short codes Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive Use of CALOMAT 6 in hazardous area and/or for measurement of flammable gases Article No. Category Operating mode ATEX II 2G (zone 1) Leakage compensation Continuous purging Flammable gases ATEX II 3G (Zone 2) Non-flammable gases Certification Additional unit Gas Purging unit Flame arrester Pressure switch Ex i isolation amplifier Ex i isolating relay 7MB8000- 7MB8000- 7MB8000- 7MB8000- Dust 7MB8000- 7MB2511***0*-0AE* X — 2BB 6BA/6BB1) 5AA1) 3AB1) 4AB1) 7MB2511***0*-1AE* X — 2BA 6BA/6BB1) 5AA1) 3AB1) 4AA1) 7MB2511***0*-0AF* X — 2CB 6BA/6BB1) o 3AB1) 4AB1) 7MB2511***0*-1AF* X — 2CA 6BA/6BB1) o 3AB1) 4AA1) 7MB2511***0*-*AJ* X X — 6BA/6BB1) o o o 7MB2511***0*-*AC* X — 2CA/2CB 6BA/6BB1) o o o 7MB2521***0*-*AB*2) X — acc. to Certificate 6BA/6BB1) o o o 7MB2511***0*-*AH* X X — o o o o 7MB2511***0*-*AB* X — o o o o o 7MB2521***0*-*AB* X — o o o o o Non-hazardous zone Non-hazardous gas zone 7MB2511***0*-*AG* — X — o o o o CLASS 1 Div 2 Flammable and non-flammable gases 7MB2511***0*-*AD* X — 1AA 6BA/6BB o o o 7MB2521***0*-*AD*2) X — 1AA 6BA/6BB o o o X Possible combination, no additional data required — Combination not allowed o Not required Ex configurations, possible combinations 1) Required under certain conditions; see table of Ex configurations, selection criteria 2) Installation in additional housing required Use of ULTRAMAT 23 in hazardous area and/or for measurement of flammable gases Article No. Category Operating mode Additional unit Gas Purging unit Flame arrestor Pressure switch Ex i isolation amplifier Ex i isolating relay 7MB8000- 7MB8000- 7MB8000- 7MB8000- Dust Ex zone Ex zone 7MB8000- *****-**** E20 — o o1) o o o Non-flammable gases2) *****-**** E20 — o o o o o Flammable and non-flammable gases2) *****-**** E20 — 1AA o1) o o o ATEX II 3G Flammable (zone 2) gases2) Class 1 Div 2 7MB233*- Certification and short codes — Combination not allowed o Not required/not defined 1) Required under certain conditions 2) Installation in additional housing required Siemens AP 01 · 2015 5/19 5 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive Ex approval ATEX CLASS I Div 2 2G - LC 2G - CP see basic MLFB ATEX 3G burn. 3G nbrn. FM CSA 3D (dust) Z + E12 Z + E11 Z + E20 Z + E20 Z + E4X except CALOMAT except CALOMAT Field device U6F ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1686 X 3016050 1526657 ATEX 2278 X U6F-S ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1686 X 3016050 1526657 — O6F ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1686 X 3016050 1526657 ATEX 2278 X O6F-S ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1686 X 3016050 1526657 — C6F ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1697 X 3018862 1526660 ATEX 2278 X C6F-S ATEX 2022 X ATEX 1708 X ATEX 1697 X ATEX 1697 X o 1526657 — C62F ATEX 2022 X ATEX 1708 X — — — — — C62F-S ATEX 2022 X ATEX 1708 X — — — — — U6E — — — 3016050 1526657 — U6E-S — — — 3016050 1526657 — O6E — — — 3016050 1526657 — O6E-S — — — 3016050 1526657 — OU6E — — — 3016050 1526657 — OU6E-S — — — 3016050 1526657 — C6E — — ATEX 1873 X 3018862 1526660 — C6E-S — — ATEX 1873 X o 1526660 — C62E — — — — — — C62E-S — — — — — — O61 — — — — — — O64 — — — — — — F6 — — — — U23 — — ATEX 0027 x 3035269 2133209 — U23 O2p — — ATEX 0027 x 3035269 2133209 — U23 H2S — — ATEX 0027 x 3035269 2133209 — 19" rack unit 5 — (SET) CP = Continuous Purging LC = Leakage compensation … -S = Special application burn. = Flammable gases nbrn. = Non-flammable gases o = In progress Possible combinations of PROFIBUS with Ex applications ATEX 2022 X PROFIBUS PA Ex-i ATEX 1708 X PROFIBUS PA Ex-i ATEX 1697 X FM/CSA PROFIBUS PA or DP PROFIBUS PA Ex-i ATEX 2278 X PROFIBUS DP - Non-hazardous gas installation (Z + E40) ATEX 1686 X PROFIBUS DP ATEX 2278 X PROFIBUS DP in combination ATEX 1686 X (Z + E41) ATEX 1873 X PROFIBUS DP ATEX 2278 X PROFIBUS PA Ex-i in combination with ATEX 1697 X (Z + E42) 5/20 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - ATEX II 2G control unit ■ Overview EEx p safety equipment (purging unit) Additional function The EEx p safety equipment to be connected to the analyzer must have at least the following features: • Adjustable pre-purging phase; purging gas flow must be approximately 50 l/min • Limitation of purging gas pressure during the pre-purging phase: < 165 hPa • "Leakage compensation" or "Continuous purging" • Connection for purging gas lines with Ø 10 mm or Ø 3/8" from and to the analyzer • Pressure after pressure reducer - 0.2 to 0.4 MPa (leakage compensation) - 0.2 to 0.3 MPa (continuous purging) • Max. permissible input pressure 0.6 MPa • Relay contacts for all-pole isolation of the analyzer supply voltage • Connection option for a key-operated switch and a pressure switch (intrinsically-safe circuits) • Device version "Leakage compensation": Connection option for a pressure switch with intrinsically-safe scan By connecting additional pressure sensors, the internal pressure of the enclosure is maintained at a pressure higher than that of the sample gas by means of a proportional valve. During the prepurging phase the purging gas flow is max. 4 100 Nl/h with an internal enclosure pressure of 50 hPa. The Bartec control unit APEX 2003.SI/B meets the requirements for "Pressured enclosure with leakage compensation or continuous purging" in accordance with EN 50016 or ATEX guidelines, and can be used as explosion-proof equipment in Zones 1 and 2. The purging unit ensures that in a closed enclosure, any explosive gases will be purged and then a pressure higher than the surrounding atmosphere will be generated and maintained. 4 programmable relay inputs (8 relay contacts) are available to separate the data lines. During the operating phase, the pressure inside the enclosure must be maintained at a level at least 50 Pa higher than that of the surrounding atmosphere. If the internal pressure drops below the defined minimum value, safety equipment must shut down the entire electical supply to the EEx-p enclosure autonomously (including the possible data lines) . Enclosures frequently contain accessories to which flammable gases or sometimes also flammable gas mixtures are fed via a separate gas path. This is the case with gas analyzers, for example. In this case, it must be ensured that the pressure of the protective gas is always more than 50 Pa higher than the pressure of the sample gas. After mounting the control device APEX 2003.SI/B on the EEx-p enclosure, and after connecting the mains power and the protective gas, the control module regulates and monitors the flow of purging gas automatically during the pre-purging phase, and the internal enclosure pressure during the operating phase. If the minimum operating pressure of the enclosure is exceeded and if flow through the pressure monitoring module is sufficient, the pressure sensors forward the sensor module signals to the control module. A non-hazardous area is thus created in the enclosure in which electrical devices can be installed that are not themselves explosion-proof. After commissioning, a distinction is made between the pre-purging phase and the operating phase: 5 The pre-purging phase is necessary to ensure that any explosive atmposphere entering during the standstill time does not become a hazard; the enclosure must therefore be purged with protective gas (air from a non-hazardous area or inert gas) before commissioning. Siemens AP 01 · 2015 5/21 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - ATEX II 2G control unit, leakage compensation ■ Overview Pressure sensors MIN A = 0 … 300 hPa, MIN B = 0 … 300 hPa, MAX = 0 … 300 hPa, MAX 1 = 0 … 300 hPa, DIFF A = 0 … 25 hPa, DIFF B = 0 … 25 hPa Prepurging time 0 … 99 min; 5 sec dropout delayed Weight 11 kg BARTEC EEx p control unit "Leakage compensation" The APEX 2003.SI/A2 control unit controls and monitors the prepurging and operating phases of gas analyzers with containment systems in hazardous zone 1. The control unit redundantly monitors the set overpressure of the purging gas. When the overpressure decreases, it is corrected to the adjustable setpoint (max. purging gas pressure 165 hPa). ■ Technical specifications Control unit ATEX IIG, compensation of losses through leaks Guidelines EC EMC directive 89/336/EEC, EC low voltage, RL 73/23/EEC, Ex directive 94/9/EC Design Explosion-protected enclosure (EEx e) with viewing window in the cover Enclosure material Glass fiber-reinforced polyester Degree of protection IP65 Terminals Electrical data Supply voltage 230 V AC (115 V AC) ± 10 %, 48 … 62 Hz Power consumption 21 W /230 V NO contacts K2/3; max. 250 V, 5 A with cos ϕ = 1, K4/K5; supply voltage or floating, max. 250 V, 5 A with cos ϕ = 1 Communication RS 485 interface Temperature switching value (option) 0 … +40 °C Explosion protection 2.5 mm, finely stranded Marking EEx e d ib [ia p] IIC T4/T6 Test certification DMT 99 ATEX E 082 Ambient temperature -20 … +40 °C ■ Dimensional drawings 5 BARTEC control unit, dimensions in mm 5/22 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - ATEX II 2G control unit, leakage compensation ■ Schematics $QDO\]HU (([SFRQWUROXQLW ,QHUWJDVRXWOHW 3UHVVXUHVHQVRU ',))& 6DPSOH FHOO 2XWOHW 6DPSOHJDV ,QOHW ,QHUWJDVLQOHW 3UHVVXUHVZLWFK UHGXQGDQW YHUVLRQ 'T 03D BARTEC control unit, leakage compensation, gas connection diagram ([WFXUUHQW VRXUFLQJ 7HPSHUDWXUH VZLWFK VHQVRU 6HQVRU PRGXOH .H\ VZLWFK 212)) VZLWFK ([WFXUUHQW VRXUFLQJ %ULGJHIRU VZLWFK FDOLEUDWLRQ 5 EEx i 1-12 5 8 13 14 15 16 17 18 &RQWUROPRGXOH L1 N PE 20 21 22 23 24 K2/K3 K4/K5 K1 25 26 27 19 28 29 2XWSXW YDOYH 31 32 + - 39 41 43 45 ,QSXW YDOYH 'DWD FDEOH 46 44 42 40 35 34 + 'DWDFDEOH 47 48 - 3RZHUVXSSO\ (([S DQDO\]HU BARTEC control unit, leakage compensation, electric connection diagram Siemens AP 01 · 2015 5/23 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - ATEX II 2G/3G control unit, continuous purging ■ Overview Terminals 2.5 mm, finely stranded Pressure sensors MIN A = 0 … 25 hPa, MIN B = 0 … 25 hPa, MAX = 0 … 25 hPa, MAX 1 = 0 … 25 hPa, DIFF A = 0 … 25 hPa, DIFF B = 0 … 25 hPa The control unit redundantly monitors the continuous flow of protective gas through the connected analyzer and thereby dilutes any escaping sample gas to below the lower explosive limit (max. purging gas pressure 25 hPa). Prepurging time 0 … 99 min; 5 sec dropout delayed Weight 10 kg At the same time, a higher pressure is maintained inside the EEx-p enclosure than in the surrounding atmosphere. If the flow of purging gas or the internal pressure falls below a determined minimum value, the supply voltage to the equipment in the pressurized enclosure is shut down. Supply voltage 230 V AC (115 V AC) ± 10 %, 48 … 62 Hz Power consumption 21 W /230 V NO contacts K2/3; max. 250 V, 5 A with cos ϕ = 1, K4/K5; supply voltage or floating, max. 250 V, 5 A with cos ϕ = 1 Communication RS 485 interface Temperature switching value (option) 0 … +40 °C BARTEC EEx p control unit "Continuous purging" The APEX 2003.SI/A4 control unit controls and monitors the prepurging and operating phases of gas analyzers with containment systems in Ex zone 1 and Ex zone 2. 4 programmable relay inputs (8 relay contacts) are available to separate the data lines. ■ Technical specifications Control unit ATEX II 2G, continuous purging Guidelines Electrical data Explosion protection EC EMC directive 89/336/EEC, EC low voltage, RL 73/23/EEC, Ex directive 94/9/EC Design Explosion-protected enclosure (EEx e) with viewing window in the cover Degree of protection IP65 Marking EEx e d ib [ia p] IIC T4/T6 Test certification DMT 99 ATEX E 082 Ambient temperature -20 … +40 °C ■ Dimensional drawings 5 BARTEC control unit, dimensions in mm 5/24 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - ATEX II 2G/3G control unit, continuous purging ■ Schematics $QDO\]HU (([SFRQWUROXQLW 2SHUDWLQJIORZQR]]OH 3XUJLQJJDV RXWOHW 2XWOHW 6DPSOH FHOO 6DPSOHJDV ,QOHW 3UHSXUJLQJ21 Dq 3XUJLQJJDV LQOHW PD[03D 6HWWLQJRSHUDWLQJIORZ BARTEC control unit, continuous purging, gas connection diagram Sensor module Temperature switch Key switch Jumper for adjustment 5 T EEx i 1-12 5 8 15 16 17 18 Control module 20 23 24 K2/K3 K4/K5 K1 25 26 27 19 28 29 Purging valve 39 41 43 45 L1 N PE Data line 46 44 42 40 35 34 + Data line 47 48 - Supply voltage EEx p analyzer BARTEC control unit, continuous purging, electric connection diagram Siemens AP 01 · 2015 5/25 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - Purging unit FM (Class I Div 2) ■ Application The Ex purging unit MiniPurge FM is used to monitor the pressure during continuous purging of an analyzer with purging or inert gas. If the pressure falls below the set value, an optical display is triggered and the relay is activated. This monitoring unit is driven by the purging gas pressure and therefore does not require an additional power supply. Switching contact Via SPCO switch approved for Class 1 Division 2 Settings Lower response limit 0.5 hPa set relative to purging gas flow of 1 … 2 l/min Prepurging time Is defined by operator, and controlled manually Enclosure pressure limitation By means of stainless steel with integrated flame arrestor; opens at 10 hPa ± 10 % ■ Technical specifications 5 Classification Class I Division 2 Enclosure dimensions (in mm) 444 x 438 x 275 Enclosure volume (I) Approx. 50 l Enclosure pressure (normal) 1 hPa FM certificate Certificate of compliance 1X8A4.AE / 0B3A3.AE Reaction upon failure of pressure Opening of switching contact, and alarm via signal indicator (red display) System type MiniPurge complete system Operating mode Continuous purging Type of enclosure Reinforced polycarbonate Enclosure surface RAL 7035 gray with transparent cover Pressure supply Dry, oil-free air or inert gas with regulated pressure of approx. 2000 hPa (30 psi) at inlet of MiniPurge Supply connections Pressure via ¼ BSPP connection, pressure hose at least ½" or 12 mm Display (signal indicator) ■ Dimensional drawings Pneumatically driven color signal: green/red MiniPurge, dimensions in mm ■ Schematics Analyzer Sample cell MiniPurge, purging unit, Class I, Div 2, gas connection diagram 5/26 Siemens AP 01 · 2015 Purging gas inlet Cable connection Sample gas Purging gas © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - Additional units ■ Overview Installation of Ex isolation modules / Ex i isolation amplifiers The mounting rail in the analyzer has a length of approximately 250 mm, with the number of installable components being limited. The maximum installation height is approximately 95 mm; however, it is less in the area of the display (88 mm). The width must not exceed 80 mm. The add-on devices must be approved for an ambient temperature of up to 60 °C; this temperature can be reached under extreme marginal conditions. The label of the analyzer shows all types of explosion protection that the device itself and the components can have. Installation must always be discussed with the competent experts. Slots in the analyzer and the purging unit. Ex i isolation amplifier Ex isolating relay 8S Comment Analyzers 2 2 Max. 2 Bartec purging unit 0 1 Max. 1 Ex i isolation amplifier, 7MB8000-3AB The analog inputs and outputs of the analyzers are not intrinsically safe in the basic version. The analog output can be supplemented later with an intrinsically-safe analog output (explosion protection type EEx ib II C or EEx ia II C). For this purpose, a suitable commercially available isolating transformer can be mounted on a rail in the device. Technical data: • Intrinsically-safe analog output • mA isolating transformer without power supply • For installing in the analyzer Isolation amplifier, rail mounting • Intrinsically-safe output EEx ia IIC • Galvanic isolation Technical data Isolating relay (signal outputs with external voltage supply) 7MB8000-4AA/-4AB If the device has to be opened, it must be isolated at all poles from the mains cable, the binary inputs, relay outputs, analog inputs/ outputs, RS 485 interface cable, and the PROFIBUS PA cables (not Ex i). For this purpose, isolating relays must be inserted. Intrinsically-safe circuits are excepted from this. An isolating relay must be explosion-proof if it is to be set up in an area subject to explosion hazard. Protective gas • The fed-in gases are not flammable. Air from an area not subject to explosion hazard can be used as the protective gas (purging gas). • Flammable gases or gas mixtures that are rarely or only briefly ignitable are fed in. The enclosure must be flooded with inert gas. • Gas mixtures that are occasionally ignitable are introduced. As with b), the enclosure must be flooded with inert gas; in addition, the sample gas inlet and outlet must be equipped with flame arrestors. • Explosive gas mixtures that are present in the long term or permanently must not be connected! Flame arrestors If the gas mixture to be measured sometimes has an explosive composition, flame arrestors must be installed in the sample gas inlet and, in certain circumstances, also in the sample gas outlet, in addition to the application already described with flammable sample gases. The material of the flame arrestors must be resistant to the flowtype sample gas mixture. For this reason, they are available in two different versions: • The detonation protection (Ex designation Ex IIG IIC) is used to prevent flashover in the case of unstable detonations and deflagrations of explosive gas or vapor/air mixtures of explosion group IIC. • The flame arrestor consists essentially of a detonation-proof enclosure with gas connections and a ceramic sinter cartridge built into the housing (max. pore width: 80 µm) to prevent flashover. It may be heated up to 150 °C and subjected to a pressure up to 3 bar (abs.). Input voltage 8.4 V + 0.02 x load x (V/Ω) Linearity < 0.1 V Temperature impact < 0.1 %/10 K Power supply 8.4 ... -30 V DC from the current loop Length 83.5 mm Diameter 32 mm Weight 160 g External thread M 30 x 1.5; 30 mm long Ambient temperature -20 °C … +60 °C Gas connections G 1/4" Relative humidity < 95 %, no condensation Material Stainless steel or Hastelloy C Max. gas operating pressure 3 bar (abs.) II (1) G D [Eex ia] IIC Max. operating temperature 150 °C (200 °C on request) EC type-examination certificate TÜV 98 ATEX 1338 Explosion group IIC Safety limits U0 ≤ 12.6 V I0 ≤ 95 mA Explosion protection Type of protection Technical data Siemens AP 01 · 2015 5/27 5 © Siemens AG 2015 General information Ex versions Continuous gas analysis, extractive - Additional units Differential pressure switch: 7MB8000-5AA There must be a fail-safe guarantee that the sample gas pressure will never exceed 5 hPa under the purging gas pressure. If this cannot be guaranteed on the plant side, a differential pressure switch must be mounted between the sample gas line and the purging gas line and connected electrically with the purging unit. The differential pressure switch always has contact with the sample gas. Technical data • Differential pressure switch with magnetic spring contact • Type 821.1 • Materials coming into contact with the sample gas: Stainless steel, mat. no. 1.4571 • Measuring range: -20 ... +20 hPa • Trigger point: adjustable 5 5/28 Siemens AP 01 · 2015 © Siemens AG 2015 General information Ex versions Continuous gas analysis, in-situ - LDS 6 ■ Overview "Intrinsic safety" is limited to circuits with relatively small capacity. To prevent sparks during closing or opening of an electrical circuit, the capacitance and inductance of an intrinsically-safe circuit are also limited depending on the maximum current and voltage values. No sparks or thermal effects which could lead to ignition of an explosive atmosphere occur either in normal operation or in the process upset. Therefore intrinsically-safe circuits may also be connected or disconnected during operation when live, since the safety is also guaranteed in the event of a shortcircuit or interruption. The following figure shows the block diagram for the type of protection "Intrinsic safety". Sensors and cables for applications of the LDS 6 in hazardous areas Intrinsic safety and intrinsically-safe circuit Principles The physical principle for the degree of protection "Intrinsic safety" is that a certain minimum ignition energy is required to ignite an explosive atmosphere. In an intrinsically-safe circuit, this minimum ignition energy is not present in the hazardous area, neither during normal operation nor in the event of an incident. The intrinsic safety of a circuit is achieved by limiting the current, voltage, power and temperature. Therefore the type of protection Safe area Hazardous area Ra La Li Limited temperature rise Ca F Ci U0 Limited spark energy Ri PA U0 Ri Li Ci F Max. output voltage Internal resistor Internal inductivity Internal capacitance Fuse PA Ra La Ca Potential compensation External resistor External inductivity External capacitance 5 Block diagram for voltage/current limiting with type of protection "Intrinsic safety" Intrinsically-safe electrical equipment and intrinsically-safe components of associated equipment are divided into two categories ("Protection levels"). A differentiation is made between the protection levels "ia" and "ib". Protection level "ib" also provides protection should one protective measure fail (fault redundancy 1). Protection level "ia" provides protection even if two protective measures should fail (fault redundancy 2). The standard refers to so-called "countable faults" instead of protective measures. These refer to protective measures, such as current limiting resistors, Zener diodes for voltage limiting, fuses, safe distances etc., i.e. all components or measures which implement an exactly defined safety function for the associated equipment. Protection level Description according to EN 50020 Installation ia The intrinsically-safe electrical equipment must not cause an ignition: Up to zone 0 • During normal operation or with the existence of non-countable safety-related faults which result in the most unfavorable condition. • During normal operation or with the existence of countable faults plus non-countable faults which result in the most unfavorable condition. • During normal operation or with the existence of two countable faults plus non-countable faults which result in the most unfavorable condition. ib The intrinsically-safe electrical equipment must not cause an ignition: Zone 2 • During normal operation or with the existence of non-countable faults which result in the most un- Zone 1 favorable condition. • During normal operation or with the existence of countable faults plus non-countable faults which result in the most unfavorable condition. Protection levels of electrical equipment and intrinsically-safe components Siemens AP 01 · 2015 5/29 © Siemens AG 2015 General information Ex versions Continuous gas analysis, in-situ - LDS 6 Minimum ignition curves Typical setup of an LDS 6 system in a hazardous area The "minimum ignition curves" are used to evaluate an intrinsically- safe circuit and to determine the maximum capacitance and inductance values. They are included in the valid intrinsically-safe standards (EN 50020 or DIN EN 50020 and IEC 60079-11 or EN 60079-11). Minimum ignition curves exist for resistive, capacitive and inductive circuits. Different minimum ignition curves are applied depending on the gas group for which an intrinsically-safe circuit is to be designed, and take into account the minimum ignition energies of the gas groups. LDS 6 is capable of measuring gases in EEx environments, provided all safety-relevant points are particularly observed. The central unit of LDS 6 must always be located outside of hazardous areas. Associated electrical equipment allow operation inside almost any EEx-classified area. Associated electrical equipment is a reference to equipment which contains one or more intrinsically-safe circuits, but in which not all circuits are intrinsically-safe. Associated electrical equipment usually has an isolating function, i.e. separating intrinsically-safe equipment from non-intrinsically-safe equipment within a signal circuit. Such devices include, for example: safety barriers, switch amplifiers, power supply units etc. For the intrinsically-safe version, an EEx barrier must be provided between the sensors and central unit. A typical version is shown in the following figure for intrinsically safe EEx ia sensors. 5 DIN/EN 60079-14 (VDE 165, Part 1) must be observed when selecting and routing the cables. Particular attention must be paid to the characteristic values, such as electric strength and minimum cross-section. In the case of intrinsically-safe circuits, the cable capacitance and inductance must be observed in addition, and must not exceed the values specified for the intrinsically-safe or associated equipment used (Co, Lo). The connection points and cables of intrinsically-safe circuits must be identified, e.g. in light blue, and be separated from the other connection points and cables of non-intrinsically-safe circuits. Transmitter Hybrid cable Cables CD6 Ex Receiver Sensor connecting cable Ex zone Ex barrier Hybrid cable Associated electrical equipment is not explosion-proof and must therefore not be installed in hazardous areas. It only contains intrinsically-safe circuits which may be routed into the hazardous area. Associated electrical equipment is identified by a square bracket enclosing "EEx" and the symbol for the type of protection, as well as absence of the temperature class (e.g. [EEx ia] IIC). Special EEx-type sensors (see explosion protection tag), certified according to • ATEX II 1G Ex ia IIC T4 and • ATEX II 1 D Ex iaD 20 IP65 T135 °C Typical setup of LDS 6 in a hazardous area 5/30 Siemens AP 01 · 2015 LDS 6 © Siemens AG 2015 General information Ex versions Continuous gas analysis, in-situ - LDS 6, EEx barrier ■ Overview ■ Technical specifications The EEx barrier is included in the scope of delivery of the CD 6 sensors in EEx ia version. It is meant for wall mounting close to the location of the LDS 6 central unit within an EEx-safe environment. The EEx barrier defines the interface between the analyzer central unit and the intrinsically-safe sensor heads and ensures under all circumstances that the total electrical energy transferred via the hybrid cable to the sensors is always less that than needed to ignite combustible gas mixtures. EEx barrier Hazardous area output • Minimum output voltage 12.5 V at 45 mA • Maximum output voltage 24 V at 170 Ω • Current limitation 45 mA Max. power consumption (45 mA output) 90 mA at 24 V, 110 mA at 20 ... 35 V DC Safety description 25 V, 170 Ω, 147 mA, Um = 250 Vrms or DC ■ Dimensional drawings 250 350 146 5 EEx barrier, dimensions in mm Siemens AP 01 · 2015 5/31 © Siemens AG 2015 General information Ex versions Continuous gas analysis, in-situ - SITRANS SL ■ Overview The SITRANS SL consists of a flameproof transmitter, a flameproof receiver, and optionally a specially certified junction box of increased safety. The complete analytical system is accommodated in the two flameproof enclosures which are connected together by a cable. A further cable is connected to the receiver, and serves as the power supply and customer interface. Both cables have a fixed connection to the flameproof enclosure. They must be connected in a suitable junction box if applicable. The receiver also has a local display. SITRANS SL can be operated by Ex-certified infrared remote control. The laser has a radiated power of 0.8 mW. The irradiance is approx. 10.9 µW/mm2. This is below the values permitted in EN 60079-28. An Ex concept with type of explosion protection "Pressurized enclosure d" is used for the SITRANS SL. The enclosure used resists an explosion caused by an explosive gas mixture in the analyzer. Ignition of an explosive atmosphere produced outside the enclosure is therefore reliably prevented. The SITRANS SL is available with ATEX or FM certificates. Connection cable Analog-I/O, Modbus Transmitter unit Connection cable PROFIBUS DP Receiver unit Connection cable ATEX cable gland 5 Remote control ATEX cable gland Mounting bracket for junction box Cable gland for Ethernet cable aP Ex-e junction box /PD[ P Special conditions Repairing of the flameproof gaps must only be carried out in accordance with the manufacturer’s design directives. 5/32 Siemens AP 01 · 2015 aP Ex-e junction box Connection conditions • Unused openings must be closed in accordance with EN 60079-1 Section 11.9. • A fixed cable must be used for the SITRANS SL gas analyzer, and routed such that it is sufficiently protected against damage. • If the temperature on the entry components is higher than 70 °C, appropriate temperature-resistant cables must be used. • The SITRANS SL gas analyzer must be included in the local equipotential bonding. • The end of the SITRANS SL gas analyzer cable must be connected in an enclosure which complies with the requirements of a recognized type of explosion protection in accordance with EN 60079-0, Section 1, if the connection is made in the hazardous area. © Siemens AG 2015 General information Ex versions Process gas chromatography ■ Overview Safety is extremely important during the storage, manufacture, processing and transportation of flammable materials, especially in the chemical and petrochemical industries, and in oil and gas production. Gas chromatographs and the sample preparation carried out in these plants must be designed such that no explosive mixtures can be ignited when applied. National directives and guidelines as well as international standards regulate the equipment prerequisites. Basic design with MAXUM edition II The electronic components are accommodated in a pressurized area. If the overpressure falls below a certain value, a control device switches off the power supply when a defined threshold is reached. The MAXUM edition II is available with certificates according to CSA/US, or ATEX certificates according to Cenelec for the EU market. The MAXUM edition II and MicroSAM chromatographs can be used in hazardous areas according to ATEX II 2G (zone 1) and ATEX II 3G (zone 2). The following individual protective measures apply: Type of protection: pressurized enclosure "p" The ignition source is enclosed by a protective gas with overpressure (at least 0.5 hPa). Air is used in most cases. The surrounding explosive atmosphere cannot penetrate. No flammable sample or flammable carrier gas may be passed into this pressurized area. Switching valves for the carrier gas H2 must therefore be mounted outside this area. The strength of the enclosure is at least 1.5 times the resistance to operating pressure. An alarm is generated in the event of failure of the purging gas or the overpressure. The electronics area must be purged prior to starting up the equipment. This purging also provides additional protection in corrosive environments. Type of protection: flameproof enclosure "d" This type of protection is used for most of our detectors. The detector is fitted in an enclosure which is resistant to the explosion of an explosive atmosphere within it. This means that the mechanical stability of the enclosure must withstand this internal explosion pressure. MAXUM edition II Basic design with MicroSAM The MicroSAM is designed such that all components (electronic and analytical) are accommodated in a flameproof enclosure. The advantage of this version is that no additional purging gases or safety monitoring systems are required. The MicroSAM is available with ATEX or FM/CSA certificates for the US market. Joints must also be so tight that hot gas escaping between two parts of the enclosure is not explosive. Ignition of an explosive atmosphere produced outside the enclosure is therefore reliably prevented. This is known as resistance to transmission of internal ignition. The FID, TCD and FPD detectors are available with this degree of protection. The maximum demands with regard to the joint parameters (width/length) are placed on enclosures of explosion group II C. MicroSAM is an example of this. MicroSAM Siemens AP 01 · 2015 5/33 5 © Siemens AG 2015 General information Tables Conversion tables ■ Overview $U 0 2 $LU &O 62 1 &O 2 + & 6 + 2 & + 3+ + & JP 0 0 0 0 0 0 0 JP 0 1H 1 +& 0 0 + 2 + & + & &2 1+ 0 &+ 0 + 5 0 00ROHFXODUZHLJKW ([DPSOHIRU62JPFRUUHVSRQGVWR9RO Conversion from g/m3 to vol. % (at 293 K and 1013 hPa) Conversion tables Component Molecular mass 1 ppm in mg/m3 1 mg/m3 in ppm hPa psia CO 28 1.250 0.800 420 6.091 NO 30 1.339 0.747 500 7.251 SO2 64 2.857 0.350 600 8.202 CO2 44 1.964 0.509 800 11.603 CH4 16 0.714 1.400 1 000 14.503 C2H4 28 1.250 0.800 1 160 16.824 C2H6 30 1.339 0.747 1 200 17.404 C4H10 58 2.589 0.386 1 300 18.854 C3H8 44 1.964 0.509 1 485 21.538 C3H6 42 1.875 0.533 1 500 21.755 2 000 29.007 psia 3 000 43.511 1013.25 14.69595 3 500 50.763 1000 14.50377 4 000 58.015 0.0145038 Conversion hPa ↔ psia Conversion ppm ↔ mg/m3 (1 atm; 0 °C), examples atm atm bar 1.01325 bar 0.9869 hPa 0.0009869 0.001 psia 0.0680 0.06894 Conversion of pressure units 5/34 Siemens AP 01 · 2015 hPa 68.94 9RO © Siemens AG 2015 General information Tables Dew point/saturation table Dew point °C °F -100 -90 Dew point Water content -148.0 -130.0 ppm (vol.) 0.014 0.008 Water content g/m3 1) °C °F ppm (vol.) g/m3 1) 0.0000103 0 +32.0 6 020 4.84 0.000119 +1 +33.8 6 480 5.2 +36.8 6 850 5.6 -80 -112.0 0.54 0.000565 +2 -70 -94.0 2.57 0.00269 +3 +37.4 7 487 6.0 -60 -78.0 10.7 0.011 +4 +39.2 8 022 6.4 -55 -67.0 20.8 0.021 +5 +41 8 595 6.8 -50 -58.0 38.4 0.038 +6 +42.8 9 216 7.3 -48 -54.4 49.6 0.049 +7 +44.6 9 875 7.8 -46 -50.8 63.0 0.061 +8 +46.4 10 584 8.3 0.067 +9 +48.2 11 318 8.8 0.076 +10 +50 12 114 9.4 0.097 +11 +51.8 12 935 10.0 0.11 +12 +53.6 13 806 10.7 +55.4 14 800 11.4 -45 -44 -42 -40 -49.0 -47.2 -43.6 -40.0 68.5 80.1 101.5 126.9 -39 -38.2 137.0 0.12 +13 -38 -36.4 158.0 0.14 +14 +57.2 15 796 12.1 -37 -34.6 174.1 0.16 +15 +59 16 791 12.8 -36 -32.8 197.8 0.17 +16 +60.8 17 885 13.6 -35 -31.0 224.0 0.19 +17 +62.6 19 030 14.5 -34 -29.2 245.0 0.22 +18 +64.4 20 396 15.4 -33 -27.4 274.0 0.24 +19 +66.2 21 641 16.3 -32 -25.6 303.4 0.26 +20 +68 23 020 17.3 0.30 +21 +69.8 24 502 18.3 0.33 +22 +71.6 26 120 19.4 0.37 +23 +73.4 27 736 20.6 +75.2 29 477 21.8 -31 -30 -29 -23.8 -22.0 -20.2 336.0 374 411 -28 -18.4 461 0.40 +24 -27 -16.8 511 0.45 +25 +77 31 219 23.0 -26 -14.3 563 0.49 +26 +78.8 33 209 24.4 -25 -13.0 623 0.55 +27 +80.6 35 200 25.8 -24 -11.2 689 0.59 +28 +82.4 37 312 27.2 -23 -9.4 759 0.66 +29 +84.2 39 551 28.7 -22 -7.3 840 0.72 +30 +86 41 791 30.3 -21 -5.8 922 0.80 +31 +87.8 44 322 32.0 0.88 +32 +89.6 46 936 33.5 0.96 +33 +91.4 49 675 35.6 1.05 +34 +93.2 52 539 37.2 1.15 +35 +95 55 472 39.6 +96.8 58 639 41.3 -20 -19 -18 -17 -4.0 -2.2 -0.4 +1.4 1 015 1 118 1 231 1 358 -16 +3.2 1 480 1.26 +36 -15 +5.0 1 630 1.38 +37 +98.6 62 001 43.8 -14 +6.8 1 779 1.51 +38 +100.4 65 487 45.8 -13 +8.8 1 953 1.65 +39 +102.2 68 973 48.4 -12 +10.4 2 140 1.79 +40 +104 71 761 50.7 -11 +12.2 2 338 1.96 +42 +107.6 81 049 56.5 -10 +14.0 2 562 2.14 +44 +111.2 89 889 62.3 -9 +15.8 2 798 2.33 +45 +113 94 527 65.3 2.54 +46 +114.8 99 600 68.7 2.76 +48 +118.4 110 681 75.5 2.99 +50 +122 120 398 82.3 +131 155 472 104.0 -8 -7 -6 +17.6 +19.4 +21.2 3 047 3 333 3 632 -5 +23.0 3 955 3.20 +55 -4 +24.8 4 303 3.51 +60 +140 196 517 129.5 -3 +26.6 4 690 3.81 +70 +158 307 212 196.5 -2 +28.4 5 100 4.13 +80 +176 467 662 290.5 -1 +30.2 5 547 4.47 +90 +194 691 542 418.0 +100 +212 1 000 980 558.0 1) 5 Reference temperature = dew point temperature. Siemens AP 01 · 2015 5/35 © Siemens AG 2015 General information Tables Dew point/saturation table Guide values for dead time (sec) per meter of sample gas line d 4 mm 6 mm 8 mm 10 mm 12 mm 14 mm 16 mm 18 mm 20 mm Q 30 l/h 1.5 3.4 6 9.4 13.5 18.4 24 30.5 37.6 60 l/h 0.8 1.7 3 4.7 6.8 9.2 12 15.3 18.8 90 l/h 0.5 1.1 2 3.1 4.5 6.1 8 10.2 12.5 120 l/h 0.4 0.9 1.5 2.4 3.4 4.6 6 7.6 9.4 150 l/h 0.3 0.7 1.2 1.9 2.7 3.7 4.8 6.1 7.5 180 l/h 0.3 0.6 1 1.6 2.3 3.1 4 5.1 6.3 210 l/h 0.2 0.5 0.9 1.3 1.9 2.6 3.4 4.3 5.4 240 l/h 0.2 0.5 0.8 1.2 1.7 2.3 3 3.8 4.7 270 l/h 0.2 0.4 0.7 1 1.5 2 2.7 3.4 4.2 300 l/h 0.15 0.34 0.6 0.9 1.4 1.8 2.4 3.1 3.8 d = Inner diameter of sample gas lines Q = Flow rate 5 5/36 Siemens AP 01 · 2015 © Siemens AG 2015 General information Tables International standards National standards also exist in most EU member states, and may be used in these countries in addition to the valid EN standards. In the Federal Republic of Germany, these are the DIN standards and the VDE regulations. Topic International However, extensive harmonization has already been carried out in the explosion protection sector, and most standards now also exist as "DIN EN ...." versions, which have also been incorporated into the VDE regulations. DIN EN standards are identical to the corresponding EN standards, were special national features, e.g. concerning areas of validity etc., are formulated in a national foreword. Europe/Germany USA FM ANSI/ISA Ex zone model ANSI/ISAS12.0.01 CSA 79-0-95 ANSI/ISAS12.26. 01 CSA-E79-6 UL Canada Ex Class Div. model Ex: General regulations IEC 60079-0 EN 50014/ VDE 0170/0171 Part 1 Oil immersion "o" IEC 60079-6 EN 50015/ DIN EN 50015, VDE 0170/0171 Part 2 Pressurized enclosure "p" IEC 60079-2 EN 50016/ DIN EN 50016, VDE 0170/0171 Part 3 Powder filling "q" IEC 60079-5 EN 50017/ DIN EN 50017, VDE 0170/0171 Part 4 Flameproof enclosure "d" IEC 60079-1 EN 50018/ DIN EN 50018, VDE 0170/0171 Part 5 Increased safety "e" IEC 60079-7 EN 50019/ DIN EN 50019, VDE 0170/0171 Part 6 Intrinsic safety "i" IEC 60079-11 EN 50020/ DIN EN 50020, VDE 0170/0171 Part 7 FM 3610 UL2279, Pt.11 pr UL 913 ANSI/ISAS12.02. 01 CSA-E79-11 CSA C22.2 No. 157 Degree of protection "n" IEC 60079-15 EN 50021/ DIN EN 50021, VDE 0170/0171 Part 8 FM 3611 UL2279, Pt.15 pr ANSI/ISA CSA-E79-15 S12.12. 01 CSA C22.2 No. 213 Encapsulation "m" IEC 60079-18 EN 50028/ DIN EN 50028, VDE 0170/0171 Part 9 Zone 0 IEC 60079-26 EN 50284/ DIN EN 50284, VDE 0170/0171 Part 12 Electrical safety IEC 61010 EN 61010-1/ DIN EN 61010-1, VDE 0411 Part 1 FM 3600 UL2279, Pt.6 FM 3620 FM 3615 (NFPA4 96) CSA-E79-2 UL2279, Pt.5 ANSI/ISAS12.25. 01 CSA-E79-5 UL2279, Pt.1 UL1203 ANSI/ISAS12.22. 01 CSA-E79-1 UL2279, Pt.7 ANSI/ISAS12.16. 01 CSA-E79-7 UL2279, Pt.18 ANSI/ISAS12.23. 01 Miscellaneous CSA TIL. E 13 A CSA C22.2 No. 30 CSA-E79-18 ANSI/ISA82. 02.01 5 CAN/CSAC22.2 No. 1010.1 Comparison of international and European standards Siemens AP 01 · 2015 5/37 © Siemens AG 2015 General information Tables International standards European standard German standard German title EN 1127 DIN EN 1127-1 Explosive atmospheres - Explosion protection - Part 1: Fundamentals and method EN 50039 DIN EN 50039, VDE 0170/0171 Part 10 Electrical equipment for hazardous areas; intrinsically-safe electrical systems "i" EN 13463-1 DIN EN 13 463-1 Non-electrical equipment for use in hazardous areas, Part 1: Fundamental method and requirements EN 50281-1-1 DIN EN 50281-1-1, VDE 0170/0171 Part 15-1-1 Electrical equipment for use in areas with combustible dust, Part 1-1: Electrical equipment with protection by enclosure EN 60079-10 DIN EN 60079-10, VDE 165 Part 101 Electrical equipment for potentially explosive gas atmospheres, Part 10: Division of potentially explosive areas EN 60079-14 DIN EN 60079-14, VDE 165 Part 1 Electrical equipment for hazardous areas, Part 14: Electrical installations in potentially explosive areas (except underground excavation) EN 60079-17 DIN EN 60079-17, VDE 0165 Part 10 Electrical equipment for potentially explosive gas atmospheres, Part 17: Testing and maintenance of electrical installations in hazardous areas (except underground excavation) EN 60950 DIN EN 60950, VDE 0805 Safety of information technology equipment, including electrical office machines Harmonized European standards for explosion protection T 1 > 450 °C 5 T 2 > 300 °C T 3 > 200 °C T 4 > 135 °C Acetone Ethane Ethyl acetate Ammonia Benzene (pure) Acetic acid Carbon monoxide Methane Methanol Propane Toluene Ethyl alcohol i-amyl acetate n-butane n-butyl alcohol Petrol Diesel fuel Aviation gasoline Fuel oil n-hexane Acetyl aldehyde Ethyl ether II B Town gas (Illuminating gas) Ethylene II C Hydrogen Acetylene I Methane II A Classification of gases and vapors into explosion groups and temperature classes 5/38 Siemens AP 01 · 2015 T 5 > 100 °C T 6 > 85 °C Carbon disulfide © Siemens AG 2015 General information Definitions ■ Overview Definitions Units of measurement Calibration gas Vol% Gas used for adjusting the sensitivity (deflection) of the detected gas. It is a gas mixture of known composition (measured component and suitable residual gas). Volume proportion in % of measured component, based on the sample gas. Sensitivity Ratio between a change in output variable observed on the measuring instrument and the change in input variable required for this. Linearity error of devices with linear characteristics Deviation of measured characteristic from a linear reference characteristic. The linearity is an important variable particularly for instruments which use a measuring effect with nonlinear characteristic and where the measured characteristic is linearized electronically. Cross-sensitivity Measure for the selectivity of a gas analyzer with regard to interfering components. It is the ratio between the displayed value of the interfering component and the displayed value of the measured component; both have the same concentration. In the case of analyzers where the total concentration of different materials is measured (e.g. total hydrocarbon concentration) and where the individual components are weighted differently in the measuring result, these factors are specified in equivalents of a master component (e.g. CH4 equivalents for the total hydrocarbon measurement) and not as cross-sensitivity. ppm (vpm) Parts per million, i.e. one proportion of the measured component per 106 proportions of the sample gas (corresponds to 10-4 %). In gas analysis technology, ppm is usually understood as volume concentrations. The dimension unit vpm is frequently used for unequivocal identification: 1 vpm = 1 cm³ / m³ Example: 1 000 vpm = 0.1 vol.% = 1 dm³ / m³ mg/m³ Mass of measured component in mg referred to 1 m³ of sample gas at 1 013 hPa and 20 °C. Example: 1 vpm = 1 cm³ / m³ corresponds to: (molecular weight of component / molecular volume of component) ⋅ (mg / m³) Weight concentration Specification of measured values in weight concentrations is not common with gas analysis. Weight concentrations can only be determined in exceptional cases. The dimension unit mg/m³ does not mean weight concentration. Dynamic response 5 The dynamic response of an analyzer is characterized by its response time and dead time. The response time is the time which passes until the output variable remains constantly within defined limits following an abrupt change in the input variable. The response time is usually understood as the time required to reach 90 (T90) or 95 % of the expected display. Siemens AP 01 · 2015 5/39 © Siemens AG 2015 General information Notes 5 5/40 Siemens AP 01 · 2015