ULTRAMAT 6 Catalog Sheet

© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
General information
■ Overview
Special versions
Special applications
Besides the standard combinations, special applications concerning material in the gas path, material in the sample cells
(e.g. Titan, Hastelloy C22) and measured components are also
available on request
TÜV version/QAL
The ULTRAMAT 6 single-channel or dual-channel gas analyzers
operate according to the NDIR two-beam alternating light principle and measure gases highly selectively whose absorption
bands lie in the infrared wavelength range from 2 to 9 μm, such
as CO, CO2, NO, SO2, NH3, H2O as well as CH4 and other
hydrocarbons.
Single-channel analyzers can simultaneously measure up to 2
gas components, while dual-channel analyzers can simultaneously measure 3 (or 4 on request) gas components.
■ Benefits
• High selectivity with double-layer detector and optical coupler
- Reliable measurements even in complex gas mixtures
• Low detection limits
- Measurements with low concentrations
• Corrosion-resistant materials in gas path (option)
- Measurement possible in highly corrosive sample gases
• Analyzer cells can be cleaned as required on site
- Cost savings due to reuse after contamination
• Electronics and physics: gas-tight isolation, purging is
possible, IP65
- Long service life even in harsh environments
• Heated versions (option)
- Use also in presence of gases condensing at low
temperature
• EEx(p) for zones 1 and 2 (according to ATEX 2G and ATEX 3G)
■ Application
Areas of application
• Measurement for boiler control in incineration plants
• Emission measurements in incineration plants
• Measurement in the automotive industry (test benches)
• Warning equipment
• Process gas concentrations in chemical plants
• Trace measurements in pure gas processes
• Environmental protection
• TLV (Threshold Limit Value) monitoring at the workplace
• Quality monitoring
• Ex versions for analyzing flammable and non-flammable
gases or vapors for use in hazardous areas
TÜV-approved versions are available for measurement of CO,
NO and SO2 according to 13th and 17th BlmSchV and TA Luft.
Smallest TÜV-approved and permitted measuring ranges:
• 1-component analyzer
CO: 0 to 50 mg/m3
NO: 0 to 100 mg/m3
SO2: 0 to 75 mg/m3
• 2-component analyzer (series connection)
CO: 0 to 75 mg/m3
NO: 0 to 200 mg/m3.
Furthermore, the TÜV-approved versions of the ULTRAMAT 6
comply with the requirements of EN 14956 and QAL 1 in accordance with EN 14181. Conformity of the analyzers with
both standards is TÜV-certified.
The analyzer drift can be determined in accordance with
EN 14181 (QAL 3) either manually or with a PC using the
SIPROM GA maintenance and servicing software. In addition,
selected manufacturers of emission evaluation computers
offer the possibility for downloading the drift data via the analyzer’s serial interface and to automatically record and process it in the evaluation computer.
Flow-type reference compartment
• The flow through the reference compartment should be
adapted to the sample gas flow
• The gas supply of the reduced flow-type reference compartment should have an upstream pressure of 3 000 to 5 000 hPa
(abs.). Then a restrictor will automatically adjust the flow to
approximately 8 ml/min
■ Design
19" rack unit
• 19" rack unit with 4 HU for installation
- in hinged frame
- in cabinets with or without telescopic rails
• Front plate for service purposes can be pivoted down (laptop
connection)
• Internal gas paths: hose made of FKM (Viton) or pipe made of
titanium or stainless steel
• Gas connections for sample gas inlet and outlet: pipe diameter 6 mm or 1/4"
• Flow indicator for sample gas on front plate (option)
• Pressure switch in sample gas path for flow monitoring
(option)
Field device
• Two-door enclosure with gas-tight separation of analyzer and
electronics sections from gas path
• Individually purgeable enclosure halves
• Parts in contact with sample gas can be heated up to 65 °C
(option)
• Gas path: hose made of FKM (Viton) or pipe made of titanium
or stainless steel (further materials possible as special applications)
• Gas connections for sample gas inlet and outlet: pipe union
for pipe diameter 6 mm or 1/4"
• Purging gas connections: pipe diameter 10 mm or 3/8"
Siemens AP 01 · 2015
1/71
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
General information
Display and control panel
• Large LCD field for simultaneous display of:
- Measured value (digital and analog displays)
- Status bar
- Measuring ranges
• Contrast of the LCD field adjustable via the menu
• Washable membrane keyboard with five softkeys
• Menu-driven operator control for parameterization, test functions, adjustment
• Operator support in plain text
• Graphic display of concentration trend; programmable time
intervals
• Bilingual operating software:
German/English, English/Spanish, French/English,
Spanish/English, Italian/English
Input and outputs
• One analog output per medium (from 0, 2, 4 to 20 mA;
NAMUR parameterizable)
• Two analog inputs freely configurable (e.g. correction of
cross-interferences or external pressure sensor)
LED backlit graphic
display and
membrane keyboard
with noticeable click
• Six binary inputs freely configurable (e.g. measurement range
changeover, processing of external signals from the sample
preparation)
• Six relay outputs freely configurable e.g. for fault, maintenance request, limit alarm, external solenoid valves)
• Expansion by eight additional binary inputs and eight additional relay outputs e.g. for autocalibration with up to four test
gases
Communication
RS 485 present in the basic unit (connection at the rear; for the
rack unit also behind the front plate).
Options
• AK interface for the automotive industry with extended
functions
• RS 485/RS 232 converter
• RS 485/Ethernet converter
• RS 485/USB converter
• Connection to networks via PROFIBUS DP/PA interface
• SIPROM GA software as the service and maintenance tool
Status line for display
of analyzer status
(programmable)
Two code levels
according to NAMUR
(maintenance and
specialist level)
Easy operation with
menu control
using five softkeys
Display of
concentrations as
numbers and bargraph
for channel 1
Display of current
measuring ranges
Display of
concentrations as
numbers and bargraph
for channel 2
Status line for channel 2
to display the unit
status (programmable)
Display of
start-of-scale and
full-scale values
ESC key
to abort inputs
Keyboard to
enter values
INFO key
for help in plain text
CLEAR key
to delete inputs
ENTER key
to accept
input values
ULTRAMAT 6, membrane keyboard and graphic display
1/72
Siemens AP 01 · 2015
MEAS key
to return to
measurement mode
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
General information
Designs – Parts wetted by sample gas, standard
Gas path
With hoses
19" rack unit
Field device
Bushing
Stainless steel, mat. no. 1.4571
Hose
FKM (e.g. Viton)
Field device Ex
-
Sample chamber:
With pipes
• Body
Aluminum
• Lining
Aluminum
• Fitting
Stainless steel, mat. no. 1.4571,
O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
• Window
CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM
(Kalrez)
Bushing
Titanium
Pipe
Titanium,
O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Sample chamber:
With pipes
• Body
Aluminum
• Lining
Tantalum (only for cell length 20 mm to 180 mm)
• Window
CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Bushing
Stainless steel, mat. no. 1.4571
Pipe
Stainless steel, mat. no. 1.4571,
O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Sample chamber:
• Body
Aluminum
• Lining
Aluminum or tantalum (tantalum only for cell length 20 mm to 180 mm)
• Window
CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Options
Gas path
Flow indicator
Pressure switch
19" rack unit
Field device
Field device Ex
Measurement pipe
Duran glass
-
-
Variable area
Duran glass
Suspension boundary
PTFE (Teflon)
Angle pieces
FKM (e.g. Viton)
Membrane
FKM (e.g. Viton)
-
-
Enclosure
PA 6.3T
Field device
Field device Ex
Versions – Parts wetted by sample gas, special applications (examples)
Gas path
With pipes
19" rack unit
Bushing
e.g. Hastelloy C22
Pipe
e.g. Hastelloy C22,
O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Sample chamber:
• Body
e.g. Hastelloy C22
• Window
CaF2, without adhesive
O-ring: FKM (e.g. Viton) or FFKM (Kalrez)
Siemens AP 01 · 2015
1/73
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
General information
Gas path (19" rack unit)
Legend for the gas path figures
1
Sample gas inlet channel 1
8
Restrictor
2
Sample gas outlet channel 1
9
Purge gas inlet
3
Reference gas outlet (option)
10
Gas inlet atmospheric pressure sensor
4
Reference gas inlet (option)
11
Atmospheric pressure sensor
5
Sample gas inlet channel 2
12
Flow indicator in sample gas path (option)
6
Sample gas outlet channel 2
13
Pressure switch in sample gas path (option)
7
IR physical system
9
10
2
1
4
3
7
8
P
P
11
F
13
12
Gas path ULTRAMAT 6, single-channel unit, 19" unit, with flow-type reference cell (option)
&KDQQHO
&KDQQHO
3
3
Gas path ULTRAMAT 6, dual-channel unit, 19" unit
1/74
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
General information
Gas path (field device)
Legend for the gas path figures
1
Sample gas inlet
7
Purging gas outlet (analyzer side)
2
Sample gas outlet
8
Purging gas inlet (analyzer side)
3
Reference gas inlet (option)
9
Connection of atmospheric pressure sensor
4
Reference gas outlet (option)
10
IR physical system
5
Purging gas inlet (electronics side)
11
Atmospheric pressure sensor
6
Purging gas outlet (electronics side)
3
(OHFWURQLFVLGH
$QDO\]LQJVLGH
Gas path ULTRAMAT 6, field unit, with flow-type reference cell (option)
Siemens AP 01 · 2015
1/75
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
General information
■ Function
Principle of operation
The ULTRAMAT 6 gas analyzer operates according to the infrared two-beam alternating light principle with double-layer detector and optical coupler.
The measuring principle is based on the molecule-specific
absorption of bands of infrared radiation. The absorbed wavelengths are characteristic to the individual gases, but may partially overlap. This results in cross-sensitivities which are reduced to a minimum in the ULTRAMAT 6 gas analyzers by the
following measures:
• Gas-filled filter cell (beam divider)
• Double-layer detector with optical coupler
• Optical filters if necessary
The figure shows the measuring principle. An IR source (1)
which is heated to approx. 700 ºC and which can be shifted to
balance the system is divided by the beam divider (3) into two
equal beams (sample and reference beams). The beam divider
also acts as a filter cell.
The reference beam passes through a reference cell (8) filled
with N2 (a non-infrared-active gas) and reaches the right-hand
side of the detector (11) practically unattenuated. The sample
beam passes through the sample chamber (7) through which
the sample gas flows and reaches the left-hand side of the detector (10) attenuated to a lesser or greater extent depending on
the concentration of the sample gas. The detector is filled with a
defined concentration of the gas component to be measured.
The detector is designed as a double-layer detector. The center
of the absorption band is preferentially absorbed in the upper
detector layer, the edges of the band are absorbed to approximately the same extent in the upper and lower layers. The upper
and lower detector layers are connected together via the microflow sensor (12). This coupling means that the spectral sensitivity has a very narrow band.
1
2
3
4
5
6
7
8
9
11
10
The optical coupler (13) lengthens the lower receiver cell layer
optically. The infrared absorption in the second detector layer is
varied by changing the slider position (14). It is thus possible to
individually minimize the influence of interfering components.
12
A chopper (5) rotates between the beam divider and the sample
chamber and interrupts the two beams alternately and periodically. If absorption takes place in the sample chamber, a pulsating flow is generated between the two detector levels which is
converted by the microflow sensor (12) into an electric signal.
The microflow sensor consists of two nickel-plated grids heated
to approximately 120 ºC, which, along with two supplementary
resistors, form a Wheatstone bridge. The pulsating flow together
with the dense arrangement of the Ni grids causes a change in
resistance. This leads to an offset in the bridge, which is dependent on the concentration of the sample gas.
Notes
The sample gases must be fed into the analyzers free of dust.
Condensation should be prevented from occurring in the sample
chambers. Therefore, the use of gas modified for the measuring
task is necessary in most application cases.
As far as possible, the ambient air of the analyzer should not
have a large concentration of the gas components to be measured.
Flow-type reference sides with reduced flow must not be operated with flammable or toxic gases.
Flow-type reference sides with reduced flow and an O2 content
> 70 % may only be used together with Y02 (Clean for O2).
1/76
Siemens AP 01 · 2015
13
14
1
2
3
4
5
6
7
IR source, adjustable
Optical filter
Beam divider
Eddy current drive
Chopper
Sample gas inlet
Sample cell
8
9
10
11
12
13
14
Reference cell
Sample gas outlet
Detector, meas. side
Detector, reference side
Microflow sensor
Optical coupler
Slider, adjustable
ULTRAMAT 6, principle of operation
Channels with electronically suppressed zero point only differ
from the standard version in the measuring range parameterization.
Physically suppressed zeros can be provided as a special application.
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
General information
Essential characteristics
• Dimension of measured value freely selectable
(e.g. vpm, mg/m3)
• Four freely-parameterizable measuring ranges per
component
• Measuring ranges with suppressed zero point possible
• Measuring range identification
• Galvanically isolated signal output 0/2/4 to 20 mA per component
• Automatic or manual measuring range switchover selectable;
remote switching is also possible
• Differential measuring ranges with flow-type reference cell
• Storage of measured values possible during adjustments
• Time constants selectable within wide limits (static/dynamic
noise suppression); i.e. the response time of the analyzer or
component can be matched to the respective measuring task
• Short response time
• Low long-term drift
• Measuring point switchover for up to 6 measuring points
(programmable)
• Measuring point identification
• Monitoring of sample gas flow (option)
• Internal pressure sensor for correction of variations in atmospheric pressure in the range 700 to 1 200 hPa absolute
• External pressure sensor can be connected for correction of
variations in the process gas pressure in the range 700 to
1 500 hPa absolute (option)
• Two control levels with separate authorization codes to prevent unintentional and unauthorized inputs
• Automatic, parameterizable measuring range calibration
• Simple handling using a numerical membrane keyboard and
operator prompting
• Operation based on NAMUR recommendation
• Customer-specific analyzer options such as:
- Customer acceptance
- TAG labels
- Drift recording
• Easy device replacement since electric connections can be
simply disconnected from the device
• Sample chambers for use in presence of highly corrosive sample gases (e.g. tantalum layer or Hastelloy C22)
Additional features, dual-channel version
• Separate design of physical unit, electronics, inputs/outputs
and power supply for each channel
• Display and operation via common LCD panel and keyboard
• Measurement channels 1 and 2 can be converted to series
connection (linking of gas connections from channel 1 to
channel 2 on rear)
Siemens AP 01 · 2015
1/77
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
■ Technical specifications
Pressure correction range
General information
Measuring ranges
4, internally and externally switchable; autoranging is also possible
Pressure sensor
• Internal
700 … 1 200 hPa absolute
Smallest possible measuring range
Dependent on the application:
e.g. CO: 0 … 10 vpm,
CO2: 0 … 5 vpm
• External
700 … 1 500 hPa absolute
Largest possible measuring span
Dependent on the application
Measuring range with suppressed
zero point
Any zero point within
0 … 100 vol.% can be implemented; smallest possible span
20 %
Operating position
Front wall, vertical
Conformity
CE mark in accordance with
EN 50081-1, EN 50082-2
Influence of interfering gases must be considered separately
Design, enclosure
Approx. 15 kg
(with one IR channel)
Approx. 21 kg
(with two IR channels)
Weight
Degree of protection
IP20 according to EN 60529
Electrical characteristics
EMC
(Electromagnetic Compatibility)
In accordance with standard
requirements of NAMUR NE21
(08/98)
Electrical safety
According to EN 61010-1,
overvoltage category III
Power supply
100 … 120 V AC (nominal range
of use 90 … 132 V), 48 … 63 Hz
or
200 … 240 V AC (nominal range
of use 180 … 264 V), 48 … 63 Hz
Power consumption
1-channel unit: Approx. 40 VA
2-channel unit: Approx. 70 VA
Fuse values
• 100 ... 120 V
1 T/250 (7MB2121)
1.6 T/250 (7MB2123)
• 200 ... 240 V
0.63 T/250 (7MB2121)
1 T/250 (7MB2123)
Measuring response (relating to sample gas pressure 1 013 hPa
absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation
< ± 1 % of the smallest possible
measuring range according to
rating plate
Zero point drift
< ± 1 % of the current measuring
range/week
Measured-value drift
< ± 1 % of the current measuring
range/week
Repeatability
≤ 1 % of the current measuring
range
Detection limit
1 % of the smallest possible
measuring range
Linearity error
< 0.5 % of the full-scale value
Influencing variables (relating to sample gas pressure 1 013 hPa
absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature
< 1 % of current measuring
range/10 K (with constant
receiver cell temperature)
Sample gas pressure
• When pressure compensation
has been switched on: < 0.15 %
of the span/1 % change in atmospheric pressure
• When pressure compensation
has been switched off: < 1.5 %
of the span/1 % change in atmospheric pressure
Sample gas flow
Negligible
Power supply
< 0.1 % of the current measuring
range with rated voltage ± 10 %
Environmental conditions
Application-specific measuring
influences possible if ambient air
contains measured components or
cross interference-sensitive gases
Electrical inputs and outputs
Analog output
0/2/4 ... 20 mA, isolated;
load ≤ 750 Ω
Relay outputs
6, with changeover contacts,
freely parameterizable, e.g. for
measuring range identification;
load: 24 V AC/DC/1 A, isolated,
non-sparking
Analog inputs
2, dimensioned for 0/2/4 ... 20 mA
for external pressure sensor and
accompanying gas influence correction (correction of cross-interference)
Binary inputs
6, designed for 24 V, isolated,
freely parameterizable, e.g. for
measuring range switchover
Dynamic response
Serial interface
RS 485
Warm-up period
At room temperature < 30 min
(the technical specification will be
met after 2 hours)
Options
Delayed display (T90-time)
Dependent on length of analyzer
chamber, sample gas line and
parameterizable damping
AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or
PROFIBUS DP
Damping (electrical time constant)
0 … 100 s, parameterizable
Dead time (purging time of the gas
path in the unit at 1 l/min)
Approximately 0.5 ... 5 s,
depending on version
Time for device-internal signal
processing
<1s
Gas inlet conditions
Permissible sample gas pressure
• With hoses
- Without pressure switch
600 … 1 500 hPa (absolute)
- With pressure switch
700 … 1 300 hPa (absolute)
• With pipes (without pressure switch) 600 … 1 500 hPa (absolute)
Sample gas flow
18 … 90 l/h (0.3 … 1.5 l/min)
Sample gas temperature
Min. 0 ... max. 50 °C, but above
the dew point
Sample gas humidity
< 90 % RH (relative humidity), or
dependent on measuring task,
non-condensing
1/78
Siemens AP 01 · 2015
Climatic conditions
Permissible ambient temperature
-30 ... +70 °C during storage and
transportation,
5 ... 45 °C during operation
Permissible humidity
< 90 % RH (relative humidity) as
annual average, during storage
and transportation (dew point
must not be undershot)
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
Single-channel 19" rack unit for installation in cabinets
7MB2121- 7 7 7 7 7 - 7 A A 7
Cannot be combined
Click on the Article No. for the online configuration in the PIA Life Cycle
Portal.
Gas connections for sample gas and reference gas
Pipe with 6 mm outer diameter
Pipe with ¼" outer diameter
0
1
0
1
Possible with measuring
range identification
CO
11 ... 30
CO highly selective (with optical filter)
12 ... 30
CO (TÜV; see table "TÜV single component", page 1/87)
A
B
X
CO2
CH4
C2H2
10 ... 30
13 ... 30
15 ... 30
C
D
E
C2H4
C2H6
C3H6
15 ... 30
14 ... 30
14 ... 30
F
G
H
C3H8
C4H6
C4H10
13 ... 30
15 ... 30
14 ... 30
J
K
L
C6H14
SO2 (TÜV; see table "TÜV single component",
page 1/87)
NO (TÜV; see table "TÜV single component",
page 1/87)
14 ... 30
13 ... 30
M
N
14 ... 20, 22
P
14 ... 30
17 ... 20, 22
13 ... 30
Q
R
S
A21
A20
Measured component
NH3 (dry)
H2O
N2O
Q
R
Smallest measuring
range
Largest measuring
range
Measuring range
identification
0 ... 5 vpm
0 ... 10 vpm
0 ... 20 vpm
0 ... 100 vpm
0 ... 200 vpm
0 ... 400 vpm
10
11
12
A
B
C
0 ... 50 vpm
0 ... 100 vpm
0 ... 300 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
13
14
15
D
E
F
0 ... 500 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
16
17
18
G
H
J
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 5 000 vpm
0 ... 30 000 vpm
0 ... 15 000 vpm
0 ... 50 000 vpm
19
20
21
K
L
M
0 ... 1 %
0 ... 1 %
0 ... 3 %
0 ... 3 %
0 ... 10 %
0 ... 10 %
22
23
24
N
P
Q
0 ... 3 %
0 ... 5 %
0 ... 5 %
0 ... 30 %
0 ... 15 %
0 ... 50 %
25
26
27
R
S
T
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 30 %
0 ... 100 %
0 ... 100 %
28
29
30
U
V
W
Internal gas paths
Sample chamber1)
(lining)
Aluminum
Aluminum
Reference chamber
(flow-type)
Non-flow-type
Flow-type
0
1
Tantalum
Tantalum
Non-flow-type
Flow-type
Stainless steel pipe
Aluminum
(mat. no. 1.4571)
Tantalum
With sample gas monitoring
Hose made of FKM
(Viton)
Hose made of FKM
(Viton)
Pipe made of titanium
Aluminum
Aluminum
0
1
0
A20, A21
4
5
4
5
A20, A21, Y02
Y02
Non-flow-type
Non-flow-type
6
8
6
8
A20, A21
A20, A21
Non-flow-type
Flow-type
2
3
2
A20, A21
2
3
Footnotes: see next page
Siemens AP 01 · 2015
1/79
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
■ Selection and ordering data
ULTRAMAT 6 gas analyzer
Single-channel 19" rack unit for installation in cabinets
Article No.
7MB2121- 7 7 7 7 7 - 7 A A 7
Add-on electronics
Without
AUTOCAL function
• With 8 additional digital inputs/outputs
• With serial interface for the automotive industry (AK)
• With 8 digital inputs/outputs, PROFIBUS PA interface
• With 8 digital inputs/outputs, PROFIBUS DP interface
0
1
3
6
7
Power supply
100 ... 120 V AC, 48 ... 63 Hz
200 ... 240 V AC, 48 ... 63 Hz
3
0
1
Operating software and documentation
German
English
French
Spanish
Italian
Additional versions
0
1
2
3
4
Order code
Add "-Z" to Article No. and specify Order code
Flow-type reference cell with reduced flow, 6 mm
A20
Flow-type reference cell with reduced flow, ¼"
A21
Telescopic rails (2 units)
A31
TAG labels (specific lettering based on customer information)
B03
Kalrez gaskets in sample gas path
B04
SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508
and IEC 61511
C20
FM/CSA certificate – Class I Div 2
E20
Clean for O2 service (specially cleaned gas path)
Y02
Measuring range indication in plain text, if different from the standard setting
Y11
Special setting (only in conjunction with an application no., e.g. extended
measuring range)
Y12
Extended special setting (only in conjunction with an application no.,
e.g. determination of cross-interferences)
Y13
TÜV version acc. to 13th and 17th BlmSchV
Y17
Accessories
Article No.
RS 485/Ethernet converter
A5E00852383
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
AUTOCAL function with serial interface for the automotive industry (AK)
C79451-A3480-D512
AUTOCAL function with 8 digital inputs/outputs
C79451-A3480-D511
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA
A5E00057307
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP
A5E00057312
Set of Torx screwdrivers
A5E34821625
1)
Only for cell length 20 to 180 mm
1/80
Siemens AP 01 · 2015
Cannot be combined
E20
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
Two-channel 19" rack unit for installation in cabinets
for measuring 2 IR components
7MB2123- 7 7 7 7 7 - 7 7 7 7
Cannot be combined
Click on the Article No. for the online configuration in the PIA Life Cycle
Portal.
Gas connections for sample gas and reference gas
Pipe with 6 mm outer diameter
Pipe with ¼" outer diameter
Possible with measuring
Channel 1
Measured component
range identification
CO
11 ... 30
CO highly selective (with optical filter)
12 ... 30
CO (TÜV; see table "TÜV single component", page 1/87)
CO2
10 ... 30
CH4
13 ... 30
C2H2
15 ... 30
C2H4
15 ... 30
C2H6
14 ... 30
C3H6
14 ... 30
C3H8
13 ... 30
C4H6
15 ... 30
C4H10
14 ... 30
C6H14
14 ... 30
SO2 (TÜV; see table "TÜV single component",
13 ... 30
page 1/87)
NO (TÜV; see table "TÜV single component",
14 ... 20, 22
page 1/87)
NH3 (dry)
14 ... 30
H2O
17 ... 20, 22
N2O
13 ... 30
0
1
0
1
A21, A41
A20, A40
A
B
X
C
D
E
F
G
H
J
K
L
M
N
P
Q
R
S
Q
R
Smallest measuring range Largest measuring
range
Measuring range
identification
0 ... 5 vpm
0 ... 10 vpm
0 ... 20 vpm
0 ... 50 vpm
0 ... 100 vpm
0 ... 300 vpm
0 ... 500 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 5 000 vpm
0 ... 100 vpm
0 ... 200 vpm
0 ... 400 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
0 ... 30 000 vpm
0 ... 15 000 vpm
0 ... 50 000 vpm
10
11
12
13
14
15
16
17
18
19
20
21
A
B
C
D
E
F
G
H
J
K
L
M
0 ... 1 %
0 ... 1 %
0 ... 3 %
0 ... 3 %
0 ... 5 %
0 ... 5 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 3 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 15 %
0 ... 50 %
0 ... 30 %
0 ... 100 %
0 ... 100 %
22
23
24
25
26
27
28
29
30
N
P
Q
R
S
T
U
V
W
Internal gas paths
Sample chamber1)
(lining)
Reference chamber
(flow-type)
Hose made of FKM
(Viton)
Aluminum
Aluminum
Non-flow-type
Flow-type
0
1
Pipe made of titanium
Tantalum
Tantalum
Non-flow-type
Flow-type
Stainless steel pipe
(mat. no. 1.4571)
Aluminum
Tantalum
0
1
0
A20, A21, A40, A41
4
5
4
5
A20, A21, A40, A41, Y02
Y02
Non-flow-type
Non-flow-type
6
8
6
8
A20, A21, A40, A41
A20, A21, A40, A41
Non-flow-type
Flow-type
2
3
2
A20, A21, A40, A41
With sample gas monitoring
Hose made of FKM
(Viton)
1)
Aluminum
Aluminum
2
3
Only for cell length 20 to 180 mm
Siemens AP 01 · 2015
1/81
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
■ Selection and ordering data
ULTRAMAT 6 gas analyzer
Two-channel 19" rack unit for installation in cabinets
for measuring 2 IR components
Article No.
7MB2123- 7 7 7 7 7 - 7 7 7 7
Cannot be combined
Add-on electronics
0
Without
AUTOCAL function
• With 8 additional digital inputs/outputs each for channel 1
• With 8 additional digital inputs/outputs each for channel 2
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
• With serial interface for the automotive industry (AK)
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
and PROFIBUS PA interface
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
and PROFIBUS DP interface
Power supply
100 ... 120 V AC, 48 ... 63 Hz
200 ... 240 V AC, 48 ... 63 Hz
1
2
3
5
6
5
7
0
1
Channel 2
Possible with measuring
Measured component
range identification
CO
11 ... 30
CO highly selective (with optical filter)
12 ... 30
CO (TÜV; see table "TÜV single component", page 1/87)
CO2
10 ... 30
CH4
13 ... 30
C2H2
15 ... 30
C2H4
15 ... 30
C2H6
14 ... 30
C3H6
14 ... 30
C3H8
13 ... 30
C4H6
15 ... 30
C4H10
14 ... 30
C6H14
14 ... 30
SO2 (TÜV; see table "TÜV single component",
13 ... 30
page 1/87)
NO (TÜV; see table "TÜV single component",
14 ... 20, 22
page 1/87)
NH3 (dry)
14 ... 30
H2O
17 ... 20, 22
N2O
13 ... 30
A
B
X
C
D
E
F
G
H
J
K
L
M
N
P
Q
R
S
Q
R
Smallest measuring range Largest measuring
range
Measuring range
identification
0 ... 5 vpm
0 ... 10 vpm
0 ... 20 vpm
0 ... 50 vpm
0 ... 100 vpm
0 ... 300 vpm
0 ... 500 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 5 000 vpm
0 ... 100 vpm
0 ... 200 vpm
0 ... 400 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
0 ... 30 000 vpm
0 ... 15 000 vpm
0 ... 50 000 vpm
10
11
12
13
14
15
16
17
18
19
20
21
A
B
C
D
E
F
G
H
J
K
L
M
0 ... 1 %
0 ... 1 %
0 ... 3 %
0 ... 3 %
0 ... 5 %
0 ... 5 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 3 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 15 %
0 ... 50 %
0 ... 30 %
0 ... 100 %
0 ... 100 %
22
23
24
25
26
27
28
29
30
N
P
Q
R
S
T
U
V
W
Operating software and documentation
German
English
French
Spanish
Italian
1/82
Siemens AP 01 · 2015
0
1
2
3
4
E20
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Selection and ordering data
Additional versions
Order code
Cannot be
combined
Add "-Z" to Article No. and specify Order codes.
Flow-type reference cell with reduced flow, 6 mm (channel 1)
A20
Flow-type reference cell with reduced flow, ¼" (channel 1)
A21
Flow-type reference cell with reduced flow, 6 mm (channel 2)
A40
Flow-type reference cell with reduced flow, ¼" (channel 2)
A41
Connection pipe
(can only be combined with the appropriate gas connection diameter and internal gas path materials)
• Made of titanium, 6 mm, complete with screwed gland, for sample gas side
A22
• Made of titanium, 6 mm, complete with screwed gland, for reference gas side
A23
• Made of titanium, ¼", complete with screwed gland, for sample gas side
A24
• Made of titanium, ¼", complete with screwed gland, for reference gas side
A25
• Made of stainless steel (mat. no. 1.4571), 6 mm, complete with screwed gland, for sample gas side
A27
• Made of stainless steel (mat. no. 1.4571), ¼", complete with screwed gland, for sample gas side
A29
Telescopic rails (2 units)
A31
TAG labels (specific lettering based on customer information)
B03
Kalrez gaskets in sample gas path (channel 1)
B04
Kalrez gaskets in sample gas path (channel 2)
B05
SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511
C20
FM/CSA certificate – Class I Div 2
E20
Clean for O2 service (specially cleaned gas path; channels 1 + 2)
Y02
Measuring range indication in plain text, if different from the standard setting
Y11
Special setting (only in conjunction with an application no., e.g. extended measuring range)
Y12
Extended special setting (only in conjunction with an application no., e.g. determination of
cross-interferences)
Y13
TÜV version acc. to 13th and 17th BlmSchV (1st channel)
Y17
TÜV version acc. to 13th and 17th BlmSchV (2nd channel)
Y18
Accessories
Article No.
RS 485/Ethernet converter
A5E00852383
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
AUTOCAL function with serial interface for the automotive industry (AK)
C79451-A3480-D33
AUTOCAL function with 8 digital inputs/outputs for channel 1 or channel 2
C79451-A3480-D511
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA for channel 1 or channel 2
A5E00057307
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP for channel 1 or channel 2
A5E00057312
Set of Torx screwdrivers
A5E34821625
Siemens AP 01 · 2015
1/83
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
Single-channel or dual-channel 19" rack unit for installation in cabinets
for measuring 2 or 3 IR components
7MB2124- 7 7 7 7 7 - 7 7 7 7
Cannot be combined
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Gas connections for sample gas and reference gas
Pipe with 6 mm outer diameter
Pipe with ¼" outer diameter
0
1
0
1
Measured component Smallest measuring
range
Largest measuring range
CO
NO
0 ... 100 vpm
0 ... 100 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
AA
CO
NO
0 ... 300 vpm
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
AB
CO
NO
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
AC
A21, A41
A20, A40
For CO/NO (TÜV; table "TÜV, 2 components in series", page 1/76)
CO2
CO
0 ... 100 vpm
0 ... 100 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
BA
CO2
CO
0 ... 300 vpm
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
BB
CO2
CO
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
BC
CO2
CO
0 ... 3 000 vpm
0 ... 3 000 vpm
0 ... 30 000 vpm
0 ... 30 000 vpm
BD
CO2
CO
0 ... 1 %
0 ... 1 %
0 ... 10 %
0 ... 10 %
BE
CO2
CO
0 ... 3 %
0 ... 3 %
0 ... 30 %
0 ... 30 %
BF
CO2
CO
0 ... 10 %
0 ... 10 %
0 ... 100 %
0 ... 100 %
BG
CO2
CH4
0 ... 10 %
0 ... 10 %
0 ... 100 %
0 ... 100 %
CG
CO2
NO
0 ... 300 vpm
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
DB
Internal gas paths
Sample chamber1)
(lining)
Aluminum
Aluminum
Reference chamber
(flow-type)
Non-flow-type
Flow-type
0
1
Pipe made of titanium Tantalum
Tantalum
Stainless steel pipe
Aluminum
(mat. no. 1.4571)
Tantalum
Non-flow-type
Flow-type
Non-flow-type
Non-flow-type
4
5
6
8
With sample gas monitoring
Hose made of FKM
Aluminum
(Viton)
Aluminum
Non-flow-type
Flow-type
2
3
Hose made of FKM
(Viton)
Add-on electronics
Without
AUTOCAL function
• With 8 additional digital inputs/outputs each for channel 1
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
• With serial interface for the automotive industry (AK), channel 1
• With serial interface for the automotive industry (AK),
channel 1 and channel 2
• With 8 additional digital inputs/outputs for channel 1
and PROFIBUS PA interface
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
and PROFIBUS PA interface
• With 8 additional digital inputs/outputs for channel 1
and PROFIBUS DP interface
• With 8 additional digital inputs/outputs each for channel 1 and channel 2
and PROFIBUS DP interface
1)
Only for cell length 20 to 180 mm
1/84
Siemens AP 01 · 2015
0
1
2
3
0
A20, A21, A40, A41
4
5
6
8
A20, A21, A40, A41, Y02
Y02
A20, A21, A40, A41
A20, A21, A40, A41
2
A20, A21, A40, A41
0
1
2
3
4
2
3
4
5
6
6
7
8
8
E20
E20
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
Single-channel or dual-channel 19" rack unit for installation in cabinets
for measuring 2 or 3 IR components
Power supply
100 ... 120 V AC, 48 ... 63 Hz
200 ... 240 V AC, 48 ... 63 Hz
7MB2124- 7 7 7 7 7 - 7 7 7 7
0
1
Channel 2
Possible with measuring
Measured component
range identification
Without channel 2
CO
11 ... 30
CO highly selective (with optical filter)
12 ... 30
CO (TÜV; see table "TÜV single component", page 1/87)
CO2
10 ... 30
CH4
13 ... 30
C2H2
15 ... 30
C2H4
15 ... 30
C2H6
14 ... 30
C3H6
14 ... 30
C3H8
13 ... 30
C4H6
15 ... 30
C4H10
14 ... 30
C6H14
14 ... 30
SO2 (TÜV; see table "TÜV single component", 13 ... 30
page 1/87)
NO (TÜV; see table "TÜV single component", 14 ... 20, 22
page 1/87)
NH3 (dry)
14 ... 30
H2O
17 ... 20, 22
N2O
13 ... 30
Smallest measuring
range
Without channel 2
Largest measuring
range
Measuring range
identification
0 ... 5 vpm
0 ... 10 vpm
0 ... 20 vpm
0 ... 50 vpm
0 ... 100 vpm
0 ... 300 vpm
0 ... 500 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 5 000 vpm
0 ... 100 vpm
0 ... 200 vpm
0 ... 400 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
0 ... 30 000 vpm
0 ... 15 000 vpm
0 ... 50 000 vpm
10
11
12
13
14
15
16
17
18
19
20
21
0 ... 1 %
0 ... 1 %
0 ... 3 %
0 ... 3 %
0 ... 5 %
0 ... 5 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 3 %
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 15 %
0 ... 50 %
0 ... 30 %
0 ... 100 %
0 ... 100 %
22
23
24
25
26
27
28
29
30
W
A
B
X
C
D
E
F
G
H
J
K
L
M
N
W
P
Q
R
S
Q
R
X
Operating software and documentation
German
English
French
Spanish
Italian
Cannot be combined
X
A40, A41, B05
A
B
C
D
E
F
G
H
J
K
L
M
N
P
Q
R
S
T
U
V
W
0
1
2
3
4
Siemens AP 01 · 2015
1/85
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
■ Selection and ordering data
Additional versions
Order code
Add "-Z" to Article No. and specify Order codes.
Flow-type reference cell with reduced flow, 6 mm (channel 1)
A20
Flow-type reference cell with reduced flow, ¼" (channel 1)
A21
Flow-type reference cell with reduced flow, 6 mm (channel 2)
A40
Flow-type reference cell with reduced flow, ¼" (channel 2)
A41
Connection pipe
(can only be combined with the appropriate gas connection diameter and internal gas path materials)
• Made of titanium, 6 mm, complete with screwed gland, for sample gas side
A22
• Made of titanium, 6 mm, complete with screwed gland, for reference gas side
A23
• Made of titanium, ¼", complete with screwed gland, for sample gas side
A24
• Made of titanium, ¼", complete with screwed gland, for reference gas side
A25
• Made of stainless steel (mat. no. 1.4571), 6 mm, complete with screwed gland, for sample gas side
A27
• Made of stainless steel (mat. no. 1.4571), ¼", complete with screwed gland, for sample gas side
A29
Telescopic rails (2 units)
A31
TAG labels (specific lettering based on customer information)
B03
Kalrez gaskets in sample gas path (channel 1)
B04
Kalrez gaskets in sample gas path (channel 2)
B05
SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511
C20
FM/CSA certificate – Class I Div 2
E20
Clean for O2 service (specially cleaned gas path; channels 1 + 2)
Y02
Measuring range indication in plain text, if different from the standard setting
Y11
Special setting (only in conjunction with an application no., e.g. extended measuring range)
Y12
Extended special setting (only in conjunction with an application no., e.g. determination of
cross-interferences)
Y13
TÜV version acc. to 13th and 17th BlmSchV (channel 1)
Y17
TÜV version acc. to 13th and 17th BlmSchV (channel 2)
Y18
Accessories
Article No.
RS 485/Ethernet converter
A5E00852383
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
AUTOCAL function with serial interface for the automotive industry (AK)
C79451-A3480-D33
AUTOCAL function with 8 digital inputs/outputs for channel 1 or channel 2
C79451-A3480-D511
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA for channel 1 or channel 2
A5E00057307
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP for channel 1 or channel 2
A5E00057312
Set of Torx screwdrivers
A5E34821625
1/86
Siemens AP 01 · 2015
Cannot be
combined
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
TÜV single component
Component
CO (TÜV)
Measuring range
identification
Smallest
measuring range
from 0 to ...
SO2 (TÜV)
Largest
measuring range
from 0 to ...
C
D
50
mg/m3
1 000
mg/m3
E
mg/m3
F
300
G
500 mg/m3
5 000 mg/m3
H
1 000 mg/m3
10 000 mg/m3
K
mg/m3
mg/m3
P
R
V
3 000
10
g/m3
30
g/m3
100
g/m3
3 000
mg/m3
30 000
100
g/m3
300
g/m3
1 160
Smallest
measuring range
from 0 to ...
Largest
measuring range
from 0 to ...
75 mg/m3
1 500 mg/m3
300
mg/m3
3 000 mg/m3
500
mg/m3
5 000 mg/m3
1 000
mg/m3
3 000 mg/m3
g/m3
NO (TÜV)
10
g/m3
30
g/m3
100
g/m3
300
g/m3
10 000
mg/m3
30 000 mg/m3
100
g/m3
300
g/m3
1 000
g/m3
2 630
g/m3
Smallest
measuring range
from 0 to ...
Largest
measuring range
from 0 to ...
100 mg/m3
2 000 mg/m3
mg/m3
3 000 mg/m3
500 mg/m3
5 000 mg/m3
1 000 mg/m3
10 000 mg/m3
mg/m3
30 000 mg/m3
300
3 000
10
g/m3
100 g/m3
30
g/m3
300 g/m3
100
g/m3
1 250 g/m3
Example for ordering
ULTRAMAT 6, TÜV
Component: CO
Measuring range: 0 to 50 / 1 000 mg/m3
with hoses, non-flow-type reference compartment
without automatic adjustment (AUTOCAL)
230 V AC; German
7MB2121-0XD00-1AA0-Z +Y17
TÜV, 2 components in series
Component
CO (TÜV)
Measuring range
identification
Smallest measuring range
from 0 to ...
Largest measuring range
from 0 to ...
Smallest measuring range
from 0 to ...
Largest measuring range
from 0 to ...
AA
75 mg/m3
1 000 mg/m3
200 mg/m3
2 000 mg/m3
mg/m3
mg/m3
3 000 mg/m3
AB
AC
300
mg/m3
1 000
mg/m3
NO (TÜV)
3 000
10 000
mg/m3
300
1 000
mg/m3
10 000 mg/m3
Example for ordering
ULTRAMAT 6, TÜV, 2-component unit
Components: CO/NO + SO2
Measuring range: CO: 0 to 75 / 1 000 mg/m3, NO: 0 to 200 / 2 000 mg/m3, SO2: 0 to 75 / 1 500 mg/m3
with hoses, non-flow-type reference compartment
without automatic adjustment (AUTOCAL)
230 V AC; German
7MB2124-0AA00-1NC0-Z +Y17+Y18
Note: for 3 components take both tables into consideration.
Ordering information measured component N2O
Certification in accordance with AM0028 and AM0034 (Kyoto Protocol) for measuring N2O, measuring range 0 ... 300 ppm /
3 000 ppm.
Version: Standard device
Siemens AP 01 · 2015
1/87
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Dimensional drawings
s
8/75$0$7
0
1/88
Siemens AP 01 · 2015
ULTRAMAT 6, 19“ unit, dimensions in mm (example: 1-channel version)
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
■ Schematics
Pin assignment (electrical and gas connections)
SUB-D 9F connector (RS 485)
M
5
9
GND
4
8
3
+5 V
7
2
6
M
1
GND
R_Level-NNC
RD/TD-N
RD/TD-P
R_Level-PNC
NC
GND
It is possible to connect
bus terminating resistors
to pins 7 and 9.
SUB-D 15F connector
M
8
15
7
14
6
13
5
12
M
4
11
3
M
10
2
9
M
1
GND
Analog output 2-P
Analog output 2-N
Analog output 1-P
Analog output 1-N
NC
NC
Analog input 2-P
Analog input 2-N
Analog input 1-P
Analog input 1-N
Binary input 6-P
Binary input 5-P
Binary input 5 to 6-N
GND
For 2-component version only
of the ULTRAMAT part
Analog outputs isolated
(also from each other), RL: ≤ 750 Ω
Pressure correction
Pressure correction
Correction of cross-interference
Correction of cross-interference
Isolated via optocoupler
"0" = 0 V (0 ... 4.5 V)
"1" = 24 V (13 ... 33 V)
Analog inputs
non-isolated,
0 ... 20 mA/500 Ω
or 0 ... 10 V
(low-resistance)
SUB-D 25F connector
13
M
25
12
24
11
23
GND
Binary input 4-P
Binary input 3-P
Binary input 2-P
Binary input 1-P
Binary input 1 to 4-N
Isolated via optocoupler
"0" = 0 V (0 ... 4.5 V)
"1" = 24 V (13 ... 33 V)
10
22
Relay 6
9
21
8
Relay 5
20
7
19
Relay 4
6
18
5
Relay 3
Contact load
max. 24 V/1 A, AC/DC; relay
contacts shown: relay coil has
zero current
17
4
Relay 2
16
3
15
2
Relay 1
1
GND
14
M
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNV
PXVWEHVKLHOGHGDQGUHVWDJDLQVWWKH
HQFORVXUHSRWHQWLDO
ULTRAMAT 6, 19“ unit, pin assignment
Siemens AP 01 · 2015
1/89
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
19" rack unit
&RQQHFWRU68%')RSWLRQ
0
*1'
1&
1&
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXWWR1
,VRODWHG
YLDRSWRFRXSOHU
99
99
5HOD\
5HOD\
5HOD\
&RQWDFWORDG
PD[9$$&'&
IRUWKHUHOD\FRQWDFWV
VKRZQ
WKHUHOD\KDV]HURFXUUHQW
5HOD\ 5HOD\
5HOD\
5HOD\
0
&RQQHFWRU68%')
352),%86'3
5HOD\
*1'
RSWLRQDO
&RQQHFWRU68%'0
352),%863$
'*1'
&1751
&1753GLUHFWLRQFRQWURO
5['7['1$
5['7['3$
3$1
3$3
939
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNVPXVW
EHVKLHOGHGDQGUHVWDJDLQVWWKHHQFORVXUHSRWHQWLDO
ULTRAMAT 6, 19“ unit, pin assignment of AUTOCAL board and PROFIBUS connectors
1/90
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
19" rack unit
&KDQQHO
&KDQQHO
&KDQQHO
6DPSOHJDVRXWOHW
6DPSOHJDVLQOHW
SLQFRQQHFWRU
%LQDU\LQSXWVDQG
DQDORJLQSXWVRXWSXWV
5HIHUHQFHJDVRXWOHW
5HIHUHQFHJDVLQOHW
&KDQQHO
SLQ
FRQQHFWRU
56
3XUJLQJJDV
LQOHW
SLQ
LQWHUIDFH
FRQQHFWRU
RSWLRQHJ
352),%86
SLQFRQQHFWRU
%LQDU\LQSXWV
DQG
UHOD\RXWSXWV
RSWLRQERDUG
3RZHUVXSSO\
DQGIXVHV
*DVFRQQHFWLRQVVWXEVPPRU~
SLQFRQQHFWRU
%LQDU\LQSXWVDQG
UHOD\RXWSXWV
ULTRAMAT 6, 19“ unit, gas and electrical connections (example: 2-channel version)
Siemens AP 01 · 2015
1/91
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
Field device
■ Technical specifications
General information
Gas inlet conditions
Measuring ranges
4, internally and externally switchable; autoranging is also possible
Smallest possible measuring range
Dependent on the application,
e.g. CO: 0 … 10 vpm,
CO2 : 0 … 5 vpm
Largest possible measuring range
Dependent on the application
Measuring range with suppressed
zero point
Any zero point within
0 … 100 vol.% can be
implemented; smallest possible
span 20 %
Permissible sample gas pressure
• With hoses (without pressure
switch)
600 … 1 500 hPa (absolute)
• With pipes (without pressure
switch)
600 … 1 500 hPa (absolute)
- Ex (leakage compensation)
600 … 1 160 hPa (absolute)
- Ex (continuous purging)
600 … 1 500 hPa (absolute)
Purging gas pressure
• Permanent
< 165 hPa above ambient pressure
Heated version
65 °C
Operating position
Front wall, vertical
• For short periods
250 hPa above ambient pressure
Conformity
CE mark in accordance with
EN 50081-1, EN 50082-2
Sample gas flow
18 … 90 l/h (0.3 … 1.5 l/min)
Sample gas temperature
Min. 0 ... max. 50 °C, but above
the dew point, for heated version
min. 0 … max. 80 °C
< 90 % RH (RH: relative humidity)
or dependent on measuring task
Influence of interfering gases must be considered separately
Design, enclosure
Weight
Approx. 32 kg
Sample gas humidity
Degree of protection
IP65 in accordance with
EN 60529, restricted breathing
enclosure to EN 50021
Dynamic response
Warm-up period
At room temperature < 30 min
(the technical specification will be
met after 2 hours)
Delayed display (T90-time)
Dependent on length of analyzer
chamber, sample gas line and
parameterizable damping
Damping (electrical time constant)
0 … 100 s, parameterizable
Approximately 0.5 ... 5 s,
depending on version
<1s
Electrical characteristics
Power supply
100 … 120 V AC (nominal range
of use 90 … 132 V), 48 … 63 Hz
or
200 … 240 V AC (nominal range
of use 180 … 264 V), 48 … 63 Hz
Power consumption
Approx. 35 VA; approx. 330 VA
with heated version
Dead time (purging time of the gas
path in the unit at 1 l/min)
EMC
(Electromagnetic Compatibility)
In accordance with standard
requirements of NAMUR NE21
(08/98)
Time for device-internal signal processing
Electrical safety
In accordance with EN 61010-1
Pressure sensor
• Heated units
Overvoltage category II
• Internal
700 … 1 200 hPa absolute
• Unheated units
Overvoltage category III
• External
700 … 1 500 hPa absolute
• 100 ... 120 V
F3: 1 T/250; F4: 1 T/250
Measuring response (relating to sample gas pressure 1 013 hPa
absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
• 200 ... 240 V
F3: 0.63 T/250; F4: 0.63 T/250
Output signal fluctuation
< ± 1 % of the smallest possible
measuring range according to
rating plate
• 100 ... 120 V
F1: 1 T/250; F2: 4 T/250
F3: 4 T/250; F4: 4 T/250
Zero point drift
< ± 1 % of the current measuring
range/week
• 200 ... 240 V
F1: 0.63 T/250; F2: 2.5 T/250
F3: 2.5 T/250; F4: 2.5 T/250
Measured-value drift
< ± 1 % of the current measuring
range/week
Repeatability
≤ 1 % of the current measuring
range
Detection limit
1 % of the smallest possible measuring range
Linearity error
< 0.5 % of the full-scale value
Fuse values (unheated unit)
Pressure correction range
Fuse values (heated unit)
1/92
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
Influencing variables (relating to sample gas pressure 1 013 hPa
absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature
< 1 % of current measuring
range/10 K (with constant
receiver cell temperature)
Sample gas pressure
When pressure compensation
has been switched on: < 0.15 %
of setpoint/1 % atmospheric pressure change
Sample gas flow
Negligible
Power supply
< 0.1 % of the current measuring
range with rated voltage ± 10 %
Environmental conditions
Application-specific measuring
influences possible if ambient air
contains measured component or
cross interference-sensitive
gases
Electrical inputs and outputs
Analog output
0/2/4 … 20 mA, isolated; load
750 Ω
Relay outputs
6, with changeover contacts,
freely parameterizable, e.g. for
measuring range identification;
load: 24 V AC/DC/1 A, isolated,
non-sparking
Analog inputs
2, dimensioned for 0/2/4 ... 20 mA
for external pressure sensor and
accompanying gas influence correction (correction of cross-interference)
Binary inputs
6, designed for 24 V, isolated,
freely parameterizable, e.g. for
measuring range switchover
Serial interface
RS 485
Options
AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or
PROFIBUS DP
Climatic conditions
Permissible ambient temperature
-30 … +70 °C during storage and
transportation; 5 … 45 °C during
operation
Permissible humidity
< 90 % RH (RH: relative humidity)
within average annual value, during storage and transportation
(dew point must not be undershot)
Siemens AP 01 · 2015
1/93
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
Field device
■ Selection and ordering data
ULTRAMAT 6 gas analyzer
For installation in the field, single-channel, 1 component
Article No.
7MB2111- 7 7 7 7 7 - 7 7 A 7
Cannot be combined
Click on the Article No. for the online configuration in the PIA Life Cycle
Portal.
Gas connections
Ferrule screw connection for pipe, outer diameter 6 mm
Ferrule screw connection for pipe, outer diameter ¼"
0
1
0
1
Measured component
Possible with measuring
range identification
CO
11 ... 30
CO highly selective (with optical filter)
12 ... 30
CO (TÜV; see table "TÜV single component", page 1/99)
A
B
X
CO2
CH4
C2H2
10 ... 30
13 ... 30
15 ... 30
D
E
C2H4
C2H6
C3H6
15 ... 30
14 ... 30
14 ... 30
G
H
C3H8
C4H6
C4H10
13 ... 30
15 ... 30
14 ... 30
K
L
C6H14
SO2 (TÜV; see table "TÜV single component",
page 1/99)
NO (TÜV; see table "TÜV single component",
page 1/99)
14 ... 30
13 ... 30
M
14 ... 20, 22
P
14 ... 30
17 ... 20; 22
(17 to 24, 26; heated)
13 ... 30
Q
NH3 (dry)
H2O
N2O
C
F
J
N
Q
R
R
S
Smallest measuring
range
Largest measuring
range
Measuring range
identification
0 ... 5 vpm
0 ... 10 vpm
0 ... 20 vpm
0 ... 100 vpm
0 ... 200 vpm
0 ... 400 vpm
10
11
12
0 ... 50 vpm
0 ... 100 vpm
0 ... 300 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
13
14
15
D
0 ... 500 vpm
0 ... 1 000 vpm
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
16
17
19
G
0 ... 3 000 vpm
0 ... 5 000 vpm
0 ... 5 000 vpm
0 ... 30 000 vpm
0 ... 15 000 vpm
0 ... 50 000 vpm
19
20
21
K
0 ... 1 %
0 ... 1 %
0 ... 3 %
0 ... 3 %
0 ... 10 %
0 ... 10 %
22
23
24
0 ... 3 %
0 ... 5 %
0 ... 5 %
0 ... 30 %
0 ... 15 %
0 ... 50 %
25
26
27
R
0 ... 10 %
0 ... 10 %
0 ... 30 %
0 ... 30 %
0 ... 100 %
0 ... 100 %
28
29
30
U
1/94
Siemens AP 01 · 2015
A
B
C
E
F
H
J
L
M
N
P
Q
S
T
V
W
A29
A28
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
For installation in the field, single-channel, 1 component
7MB2111- 7 7 7 7 7 - 7 7 A 7
Cannot be combined
Internal gas paths
Sample chamber
(lining)
Reference chamber
(flow-type)
Hose made of FKM
(Viton)
Aluminum
Aluminum
Non-flow-type
Flow-type
0
0
0
1
1
1
Pipe made of titanium
Tantalum1)
Tantalum1)
Non-flow-type
Flow-type
Aluminum
Tantalum1)
Non-flow-type
Non-flow-type
6
8
Stainless steel pipe
(mat. no. 1.4571)
Add-on electronics
Without
AUTOCAL function
• With 8 additional digital inputs/outputs
• With 8 digital inputs/outputs and PROFIBUS PA interface
• With 8 digital inputs/outputs and PROFIBUS DP interface
• With 8 digital inputs/outputs and PROFIBUS PA Ex i
0
A28, A29
2
2
A28, A29, Y02
3
3
6
Y02
A28, A29
8
A28, A29
0
1
6
7
8
6
7
E12
E12
8
Power supply
Standard unit and acc. to ATEX II 3G version (Zone 2)
• 100 ... 120 V AC, 48 ... 63 Hz
• 200 ... 240 V AC, 48 ... 63 Hz
ATEX II 2G versions (Zone 1), incl. certificate
• 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: leakage compensation)
• 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: leakage compensation)
• 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: continuous purging)
• 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: continuous purging)
Heating of internal gas paths and analyzer unit
Without
With (max. 65 °C)
Language (supplied documentation, software)
German
English
French
Spanish
Italian
1)
Only for cell length 20 to 180 mm
2)
Only in connection with an approved purging unit
0
1
0
1
2
2
2
3
3
3
6
6
6
7
7
7
A
B
0
1
2
3
4
Siemens AP 01 · 2015
1/95
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
Field device
■ Selection and ordering data
Additional versions
Order code
Add "-Z" to Article No. and specify Order codes.
Flow-type reference cell with reduced flow, 6 mm
A28
Flow-type reference cell with reduced flow, ¼"
A29
TAG labels (specific lettering based on customer information)
B03
Kalrez gaskets in sample gas path
B04
SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511
C20
Ex versions
Possible combinations: see: Table "Ex configurations – principle selection criteria", page 5/16
ATEX II 3G certificate; restricted breathing enclosure, non-flammable gases
E11
ATEX II 3G certificate; flammable gases
E12
FM/CSA certificate – Class I Div 2
E20
ATEX II 3D certificate; potentially explosive dust atmospheres
• In non-hazardous gas zone
E40
• In Ex zone acc. to ATEX II 3G, non-flammable gases
E41
• In Ex zone acc. to ATEX II 3G, flammable gases1)
E42
BARTEC EEx p control unit "Leakage compensation"
E71
BARTEC EEx p control unit "Continuous purging"
E72
Clean for O2 service (specially cleaned gas path)
Y02
Measuring range indication in plain text, if different from the standard setting
Y11
Special setting (only in conjunction with an application no., e.g. extended measuring range)
Y12
Extended special setting
(only in conjunction with an application no., e.g. determination of cross-interferences)
Y13
TÜV version acc. to 13th and 17th BlmSchV
Y17
Additional units for Ex versions
Article No.
Category ATEX II 2G (zone 1)
BARTEC EEx p control unit, 230 V, "leakage compensation"
7MB8000-2BA
BARTEC EEx p control unit, 115 V, "leakage compensation"
7MB8000-2BB
BARTEC EEx p control unit, 230 V, "continuous purging"
7MB8000-2CA
BARTEC EEx p control unit, 115 V, "continuous purging"
7MB8000-2CB
Ex isolation amplifier
7MB8000-3AB
Ex isolating relay, 230 V
7MB8000-4AA
Ex isolating relay, 110 V
7MB8000-4AB
Differential pressure switch for corrosive and non-corrosive gases
7MB8000-5AA
Stainless steel flame arrestor
7MB8000-6BA
Hastelloy flame arrestor
7MB8000-6BB
Category ATEX II 3G (Zone 2)
BARTEC EEx p control unit, 230 V, "continuous purging"
7MB8000-2CA
BARTEC EEx p control unit, 115 V, "continuous purging"
7MB8000-2CB
FM/CSA (Class I Div. 2)
Ex purging unit MiniPurge FM
7MB8000-1AA
Accessories
Article No.
RS 485/Ethernet converter
A5E00852383
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
AUTOCAL function with 8 digital inputs/outputs
A5E00064223
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA
A5E00057315
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP
A5E00057318
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA Ex i (firmware 4.1.10 required)
A5E00057317
Set of Torx screwdrivers
A5E34821625
1)
Only in connection with an approved purging unit
1/96
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
■ Selection and ordering data
Article No.
ULTRAMAT 6 gas analyzer
For installation in the field, single-channel, 2 components
7MB2112- 7 7 7 7 7 - 7 7 A 7
Cannot be combined
Click on the Article No. for the online configuration in the PIA Life Cycle
Portal.
Gas connections
Ferrule screw connection for pipe, outer diameter 6 mm
Ferrule screw connection for pipe, outer diameter ¼"
0
1
0
1
Measured component
Smallest measuring
range
Largest measuring range
CO
NO
0 ... 100 vpm
0 ... 100 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
AA
CO
NO
0 ... 300 vpm
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
AB
CO
0 ... 1 000 vpm
0 ... 10 000 vpm
NO
0 ... 1 000 vpm
0 ... 10 000 vpm
For CO/NO (TÜV; see table "TÜV, 2 components in series", page 1/99)
AC
CO2
CO
0 ... 100 vpm
0 ... 100 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
BA
CO2
CO
0 ... 300 vpm
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 3 000 vpm
BB
CO2
CO
0 ... 1 000 vpm
0 ... 1 000 vpm
0 ... 10 000 vpm
0 ... 10 000 vpm
BC
CO2
CO
0 ... 3 000 vpm
0 ... 3 000 vpm
0 ... 30 000 vpm
0 ... 30 000 vpm
BD
CO2
CO
0 ... 1 %
0 ... 1 %
0 ... 10 %
0 ... 10 %
BE
CO2
CO
0 ... 3 %
0 ... 3 %
0 ... 30 %
0 ... 30 %
BF
CO2
CO
0 ... 10 %
0 ... 10 %
0 ... 100 %
0 ... 100 %
BG
CO2
CH4
0 ... 10 %
0 ... 10 %
0 ... 100 %
0 ... 100 %
CG
CO2
NO
0 ... 100 vpm
0 ... 100 vpm
0 ... 1 000 vpm
0 ... 1 000 vpm
DA
CO2
NO
Internal gas paths
0 ... 300 vpm
0 ... 3 000 vpm
0 ... 300 vpm
0 ... 3 000 vpm
Sample chamber (lining) Reference chamber
(flow-type)
Aluminum
Non-flow-type
Aluminum
Flow-type
Hose made of FKM
(Viton)
Pipe made of titanium
Tantalum1)
Tantalum1)
Aluminum
Tantalum1)
Non-flow-type
Flow-type
Non-flow-type
Non-flow-type
Stainless steel pipe
(mat. no. 1.4571)
Add-on electronics
Without
AUTOCAL function
• With 8 additional digital inputs/outputs
• With 8 digital inputs/outputs and PROFIBUS PA interface
• With 8 digital inputs/outputs and PROFIBUS DP interface
• With 8 digital inputs/outputs and PROFIBUS PA Ex i
Power supply
Standard unit and acc. to ATEX II 3G version (Zone 2)
• 100 ... 120 V AC, 48 ... 63 Hz
• 200 ... 240 V AC, 48 ... 63 Hz
ATEX II 2G versions (Zone 1), incl. certificate
• 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: leakage compensation)
• 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: leakage compensation)
• 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: continuous purging)
• 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2)
(operating mode: continuous purging)
Heating of internal gas paths and analyzer unit
none
With (max. 65 °C)
A29
A28
DB
0
1
0
1
2
3
6
8
0
A28, A29
2
3
6
8
A28, A29, Y02
Y02
A28, A29
A28, A29
0
1
6
7
8
6
7
8
0
1
0
1
2
2
2
3
3
3
6
6
6
7
7
7
A
B
Siemens AP 01 · 2015
1/97
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
Field device
■ Selection and ordering data
ULTRAMAT 6 gas analyzer
For installation in the field, single-channel, 2 components
Article No.
7MB2112- 7 7 7 7 7 - 7 7 A 7
Cannot be combined
Language (supplied documentation, software)
German
0
English
1
French
2
Spanish
3
Italian
4
1)
Only for cell length 20 to 180 mm.
2)
See also next page "Additional units for Ex versions".
Additional versions
Add "-Z" to Article No. and specify Order codes.
Flow-type reference cell with reduced flow, 6 mm
Flow-type reference cell with reduced flow, ¼"
TAG labels (specific lettering based on customer information)
Kalrez gaskets in sample gas path
SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511
Ex versions
Possible combinations: see: Table "Ex configurations – principle selection criteria", page 5/16
ATEX II 3G certificate; restricted breathing enclosure, non-flammable gases
ATEX II 3G certificate; flammable gases
CSA certificate – Class I Div 2
ATEX II 3D certificate; potentially explosive dust atmospheres
• In non-hazardous gas zone
• In Ex zone acc. to ATEX II 3G, non-flammable gases
• In Ex zone acc. to ATEX II 3G, flammable gases
BARTEC EEx p control unit "Leakage compensation"
BARTEC EEx p control unit "Continuous purging"
Order code
A28
A29
B03
B04
C20
E11
E12
E20
E40
E41
E42
E71
E72
Clean for O2 service (specially cleaned gas path)
Measuring range indication in plain text, if different from the standard setting
Special setting (only in conjunction with an application no., e.g. extended measuring range)
Extended special setting
(only in conjunction with an application no., e.g. determination of cross-interferences)
TÜV version acc. to 13th and 17th BlmSchV
Y02
Y11
Y12
Y13
Additional units for Ex versions
Category ATEX II 2G (zone 1)
BARTEC EEx p control unit, 230 V, "leakage compensation"
BARTEC EEx p control unit, 115 V, "leakage compensation"
BARTEC EEx p control unit, 230 V, "continuous purging"
BARTEC EEx p control unit, 115 V, "continuous purging"
Ex isolation amplifier
Ex isolating relay, 230 V
Ex isolating relay, 110 V
Differential pressure switch for corrosive and non-corrosive gases
Stainless steel flame arrestor
Hastelloy flame arrestor
Category ATEX II 3G (Zone 2)
BARTEC EEx p control unit, 230 V, "continuous purging"
BARTEC EEx p control unit, 115 V, "continuous purging"
FM/CSA (Class I Div. 2)
Ex purging unit MiniPurge FM
Article No.
Accessories
RS 485/Ethernet converter
RS 485/RS 232 converter
RS 485/USB converter
AUTOCAL function with 8 digital inputs/outputs
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP
AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA Ex i (firmware 4.1.10 required)
Set of Torx screwdrivers
Article No.
A5E00852383
C79451-Z1589-U1
A5E00852382
A5E00064223
A5E00057315
A5E00057318
A5E00057317
A5E34821625
1/98
Siemens AP 01 · 2015
Y17
7MB8000-2BA
7MB8000-2BB
7MB8000-2CA
7MB8000-2CB
7MB8000-3AB
7MB8000-4AA
7MB8000-4AB
7MB8000-5AA
7MB8000-6BA
7MB8000-6BB
7MB8000-2CA
7MB8000-2CB
7MB8000-1AA
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
TÜV, single component
(only with additional suffix Z (Y17, Y18))
Component
CO (TÜV)
Measuring range
identification
Smallest
measuring range
from 0 to ...
SO2 (TÜV)
Largest
measuring range
from 0 to ...
C
D
50
mg/m3
1 000
mg/m3
E
F
300 mg/m3
3 000 mg/m3
G
mg/m3
mg/m3
H
K
500
1 000
mg/m3
3 000
mg/m3
10
g/m3
R
30
g/m3
V
100 g/m3
P
5 000
10 000
mg/m3
30 000
mg/m3
100
g/m3
300
g/m3
Largest
measuring range
from 0 to ...
75 mg/m3
1 500 mg/m3
Smallest
measuring range
from 0 to ...
Largest
measuring range
from 0 to ...
mg/m3
3 000 mg/m3
500 mg/m3
5 000 mg/m3
100 mg/m3
2 000 mg/m3
1 000 mg/m3
10 000 mg/m3
300 mg/m3
3 000 mg/m3
mg/m3
5 000 mg/m3
300
500
3 000
mg/m3
10
g/m3
30
g/m3
100
1 160 g/m3
NO (TÜV)
Smallest
measuring range
from 0 to ...
g/m3
300 g/m3
30 000
mg/m3
100
g/m3
300
g/m3
1 000
g/m3
2 630 g/m3
1 000
mg/m3
10 000 mg/m3
3 000
mg/m3
30 000 mg/m3
10
g/m3
100 g/m3
30
g/m3
300 g/m3
100 g/m3
1 250 g/m3
Example for ordering
ULTRAMAT 6, TÜV (1-component unit)
Component: CO
Measuring range: 0 to 50 / 1 000 mg/m3
with hoses, non-flow-type reference compartment
without automatic adjustment (AUTOCAL)
230 V AC; without heating, German
7MB2111-0XD00-1AA0-Z +Y17
TÜV, 2 components in series
Component
CO (TÜV)
Measuring range
identification
Smallest measuring range
from 0 to ...
Largest measuring range
from 0 to ...
Smallest measuring range
from 0 to ...
Largest measuring range
from 0 to ...
AA
75 mg/m3
1 000 mg/m3
200 mg/m3
2 000 mg/m3
mg/m3
mg/m3
3 000 mg/m3
mg/m3
AB
300
AC
1 000 mg/m3
NO (TÜV)
3 000
10 000 mg/m3
300
1 000 mg/m3
10 000 mg/m3
Example for ordering
ULTRAMAT 6, TÜV (2 components in series)
Components: CO/NO
Measuring range CO: 0 to 75 / 1 000 mg/m3, NO: 0 to 200 / 2 000 mg/m3
with hoses, non-flow-type reference compartment
without automatic adjustment (AUTOCAL)
230 V AC; without heating, German
7MB2112-0AA00-1AA0-Z +Y17
Note: for 3 components take both tables into consideration.
Ordering information measured component N2O
Certification in accordance with AM0028 and AM0034 (Kyoto Protocol) for measuring N2O,
measuring range 0 to 300 ppm / 3 000 ppm.
Version: Standard device
Siemens AP 01 · 2015
1/99
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
418
394
250
259
184
148
112
77
42
20
■ Dimensional drawings
0
271
253
237
222
201
127
65
58
311
438
328
26.5
14
ULTRAMAT 6
14
ULTRAMAT 6, field unit, dimensions in mm
1/100
Siemens AP 01 · 2015
480
s
440
1
20
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
■ Schematics
Pin assignment (electrical and gas connections)
&RQQHFWRU68%')56
0
*1'
*1'
1&
5'7'1
5'7'3
5B/HYHO1
9
0
,WLVSRVVLEOHWRFRQQHFWEXVWHUPLQDWLQJ
UHVLVWRUVWRSLQVDQG
5B/HYHO3
1&
1&
*1'
7HUPLQDOEORFN$
0
0
0
0
0
Ω
0
*1'
5HOD\
5HOD\
5HOD\
5HOD\
&RQWDFWORDGPD[
9$$&'&UHOD\FRQWDFWV
VKRZQUHOD\FRLOKDV]HURFXUUHQW
5HOD\
5HOD\
%LQDU\LQSXWWR1
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
,VRODWHGYLDRSWRFRXSOHU
99
99
*1'
*1'
%LQDU\LQSXWWR1
%LQDU\LQSXW3
%LQDU\LQSXW3
$QDORJLQSXW1
$QDORJLQSXW3
$QDORJLQSXW1
$QDORJLQSXW3
,VRODWHGYLDRSWRFRXSOHU
99
99
$QDORJLQSXWV
&RUUHFWLRQRI
QRQLVRODWHG
FURVVLQWHUIHUHQFH
P$RU9
3UHVVXUHFRUUHFWLRQ
LQWHUQDOUHVLVWDQFHืΩ
1&
$QDORJRXWSXW1
$QDORJRXWSXW3
$QDORJRXWSXW
$QDORJRXWSXW
*1'
&RPSRQHQW
&RPSRQHQW
LIDYDLODEOH
$QDORJRXWSXWVLVRODWHG
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNVPXVWEH
VKLHOGHGDQGUHVWDJDLQVWWKHHQFORVXUHSRWHQWLDO
ULTRAMAT 6, field device, pin and terminal assignment
Siemens AP 01 · 2015
1/101
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
1
Field device
7HUPLQDOEORFN%RSWLRQ
0
0
&RQQHFWRU68%');
352),%86'3
*1'
5HOD\
5HOD\
5HOD\
5HOD\
&RQWDFWORDG
PD[9$$&'&
5HOD\
IRUVKRZQ
UHOD\FRQWDFWSRVLWLRQ
WKHUHOD\KDV]HURFXUUHQW
5HOD\
5HOD\
5HOD\
1&
%LQDU\LQSXWWR1
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
1&
*1'
1&
1&
RSWLRQDO
'*1'
&1751
&1753GLUHFWLRQFRQWURO
5['7['1$
5['7['3%
&RQQHFWRU68%'0;
352),%863$
3$1
3$3
939
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNVPXVW
EHVKLHOGHGDQGUHVWDJDLQVWWKHHQFORVXUHSRWHQWLDO
ULTRAMAT 6, field device, pin and terminal assignment of the AUTOCAL board and PROFIBUS connectors
1/102
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Field device
a
b
c
d
7
5
1
3
2
4
9
e
8
6
Gas connections
Electrical connections
1
Sample gas inlet
2
Sample gas outlet
3
Reference gas inlet (option)
4
Reference gas outlet (option)
Clamping
gland for pipe
Ø 6 mm or ¼"
a-c
Signal cable (Ø 10 ... 14 mm)
(analog + digital): cable gland M20x1.5
d
Interface connection: (Ø 7 ... 12 mm)
cable gland M20x1.5
e
Power supply: (Ø 7 ... 12 mm)
cable gland M20x1.5
5 - 8 Purging gas inlets/outlets, stubs Ø 10 mm or 3/8"
9
Connection atmospheric pressure sensor, stubs Ø ¼"
ULTRAMAT 6, field device, gas connections and electrical connections
Documentation
DDocumentationocumentation
■ Selection and ordering data
Operating instructions
Article No.
ULTRAMAT 6 / OXYMAT 6
Gas analyzer for IR-absorbing gases
and oxygen
• German
C79000-G5200-C143
• English
C79000-G5276-C143
• French
C79000-G5277-C143
• Spanish
C79000-G5278-C143
• Italian
C79000-G5272-C143
Siemens AP 01 · 2015
1/103
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
ULTRAMAT 6
Suggestions for spare parts
7MB-2124
7MB-2111
7MB-2112
7MB-2111/2 Ex
2 years
(quantity)
7MB-2123
■ Selection and ordering data
7MB-2121
1
O-ring for cover (window)
x
x
x
x
x
x
2
4
C79121-Z100-A24
Cover (cell length 20 ... 180 mm)
x
x
x
x
x
x
2
2
C79451-A3462-B151
Cover (cell length 0.2 ... 6 mm)
x
x
x
x
x
x
2
2
C79451-A3462-B152
O-rings, set
x
x
x
x
x
x
1
C79451-A3462-D501
x
x
x
2
4
C71121-Z100-A159
Pressure switch
x
x
x
1
2
C79302-Z1210-A2
Flow indicator
x
x
x
1
Hose clip
x
x
x
Description
5 years
(quantity)
Article No.
Analyzer unit
Sample gas path
O-ring (hose clip)
Heating cartridge (heated unit)
2
C79402-Z560-T1
x
x
x
1
C79451-A3478-C9
x
x
x
1
W75083-A1004-F120
x
x
1
W75054-T1001-A150
2
A5E00061505
Electronics
Temperature fuse (heated unit)
Fuse (device fuse)
x
1
Temperature controller electronics, 230 V AC
x
x
x
1
A5E00118527
Temperature controller electronics, 115 V AC
x
x
x
1
A5E00118530
x
x
x
1
A5E00302916
1
C79165-A3042-B504
1
C79165-A3044-B176
1
C79451-A3474-B605
Fan, 24 V DC (heated unit)
Front plate with keyboard
x
x
x
Temperature sensor
1
x
x
x
Adapter plate, LCD/keyboard
x
x
x
x
x
Motherboard, with firmware: see
spare parts list
x
x
x
x
x
LC display
x
x
x
x
x
Connector filter
x
x
x
x
x
Fusible element, T 0.63 A/250 V
x
x
x
x
x
Fusible element, T 1 A/250 V
x
x
x
x
x
x
x
x
Fusible element, T 1.6 A/250 V
Fusible element, T 2.5 A/250 V
x
x
1
x
1
1
x
1
W75025-B5001-B1
1
W75041-E5602-K2
2
3
W79054-L1010-T630
2
3
W79054-L1011-T100
2
3
W79054-L1011-T160
2
3
W79054-L1011-T250
If the ULTRAMAT 6 is supplied with a specially cleaned gas path for high oxygen content ("Cleaned for O2 service"), please ensure
that you specify this when ordering spare parts. This is the only way to guarantee that the gas path will continue to comply with the
special requirements for this version.
1/104
Siemens AP 01 · 2015