© Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 General information ■ Overview Special versions Special applications Besides the standard combinations, special applications concerning material in the gas path, material in the sample cells (e.g. Titan, Hastelloy C22) and measured components are also available on request TÜV version/QAL The ULTRAMAT 6 single-channel or dual-channel gas analyzers operate according to the NDIR two-beam alternating light principle and measure gases highly selectively whose absorption bands lie in the infrared wavelength range from 2 to 9 μm, such as CO, CO2, NO, SO2, NH3, H2O as well as CH4 and other hydrocarbons. Single-channel analyzers can simultaneously measure up to 2 gas components, while dual-channel analyzers can simultaneously measure 3 (or 4 on request) gas components. ■ Benefits • High selectivity with double-layer detector and optical coupler - Reliable measurements even in complex gas mixtures • Low detection limits - Measurements with low concentrations • Corrosion-resistant materials in gas path (option) - Measurement possible in highly corrosive sample gases • Analyzer cells can be cleaned as required on site - Cost savings due to reuse after contamination • Electronics and physics: gas-tight isolation, purging is possible, IP65 - Long service life even in harsh environments • Heated versions (option) - Use also in presence of gases condensing at low temperature • EEx(p) for zones 1 and 2 (according to ATEX 2G and ATEX 3G) ■ Application Areas of application • Measurement for boiler control in incineration plants • Emission measurements in incineration plants • Measurement in the automotive industry (test benches) • Warning equipment • Process gas concentrations in chemical plants • Trace measurements in pure gas processes • Environmental protection • TLV (Threshold Limit Value) monitoring at the workplace • Quality monitoring • Ex versions for analyzing flammable and non-flammable gases or vapors for use in hazardous areas TÜV-approved versions are available for measurement of CO, NO and SO2 according to 13th and 17th BlmSchV and TA Luft. Smallest TÜV-approved and permitted measuring ranges: • 1-component analyzer CO: 0 to 50 mg/m3 NO: 0 to 100 mg/m3 SO2: 0 to 75 mg/m3 • 2-component analyzer (series connection) CO: 0 to 75 mg/m3 NO: 0 to 200 mg/m3. Furthermore, the TÜV-approved versions of the ULTRAMAT 6 comply with the requirements of EN 14956 and QAL 1 in accordance with EN 14181. Conformity of the analyzers with both standards is TÜV-certified. The analyzer drift can be determined in accordance with EN 14181 (QAL 3) either manually or with a PC using the SIPROM GA maintenance and servicing software. In addition, selected manufacturers of emission evaluation computers offer the possibility for downloading the drift data via the analyzer’s serial interface and to automatically record and process it in the evaluation computer. Flow-type reference compartment • The flow through the reference compartment should be adapted to the sample gas flow • The gas supply of the reduced flow-type reference compartment should have an upstream pressure of 3 000 to 5 000 hPa (abs.). Then a restrictor will automatically adjust the flow to approximately 8 ml/min ■ Design 19" rack unit • 19" rack unit with 4 HU for installation - in hinged frame - in cabinets with or without telescopic rails • Front plate for service purposes can be pivoted down (laptop connection) • Internal gas paths: hose made of FKM (Viton) or pipe made of titanium or stainless steel • Gas connections for sample gas inlet and outlet: pipe diameter 6 mm or 1/4" • Flow indicator for sample gas on front plate (option) • Pressure switch in sample gas path for flow monitoring (option) Field device • Two-door enclosure with gas-tight separation of analyzer and electronics sections from gas path • Individually purgeable enclosure halves • Parts in contact with sample gas can be heated up to 65 °C (option) • Gas path: hose made of FKM (Viton) or pipe made of titanium or stainless steel (further materials possible as special applications) • Gas connections for sample gas inlet and outlet: pipe union for pipe diameter 6 mm or 1/4" • Purging gas connections: pipe diameter 10 mm or 3/8" Siemens AP 01 · 2015 1/71 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 General information Display and control panel • Large LCD field for simultaneous display of: - Measured value (digital and analog displays) - Status bar - Measuring ranges • Contrast of the LCD field adjustable via the menu • Washable membrane keyboard with five softkeys • Menu-driven operator control for parameterization, test functions, adjustment • Operator support in plain text • Graphic display of concentration trend; programmable time intervals • Bilingual operating software: German/English, English/Spanish, French/English, Spanish/English, Italian/English Input and outputs • One analog output per medium (from 0, 2, 4 to 20 mA; NAMUR parameterizable) • Two analog inputs freely configurable (e.g. correction of cross-interferences or external pressure sensor) LED backlit graphic display and membrane keyboard with noticeable click • Six binary inputs freely configurable (e.g. measurement range changeover, processing of external signals from the sample preparation) • Six relay outputs freely configurable e.g. for fault, maintenance request, limit alarm, external solenoid valves) • Expansion by eight additional binary inputs and eight additional relay outputs e.g. for autocalibration with up to four test gases Communication RS 485 present in the basic unit (connection at the rear; for the rack unit also behind the front plate). Options • AK interface for the automotive industry with extended functions • RS 485/RS 232 converter • RS 485/Ethernet converter • RS 485/USB converter • Connection to networks via PROFIBUS DP/PA interface • SIPROM GA software as the service and maintenance tool Status line for display of analyzer status (programmable) Two code levels according to NAMUR (maintenance and specialist level) Easy operation with menu control using five softkeys Display of concentrations as numbers and bargraph for channel 1 Display of current measuring ranges Display of concentrations as numbers and bargraph for channel 2 Status line for channel 2 to display the unit status (programmable) Display of start-of-scale and full-scale values ESC key to abort inputs Keyboard to enter values INFO key for help in plain text CLEAR key to delete inputs ENTER key to accept input values ULTRAMAT 6, membrane keyboard and graphic display 1/72 Siemens AP 01 · 2015 MEAS key to return to measurement mode © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 General information Designs – Parts wetted by sample gas, standard Gas path With hoses 19" rack unit Field device Bushing Stainless steel, mat. no. 1.4571 Hose FKM (e.g. Viton) Field device Ex - Sample chamber: With pipes • Body Aluminum • Lining Aluminum • Fitting Stainless steel, mat. no. 1.4571, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) • Window CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Bushing Titanium Pipe Titanium, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Sample chamber: With pipes • Body Aluminum • Lining Tantalum (only for cell length 20 mm to 180 mm) • Window CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Bushing Stainless steel, mat. no. 1.4571 Pipe Stainless steel, mat. no. 1.4571, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Sample chamber: • Body Aluminum • Lining Aluminum or tantalum (tantalum only for cell length 20 mm to 180 mm) • Window CaF2, adhesive: E353, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Options Gas path Flow indicator Pressure switch 19" rack unit Field device Field device Ex Measurement pipe Duran glass - - Variable area Duran glass Suspension boundary PTFE (Teflon) Angle pieces FKM (e.g. Viton) Membrane FKM (e.g. Viton) - - Enclosure PA 6.3T Field device Field device Ex Versions – Parts wetted by sample gas, special applications (examples) Gas path With pipes 19" rack unit Bushing e.g. Hastelloy C22 Pipe e.g. Hastelloy C22, O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Sample chamber: • Body e.g. Hastelloy C22 • Window CaF2, without adhesive O-ring: FKM (e.g. Viton) or FFKM (Kalrez) Siemens AP 01 · 2015 1/73 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 General information Gas path (19" rack unit) Legend for the gas path figures 1 Sample gas inlet channel 1 8 Restrictor 2 Sample gas outlet channel 1 9 Purge gas inlet 3 Reference gas outlet (option) 10 Gas inlet atmospheric pressure sensor 4 Reference gas inlet (option) 11 Atmospheric pressure sensor 5 Sample gas inlet channel 2 12 Flow indicator in sample gas path (option) 6 Sample gas outlet channel 2 13 Pressure switch in sample gas path (option) 7 IR physical system 9 10 2 1 4 3 7 8 P P 11 F 13 12 Gas path ULTRAMAT 6, single-channel unit, 19" unit, with flow-type reference cell (option) &KDQQHO &KDQQHO 3 3 Gas path ULTRAMAT 6, dual-channel unit, 19" unit 1/74 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 General information Gas path (field device) Legend for the gas path figures 1 Sample gas inlet 7 Purging gas outlet (analyzer side) 2 Sample gas outlet 8 Purging gas inlet (analyzer side) 3 Reference gas inlet (option) 9 Connection of atmospheric pressure sensor 4 Reference gas outlet (option) 10 IR physical system 5 Purging gas inlet (electronics side) 11 Atmospheric pressure sensor 6 Purging gas outlet (electronics side) 3 (OHFWURQLFVLGH $QDO\]LQJVLGH Gas path ULTRAMAT 6, field unit, with flow-type reference cell (option) Siemens AP 01 · 2015 1/75 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 General information ■ Function Principle of operation The ULTRAMAT 6 gas analyzer operates according to the infrared two-beam alternating light principle with double-layer detector and optical coupler. The measuring principle is based on the molecule-specific absorption of bands of infrared radiation. The absorbed wavelengths are characteristic to the individual gases, but may partially overlap. This results in cross-sensitivities which are reduced to a minimum in the ULTRAMAT 6 gas analyzers by the following measures: • Gas-filled filter cell (beam divider) • Double-layer detector with optical coupler • Optical filters if necessary The figure shows the measuring principle. An IR source (1) which is heated to approx. 700 ºC and which can be shifted to balance the system is divided by the beam divider (3) into two equal beams (sample and reference beams). The beam divider also acts as a filter cell. The reference beam passes through a reference cell (8) filled with N2 (a non-infrared-active gas) and reaches the right-hand side of the detector (11) practically unattenuated. The sample beam passes through the sample chamber (7) through which the sample gas flows and reaches the left-hand side of the detector (10) attenuated to a lesser or greater extent depending on the concentration of the sample gas. The detector is filled with a defined concentration of the gas component to be measured. The detector is designed as a double-layer detector. The center of the absorption band is preferentially absorbed in the upper detector layer, the edges of the band are absorbed to approximately the same extent in the upper and lower layers. The upper and lower detector layers are connected together via the microflow sensor (12). This coupling means that the spectral sensitivity has a very narrow band. 1 2 3 4 5 6 7 8 9 11 10 The optical coupler (13) lengthens the lower receiver cell layer optically. The infrared absorption in the second detector layer is varied by changing the slider position (14). It is thus possible to individually minimize the influence of interfering components. 12 A chopper (5) rotates between the beam divider and the sample chamber and interrupts the two beams alternately and periodically. If absorption takes place in the sample chamber, a pulsating flow is generated between the two detector levels which is converted by the microflow sensor (12) into an electric signal. The microflow sensor consists of two nickel-plated grids heated to approximately 120 ºC, which, along with two supplementary resistors, form a Wheatstone bridge. The pulsating flow together with the dense arrangement of the Ni grids causes a change in resistance. This leads to an offset in the bridge, which is dependent on the concentration of the sample gas. Notes The sample gases must be fed into the analyzers free of dust. Condensation should be prevented from occurring in the sample chambers. Therefore, the use of gas modified for the measuring task is necessary in most application cases. As far as possible, the ambient air of the analyzer should not have a large concentration of the gas components to be measured. Flow-type reference sides with reduced flow must not be operated with flammable or toxic gases. Flow-type reference sides with reduced flow and an O2 content > 70 % may only be used together with Y02 (Clean for O2). 1/76 Siemens AP 01 · 2015 13 14 1 2 3 4 5 6 7 IR source, adjustable Optical filter Beam divider Eddy current drive Chopper Sample gas inlet Sample cell 8 9 10 11 12 13 14 Reference cell Sample gas outlet Detector, meas. side Detector, reference side Microflow sensor Optical coupler Slider, adjustable ULTRAMAT 6, principle of operation Channels with electronically suppressed zero point only differ from the standard version in the measuring range parameterization. Physically suppressed zeros can be provided as a special application. © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 General information Essential characteristics • Dimension of measured value freely selectable (e.g. vpm, mg/m3) • Four freely-parameterizable measuring ranges per component • Measuring ranges with suppressed zero point possible • Measuring range identification • Galvanically isolated signal output 0/2/4 to 20 mA per component • Automatic or manual measuring range switchover selectable; remote switching is also possible • Differential measuring ranges with flow-type reference cell • Storage of measured values possible during adjustments • Time constants selectable within wide limits (static/dynamic noise suppression); i.e. the response time of the analyzer or component can be matched to the respective measuring task • Short response time • Low long-term drift • Measuring point switchover for up to 6 measuring points (programmable) • Measuring point identification • Monitoring of sample gas flow (option) • Internal pressure sensor for correction of variations in atmospheric pressure in the range 700 to 1 200 hPa absolute • External pressure sensor can be connected for correction of variations in the process gas pressure in the range 700 to 1 500 hPa absolute (option) • Two control levels with separate authorization codes to prevent unintentional and unauthorized inputs • Automatic, parameterizable measuring range calibration • Simple handling using a numerical membrane keyboard and operator prompting • Operation based on NAMUR recommendation • Customer-specific analyzer options such as: - Customer acceptance - TAG labels - Drift recording • Easy device replacement since electric connections can be simply disconnected from the device • Sample chambers for use in presence of highly corrosive sample gases (e.g. tantalum layer or Hastelloy C22) Additional features, dual-channel version • Separate design of physical unit, electronics, inputs/outputs and power supply for each channel • Display and operation via common LCD panel and keyboard • Measurement channels 1 and 2 can be converted to series connection (linking of gas connections from channel 1 to channel 2 on rear) Siemens AP 01 · 2015 1/77 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit ■ Technical specifications Pressure correction range General information Measuring ranges 4, internally and externally switchable; autoranging is also possible Pressure sensor • Internal 700 … 1 200 hPa absolute Smallest possible measuring range Dependent on the application: e.g. CO: 0 … 10 vpm, CO2: 0 … 5 vpm • External 700 … 1 500 hPa absolute Largest possible measuring span Dependent on the application Measuring range with suppressed zero point Any zero point within 0 … 100 vol.% can be implemented; smallest possible span 20 % Operating position Front wall, vertical Conformity CE mark in accordance with EN 50081-1, EN 50082-2 Influence of interfering gases must be considered separately Design, enclosure Approx. 15 kg (with one IR channel) Approx. 21 kg (with two IR channels) Weight Degree of protection IP20 according to EN 60529 Electrical characteristics EMC (Electromagnetic Compatibility) In accordance with standard requirements of NAMUR NE21 (08/98) Electrical safety According to EN 61010-1, overvoltage category III Power supply 100 … 120 V AC (nominal range of use 90 … 132 V), 48 … 63 Hz or 200 … 240 V AC (nominal range of use 180 … 264 V), 48 … 63 Hz Power consumption 1-channel unit: Approx. 40 VA 2-channel unit: Approx. 70 VA Fuse values • 100 ... 120 V 1 T/250 (7MB2121) 1.6 T/250 (7MB2123) • 200 ... 240 V 0.63 T/250 (7MB2121) 1 T/250 (7MB2123) Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) Output signal fluctuation < ± 1 % of the smallest possible measuring range according to rating plate Zero point drift < ± 1 % of the current measuring range/week Measured-value drift < ± 1 % of the current measuring range/week Repeatability ≤ 1 % of the current measuring range Detection limit 1 % of the smallest possible measuring range Linearity error < 0.5 % of the full-scale value Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) Ambient temperature < 1 % of current measuring range/10 K (with constant receiver cell temperature) Sample gas pressure • When pressure compensation has been switched on: < 0.15 % of the span/1 % change in atmospheric pressure • When pressure compensation has been switched off: < 1.5 % of the span/1 % change in atmospheric pressure Sample gas flow Negligible Power supply < 0.1 % of the current measuring range with rated voltage ± 10 % Environmental conditions Application-specific measuring influences possible if ambient air contains measured components or cross interference-sensitive gases Electrical inputs and outputs Analog output 0/2/4 ... 20 mA, isolated; load ≤ 750 Ω Relay outputs 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking Analog inputs 2, dimensioned for 0/2/4 ... 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference) Binary inputs 6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover Dynamic response Serial interface RS 485 Warm-up period At room temperature < 30 min (the technical specification will be met after 2 hours) Options Delayed display (T90-time) Dependent on length of analyzer chamber, sample gas line and parameterizable damping AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP Damping (electrical time constant) 0 … 100 s, parameterizable Dead time (purging time of the gas path in the unit at 1 l/min) Approximately 0.5 ... 5 s, depending on version Time for device-internal signal processing <1s Gas inlet conditions Permissible sample gas pressure • With hoses - Without pressure switch 600 … 1 500 hPa (absolute) - With pressure switch 700 … 1 300 hPa (absolute) • With pipes (without pressure switch) 600 … 1 500 hPa (absolute) Sample gas flow 18 … 90 l/h (0.3 … 1.5 l/min) Sample gas temperature Min. 0 ... max. 50 °C, but above the dew point Sample gas humidity < 90 % RH (relative humidity), or dependent on measuring task, non-condensing 1/78 Siemens AP 01 · 2015 Climatic conditions Permissible ambient temperature -30 ... +70 °C during storage and transportation, 5 ... 45 °C during operation Permissible humidity < 90 % RH (relative humidity) as annual average, during storage and transportation (dew point must not be undershot) © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer Single-channel 19" rack unit for installation in cabinets 7MB2121- 7 7 7 7 7 - 7 A A 7 Cannot be combined Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections for sample gas and reference gas Pipe with 6 mm outer diameter Pipe with ¼" outer diameter 0 1 0 1 Possible with measuring range identification CO 11 ... 30 CO highly selective (with optical filter) 12 ... 30 CO (TÜV; see table "TÜV single component", page 1/87) A B X CO2 CH4 C2H2 10 ... 30 13 ... 30 15 ... 30 C D E C2H4 C2H6 C3H6 15 ... 30 14 ... 30 14 ... 30 F G H C3H8 C4H6 C4H10 13 ... 30 15 ... 30 14 ... 30 J K L C6H14 SO2 (TÜV; see table "TÜV single component", page 1/87) NO (TÜV; see table "TÜV single component", page 1/87) 14 ... 30 13 ... 30 M N 14 ... 20, 22 P 14 ... 30 17 ... 20, 22 13 ... 30 Q R S A21 A20 Measured component NH3 (dry) H2O N2O Q R Smallest measuring range Largest measuring range Measuring range identification 0 ... 5 vpm 0 ... 10 vpm 0 ... 20 vpm 0 ... 100 vpm 0 ... 200 vpm 0 ... 400 vpm 10 11 12 A B C 0 ... 50 vpm 0 ... 100 vpm 0 ... 300 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 13 14 15 D E F 0 ... 500 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm 16 17 18 G H J 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 5 000 vpm 0 ... 30 000 vpm 0 ... 15 000 vpm 0 ... 50 000 vpm 19 20 21 K L M 0 ... 1 % 0 ... 1 % 0 ... 3 % 0 ... 3 % 0 ... 10 % 0 ... 10 % 22 23 24 N P Q 0 ... 3 % 0 ... 5 % 0 ... 5 % 0 ... 30 % 0 ... 15 % 0 ... 50 % 25 26 27 R S T 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 30 % 0 ... 100 % 0 ... 100 % 28 29 30 U V W Internal gas paths Sample chamber1) (lining) Aluminum Aluminum Reference chamber (flow-type) Non-flow-type Flow-type 0 1 Tantalum Tantalum Non-flow-type Flow-type Stainless steel pipe Aluminum (mat. no. 1.4571) Tantalum With sample gas monitoring Hose made of FKM (Viton) Hose made of FKM (Viton) Pipe made of titanium Aluminum Aluminum 0 1 0 A20, A21 4 5 4 5 A20, A21, Y02 Y02 Non-flow-type Non-flow-type 6 8 6 8 A20, A21 A20, A21 Non-flow-type Flow-type 2 3 2 A20, A21 2 3 Footnotes: see next page Siemens AP 01 · 2015 1/79 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit ■ Selection and ordering data ULTRAMAT 6 gas analyzer Single-channel 19" rack unit for installation in cabinets Article No. 7MB2121- 7 7 7 7 7 - 7 A A 7 Add-on electronics Without AUTOCAL function • With 8 additional digital inputs/outputs • With serial interface for the automotive industry (AK) • With 8 digital inputs/outputs, PROFIBUS PA interface • With 8 digital inputs/outputs, PROFIBUS DP interface 0 1 3 6 7 Power supply 100 ... 120 V AC, 48 ... 63 Hz 200 ... 240 V AC, 48 ... 63 Hz 3 0 1 Operating software and documentation German English French Spanish Italian Additional versions 0 1 2 3 4 Order code Add "-Z" to Article No. and specify Order code Flow-type reference cell with reduced flow, 6 mm A20 Flow-type reference cell with reduced flow, ¼" A21 Telescopic rails (2 units) A31 TAG labels (specific lettering based on customer information) B03 Kalrez gaskets in sample gas path B04 SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511 C20 FM/CSA certificate – Class I Div 2 E20 Clean for O2 service (specially cleaned gas path) Y02 Measuring range indication in plain text, if different from the standard setting Y11 Special setting (only in conjunction with an application no., e.g. extended measuring range) Y12 Extended special setting (only in conjunction with an application no., e.g. determination of cross-interferences) Y13 TÜV version acc. to 13th and 17th BlmSchV Y17 Accessories Article No. RS 485/Ethernet converter A5E00852383 RS 485/RS 232 converter C79451-Z1589-U1 RS 485/USB converter A5E00852382 AUTOCAL function with serial interface for the automotive industry (AK) C79451-A3480-D512 AUTOCAL function with 8 digital inputs/outputs C79451-A3480-D511 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA A5E00057307 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP A5E00057312 Set of Torx screwdrivers A5E34821625 1) Only for cell length 20 to 180 mm 1/80 Siemens AP 01 · 2015 Cannot be combined E20 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer Two-channel 19" rack unit for installation in cabinets for measuring 2 IR components 7MB2123- 7 7 7 7 7 - 7 7 7 7 Cannot be combined Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections for sample gas and reference gas Pipe with 6 mm outer diameter Pipe with ¼" outer diameter Possible with measuring Channel 1 Measured component range identification CO 11 ... 30 CO highly selective (with optical filter) 12 ... 30 CO (TÜV; see table "TÜV single component", page 1/87) CO2 10 ... 30 CH4 13 ... 30 C2H2 15 ... 30 C2H4 15 ... 30 C2H6 14 ... 30 C3H6 14 ... 30 C3H8 13 ... 30 C4H6 15 ... 30 C4H10 14 ... 30 C6H14 14 ... 30 SO2 (TÜV; see table "TÜV single component", 13 ... 30 page 1/87) NO (TÜV; see table "TÜV single component", 14 ... 20, 22 page 1/87) NH3 (dry) 14 ... 30 H2O 17 ... 20, 22 N2O 13 ... 30 0 1 0 1 A21, A41 A20, A40 A B X C D E F G H J K L M N P Q R S Q R Smallest measuring range Largest measuring range Measuring range identification 0 ... 5 vpm 0 ... 10 vpm 0 ... 20 vpm 0 ... 50 vpm 0 ... 100 vpm 0 ... 300 vpm 0 ... 500 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 5 000 vpm 0 ... 100 vpm 0 ... 200 vpm 0 ... 400 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm 0 ... 30 000 vpm 0 ... 15 000 vpm 0 ... 50 000 vpm 10 11 12 13 14 15 16 17 18 19 20 21 A B C D E F G H J K L M 0 ... 1 % 0 ... 1 % 0 ... 3 % 0 ... 3 % 0 ... 5 % 0 ... 5 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 3 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 15 % 0 ... 50 % 0 ... 30 % 0 ... 100 % 0 ... 100 % 22 23 24 25 26 27 28 29 30 N P Q R S T U V W Internal gas paths Sample chamber1) (lining) Reference chamber (flow-type) Hose made of FKM (Viton) Aluminum Aluminum Non-flow-type Flow-type 0 1 Pipe made of titanium Tantalum Tantalum Non-flow-type Flow-type Stainless steel pipe (mat. no. 1.4571) Aluminum Tantalum 0 1 0 A20, A21, A40, A41 4 5 4 5 A20, A21, A40, A41, Y02 Y02 Non-flow-type Non-flow-type 6 8 6 8 A20, A21, A40, A41 A20, A21, A40, A41 Non-flow-type Flow-type 2 3 2 A20, A21, A40, A41 With sample gas monitoring Hose made of FKM (Viton) 1) Aluminum Aluminum 2 3 Only for cell length 20 to 180 mm Siemens AP 01 · 2015 1/81 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit ■ Selection and ordering data ULTRAMAT 6 gas analyzer Two-channel 19" rack unit for installation in cabinets for measuring 2 IR components Article No. 7MB2123- 7 7 7 7 7 - 7 7 7 7 Cannot be combined Add-on electronics 0 Without AUTOCAL function • With 8 additional digital inputs/outputs each for channel 1 • With 8 additional digital inputs/outputs each for channel 2 • With 8 additional digital inputs/outputs each for channel 1 and channel 2 • With serial interface for the automotive industry (AK) • With 8 additional digital inputs/outputs each for channel 1 and channel 2 and PROFIBUS PA interface • With 8 additional digital inputs/outputs each for channel 1 and channel 2 and PROFIBUS DP interface Power supply 100 ... 120 V AC, 48 ... 63 Hz 200 ... 240 V AC, 48 ... 63 Hz 1 2 3 5 6 5 7 0 1 Channel 2 Possible with measuring Measured component range identification CO 11 ... 30 CO highly selective (with optical filter) 12 ... 30 CO (TÜV; see table "TÜV single component", page 1/87) CO2 10 ... 30 CH4 13 ... 30 C2H2 15 ... 30 C2H4 15 ... 30 C2H6 14 ... 30 C3H6 14 ... 30 C3H8 13 ... 30 C4H6 15 ... 30 C4H10 14 ... 30 C6H14 14 ... 30 SO2 (TÜV; see table "TÜV single component", 13 ... 30 page 1/87) NO (TÜV; see table "TÜV single component", 14 ... 20, 22 page 1/87) NH3 (dry) 14 ... 30 H2O 17 ... 20, 22 N2O 13 ... 30 A B X C D E F G H J K L M N P Q R S Q R Smallest measuring range Largest measuring range Measuring range identification 0 ... 5 vpm 0 ... 10 vpm 0 ... 20 vpm 0 ... 50 vpm 0 ... 100 vpm 0 ... 300 vpm 0 ... 500 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 5 000 vpm 0 ... 100 vpm 0 ... 200 vpm 0 ... 400 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm 0 ... 30 000 vpm 0 ... 15 000 vpm 0 ... 50 000 vpm 10 11 12 13 14 15 16 17 18 19 20 21 A B C D E F G H J K L M 0 ... 1 % 0 ... 1 % 0 ... 3 % 0 ... 3 % 0 ... 5 % 0 ... 5 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 3 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 15 % 0 ... 50 % 0 ... 30 % 0 ... 100 % 0 ... 100 % 22 23 24 25 26 27 28 29 30 N P Q R S T U V W Operating software and documentation German English French Spanish Italian 1/82 Siemens AP 01 · 2015 0 1 2 3 4 E20 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Selection and ordering data Additional versions Order code Cannot be combined Add "-Z" to Article No. and specify Order codes. Flow-type reference cell with reduced flow, 6 mm (channel 1) A20 Flow-type reference cell with reduced flow, ¼" (channel 1) A21 Flow-type reference cell with reduced flow, 6 mm (channel 2) A40 Flow-type reference cell with reduced flow, ¼" (channel 2) A41 Connection pipe (can only be combined with the appropriate gas connection diameter and internal gas path materials) • Made of titanium, 6 mm, complete with screwed gland, for sample gas side A22 • Made of titanium, 6 mm, complete with screwed gland, for reference gas side A23 • Made of titanium, ¼", complete with screwed gland, for sample gas side A24 • Made of titanium, ¼", complete with screwed gland, for reference gas side A25 • Made of stainless steel (mat. no. 1.4571), 6 mm, complete with screwed gland, for sample gas side A27 • Made of stainless steel (mat. no. 1.4571), ¼", complete with screwed gland, for sample gas side A29 Telescopic rails (2 units) A31 TAG labels (specific lettering based on customer information) B03 Kalrez gaskets in sample gas path (channel 1) B04 Kalrez gaskets in sample gas path (channel 2) B05 SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511 C20 FM/CSA certificate – Class I Div 2 E20 Clean for O2 service (specially cleaned gas path; channels 1 + 2) Y02 Measuring range indication in plain text, if different from the standard setting Y11 Special setting (only in conjunction with an application no., e.g. extended measuring range) Y12 Extended special setting (only in conjunction with an application no., e.g. determination of cross-interferences) Y13 TÜV version acc. to 13th and 17th BlmSchV (1st channel) Y17 TÜV version acc. to 13th and 17th BlmSchV (2nd channel) Y18 Accessories Article No. RS 485/Ethernet converter A5E00852383 RS 485/RS 232 converter C79451-Z1589-U1 RS 485/USB converter A5E00852382 AUTOCAL function with serial interface for the automotive industry (AK) C79451-A3480-D33 AUTOCAL function with 8 digital inputs/outputs for channel 1 or channel 2 C79451-A3480-D511 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA for channel 1 or channel 2 A5E00057307 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP for channel 1 or channel 2 A5E00057312 Set of Torx screwdrivers A5E34821625 Siemens AP 01 · 2015 1/83 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer Single-channel or dual-channel 19" rack unit for installation in cabinets for measuring 2 or 3 IR components 7MB2124- 7 7 7 7 7 - 7 7 7 7 Cannot be combined Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections for sample gas and reference gas Pipe with 6 mm outer diameter Pipe with ¼" outer diameter 0 1 0 1 Measured component Smallest measuring range Largest measuring range CO NO 0 ... 100 vpm 0 ... 100 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm AA CO NO 0 ... 300 vpm 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm AB CO NO 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm AC A21, A41 A20, A40 For CO/NO (TÜV; table "TÜV, 2 components in series", page 1/76) CO2 CO 0 ... 100 vpm 0 ... 100 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm BA CO2 CO 0 ... 300 vpm 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm BB CO2 CO 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm BC CO2 CO 0 ... 3 000 vpm 0 ... 3 000 vpm 0 ... 30 000 vpm 0 ... 30 000 vpm BD CO2 CO 0 ... 1 % 0 ... 1 % 0 ... 10 % 0 ... 10 % BE CO2 CO 0 ... 3 % 0 ... 3 % 0 ... 30 % 0 ... 30 % BF CO2 CO 0 ... 10 % 0 ... 10 % 0 ... 100 % 0 ... 100 % BG CO2 CH4 0 ... 10 % 0 ... 10 % 0 ... 100 % 0 ... 100 % CG CO2 NO 0 ... 300 vpm 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm DB Internal gas paths Sample chamber1) (lining) Aluminum Aluminum Reference chamber (flow-type) Non-flow-type Flow-type 0 1 Pipe made of titanium Tantalum Tantalum Stainless steel pipe Aluminum (mat. no. 1.4571) Tantalum Non-flow-type Flow-type Non-flow-type Non-flow-type 4 5 6 8 With sample gas monitoring Hose made of FKM Aluminum (Viton) Aluminum Non-flow-type Flow-type 2 3 Hose made of FKM (Viton) Add-on electronics Without AUTOCAL function • With 8 additional digital inputs/outputs each for channel 1 • With 8 additional digital inputs/outputs each for channel 1 and channel 2 • With serial interface for the automotive industry (AK), channel 1 • With serial interface for the automotive industry (AK), channel 1 and channel 2 • With 8 additional digital inputs/outputs for channel 1 and PROFIBUS PA interface • With 8 additional digital inputs/outputs each for channel 1 and channel 2 and PROFIBUS PA interface • With 8 additional digital inputs/outputs for channel 1 and PROFIBUS DP interface • With 8 additional digital inputs/outputs each for channel 1 and channel 2 and PROFIBUS DP interface 1) Only for cell length 20 to 180 mm 1/84 Siemens AP 01 · 2015 0 1 2 3 0 A20, A21, A40, A41 4 5 6 8 A20, A21, A40, A41, Y02 Y02 A20, A21, A40, A41 A20, A21, A40, A41 2 A20, A21, A40, A41 0 1 2 3 4 2 3 4 5 6 6 7 8 8 E20 E20 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer Single-channel or dual-channel 19" rack unit for installation in cabinets for measuring 2 or 3 IR components Power supply 100 ... 120 V AC, 48 ... 63 Hz 200 ... 240 V AC, 48 ... 63 Hz 7MB2124- 7 7 7 7 7 - 7 7 7 7 0 1 Channel 2 Possible with measuring Measured component range identification Without channel 2 CO 11 ... 30 CO highly selective (with optical filter) 12 ... 30 CO (TÜV; see table "TÜV single component", page 1/87) CO2 10 ... 30 CH4 13 ... 30 C2H2 15 ... 30 C2H4 15 ... 30 C2H6 14 ... 30 C3H6 14 ... 30 C3H8 13 ... 30 C4H6 15 ... 30 C4H10 14 ... 30 C6H14 14 ... 30 SO2 (TÜV; see table "TÜV single component", 13 ... 30 page 1/87) NO (TÜV; see table "TÜV single component", 14 ... 20, 22 page 1/87) NH3 (dry) 14 ... 30 H2O 17 ... 20, 22 N2O 13 ... 30 Smallest measuring range Without channel 2 Largest measuring range Measuring range identification 0 ... 5 vpm 0 ... 10 vpm 0 ... 20 vpm 0 ... 50 vpm 0 ... 100 vpm 0 ... 300 vpm 0 ... 500 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 5 000 vpm 0 ... 100 vpm 0 ... 200 vpm 0 ... 400 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm 0 ... 30 000 vpm 0 ... 15 000 vpm 0 ... 50 000 vpm 10 11 12 13 14 15 16 17 18 19 20 21 0 ... 1 % 0 ... 1 % 0 ... 3 % 0 ... 3 % 0 ... 5 % 0 ... 5 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 3 % 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 15 % 0 ... 50 % 0 ... 30 % 0 ... 100 % 0 ... 100 % 22 23 24 25 26 27 28 29 30 W A B X C D E F G H J K L M N W P Q R S Q R X Operating software and documentation German English French Spanish Italian Cannot be combined X A40, A41, B05 A B C D E F G H J K L M N P Q R S T U V W 0 1 2 3 4 Siemens AP 01 · 2015 1/85 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit ■ Selection and ordering data Additional versions Order code Add "-Z" to Article No. and specify Order codes. Flow-type reference cell with reduced flow, 6 mm (channel 1) A20 Flow-type reference cell with reduced flow, ¼" (channel 1) A21 Flow-type reference cell with reduced flow, 6 mm (channel 2) A40 Flow-type reference cell with reduced flow, ¼" (channel 2) A41 Connection pipe (can only be combined with the appropriate gas connection diameter and internal gas path materials) • Made of titanium, 6 mm, complete with screwed gland, for sample gas side A22 • Made of titanium, 6 mm, complete with screwed gland, for reference gas side A23 • Made of titanium, ¼", complete with screwed gland, for sample gas side A24 • Made of titanium, ¼", complete with screwed gland, for reference gas side A25 • Made of stainless steel (mat. no. 1.4571), 6 mm, complete with screwed gland, for sample gas side A27 • Made of stainless steel (mat. no. 1.4571), ¼", complete with screwed gland, for sample gas side A29 Telescopic rails (2 units) A31 TAG labels (specific lettering based on customer information) B03 Kalrez gaskets in sample gas path (channel 1) B04 Kalrez gaskets in sample gas path (channel 2) B05 SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511 C20 FM/CSA certificate – Class I Div 2 E20 Clean for O2 service (specially cleaned gas path; channels 1 + 2) Y02 Measuring range indication in plain text, if different from the standard setting Y11 Special setting (only in conjunction with an application no., e.g. extended measuring range) Y12 Extended special setting (only in conjunction with an application no., e.g. determination of cross-interferences) Y13 TÜV version acc. to 13th and 17th BlmSchV (channel 1) Y17 TÜV version acc. to 13th and 17th BlmSchV (channel 2) Y18 Accessories Article No. RS 485/Ethernet converter A5E00852383 RS 485/RS 232 converter C79451-Z1589-U1 RS 485/USB converter A5E00852382 AUTOCAL function with serial interface for the automotive industry (AK) C79451-A3480-D33 AUTOCAL function with 8 digital inputs/outputs for channel 1 or channel 2 C79451-A3480-D511 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA for channel 1 or channel 2 A5E00057307 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP for channel 1 or channel 2 A5E00057312 Set of Torx screwdrivers A5E34821625 1/86 Siemens AP 01 · 2015 Cannot be combined © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit TÜV single component Component CO (TÜV) Measuring range identification Smallest measuring range from 0 to ... SO2 (TÜV) Largest measuring range from 0 to ... C D 50 mg/m3 1 000 mg/m3 E mg/m3 F 300 G 500 mg/m3 5 000 mg/m3 H 1 000 mg/m3 10 000 mg/m3 K mg/m3 mg/m3 P R V 3 000 10 g/m3 30 g/m3 100 g/m3 3 000 mg/m3 30 000 100 g/m3 300 g/m3 1 160 Smallest measuring range from 0 to ... Largest measuring range from 0 to ... 75 mg/m3 1 500 mg/m3 300 mg/m3 3 000 mg/m3 500 mg/m3 5 000 mg/m3 1 000 mg/m3 3 000 mg/m3 g/m3 NO (TÜV) 10 g/m3 30 g/m3 100 g/m3 300 g/m3 10 000 mg/m3 30 000 mg/m3 100 g/m3 300 g/m3 1 000 g/m3 2 630 g/m3 Smallest measuring range from 0 to ... Largest measuring range from 0 to ... 100 mg/m3 2 000 mg/m3 mg/m3 3 000 mg/m3 500 mg/m3 5 000 mg/m3 1 000 mg/m3 10 000 mg/m3 mg/m3 30 000 mg/m3 300 3 000 10 g/m3 100 g/m3 30 g/m3 300 g/m3 100 g/m3 1 250 g/m3 Example for ordering ULTRAMAT 6, TÜV Component: CO Measuring range: 0 to 50 / 1 000 mg/m3 with hoses, non-flow-type reference compartment without automatic adjustment (AUTOCAL) 230 V AC; German 7MB2121-0XD00-1AA0-Z +Y17 TÜV, 2 components in series Component CO (TÜV) Measuring range identification Smallest measuring range from 0 to ... Largest measuring range from 0 to ... Smallest measuring range from 0 to ... Largest measuring range from 0 to ... AA 75 mg/m3 1 000 mg/m3 200 mg/m3 2 000 mg/m3 mg/m3 mg/m3 3 000 mg/m3 AB AC 300 mg/m3 1 000 mg/m3 NO (TÜV) 3 000 10 000 mg/m3 300 1 000 mg/m3 10 000 mg/m3 Example for ordering ULTRAMAT 6, TÜV, 2-component unit Components: CO/NO + SO2 Measuring range: CO: 0 to 75 / 1 000 mg/m3, NO: 0 to 200 / 2 000 mg/m3, SO2: 0 to 75 / 1 500 mg/m3 with hoses, non-flow-type reference compartment without automatic adjustment (AUTOCAL) 230 V AC; German 7MB2124-0AA00-1NC0-Z +Y17+Y18 Note: for 3 components take both tables into consideration. Ordering information measured component N2O Certification in accordance with AM0028 and AM0034 (Kyoto Protocol) for measuring N2O, measuring range 0 ... 300 ppm / 3 000 ppm. Version: Standard device Siemens AP 01 · 2015 1/87 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Dimensional drawings s 8/75$0$7 0 1/88 Siemens AP 01 · 2015 ULTRAMAT 6, 19“ unit, dimensions in mm (example: 1-channel version) 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit ■ Schematics Pin assignment (electrical and gas connections) SUB-D 9F connector (RS 485) M 5 9 GND 4 8 3 +5 V 7 2 6 M 1 GND R_Level-NNC RD/TD-N RD/TD-P R_Level-PNC NC GND It is possible to connect bus terminating resistors to pins 7 and 9. SUB-D 15F connector M 8 15 7 14 6 13 5 12 M 4 11 3 M 10 2 9 M 1 GND Analog output 2-P Analog output 2-N Analog output 1-P Analog output 1-N NC NC Analog input 2-P Analog input 2-N Analog input 1-P Analog input 1-N Binary input 6-P Binary input 5-P Binary input 5 to 6-N GND For 2-component version only of the ULTRAMAT part Analog outputs isolated (also from each other), RL: ≤ 750 Ω Pressure correction Pressure correction Correction of cross-interference Correction of cross-interference Isolated via optocoupler "0" = 0 V (0 ... 4.5 V) "1" = 24 V (13 ... 33 V) Analog inputs non-isolated, 0 ... 20 mA/500 Ω or 0 ... 10 V (low-resistance) SUB-D 25F connector 13 M 25 12 24 11 23 GND Binary input 4-P Binary input 3-P Binary input 2-P Binary input 1-P Binary input 1 to 4-N Isolated via optocoupler "0" = 0 V (0 ... 4.5 V) "1" = 24 V (13 ... 33 V) 10 22 Relay 6 9 21 8 Relay 5 20 7 19 Relay 4 6 18 5 Relay 3 Contact load max. 24 V/1 A, AC/DC; relay contacts shown: relay coil has zero current 17 4 Relay 2 16 3 15 2 Relay 1 1 GND 14 M 1RWH $OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNV PXVWEHVKLHOGHGDQGUHVWDJDLQVWWKH HQFORVXUHSRWHQWLDO ULTRAMAT 6, 19“ unit, pin assignment Siemens AP 01 · 2015 1/89 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 19" rack unit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“ unit, pin assignment of AUTOCAL board and PROFIBUS connectors 1/90 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 19" rack unit &KDQQHO &KDQQHO &KDQQHO 6DPSOHJDVRXWOHW 6DPSOHJDVLQOHW SLQFRQQHFWRU %LQDU\LQSXWVDQG DQDORJLQSXWVRXWSXWV 5HIHUHQFHJDVRXWOHW 5HIHUHQFHJDVLQOHW &KDQQHO SLQ FRQQHFWRU 56 3XUJLQJJDV LQOHW SLQ LQWHUIDFH FRQQHFWRU RSWLRQHJ 352),%86 SLQFRQQHFWRU %LQDU\LQSXWV DQG UHOD\RXWSXWV RSWLRQERDUG 3RZHUVXSSO\ DQGIXVHV *DVFRQQHFWLRQVVWXEVPPRU~ SLQFRQQHFWRU %LQDU\LQSXWVDQG UHOD\RXWSXWV ULTRAMAT 6, 19“ unit, gas and electrical connections (example: 2-channel version) Siemens AP 01 · 2015 1/91 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 Field device ■ Technical specifications General information Gas inlet conditions Measuring ranges 4, internally and externally switchable; autoranging is also possible Smallest possible measuring range Dependent on the application, e.g. CO: 0 … 10 vpm, CO2 : 0 … 5 vpm Largest possible measuring range Dependent on the application Measuring range with suppressed zero point Any zero point within 0 … 100 vol.% can be implemented; smallest possible span 20 % Permissible sample gas pressure • With hoses (without pressure switch) 600 … 1 500 hPa (absolute) • With pipes (without pressure switch) 600 … 1 500 hPa (absolute) - Ex (leakage compensation) 600 … 1 160 hPa (absolute) - Ex (continuous purging) 600 … 1 500 hPa (absolute) Purging gas pressure • Permanent < 165 hPa above ambient pressure Heated version 65 °C Operating position Front wall, vertical • For short periods 250 hPa above ambient pressure Conformity CE mark in accordance with EN 50081-1, EN 50082-2 Sample gas flow 18 … 90 l/h (0.3 … 1.5 l/min) Sample gas temperature Min. 0 ... max. 50 °C, but above the dew point, for heated version min. 0 … max. 80 °C < 90 % RH (RH: relative humidity) or dependent on measuring task Influence of interfering gases must be considered separately Design, enclosure Weight Approx. 32 kg Sample gas humidity Degree of protection IP65 in accordance with EN 60529, restricted breathing enclosure to EN 50021 Dynamic response Warm-up period At room temperature < 30 min (the technical specification will be met after 2 hours) Delayed display (T90-time) Dependent on length of analyzer chamber, sample gas line and parameterizable damping Damping (electrical time constant) 0 … 100 s, parameterizable Approximately 0.5 ... 5 s, depending on version <1s Electrical characteristics Power supply 100 … 120 V AC (nominal range of use 90 … 132 V), 48 … 63 Hz or 200 … 240 V AC (nominal range of use 180 … 264 V), 48 … 63 Hz Power consumption Approx. 35 VA; approx. 330 VA with heated version Dead time (purging time of the gas path in the unit at 1 l/min) EMC (Electromagnetic Compatibility) In accordance with standard requirements of NAMUR NE21 (08/98) Time for device-internal signal processing Electrical safety In accordance with EN 61010-1 Pressure sensor • Heated units Overvoltage category II • Internal 700 … 1 200 hPa absolute • Unheated units Overvoltage category III • External 700 … 1 500 hPa absolute • 100 ... 120 V F3: 1 T/250; F4: 1 T/250 Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) • 200 ... 240 V F3: 0.63 T/250; F4: 0.63 T/250 Output signal fluctuation < ± 1 % of the smallest possible measuring range according to rating plate • 100 ... 120 V F1: 1 T/250; F2: 4 T/250 F3: 4 T/250; F4: 4 T/250 Zero point drift < ± 1 % of the current measuring range/week • 200 ... 240 V F1: 0.63 T/250; F2: 2.5 T/250 F3: 2.5 T/250; F4: 2.5 T/250 Measured-value drift < ± 1 % of the current measuring range/week Repeatability ≤ 1 % of the current measuring range Detection limit 1 % of the smallest possible measuring range Linearity error < 0.5 % of the full-scale value Fuse values (unheated unit) Pressure correction range Fuse values (heated unit) 1/92 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) Ambient temperature < 1 % of current measuring range/10 K (with constant receiver cell temperature) Sample gas pressure When pressure compensation has been switched on: < 0.15 % of setpoint/1 % atmospheric pressure change Sample gas flow Negligible Power supply < 0.1 % of the current measuring range with rated voltage ± 10 % Environmental conditions Application-specific measuring influences possible if ambient air contains measured component or cross interference-sensitive gases Electrical inputs and outputs Analog output 0/2/4 … 20 mA, isolated; load 750 Ω Relay outputs 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking Analog inputs 2, dimensioned for 0/2/4 ... 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference) Binary inputs 6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover Serial interface RS 485 Options AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP Climatic conditions Permissible ambient temperature -30 … +70 °C during storage and transportation; 5 … 45 °C during operation Permissible humidity < 90 % RH (RH: relative humidity) within average annual value, during storage and transportation (dew point must not be undershot) Siemens AP 01 · 2015 1/93 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 Field device ■ Selection and ordering data ULTRAMAT 6 gas analyzer For installation in the field, single-channel, 1 component Article No. 7MB2111- 7 7 7 7 7 - 7 7 A 7 Cannot be combined Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections Ferrule screw connection for pipe, outer diameter 6 mm Ferrule screw connection for pipe, outer diameter ¼" 0 1 0 1 Measured component Possible with measuring range identification CO 11 ... 30 CO highly selective (with optical filter) 12 ... 30 CO (TÜV; see table "TÜV single component", page 1/99) A B X CO2 CH4 C2H2 10 ... 30 13 ... 30 15 ... 30 D E C2H4 C2H6 C3H6 15 ... 30 14 ... 30 14 ... 30 G H C3H8 C4H6 C4H10 13 ... 30 15 ... 30 14 ... 30 K L C6H14 SO2 (TÜV; see table "TÜV single component", page 1/99) NO (TÜV; see table "TÜV single component", page 1/99) 14 ... 30 13 ... 30 M 14 ... 20, 22 P 14 ... 30 17 ... 20; 22 (17 to 24, 26; heated) 13 ... 30 Q NH3 (dry) H2O N2O C F J N Q R R S Smallest measuring range Largest measuring range Measuring range identification 0 ... 5 vpm 0 ... 10 vpm 0 ... 20 vpm 0 ... 100 vpm 0 ... 200 vpm 0 ... 400 vpm 10 11 12 0 ... 50 vpm 0 ... 100 vpm 0 ... 300 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 13 14 15 D 0 ... 500 vpm 0 ... 1 000 vpm 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm 16 17 19 G 0 ... 3 000 vpm 0 ... 5 000 vpm 0 ... 5 000 vpm 0 ... 30 000 vpm 0 ... 15 000 vpm 0 ... 50 000 vpm 19 20 21 K 0 ... 1 % 0 ... 1 % 0 ... 3 % 0 ... 3 % 0 ... 10 % 0 ... 10 % 22 23 24 0 ... 3 % 0 ... 5 % 0 ... 5 % 0 ... 30 % 0 ... 15 % 0 ... 50 % 25 26 27 R 0 ... 10 % 0 ... 10 % 0 ... 30 % 0 ... 30 % 0 ... 100 % 0 ... 100 % 28 29 30 U 1/94 Siemens AP 01 · 2015 A B C E F H J L M N P Q S T V W A29 A28 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer For installation in the field, single-channel, 1 component 7MB2111- 7 7 7 7 7 - 7 7 A 7 Cannot be combined Internal gas paths Sample chamber (lining) Reference chamber (flow-type) Hose made of FKM (Viton) Aluminum Aluminum Non-flow-type Flow-type 0 0 0 1 1 1 Pipe made of titanium Tantalum1) Tantalum1) Non-flow-type Flow-type Aluminum Tantalum1) Non-flow-type Non-flow-type 6 8 Stainless steel pipe (mat. no. 1.4571) Add-on electronics Without AUTOCAL function • With 8 additional digital inputs/outputs • With 8 digital inputs/outputs and PROFIBUS PA interface • With 8 digital inputs/outputs and PROFIBUS DP interface • With 8 digital inputs/outputs and PROFIBUS PA Ex i 0 A28, A29 2 2 A28, A29, Y02 3 3 6 Y02 A28, A29 8 A28, A29 0 1 6 7 8 6 7 E12 E12 8 Power supply Standard unit and acc. to ATEX II 3G version (Zone 2) • 100 ... 120 V AC, 48 ... 63 Hz • 200 ... 240 V AC, 48 ... 63 Hz ATEX II 2G versions (Zone 1), incl. certificate • 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: leakage compensation) • 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: leakage compensation) • 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: continuous purging) • 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: continuous purging) Heating of internal gas paths and analyzer unit Without With (max. 65 °C) Language (supplied documentation, software) German English French Spanish Italian 1) Only for cell length 20 to 180 mm 2) Only in connection with an approved purging unit 0 1 0 1 2 2 2 3 3 3 6 6 6 7 7 7 A B 0 1 2 3 4 Siemens AP 01 · 2015 1/95 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 Field device ■ Selection and ordering data Additional versions Order code Add "-Z" to Article No. and specify Order codes. Flow-type reference cell with reduced flow, 6 mm A28 Flow-type reference cell with reduced flow, ¼" A29 TAG labels (specific lettering based on customer information) B03 Kalrez gaskets in sample gas path B04 SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511 C20 Ex versions Possible combinations: see: Table "Ex configurations – principle selection criteria", page 5/16 ATEX II 3G certificate; restricted breathing enclosure, non-flammable gases E11 ATEX II 3G certificate; flammable gases E12 FM/CSA certificate – Class I Div 2 E20 ATEX II 3D certificate; potentially explosive dust atmospheres • In non-hazardous gas zone E40 • In Ex zone acc. to ATEX II 3G, non-flammable gases E41 • In Ex zone acc. to ATEX II 3G, flammable gases1) E42 BARTEC EEx p control unit "Leakage compensation" E71 BARTEC EEx p control unit "Continuous purging" E72 Clean for O2 service (specially cleaned gas path) Y02 Measuring range indication in plain text, if different from the standard setting Y11 Special setting (only in conjunction with an application no., e.g. extended measuring range) Y12 Extended special setting (only in conjunction with an application no., e.g. determination of cross-interferences) Y13 TÜV version acc. to 13th and 17th BlmSchV Y17 Additional units for Ex versions Article No. Category ATEX II 2G (zone 1) BARTEC EEx p control unit, 230 V, "leakage compensation" 7MB8000-2BA BARTEC EEx p control unit, 115 V, "leakage compensation" 7MB8000-2BB BARTEC EEx p control unit, 230 V, "continuous purging" 7MB8000-2CA BARTEC EEx p control unit, 115 V, "continuous purging" 7MB8000-2CB Ex isolation amplifier 7MB8000-3AB Ex isolating relay, 230 V 7MB8000-4AA Ex isolating relay, 110 V 7MB8000-4AB Differential pressure switch for corrosive and non-corrosive gases 7MB8000-5AA Stainless steel flame arrestor 7MB8000-6BA Hastelloy flame arrestor 7MB8000-6BB Category ATEX II 3G (Zone 2) BARTEC EEx p control unit, 230 V, "continuous purging" 7MB8000-2CA BARTEC EEx p control unit, 115 V, "continuous purging" 7MB8000-2CB FM/CSA (Class I Div. 2) Ex purging unit MiniPurge FM 7MB8000-1AA Accessories Article No. RS 485/Ethernet converter A5E00852383 RS 485/RS 232 converter C79451-Z1589-U1 RS 485/USB converter A5E00852382 AUTOCAL function with 8 digital inputs/outputs A5E00064223 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA A5E00057315 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP A5E00057318 AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA Ex i (firmware 4.1.10 required) A5E00057317 Set of Torx screwdrivers A5E34821625 1) Only in connection with an approved purging unit 1/96 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device ■ Selection and ordering data Article No. ULTRAMAT 6 gas analyzer For installation in the field, single-channel, 2 components 7MB2112- 7 7 7 7 7 - 7 7 A 7 Cannot be combined Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections Ferrule screw connection for pipe, outer diameter 6 mm Ferrule screw connection for pipe, outer diameter ¼" 0 1 0 1 Measured component Smallest measuring range Largest measuring range CO NO 0 ... 100 vpm 0 ... 100 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm AA CO NO 0 ... 300 vpm 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm AB CO 0 ... 1 000 vpm 0 ... 10 000 vpm NO 0 ... 1 000 vpm 0 ... 10 000 vpm For CO/NO (TÜV; see table "TÜV, 2 components in series", page 1/99) AC CO2 CO 0 ... 100 vpm 0 ... 100 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm BA CO2 CO 0 ... 300 vpm 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 3 000 vpm BB CO2 CO 0 ... 1 000 vpm 0 ... 1 000 vpm 0 ... 10 000 vpm 0 ... 10 000 vpm BC CO2 CO 0 ... 3 000 vpm 0 ... 3 000 vpm 0 ... 30 000 vpm 0 ... 30 000 vpm BD CO2 CO 0 ... 1 % 0 ... 1 % 0 ... 10 % 0 ... 10 % BE CO2 CO 0 ... 3 % 0 ... 3 % 0 ... 30 % 0 ... 30 % BF CO2 CO 0 ... 10 % 0 ... 10 % 0 ... 100 % 0 ... 100 % BG CO2 CH4 0 ... 10 % 0 ... 10 % 0 ... 100 % 0 ... 100 % CG CO2 NO 0 ... 100 vpm 0 ... 100 vpm 0 ... 1 000 vpm 0 ... 1 000 vpm DA CO2 NO Internal gas paths 0 ... 300 vpm 0 ... 3 000 vpm 0 ... 300 vpm 0 ... 3 000 vpm Sample chamber (lining) Reference chamber (flow-type) Aluminum Non-flow-type Aluminum Flow-type Hose made of FKM (Viton) Pipe made of titanium Tantalum1) Tantalum1) Aluminum Tantalum1) Non-flow-type Flow-type Non-flow-type Non-flow-type Stainless steel pipe (mat. no. 1.4571) Add-on electronics Without AUTOCAL function • With 8 additional digital inputs/outputs • With 8 digital inputs/outputs and PROFIBUS PA interface • With 8 digital inputs/outputs and PROFIBUS DP interface • With 8 digital inputs/outputs and PROFIBUS PA Ex i Power supply Standard unit and acc. to ATEX II 3G version (Zone 2) • 100 ... 120 V AC, 48 ... 63 Hz • 200 ... 240 V AC, 48 ... 63 Hz ATEX II 2G versions (Zone 1), incl. certificate • 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: leakage compensation) • 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: leakage compensation) • 100 ... 120 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: continuous purging) • 200 ... 240 V AC, 48 ... 63 Hz, according to ATEX II 2G2) (operating mode: continuous purging) Heating of internal gas paths and analyzer unit none With (max. 65 °C) A29 A28 DB 0 1 0 1 2 3 6 8 0 A28, A29 2 3 6 8 A28, A29, Y02 Y02 A28, A29 A28, A29 0 1 6 7 8 6 7 8 0 1 0 1 2 2 2 3 3 3 6 6 6 7 7 7 A B Siemens AP 01 · 2015 1/97 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 Field device ■ Selection and ordering data ULTRAMAT 6 gas analyzer For installation in the field, single-channel, 2 components Article No. 7MB2112- 7 7 7 7 7 - 7 7 A 7 Cannot be combined Language (supplied documentation, software) German 0 English 1 French 2 Spanish 3 Italian 4 1) Only for cell length 20 to 180 mm. 2) See also next page "Additional units for Ex versions". Additional versions Add "-Z" to Article No. and specify Order codes. Flow-type reference cell with reduced flow, 6 mm Flow-type reference cell with reduced flow, ¼" TAG labels (specific lettering based on customer information) Kalrez gaskets in sample gas path SIL conformity declaration (SIL 2) Functional Safety according to IEC 61508 and IEC 61511 Ex versions Possible combinations: see: Table "Ex configurations – principle selection criteria", page 5/16 ATEX II 3G certificate; restricted breathing enclosure, non-flammable gases ATEX II 3G certificate; flammable gases CSA certificate – Class I Div 2 ATEX II 3D certificate; potentially explosive dust atmospheres • In non-hazardous gas zone • In Ex zone acc. to ATEX II 3G, non-flammable gases • In Ex zone acc. to ATEX II 3G, flammable gases BARTEC EEx p control unit "Leakage compensation" BARTEC EEx p control unit "Continuous purging" Order code A28 A29 B03 B04 C20 E11 E12 E20 E40 E41 E42 E71 E72 Clean for O2 service (specially cleaned gas path) Measuring range indication in plain text, if different from the standard setting Special setting (only in conjunction with an application no., e.g. extended measuring range) Extended special setting (only in conjunction with an application no., e.g. determination of cross-interferences) TÜV version acc. to 13th and 17th BlmSchV Y02 Y11 Y12 Y13 Additional units for Ex versions Category ATEX II 2G (zone 1) BARTEC EEx p control unit, 230 V, "leakage compensation" BARTEC EEx p control unit, 115 V, "leakage compensation" BARTEC EEx p control unit, 230 V, "continuous purging" BARTEC EEx p control unit, 115 V, "continuous purging" Ex isolation amplifier Ex isolating relay, 230 V Ex isolating relay, 110 V Differential pressure switch for corrosive and non-corrosive gases Stainless steel flame arrestor Hastelloy flame arrestor Category ATEX II 3G (Zone 2) BARTEC EEx p control unit, 230 V, "continuous purging" BARTEC EEx p control unit, 115 V, "continuous purging" FM/CSA (Class I Div. 2) Ex purging unit MiniPurge FM Article No. Accessories RS 485/Ethernet converter RS 485/RS 232 converter RS 485/USB converter AUTOCAL function with 8 digital inputs/outputs AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA AUTOCAL function with 8 digital inputs/outputs and PROFIBUS DP AUTOCAL function with 8 digital inputs/outputs and PROFIBUS PA Ex i (firmware 4.1.10 required) Set of Torx screwdrivers Article No. A5E00852383 C79451-Z1589-U1 A5E00852382 A5E00064223 A5E00057315 A5E00057318 A5E00057317 A5E34821625 1/98 Siemens AP 01 · 2015 Y17 7MB8000-2BA 7MB8000-2BB 7MB8000-2CA 7MB8000-2CB 7MB8000-3AB 7MB8000-4AA 7MB8000-4AB 7MB8000-5AA 7MB8000-6BA 7MB8000-6BB 7MB8000-2CA 7MB8000-2CB 7MB8000-1AA © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device TÜV, single component (only with additional suffix Z (Y17, Y18)) Component CO (TÜV) Measuring range identification Smallest measuring range from 0 to ... SO2 (TÜV) Largest measuring range from 0 to ... C D 50 mg/m3 1 000 mg/m3 E F 300 mg/m3 3 000 mg/m3 G mg/m3 mg/m3 H K 500 1 000 mg/m3 3 000 mg/m3 10 g/m3 R 30 g/m3 V 100 g/m3 P 5 000 10 000 mg/m3 30 000 mg/m3 100 g/m3 300 g/m3 Largest measuring range from 0 to ... 75 mg/m3 1 500 mg/m3 Smallest measuring range from 0 to ... Largest measuring range from 0 to ... mg/m3 3 000 mg/m3 500 mg/m3 5 000 mg/m3 100 mg/m3 2 000 mg/m3 1 000 mg/m3 10 000 mg/m3 300 mg/m3 3 000 mg/m3 mg/m3 5 000 mg/m3 300 500 3 000 mg/m3 10 g/m3 30 g/m3 100 1 160 g/m3 NO (TÜV) Smallest measuring range from 0 to ... g/m3 300 g/m3 30 000 mg/m3 100 g/m3 300 g/m3 1 000 g/m3 2 630 g/m3 1 000 mg/m3 10 000 mg/m3 3 000 mg/m3 30 000 mg/m3 10 g/m3 100 g/m3 30 g/m3 300 g/m3 100 g/m3 1 250 g/m3 Example for ordering ULTRAMAT 6, TÜV (1-component unit) Component: CO Measuring range: 0 to 50 / 1 000 mg/m3 with hoses, non-flow-type reference compartment without automatic adjustment (AUTOCAL) 230 V AC; without heating, German 7MB2111-0XD00-1AA0-Z +Y17 TÜV, 2 components in series Component CO (TÜV) Measuring range identification Smallest measuring range from 0 to ... Largest measuring range from 0 to ... Smallest measuring range from 0 to ... Largest measuring range from 0 to ... AA 75 mg/m3 1 000 mg/m3 200 mg/m3 2 000 mg/m3 mg/m3 mg/m3 3 000 mg/m3 mg/m3 AB 300 AC 1 000 mg/m3 NO (TÜV) 3 000 10 000 mg/m3 300 1 000 mg/m3 10 000 mg/m3 Example for ordering ULTRAMAT 6, TÜV (2 components in series) Components: CO/NO Measuring range CO: 0 to 75 / 1 000 mg/m3, NO: 0 to 200 / 2 000 mg/m3 with hoses, non-flow-type reference compartment without automatic adjustment (AUTOCAL) 230 V AC; without heating, German 7MB2112-0AA00-1AA0-Z +Y17 Note: for 3 components take both tables into consideration. Ordering information measured component N2O Certification in accordance with AM0028 and AM0034 (Kyoto Protocol) for measuring N2O, measuring range 0 to 300 ppm / 3 000 ppm. Version: Standard device Siemens AP 01 · 2015 1/99 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device 418 394 250 259 184 148 112 77 42 20 ■ Dimensional drawings 0 271 253 237 222 201 127 65 58 311 438 328 26.5 14 ULTRAMAT 6 14 ULTRAMAT 6, field unit, dimensions in mm 1/100 Siemens AP 01 · 2015 480 s 440 1 20 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device ■ Schematics Pin assignment (electrical and gas connections) &RQQHFWRU68%')56 0 *1' *1' 1& 5'7'1 5'7'3 5B/HYHO1 9 0 ,WLVSRVVLEOHWRFRQQHFWEXVWHUPLQDWLQJ UHVLVWRUVWRSLQVDQG 5B/HYHO3 1& 1& *1' 7HUPLQDOEORFN$ 0 0 0 0 0 Ω 0 *1' 5HOD\ 5HOD\ 5HOD\ 5HOD\ &RQWDFWORDGPD[ 9$$&'&UHOD\FRQWDFWV VKRZQUHOD\FRLOKDV]HURFXUUHQW 5HOD\ 5HOD\ %LQDU\LQSXWWR1 %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXW3 ,VRODWHGYLDRSWRFRXSOHU 99 99 *1' *1' %LQDU\LQSXWWR1 %LQDU\LQSXW3 %LQDU\LQSXW3 $QDORJLQSXW1 $QDORJLQSXW3 $QDORJLQSXW1 $QDORJLQSXW3 ,VRODWHGYLDRSWRFRXSOHU 99 99 $QDORJLQSXWV &RUUHFWLRQRI QRQLVRODWHG FURVVLQWHUIHUHQFH P$RU9 3UHVVXUHFRUUHFWLRQ LQWHUQDOUHVLVWDQFHืΩ 1& $QDORJRXWSXW1 $QDORJRXWSXW3 $QDORJRXWSXW $QDORJRXWSXW *1' &RPSRQHQW &RPSRQHQW LIDYDLODEOH $QDORJRXWSXWVLVRODWHG 1RWH $OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNVPXVWEH VKLHOGHGDQGUHVWDJDLQVWWKHHQFORVXUHSRWHQWLDO ULTRAMAT 6, field device, pin and terminal assignment Siemens AP 01 · 2015 1/101 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 1 Field device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field device, pin and terminal assignment of the AUTOCAL board and PROFIBUS connectors 1/102 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Field device a b c d 7 5 1 3 2 4 9 e 8 6 Gas connections Electrical connections 1 Sample gas inlet 2 Sample gas outlet 3 Reference gas inlet (option) 4 Reference gas outlet (option) Clamping gland for pipe Ø 6 mm or ¼" a-c Signal cable (Ø 10 ... 14 mm) (analog + digital): cable gland M20x1.5 d Interface connection: (Ø 7 ... 12 mm) cable gland M20x1.5 e Power supply: (Ø 7 ... 12 mm) cable gland M20x1.5 5 - 8 Purging gas inlets/outlets, stubs Ø 10 mm or 3/8" 9 Connection atmospheric pressure sensor, stubs Ø ¼" ULTRAMAT 6, field device, gas connections and electrical connections Documentation DDocumentationocumentation ■ Selection and ordering data Operating instructions Article No. ULTRAMAT 6 / OXYMAT 6 Gas analyzer for IR-absorbing gases and oxygen • German C79000-G5200-C143 • English C79000-G5276-C143 • French C79000-G5277-C143 • Spanish C79000-G5278-C143 • Italian C79000-G5272-C143 Siemens AP 01 · 2015 1/103 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive ULTRAMAT 6 Suggestions for spare parts 7MB-2124 7MB-2111 7MB-2112 7MB-2111/2 Ex 2 years (quantity) 7MB-2123 ■ Selection and ordering data 7MB-2121 1 O-ring for cover (window) x x x x x x 2 4 C79121-Z100-A24 Cover (cell length 20 ... 180 mm) x x x x x x 2 2 C79451-A3462-B151 Cover (cell length 0.2 ... 6 mm) x x x x x x 2 2 C79451-A3462-B152 O-rings, set x x x x x x 1 C79451-A3462-D501 x x x 2 4 C71121-Z100-A159 Pressure switch x x x 1 2 C79302-Z1210-A2 Flow indicator x x x 1 Hose clip x x x Description 5 years (quantity) Article No. Analyzer unit Sample gas path O-ring (hose clip) Heating cartridge (heated unit) 2 C79402-Z560-T1 x x x 1 C79451-A3478-C9 x x x 1 W75083-A1004-F120 x x 1 W75054-T1001-A150 2 A5E00061505 Electronics Temperature fuse (heated unit) Fuse (device fuse) x 1 Temperature controller electronics, 230 V AC x x x 1 A5E00118527 Temperature controller electronics, 115 V AC x x x 1 A5E00118530 x x x 1 A5E00302916 1 C79165-A3042-B504 1 C79165-A3044-B176 1 C79451-A3474-B605 Fan, 24 V DC (heated unit) Front plate with keyboard x x x Temperature sensor 1 x x x Adapter plate, LCD/keyboard x x x x x Motherboard, with firmware: see spare parts list x x x x x LC display x x x x x Connector filter x x x x x Fusible element, T 0.63 A/250 V x x x x x Fusible element, T 1 A/250 V x x x x x x x x Fusible element, T 1.6 A/250 V Fusible element, T 2.5 A/250 V x x 1 x 1 1 x 1 W75025-B5001-B1 1 W75041-E5602-K2 2 3 W79054-L1010-T630 2 3 W79054-L1011-T100 2 3 W79054-L1011-T160 2 3 W79054-L1011-T250 If the ULTRAMAT 6 is supplied with a specially cleaned gas path for high oxygen content ("Cleaned for O2 service"), please ensure that you specify this when ordering spare parts. This is the only way to guarantee that the gas path will continue to comply with the special requirements for this version. 1/104 Siemens AP 01 · 2015