Keeping Business in Business • • • • • • • • Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C Dedicated, equipment-matched cooling supplies the proper temperature and water flow year round. Factory-tested, piped, wired and charged for hassle-free startup and operation. Clean, non-ferrous cooling loop, required for sensitive medical equipment. Local service and parts for quick and knowledgeable maintenance and service. Optional switch-over to alternate water source provides redundant, continuous operation. Optional remote monitoring systems keep you informed of chiller operation. UL/CSA listing on standard models. ISO 9001 certified company. CLINAC 4/100, 6/100 & 600C Specifications Fluid Specified Water Heat Dissipation—BTUH (kW) 23,900 (7) Fluid Temperature Range—°F (°C) Fluid Flow Rate—GPM (l/m) 50-70 (10-25) 1 (3.8) Fluid Pressure Drop—PSI (kPa) 25-40 (172-276) Note: Performance is based on 95°F condenser ambient. Process Fluid Chiller Specifications Model Number PS024A PS036A PS036A Fluid Circulated Water Ethylene Glycol Propylene Glycol Net Cooling Capacity—BTUH (kW) 24,300 (7.1) 24,400 (7.2) 23,900 (7.0) Leaving Fluid Temperature—°F (°C) 61 (16.1) 47 (8.3) 50 (10) Fluid Flow Rate—GPM (l/m) 1.3 (4.9) 1.5 (5.7) 1.6 (6.1) * (*) * (*) * (*) Chiller Pressure Drop—PSI (kPa) 6.3 (43.4) 2.3 (15.9) 2.8 (19.3) Pump Total Head—PSI (kPa) 41 (282.7) 41 (282.7) 41 (282.7) 3/4 (.56) 3/4 (.56) 3/4 (.56) CLINAC Pressure Drop—PSI (kPa) Pump HP (kW) *Data not available Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C 1 A AIRFLOW IN B D AIRFLOW OUT C* SHADED AREAS INDICATE A RECOMMENDED CLEARANCE FOR COMPONENT REMOVAL, ACCESS, AND AIR FLOW. OPTIONAL REFRIGERANT PRESSURE GAUGES SHADED AREAS INDICATE A RECOMMENDED CLEARANCE FOR COMPONENT REMOVAL, ACCESS, AND AIR FLOW. 2" 18" (457mm) (51mm) 36" (914mm) 36" (914mm) I I H 1" MAX. (25mm) 36" (914mm) H UNIT DIMENSIONAL DATA 1" MAX. (25mm) UNIT DIMENSIONAL DATA 1" NPT Model No. A-Depth in. (mm) B-Width in. (mm) C-Height in. (mm) Weight lbs. (kg) Voltage (60 Hz) Voltage (50 Hz) PS018A 36 (914) 40 (1016) 25.25 (641)) 394 (178) 208/230, 1-phase 200/230, 1-phase PS024A 36 (914) 40 (1016) 25.25 (641) 404 (183) 208/230, 1-phase 200/230, 1-phase PS036A 36 (914) 48 (1219) 32.5 (825) 515 (233) 460, 3-phase 208/230, 1- or 3-phase 400, 3-phase 200/230, 1- or 3-phase PS048A 36 (914) 53 (1346) 38.5 (978) 592 (268) 460, 3-phase 208/230, 3-phase 400, 3-phase 200/230, 1- or 3-phase PS060A 36 (914) 53 (1346) 38.5 (978) 642 (291) 460, 3-phase 208/230, 3-phase 400, 3-phase 200/230, 1- or 3-phase Standard Features Accessories • • • • • • • • • • • • • • • • • • • • Indoor/outdoor cabinet Hermetic compressor Refrigerant dehydrator Expansion valve Refrigerant receiver Stainless steel fluid reservoir expansion tank Pressure relief valve Hot gas bypass valve capacity control Optional Features • Refrigerant gauges and reservoir heater • Stainless steel pump 1050 Dearborn Drive P.O. Box 29186 Columbus, OH 43229 Phone: 800.877.9222 Fax: 614.841.6022 Via Leonardo Da Vinci 8 Zona Industriale Tognana 35028 Piove Di Sacco Italy Phone: 39.049.9719.111 Fax: 39.049.5841257 Liebert Web Site—http://www.liebert.com Emergency water bypass Enviro-source emergency bypass Water level control and alarm system Fluid pressure gauge Check valve Dial thermometer Relief valve Circuit setter Ball valve High temperature thermostat Low flow switch Liebert RCM8 monitor 23/F Allied Kajima Bldg. 138 Gloucester Road Wanchai Hong Kong Phone: 852.2.572.2201 Fax: 852.2.572.1182 ©2000 Liebert Corporation. All rights reserved throughout the world. Specifications subject to change without notice. ®Liebert and the Liebert logo are registered trademarks of Liebert Corporation. All names referred to are trademarks or regisered trademarks of their respective owners. SL-16548 (07/00) Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C 2