ASM-SENSOR 3193

3193 POWER HiTESTER
Power measuring instruments
Printer is optional unit
DC/0.5Hz to 1MHz broad-band POWER HiTESTER measures up to 6 systems simultaneously.
Wide Spectrum Power Meter for
Comprehensive Device Assessment
3194
MOTOR / HARMONIC
HiTESTER
3194 MOTOR / HARMONIC HiTESTER
provides analysis of high-order harmonics
up to the 3000th order. Additionally, with the
optional 9603-01 EXTERNAL SIGNAL
INPUT UNIT installed, the HiTESTER can
directly measure torque and rotation speed, an
essential feature for evaluating the performance
of inverter motors. This makes it easy to
construct measurement systems.
The 3193 POWER HiTESTER is a multi-function power meter
for use with single phase power lines to 3-phase, 4-wire circuits.
Accommodating up to 6 units, it is not only capable of measuring up to
6 single phase systems, but can simultaneously measure the input and
output of a 3-phase inverter and provide effective power measurements.
Additionally, it supports harmonic analysis and flicker measurement
(optional), features which are essential for overall device assessment.
Standard features include a GP-IB/RS-232C interface and 3.5-inch
floppy disk drive, making it easy to feed data to a personal computer
for processing and analysis. This unit is ideal for those requiring greater
efficiency in electrical device assessment.
1
Broad coverage, high accuracy, and well-developed interface
Complete with functions that answer
all your power measurement needs.
Features
■ Wide range of measurement functions
■ High Basic Accuracy of ±0.2%
Capable of measuring voltage, current, active /reactive /apparent
power, power factor, phase, frequency, and current, and of integrating
power according to polarity, the 3193 (or 3194) also provides wave peak
and efficiency measurements that are essential to device assessment.
Measurements of even greater precision can be obtained using the
optional 9600 to 9602 input unit, which provides a basic accuracy
of ±0.1% rdg,±0.1% f.s. (With the 9602, the accuracy of the clamp-on sensor is a factor
affecting total accuracy during power measurement.)
■ Can Measure Motor Output
■ A Variety of Interfaces for Differing Needs
With the optional 9603 EXTERNAL SIGNAL INPUT UNIT , the
★ Connecting to a PC
HiTESTER can take analog input in from torque and revolution
measurements and use that information to calculate motor output.
The 3.5 inch floppy disk drive and RS-232C / GP-IB interface,
provided as standard features, make it possible to connect the power
meter directly to a PC, allowing efficient
measurement, management and analysis of data.
■ Measurement for Minute Stand-by Power also Available
(by special-order product)
The 9600 and 9601 input units have 10-times improved
current sensitivity, and currents starting from the 20.000mA
range can be measured. (Please ask for further information.)
★ Connecting to a Recorder
■ High Visibility Color LCD
monitor output (current and voltage only) as standard features, the
Featuring a wide viewing angle, the
color LCD displays a variety of items
simultaneously, making it ideal for
quickly grasping power usage on the
system being measured. Expanded
display is possible for any four
selected items.
HiTESTER allows recording of changes and
Enlarged Indication
■ Harmonic and Flicker Analysis
Harmonic and flicker analysis are possible when using the optional
9605 HARMONIC / FLICKER MEASUREMENTS UNIT(3193 only).
With 8 selectable D/A outputs and voltage, current and power analog/
transient fluctuations in waveforms using a
recording unit.
★ Connecting to a Printer
Data can be output to the optional 9604 PRINTER UNIT.
Print type
Paper width
Main
functions
: Thermal line dot printing
: 72mm
: printing of items measured, hard copy output
of displayed screens, printout of meter settings,
printout of various times (such as interval
time, timer time, and realtime control time).
Printouts are performed either automatically,
upon input of an external control signal, or
synchronized with an integrator.
2
Other Analysis Functions
Integration According to Polarity
Peak Measurements Function
Voltage and current wave peaks
can be measured. The Peak
Hold function can be used to
find peak values and effective
maximums for motor rush
current waves.
Positive, negative, and total
current and power can be
integrated simultaneously for
all channels. This makes it
possible to grasp the income
and outflow of power at a
glance.
Provides Analog and D/A output.
Analog (voltage, current, and
effective power) and D/A
outputs (any selected eight
items) are output as a 5V range
full scale value. (Except for the
1000V range), 100ms response
time can be obtained by using
the FAST setting.
peak
value
Wave Monitor Output
With the voltage and current
ranges, waveforms are output
as 1V full scale values,
allowing waveforms to be
monitored using devices such
as recorders or synchroscopes.
Input wave
60Hz, sinwave
Active power
Analog output
Input wave
50kHz
Pulse
Voltage
Monitor Output
current
Monitor Output
Active power
D/A output
20μs
100ms
● 3 Types of Averaging Function
● 3 Types of Built-in Low Pass Filters
Select from time average, moving average and exponential average.
With the frequency ranges, LPF and HPF can be used in combination, allowing
Selectable cutoff frequencies (500/5k/300kHz) allow extraction of the
frequency component of fundamental inverter waveforms and provide
data compatibility with previous instruments.
measurement of fundamental waveforms and carrier waveforms of inverters.
● Choose from Three Types of Calculation Algorithms
● 3ch Frequency Measurement Function
● Efficiency Calculation Function
Three efficiency calculations can be obtained simultaneously from measured voltage values.
Three selectable algorithms are provided for calculating apparent power
and reactive power, providing compatibility with previous devices
■ Choose from a variety of input units according to application
■ Simultaneous Measurement of Multiple Systems
Three types of input units are available, including the 9602 clamp
input meter, which can be used with current levels exceeding 50A
under live circuit conditions, as well as the 9600 and 9601 which
accept direct input of up to 1000V/50A.
Since all units are mutually isolated, the primary and
secondary sides of devices or disparate power lines can be
measured simultaneously. Simultaneous measurement of single
phase 6 wire or 3 phase 2 wire systems which previously
required multiple units, can now be handled with one. What's
more, measurements of all devices can be taken at the same
instant, providing a powerful tool for integrated, all-round
device assessment.
● 9600 · · · DC/0.5Hz to 1MHz wide band
● 9601 · · · 5Hz to 100kHz, for AC only
● 9602 · · · DC/0.5Hz to 200kHz clamp input
A variety of clamp on sensors are available, including models 9270 to 9272 for
AC only, and models 9277 to 9279 for AC/DC.
1ch
2ch
3ch
4ch
5ch
6ch
1 ø2 W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 2W
1 ø 3W / 3 ø 3W
1 ø 3W / 3 ø 3W
1 ø 2W
1 ø 2W
1 ø 3W / 3 ø 3W
1 ø 3W / 3 ø 3W
1 ø 3W / 3 ø 3W
1 ø 3W / 3 ø 3W
3V3A (3 ø 3W) / 3 ø 4W
1 ø 2W
1 ø 2W
1 ø 2W
3V3A (3 ø 3W) / 3 ø 4W
1 ø 3W / 3 ø 3W
1 ø 2W
3V3A (3 ø 3W) / 3 ø 4W
3V3A (3 ø 3W) / 3 ø 4W
Effect of common mode voltage
Error (%)
Influence (%)
9600 Frequency characteristics (at 15W[15V✕1A] range, response slow)
frequency (Hz)
frequency (Hz)
3
Can measure all items at same point in time
The Power Analysis Station
■ Applications of 3193
Example of Application With Power Converter
Example of Assessment Trial of EV ( Electric Vehicle)
Separate charge/generation
integration capability.
Measurement under live
circuit conditions.
(Clamp input)
Measure mixed AC and DC components with a single unit.
3ch
U /I /P
Separate charge / generation integration ±Wh
1, 2 ch
3 phase after DC/AC, AC U /I /P /Wh
1, 3, 4ch
AC U /I /P, Primary to secondary
efficiency
2, 5 ch
Monitoring of DC U /I /P after DC/AC
■ 3194 Performs Comprehensive Evaluation of 3-phase Inverter Motors
Capable of measuring carrier frequencies on the secondary side of inverters. Also allows analysis to be synchronized with motor rotation.
1, 2, 3 ch (Primary)
V/A/W/PF,
Harmonic analysis
4, 5, 6 ch (Secondary)
V/A/W/PF and Motor
output, Efficiency,
Harmonic analysis
3194 Specifications
(Harmonic Waveform Analysis Function)
Measurement lines : Single-phase 2- and 3- wire, three-phase 3- and 4-wire
No. of channels
: Up to 3 channels from channels 1 to 6
Measurement range : Fundamental frequency: 10 Hz to 4.5 kHz
Harmonic Waveform Analysis Functions
Harmonic analysis to the 3000th order
9603-01 Specifications
Torque, rotation speed, and motor power can be measured by inputting analog
torque and r/m signals to the 9603-01 EXTERNAL INPUT UNIT. Further, an
input terminal is provided for use with a strain gauge-type torque sensor.
9603-01 EXTERNAL INPUT UNIT
: 2 channels
Input channels
: Strain gauge sensor or DC voltage input (BNC)
Ch A
: Pulse or DC voltage input (BNC)
Ch B
Strain gauge input specifications (dedicated connector)
Applicable converter : Strain gauge type converter (bridge resistance 350Ω - 1.5kΩ)
Measurement range : 1 mV/V / 1.5 mV/V / 2 mV/V
Measurement accuracy : ±0.1%rdg. ±0.065%f.s.
: PRC03-23A10-7F (manufactured by Tajimi)
Connector
DC voltage input specifications (ch A / ch B common BNC connector)
: 200kΩ ±5% (differential)
Input resistance
Measurement range : ±1.0000 / ±5.0000 / ±10.000V
Measurement accuracy : ±0.1%rdg. ±0.1%f.s. (23˚C ±5˚C, not higher than 80%rh)
Pulse input specifications (ch B, BNC connector)
Frequency measurement : 1Hz to 100 kHz
Common specifications
: ±5V f.s. (Output accuracy : Measurement accuracy ±0.2%f.s)
Analog output
Measurement method
A/D resolution
Windowing type
Crest factor
PLL source
Ext. synchronization
signal
Measurement items
Harmonic wave
measurement items
Screen displays
Wiring conversion
Sort function
Averaging function
Frequency ranges
: PLL or external clock
: 12 bits
: Rectangular tiling (with gap between windows)
: Up to 2.5 (voltage, current)
: U, I or external synchronization
: Input to a rear panel control terminal on the 3194 main unit
Input level: 1V to 10 Vrms, Division function: 1/1 to 1/255
: RMS voltage, RMS current, effective power value,
frequency, ±U peak, ±I peak
: Harmonic level, percentage and phase angle of harmonic
wave, Total harmonic distortion (THD-F, THD-R)
: List, graph, vector and waveform displays
: Δ-Y voltage conversion, Y-Δ voltage conversion
: Sorts according to decreasing order of analysis, displays up to 50th order
: Index average with time constant of 1.5 sec
:
Fundamental Sampling
Frequency
speed (Hz)
10 - 17.5 Hz f × 8192
PLL
Synchronization
Ranges
Fixed clock
Window
Width
1 waveform
Analysis Order
3000 (
10kHz or less)
17.5 - 35 Hz
f × 8192
1 waveform
3000 (
10kHz or less)
35 - 70 Hz
f × 8192
1 waveform
3000 (100kHz or less)
70 - 140 Hz
f × 4096
2 waveforms 1500 (100kHz or less)
140 - 280 Hz
f × 2048
4 waveforms
800 (100kHz or less)
280 - 560 Hz
f × 1024
8 waveforms
400 (100kHz or less)
560 - 1120 Hz
f × 512
16 waveforms
200 (100kHz or less)
1120 - 2240 Hz
f × 256
32 waveforms
100 (100kHz or less)
2240 - 4500 Hz
f × 128
64 waveforms
50 (100kHz or less)
----
50 × 8192
Fixed
2 waveforms 3000 (100kHz or less)
Note 1: Analysis order accuracy is restricted to the frequency in brackets.
Note 2: With PLL synchronization in the range of 10 to 35 Hz, an anti-aliasing filter of about
15 kHz is used, and with PLL synchronization in the range of 35 Hz to 4.5kHz, an
anti-aliasing filter of about 120 kHz is used.
4
■ Applications of 3193
Support up to Harmonic / Flicker analysis
when using the optional 9605 HARMONIC / FLICKER MEASUREMENTS UNIT .
Graph Display of Harmonics
List Display of Harmonics
Voltage, current and power
can be analyzed and displayed
by bargraphs of harmonic
amplitude, content and phase
angle.Voltage, current and power
can be displayed simultaneously
for a single channel, or a single
parameter can be displayed
simultaneously for each of three
channels.
The harmonic list display
shows the amplitude, proportion, phase angle and
distortion for each harmonic
of voltage, current and power.
Displaying only proportion,
or two parameters simultaneously, such as amplitude
and phase angle, is selectable.
Vector Display of Harmonics
Waveform Display
The harmonic vector display
shows the voltage, current and
phase angle for each harmonic,
making clear the voltagecurrent phase relationship.
The waveform display shows
one cycle of the voltage and
current waveforms. RMS and
peak values can be displayed
along with voltage and current
waveforms, or voltage and
current waveforms for up to
three channels can be displayed
at the same time.
Monitor Display
Flicker Measurement Display
Displays data during measurement
in real-time. Display can also be
switched to D measurement and
"Pst" value.
9605 Specifications
The relative "d" voltage change
∆V/V and the instantaneous
flicker value "S(t)" can be
displayed in a time series,
so past variations are clearly
displayed.
(optional)
: Installs in the 3193 main unit
Installation
Measurement lines : Single-phase 2- and 3-wire, three-phase 4-wire
No. of channels : Up to 3 channels within channels 1 to 6, depending
on 3193 wiring mode
Output functions : Floppy disk, RS-232C, GP-IB, printer
● Harmonic Waveform Analysis Functions (IEC61000-3-2*1)
Measurement range : Fundamental frequency: 1 to 440 Hz
PLL system (5 to 440 Hz), external clock system (1 to 5 Hz)
Orders analyzed : Up to 50th harmonic (with 1 to 250 Hz fundamental)
Window width : 16 cycles (for 40 to 70 Hz fundamental)
Windowing type : Rectangular tiling (no gap between or overlap of windows)
Amount of data analyzed : 512 points (for 40 to 70 Hz fundamental)
: Up to 4 (current), and up to 3 (voltage)
Crest factor
Measurement items : Harmonic level, percentage and phase angle of each order
of harmonic wave for each of voltage, current and power.
Total up to 50th harmonic (of 40 to 70 Hz fundamental) for
voltage, current and power. Total harmonic distortion for
voltage and current (THD-F and THD-R)
Measurement of voltage, current, active power, peak voltage
and peak current values of the fundamental
*1. The 9605 does not support limit values, so pass/fail decisions based on limit values, classification
of special waveforms (A to D) and 2.5-minute measurement function for measuring transient harmonic
currents are not supported.
: Every 1 window (except during communications with other devices)
Update rate
Screen displays : List, graph, vector and waveform
: Harmonic levels; at 45 to 66 Hz
Accuracy*3
Voltage/current: ±0.5%rdg.±0.05% f.s.
Active power: ±1.0%rdg.±0.1% f.s.
For 45 to 66-Hz fundamental, effective input is 0.1 to 110%
of range, total error of harmonic current when PLL is locked
is 5% of the limit value or within 0.2% of the rated current
of the device under test (no limitation with 9602).
● Flicker Measurement Function (IEC61000-3-3*2)
Measurement range : Fundamental frequency; 45 to 66 Hz, PLL synchronization system
Analysis items : dc (relative constant voltage change), d max (max. relative constant
voltage change), d(t)200ms (relative voltage change per time), P0.1/
P1s/ P3s/ P10s and 30s (cumulative probability), Pst (short-term flicker
value), Plt (long-term flicker value)
Screen displays : Measured value, CPF, Pst list, Monitor
Accuracy*3
: Within ±5% rdg. of limit on the limit curve line (Pst = 1) specified
for voltage fluctuation in IEC60868-0 (d measurement is the same)
RMS voltage ±0.5%rdg. ±0.05%f.s. (45Hz~66Hz)
*2. Decision function not included.
*3. The reading accuracy of the input unit must combined with the analysis accuracy shown above. When
used with a clamp sensor, accuracy and frequency characteristics of the clamp must be added to the
analysis accuracy above.
5
■ Optional Input Unit Specifications
The 9600 and 9601 are also the products for stand-by power measurement with 10-times
improved sensitivity for current ranges. Please ask for detailed specs.
9600 AC/DC DIRECT INPUT UNIT
Voltage
Current
Active power
Measurement
range
Max.operating input (55Hz)
Crest factor
Input resistance
Accuracy
0.5
9602 AC/DC CLAMP INPUT UNIT
Voltage
Current
Active power
6.0000/15.000/30.000/
200.00/500.00mA/
Depends on
6.0000/15.000/30.000/
60.000/150.00/300.00/ 1. 0000/2.0000/5.0000/ combination of voltage 60.000/150.00/300.00/
600.00V/1.0000 kV 10.000/20.000/50.000 A
and current ranges
600.00V
1000Vrms/1500 V peak
65Arms/100 A peak
650Vrms/850Vpeak
Lower of either (measured range ✕ 6) / measured value
or maximum permissible rated peak / measured value
2MΩ±5%
500.00mA to 500.00A
Depends on
(Depends on clamp-on combination of voltage
sensor)
and current ranges
(Depends on clamp-on sensor)
Lower of either (measured range ✕ 6)/ measured value or
maximum permissible rated peak/measured value
1mΩ max.
2MΩ±5%
200kΩ±5%
(Accuracy assured at 23˚C±5˚C (73˚F ±9˚F) at 80% R.H., power factor = 1, sine wave input, in-phase voltage 0, after DMAG)
DC
±0.1%rdg.±0.2%f.s.
←
←
±0.1%rdg.±0.2%f.s.
←
←
to 1Hz
±0.5%rdg.±0.5%f.s.
←
←
±0.5%rdg.±0.5%f.s.
←
←
1
to 10Hz
±0.2%rdg.±0.2%f.s.
←
←
±0.2%rdg.±0.2%f.s.
←
←
10
to 45Hz
±0.1%rdg.±0.2%f.s.
←
←
±0.1%rdg.±0.2%f.s.
←
←
45
to 66Hz
±0.1%rdg.±0.1%f.s.
←
←
±0.1%rdg.±0.1%f.s.
←
←
66Hz to 10kHz
±0.1%rdg.±0.2%f.s.
←
←
±0.1%rdg.±0.2%f.s.
←
←
10 k to 50kHz
±0.3%rdg.±0.3%f.s.
←
←
±0.5%rdg.±0.5%f.s.
←
←
±0.5%rdg.±0.5%f.s.
←
±0.3%rdg.±0.5%f.s.
±5%f.s. ±1.0%rdg.
±1.5%f.s. ±10.0%f.s. ±15%f.s.(up to 200kHz)
←
±30%f.s.(up to 200kHz)
Less than 5 A Greater than 5 A Less than 5 A Greater than 5 A
50 k to 100kHz
±0.5%rdg.±0.5%f.s.
±0.5%rdg.
±0.5%rdg.
±0.5%f.s. ±2.5%f.s. ±0.5%f.s. ±5.0%f.s.
100k to 300kHz
±0.5%rdg.±0.5%f.s.
±0.5%rdg.
±0.5%f.s.
±1.5%rdg.±0.5%f.s.
±1.0%rdg.
±0.5%f.s.
±1.0%rdg.
±2.5%f.s.
±2.0%rdg.±1.0%f.s.
±2.0%rdg.
±1.0%f.s.
±2.0%rdg.
±2.5%f.s.
500k to 700kHz
±10.0%f.s.
±10.0%f.s.
±15.0%f.s.
700k to 1 MHz
±15.0%f.s.
±15.0%f.s.
±30.0%f.s.
300k to 400kHz
400k to 500kHz
■ Compatible
Clamp (Optional)
9270· 9271
9279
9272
9709
9277/9278
With the 9602, the accuracy of the clamp-on sensor is a factor affecting
total accuracy during current measurement, as are phase and frequency
characteristics. See page 7 for details on the clamp-on sensor specifications.
Measurement
range
Max.operating input (55Hz)
Crest factor
Input resistance
200.00/500.00mA/
Depends on
60.000/150.00/300.00/6
1. 0000/2.0000/5.0000/ combination of voltage
00.00V/1.0000 kV
10.000/20.000/50.000 A
and current ranges
1000Vrms/1500 V peak
65Arms/100 A peak
Input voltage (V)
9601 AC DIRECT INPUT UNIT
Assured Accuracy Range for Input Frequency of the 9600
9601
and 9602 each have assured ranges for input frequency.
Voltage
Current
Active power
Lower of either (measured range ✕ 6) / measured value
or maximum permissible rated peak / measured value
2MΩ±5%
1mΩ max.
±2.5%f.s.
←
←
10 to 20Hz
±1.0%f.s.
←
←
20 to 45Hz
±0.1%rdg.±0.2%f.s.
←
←
45 to 66Hz
±0.1%rdg.±0.1%f.s.
←
←
66Hz to 5kHz
±0.1%rdg.±0.2%f.s.
←
←
5k to 10kHz
±0.2%rdg.±0.4%f.s.
←
←
10k to 20kHz
±1.0%f.s.
←
←
20k to 50kHz
±2.5%f.s.
←
←
50k to 100kHz
±10.0%f.s.
←
←
←
5
to 10Hz
Input current (A)
Accuracy (Accuracy assured at 23˚C ±5˚C(73˚F ±9˚F) at 80% R.H., power factor = 1, sine wave input, in-phase voltage 0)
Frequency (Hz)
Note 1: Assured accuracy ranges for different response settings are as follows: FAST (0.1 sec) to DC and greater than 50 Hz, MID (0.8 sec) to DC and greater than 10 Hz,
SLOW (5.0 sec) to DC or greater than 0.5 Hz
Note 2: Assured accuracy ranges for combined mode measurement are 10 Hz or greater for the AC mode, and DC only for the AC+DC mode or DC mode.
■ Calculation algorithm
1ø2W
S
U
M
3ø3W
(3V3A)
(Indicated only for single phase, 2 wire and 3 phase, 3 wire (3V3A).Two additional calculation algorithms can be selected for apparent/reactive power)
Voltage
U1
U1·2·3=
U 1+ U 2 + U 3
3
Current
I1
I1·2·3=
I 1+ I 2 + I 3
3
Active power Apparent power Reactive power Power factor
λ 1 =s1 P1/S1
P1
S1 =U1✕I1
P1·2·3
P1·2·3 =P1+P2
λ 1·2·3 = su
S1·2 ·3
Phase
ø1 =s1 cos -1 λ 1
ø1·2·3=su cos -1 λ 1·2·3
Note 1: The above calculation algorithm is for a single phase, 2 wire input to ch 1, and 3 phase, 3 wire input to ch 1/2/3 (3 voltage, 3 current).
Note 2: The "s" before each power factor or phase operation indicates the lead or lag of current phase in relation to voltage. The "-" sign means current phase leads voltage
and when there is no symbol, it lags. "su" is "-" when the sum of reactive power is negative and "+" (but unsigned) when it is positive.
6
■ Basic specifications
Measurement line : Single phase 2 wire, single phase 3 wire, 3 phase 3 wire
(3V3A is possible), 3 phase 4 wire
Measurement item : When using 9600, 9601, 9602 (optional)
Voltage, current, voltage/current peak, effective/reactive/apparent power,
power factor, phase, frequency, current/power integration, load rate, efficiency.
When using the 9603 (optional)
Voltage, torque, r/min, frequency, motor output.
When using the 9605 (optional)
Harmonic, waveform, voltage fluctuation / flicker measurement function.
Display indication range : At the lowest range in the DC mode of Models 9600 & 9602: 0.2% to 130%
At the lowest range in the AC+DC mode of Models 9600 & 9602: 0.5% to 130%
At the 200mA range of Model 9601: 0.5% to 130%
At 0.1% to 130% of all other ranges.
All range is zero suppressed at less than lower % value.
Valid input range for voltage, current, and power is 0.5% to 110%
: 6.4 inch TFT color LCD (640 × 480 dot)
Display
Display resolution : 99999 count (except with integration), 9999999 count (with integration)
Rectification method : Switchable between RMS (true root mean square value) and MEAN
(average rectified RMS indication). When combined mode DC is
selected, its not possible to switch between them.
Display update rate : 8 times/sec.
Combined mode : DC, AC + DC, AC (AC only when used in combination
with 9601 or 9602 + AC clamp-on sensor)
Analog response time : FAST (0.1 sec.), MID (0.8 sec.), SLOW (5.0 sec.)
(Time required for stabilization to within ±1% when input is changed
from 0% to 90%, or 100% to 10% of range.)
[Motor output (Pm) measured]
Measurement method
Effective when 9603 ch A is torque (any of N/ m, mN/m, kN/m, kgf/m,
kgf/cm), and ch B is r/m
Display indication range : 0.1% to 130% of 9603 voltage range (polarity not indicated)
Calculation algorithm : Set to required format
[Efficiency measurement]
Calculable factors : Maximum of 3 formats
Calculated items : P for each input unit, Pm when combined with 9603
(Items measured with 9605 not allowed)
[D/A output]
Number of channels
Output impedance
Output items
Output voltage
Output update rate
:
:
:
:
:
8ch (12 bit D/A converter polarity + 11bit)
100Ω±5%
Outputs 8 arbitrarily selected items
DC±5V/f.s.
16 times/sec
[FDD]
Compatible media : 3.5-inch, 2HD (1.2MB/1.44MB)
: MS-DOS
Format
: Function for saving and loading settings, and saving
Functions
measured values. Function for outputting measured values.
Floppy disk formatting function, file renaming and deleting
[Interface]
: Conforms to IEEE-488.1 1987, with reference to IEEE-488.2 1987
: Start-stop synchronous, with baud rate of 2400 or 9600 bits/sec
Low pass filter : OFF / 500Hz / 5kHz / 300kHz (-3dB) For 9601, 5k
/ 300kHz not available.
Polarity detection regulation filter : OFF / 200Hz (-3dB)
Analog output : Voltage / current / active power
DC ±5V f.s (1000V range is DC ±3.333 V f.s.)
Monitor output : Voltage / current: 1Vrms f.s. (1000V range is 0.6667 Vrms f.s.)
GP-IB
RS-232C
[Voltage/ Current/ Active power measurement]
Control signal level
Measurement range : See Page 5 specifications for individual input units
[Other functions]
[Integration measurement]
Scaling
Averaging
Number of measurements : 64 times/sec
Measurement range : 0 to±9999999 TAh / TWh (integration time up to 10,000 hours)
[Power factor/ Phase angle measurement]
Measurement range
: -1.0000 (lead) to 0.0000 to 1.0000 (lag)
-180˚(lead) to 0.00˚to 180.00˚(lag)
[External Control]
Functions
[Wave peak measurement]
: Integration start / stop control, Integration data reset, External
A/D (For display update when power meter display is in hold mode),
Manual print control, FD save control
: From 0 / 5V logic signal or short//open contact signal
: PT/CT ratio Set range 0.0001 to 10000
: Time average (set interval time, timer time, and average of
realtime control time)
Moving average (number of samples: 8/16/32/64)
Exponential average (attenuation factor: 8/16/32/64)
Multilingual display : Japanese/English screen display switching
Set time (all types) : Interval control time (10 sec to 100 hours) in 10 sec increments
(When used in combination with FDD or printer, auto select increments)
[Frequency measurement]
Number of channels : Max. 3ch (selection of voltage or current for arbitrary channel)
Effective input range : 0.5 Hz to 2 MHz
Measurement range : Auto / 50Hz/ 500Hz / 50 kHz / 2 MHz
9603 (optional) external input unit required
: Digital calculation from measured voltage or pulse signal
Timer control time (1 min to 10000 hours) in 1 minute increments
Realtime control time, 1 minute increments
[Harmonic / Flicker measurement]
9605 (optional) required
Measurement item : See Page 4 specifications for 9605
Measurement items : Select either voltage or current for each unit (Shows absolute value of max.)
Effective Input Range : Effective value of sine wave is within effective input permissible in the range
■ Measurement accuracy (23˚C±5˚C (73˚F ±9˚F), Less than 80% rh, warm up time greater than 1 hour, sine wave input, power factor = 1, in-phase voltage = 0)
:
V, A, W
Apparent / reactive :
power
:
Integration
:
Power factor
:
Phase angle
:
Frequency
Wave peak
Per accuracy table on page 5
±1 dgt. with respect to calculation from measured value (U, I, P)
sum value is max. ±3dgt.
±1dgt with respect to values calculated from measurements ( I, P)
Max. ±3dgt with respect to values calculated from measurements (U, I, P)
Max. ±3dgt with respect to values calculated from measurements (U, I, P)
±0.1% rdg. ±1 dgt. (0˚C to 40˚C(32˚F to 104˚F), for sine wave
input between 10% to 130% of U/I range)
: ±1% (at 0.5Hz to 1kHz), ±2% (at 1kHz to 10kHz), ±10%
(at 10kHz to 100kHz)
■ General Specifications
Location for use : Indoors, altitude to 2000 m
Ambient use humidity : Power meter, 0˚Cto 40˚C (32˚F to104˚F), rh below 80%(no condensation)
When using FDD / printer, 5˚C to 40˚C (41˚F to 104˚F), rh below 80%
Ambient storage humidity : -10˚C to 50˚C (14˚F to 122˚F), rh below 80% and no condensation
Insulation resistance : More than 100MΩ from DC500V
Greater than 50M( at DC500V between U/I input terminal and
unit case, between U/I input terminal and power supply plug
(for the 9600/9601), between U input terminal and clamp input
terminal, between U input terminal and unit case, and between
U input terminal and power supply plug (for 9602)
Withstand voltage : AC5.55kV between U/I terminal and unit case, between U/I
terminal and power supply plug (for 9600 and 9601), between U
(50/60 Hz, 1 minute)
terminal and clamp input terminal, between U terminal and unit
case, between U input terminal and power supply plug (for 9602)
Motor output
Efficiency
: ±1dgt for calculations of each measured value
: Max. ±7 dgt with respect to values calculated from measurements of items substituted into algorithm
Thermal coefficient : Within ±0.03% f.s/ ˚C
Effect of in-phase : Within ±0.05% f.s
(1000 Vrms, 50/60Hz, between shorted voltage input terminals and case)
voltage
Effect of power factor : ±0.15% f.s (power factor = 0)
: ±25ppm ±1dgt. (0 to 40˚C (73˚F to 41˚F))
Actual time
: Display accuracy - ±0.2% f.s
D/A output
Analog output : Display accuracy - ±0.2% f.s
Monitor output : Display accuracy - ±0.2% f.s (less than 100kHz)
Display accuracy - ±3dB (100k to 1MHz)
: Safety
EN61010-1: 2001
: EMC
EN61326: 1997+A1: 1998+A2: 2001 class A
EN61000-3-2: 2000
EN61000-3-3: 1995+A1: 2001
: AC100V/120V/200V/230V (switched automatically), 50/60Hz
Power supply
Maximum rated power : 150VA max.
Dimensions, mass : Approx 430 W ✕150 H ✕ 370 D mm, Approx13 kg
Certifications
(Approx 16.93ʺ(W ) 5.91ʺ (H)14.57ʺ (D), Approx 458.6 oz.)
(in configuration including 9600 5 6ch, 9603, 9604)
(Not including projections such as terminals, feet, and handles)
Accessories
: Power cord 1, ground adapter (3P to 2P) 1,
connector 1
7
■ CLAMP-ON SENSOR(Optional) Specifications
To use the Clamp-On sensor, be sure to order the factory option 9602 AC/DC clamp input unit.
AC / DC
AC
9270
9290
1500A
ø40mm
80mm-wide busbar
20Hz to 4kHz
CT ratio 10 : 1
9271
9272
3193 or 3194
20A
ø20mm
5Hz to 50kHz
200A
20 / 200A
ø20mm
DC to 100kHz
+
ø20mm
5Hz to 50kHz
9602
20A/200A
ø46mm , 50✕20mm busbar
5Hz to 10kHz
9277 / 9278
9277
500A
ø40mm
DC to 20kHz
9279
500A
ø36mm
DC to 100kHz
9709
Ordering information
3193 POWER HiTESTER (main unit only)
3194 MOTOR / HARMONIC HiTESTER
(main unit only)
Measurements cannot be taken with a 3193 POWER HiTESTER and 3194 MOTOR / HARMONIC
HiTESTER unit only.
A factory option input unit must be purchased.
Pattern A
Pattern B
Pattern C
Pattern D
Pattern E
Pattern F
Pattern G
1ch
2ch
3ch
4ch
5ch
6ch
Notes on input unit selection
1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
)
1 ø 3W / 3 ø 3W (
×2 ) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) ● Use the same input unit for a particular
measurement line.
1 ø 3W / 3 ø 3W (
×2 ) 1 ø 3W / 3 ø 3W (
×2 ) 1 ø 2W (
) 1 ø 2W (
) ● Install units in succession starting
1 ø 3W / 3 ø 3W (
×2 ) 1 ø 3W / 3 ø 3W (
×2 ) 1 ø 3W / 3 ø 3W (
×2 )
from channel 1.
3V3A (3 ø 3W) / 3 ø 4W (
×3 ) 1 ø 2W (
) 1 ø 2W (
) 1 ø 2W (
) ● For the 9603 or 9603-01, only one unit
3V3A (3 ø 3W) / 3 ø 4W (
×3 ) 1 ø 3W / 3 ø 3W (
×2 ) 1 ø 2W (
) can be installed.
3V3A (3 ø 3W) / 3 ø 4W (
×3 )
3V3A (3 ø 3W) / 3 ø 4W (
×3 )
● When the 9602 is selected, use the
(
Options that can be installed at the factory
optional clamp-on sensor.
Options
(specify at time of order)
9600
9601
9602
9603
) : 9600, 9601, 9602 can be selected.
*
AC/DC DIRECT INPUT UNIT
AC DIRECT INPUT UNIT
AC/DC CLAMP INPUT UNIT
EXTERNAL SIGNAL INPUT UNIT (3193 only)
*
*
9603-01 EXTERNAL SIGNAL INPUT UNIT (3194 only)
9604
PRINTER UNIT
9605
HARMONIC / FLICKER MEASUREMENTS UNIT
(3193 only)
*The voltage cable is not supplied. Contact your dealer when it is
necessary to use the clip type leads.
*
*
9270
9271
9272
9277
9278
9279
9709
9290
9232
CLAMP ON SENSOR (20A AC)
CLAMP ON SENSOR (200A AC)
CLAMP ON SENSOR (20/200A AC)
UNIVERSAL CLAMP ON CT (20A AC/DC)
UNIVERSAL CLAMP ON CT (200A AC/DC)
UNIVERSAL CLAMP ON CT (500A AC/DC)
AC/DC CURRENT SENSOR (500A AC/DC)
CLAMP ON ADAPTER (1500A AC, ratio10 :1)
RECORDING PAPER (10 m, 10 roll, For 9604)
*No CE marking
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@ hiokiusa.com
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090/0092
FAX +86-21-6391-0360
E-mail: [email protected]
All information correct as of Feb.13, 2006. All specifications are subject to change without notice.