FUJI EDS10-72B

ALARM SETTER WITH DIGITAL DISPLAY
DATA SHEET
PCP
This product is a digital type alarming indicator whose front
face measures 48 x 48 mm.
FEATURES
1. Eight alarm modes are selectable for a variety of
applications.
2. Front-face waterproof structure (NEMA-4X) is available.
3. SPDT contact relay output is provided in a pair.
4. Input signal and measurable range are changeable by
key operation.
SPECIFICATIONS
Operational specifications
Input section: No. of inputs: 1 point
Input signal: See table 1.
Measuring range: See table 1.
Burnout function: Upscale or downscale burnout settable
The instrument with burnout releasing function can
also be specified.
Input sampling period: 0.5 sec
Input impedance:
Voltage input; 1MΩ or 450kΩ or more
Current input; 250Ω (external mounted
resistor)
Thermocouple; 1MΩ or more
Allowable resistance of signal supply:
Thermocouple input; 100Ω or less
Voltage input; 1kΩ or less
Alarm output: No. of outputs 2 points of SPDT contact
Kinds of alarm: See table 2. each point independently settable
Contact capacity: 220V AC/30V DC, 3A (resistance load) 220V AC/30V
DC,1A (inductive load)
Alarm value settable range:
-5 to 105% FS (each point independently settable)
Hysteresis settable range:
0 to 102% FS (each point independently settable)
Alarm action delay time:
Settable within 1 to 10 sec
Alarm action delay time at power-on:
Settable within 0 to 20 sec
Indicating section:
Numerical indication:
7-segment, 4-digit LED
(upper stage red, lower stage green)
Indicated contents: Alarm set value 1, alarm
set value 2, measured
value
Status indication: Red/green LED lamp. alarm
occurrence, power ON,
measured value indication
Indicating accuracy:
Thermocouple input: ± 0.5% FS ± 1 digit ± 3˚C
(For B thermocouple 0 to 400˚C, ± 5% FS ±1 digit ±3˚C)
(For R thermocouple 0 to 500˚C, ± 1% FS ±1 digit ±3˚C)
Resistance bulb input: ± 0.5% FS ± 1 digit
Voltage/current input: ± 0.5% FS ± 1 digit
Rated voltage:
100 (-15%) to 240 (+10%)V AC 50/60Hz
24V AC (± 10%) 50/60Hz
24V DC (± 10%)
Power consumption: 6VA max. (at 100V AC or 24V AC)
8VA max. (at 200V AC)
3W max. (at 24V DC)
Ambient temperature: -10 to 50˚C
External dimensions: 48 (W) x 48 (H) x 93 (D) mm
Casing color: White or black (when front is waterproof)
Mass :
Approx.150 g
Mounting method: Panel flush mounting or rail mounting
(According to specified socket)
(Note) For socket, see OUTLINE DIAGRAM
Input signal and measuring range
• Table 1
Input signal
Resistance
bulb
JIS (IEC)
Resistance
bulb
JIS (IEC)
Group I
(Note 1)
Group II
Measurable
range (℃)(Note 2)
Minimum
span (℃)
Pt100Ω
-150 to 800
150
JPt100Ω
-150 to 600
150
J
0 to 800
ThermoK
0 to 1200
couple
R
0 to 1600
B
0 to 1800
T
0 to 1600
E
-199 to 400
S
-199 to 800
N
0 to 1300
PL-II
0 to 1300
DC voltage DC1 to 5V
Scalable range
DC0 to 5V
-1999 to 9999
DC current DC4 to 20mA
400
400
1600
1800
1600
599
800
1300
1300
Group III Thermistor Fuji-made
thermistor
-50 to 100
0 to 150
150
150
Note 1) Input signal can be changed by key operation within the same group,
but cannot be changed between different groups.
Note 2) When the measuring range exceed 1000˚C, decimal point cannot be
used.
EDS10-72b
Date
Nov. 27, 2003
PCP
• Table 2 Kind of alarm
Hold value Relay and
Alarm
method Set value at startup LED action
CODE SYMBOLS
Action diagram
Not
provided
LED lit,relay excited
Upper
limit
Absolute
value
provided
LED lit,relay excited
Absolute
value
Not
provided
LED lit,relay unexcited
Lower
limit
Absolute
value
Not
provided
LED lit,relay unexcited
Upper
limit
Absolute
value
provided
Lower
limit
Absolute
value
provided
LED lit,relay unexcited
LED lit,relay unexcited
4
Note 1
Input signal
Thermocouple
Resistance bulb Pt100, 3-wire type
Resistance bulb JPt, 3-wire type
Note 2
Thermistor
DC1 to 5V (Input impedance: 1MΩ)
Note 3
DC4 to 20mA
DC1 to 5V (Input impedance: 450Ω)
T
N
P
H
C
B
A
1
LED lit,relay excited
△
△
Revision No.
Alarm
With 2 alarm points
F
Instruction manual/power supply
Not provided/100 to 240 V AC,
50/60 Hz
Japanese version/100 to 240 V AC,
50/60 Hz
English version/100 to 240 V AC,
50/60 Hz
Not provided/ 24V DC
Japanese version/ 24V DC
English version/ 24V DC
N
Y
△
△
△
△
V
C
A
B
Socket
Not provided
0 0 0
4 0 0
For rail mounting, type: TP411X
5 0 0
For panel flush mounting, type:
Note 4
TP411SBA
Note 4
▲
Upper
limit
Description
▲
provided
14
Front size (W x H)
48 x 48mm
▲
Absolute
value
9 10 11 12 13
F
▲
Lower
limit
△
▲
Absolute
value
▲
Lower
limit
△
▲
LED lit,relay excited
▲
Not
provided
▲
△
Absolute
value
Y Y 1
PCP
Without
alarm
Upper
limit
1 2 3 4 5 6 7 8
With burnout releasing function
For rail mounting, type: TP411X
P 0 4
How to read action diagram
Note 4
For panel flush mounting,
Note 4
type: TP411SBA
Without socket
P 0 5
area : Range in which front LED is lit
area : Range in which alarm relay is excited
P 0 6
△ : Alarm set value
Y
D
Non-standard specification
White casing without front
waterproof structure
Black casing with front waterproof
Note 5
structure (NEMA-4X)
Note 1) Input signal has been factory-set as follows.
Thermocouple input: K thermocouple 0 to 400˚C
Resistance bulb input: 0 to 150˚C
Thermistor input: 0 to 150˚C
Voltage/current input: Scaling 0 to 100%
Note 2) Use a thermistor that matches Fuji Electric's specifications.
Note 3) Before use, attach the accessory 250Ω resistor between input
terminals.
Note 4) If a socket of any type other than specified is used, input
accuracy cannot be guaranteed.
Note 5) Contact us to select P for the 11th digit, because it is
nonstandard.
OUTLINE DIAGRAM (Unit:mm)
External dimensions
48
85.7
71.5
1
45
Panel
+0.5
0
+0.5
0
7
Panel cutout
57
63 MIN.
45
48
Mounting frame
48
44.8
57
63 MIN.
Lateral installation of multiple
indicators in close contact
t
Packing
1≦t≦8
(used for front waterproof structure)
Mounting frame
45
9.5
+0.5
0
+0.5
0
(48×N-3)
Note) Impossible for front
waterproof structure
2
CONNECTION DIAGRAM
• For AC power supply
<Measured value input>
Note 1
6
6
Thermocouple T.C.
B
6
250Ω
5
5
4 to 20mA DC
A
5
B
Thermistor
1 to 5V DC
6
+
5
6
4
SET2
4
Resistance bulb
5
8
SET1
9
3
Alarm output 2
7
2
10
1
Power supply
100 to 240V AC
50/60Hz
Alarm output 1
11
• For 24V DC power supply
<Measured value input>
Note 1
6
6
Thermocouple T.C.
B
6
250Ω
5
5
4 to 20mA DC
A
5
B
Thermistor
1 to 5V DC
6
+
5
6
4
SET2
4
Resistance bulb
5
3
Alarm output 2
7
8
SET1
9
2
10
1
Power supply
24V DC / 24V AC
50/60Hz
Alarm output 1
11
Note 1) For 4 to 20mA DC input, attach the accessory 250 Ω resistor between terminals t and y.
APPLICABLE SOCKET (Unit:mm)
TP411X type (rail mounting)
Terminal screw M3.5×8
TP411SBA type (panel mounting)
4.5
Terminal screw M3.5×7
φ27
5
5
11
1
3
6
7
8
9
1 11
2
45
2
7.8
45
3
4
4.5
16.3
6.2
10
7
4
Two φ4.5
9
holes
10
25.6
4
35.4
6
70
8 7
50
31.2 or less
40±0.2
Two φ4.5 mounting holes
3
PCP
SCOPE OF DELIVERY
Indicator main frame, fixture set, instruction manual (as
specified in code), soket (as specified in code), shunt resistor (250Ω) (for current input), panel face packing (for front
waterproof structure)
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan