MAGNETROL 941-722E-004

Echotel® Model 940/941
Ultrasonic Level Switch
DESCRIPTION
Echotel Model 940 and 941 Ultrasonic Level Switches
are compact integral units that utilize pulsed signal
technology to perform high or low level measurement
in a wide variety of liquid applications. These switches
feature a 316 stainless steel tip-sensitive transducer that
is offered in a variety of NPT, flanged, and sanitary
process connections. The compact electronics are
completely encapsulated just above the process fitting.
The Model 940 offers a 1-amp SPDT relay output. The
Model 941 has a mA current shift output.
Model 940/941
with Tri-Clamp® fitting
FEATURES
• Pulsed electronics for excellent performance in difficult process conditions and superior immunity from
sources of electromagnetic noise
• Tip-sensitive transducer gap provides reliable operation by draining viscous liquids and shedding foam
in turbulent applications
• Safety Integrity Level (SIL) data (FMEDA analysis) is
available. Model 940 is suitable for SIL 2 loops.
Model 941 is suitable for SIL 1 loops.
• Extremely compact unit is easily installed in areas
where access space is limited
• 1 amp SPDT relay output (940), or
mA current shift (941) output
• No calibration or configuration necessary
Model 940/941
with 3⁄4" NPT
Model 940/941
with ANSI flange
APPLICATIONS
The Model 940/941 can be applied in a wide variety of
high or low liquid level applications, pump protection,
and fill line monitoring as shown in the diagram at the
left. These compact units can also be installed in the
interstitial space between two tank walls for fluid leak
detection.
Filling line
High alarm
Small size and simplicity of installation make these units
ideal for OEM skids as a low cost, yet high performance
level measurement solution. They are also the perfect
replacement for older floats, conductivity switches, and
tuning forks.
Low alarm
High alarm
Low alarm
Pump
Pump
protection
Typical Mounting Orientations
TECHNOLOGY
Ultrasonic energy detects the presence or absence of liquid in a
tip sensitive transducer gap. The principle behind contact ultrasonic technology is that high frequency sound waves are easily
transmitted across a transducer gap in the presence of liquid, but
are attenuated when the gap is dry.
The transducer uses a pair of piezoelectric crystals that are
encapsulated in epoxy at the tip of the transducer. The crystals
are made of a ceramic material that vibrates at a given frequency
when subjected to an applied voltage. The transmit crystal converts the applied voltage from the electronics into an ultrasonic
signal. When liquid is present in the gap, the receive crystal is
able to sense the ultrasonic signal from the transmit crystal and
convert it back to an electrical signal. This signal is sent to the
electronics to indicate the presence of liquid in the transducer
gap. When there is no liquid present, the ultrasonic signal is
attenuated and is not detected by the receive crystal.
Customer
provided
conduit
Transmit
crystal
Ultrasonic
transducer
Receive
crystal
Figure 1
Ultrasonic signal transmission
across transducer gap
SPECIFICATIONS
Electrical:
Power Supply:
12–35 VDC
Power Consumption:
Less than 1 Watt
Signal Output:
Model 940:
SPDT relay, 1 amp @ 30 VDC, 0.5 amp @ 125 VDC, 0.5 amp @ 150 VAC
Model 941:
Dry Gap: 8 mA (±1 mA), Wet Gap: 16 mA (±1 mA)
Cabling:
18" (457 mm) flying leads of 18 AWG wires
Repeatability:
0.078" (2 millimeters)
Response Time:
1
⁄2 second typical
Environmental:
Ambient Temperature:
-40° to +185° F (-40° to +85° C)
Process Temperature:
-40° to +185° F (-40° to +85° C)
Maximum Pressure:
2000 psig (138 bar) for 1" (3 cm) and 2" (5 cm) transducers
1500 psig (103 bar) for transducers longer than 2" (5 cm)
Shock
ANSI/ISA-S71.03 Class SA1
Vibration
ANSI/ISA-S71.03 Class VC2
AGENCY
APPROVALS
AGENCY
APPROVED MODELS
PROTECTION METHOD
AREA CLASSIFICATION
FM & CSA
94X-XXXX-XXX
Non-Incendive
Class I, Div. 2,Groups A,B,C, & D
Class II, Div. 2, Groups E,F, & G
Class III, Type 4X, IP 66, T4
941-XXXX-XXX
Intrinsically Safe
Class I, Div. 1,Groups A,B,C, & D
Class II, Div. 1, Groups E,F, & G
Class III, Type 4X, IP 66, T4
These units have been tested to EN 50081-2
and EN 50082-2 and are in compliance with
the EMC Directive 89/336/EEC.
2
SPECIFICATIONS
Inches
(mm)
3/4" NPT
3/4" NPT
3/4" NPT
4.13 (105)
3.9 (99)
3.9 (99)
Actuation length
for NPT process
connection
Actuation length
for Tri-Clamp fitting
0.25 (6)
Actuation length
for ANSI flange
0.25 (6)
0.25 (6)
∅ .875 (22)
∅ .875 (22)
∅ .875 (22)
Model 940/941 with Tri-Clamp fitting
Model 940/941 with 3⁄4" NPT
Model 940/941 with ANSI flange
MODEL
NUMBER
Models available for quick shipment, usually within one week after factory
receipt of a purchase order, through the Expedite Ship Plan (ESP)
BASIC MODEL NUMBER
940
941
Low cost compact ultrasonic level switch with SPDT relay output
Low cost compact ultrasonic level switch with current shift output
AGENCY APPROVAL
1
7
FM/CSA Intrinsically Safe & Non-Incendive (Model 941 only)
FM/CSA Non-Incendive (Model 940 only)
PROCESS CONNECTIONS
11
22
23
33
43
3T
4T
3
⁄4" NPT
1" BSP (G1)*
1" 150# ANSI RF flange*
1.5" 150# ANSI RF flange*
2" 150# ANSI RF flange*
1"/1.5" Tri-Clamp 16 AMP fitting*
2" Tri-Clamp 16 AMP fitting*
* Minimum actuation length of 2 inches or 5 centimeters
ACTUATION UNITS
E
M
English (inches)
Metric (centimeters)
ACTUATION LENGTH
1 to 100 inches (Example: 4 inches = 004)
3 to 250 centimeters (Example: 6 cm = 006)
Available ESP length = 1" (001)
9
4
Note: Optional NEMA 4 cast aluminum junction
box assembly 052-7012-XXX is available
from the factory.
3
QUALITY
The quality assurance system in place at
Magnetrol guarantees the highest level of
Magnetrol’s quality assurance system is registered to ISO 9001 affirming its commitment
quality throughout the company. Magnetrol
is committed to providing full customer
satisfaction both in quality products and
to known international quality standards
providing the strongest assurance of product/service quality available.
quality service.
ESP
Several Echotel switches are available for
quick shipment, usually within one week
after factory receipt of a purchase order,
through the Expedite Ship Plan (ESP).
Expedite
Ship
Plan
Models covered by ESP service are color
coded in the selection data charts.
color coded model number codes (standard
dimensions apply).
ESP service may not apply to orders of ten
units or more. Contact your local representative for lead times on larger volume orders,
as well as other products and options.
To take advantage of ESP, simply match the
WARRANTY
All Magnetrol electronic level and flow controls
are warranted free of defects in materials or
workmanship for one full year from the date
of original factory shipment.
If returned within the warranty period; and, upon
factory inspection of the control, the cause of the
claim is determined to be covered under the warranty; then, Magnetrol will repair or replace the
control at no cost to the purchaser (or owner)
other than transportation.
Magnetrol shall not be liable for misapplication,
labor claims, direct or consequential damage
or expense arising from the installation or use of
equipment. There are no other warranties
expressed or implied, except special written
warranties covering some Magnetrol products.
5300 Belmont Road • Downers Grove, Illinois 60515-4499 • 630-969-4000 • Fax 630-969-9489 • www.magnetrol.com
145 Jardin Drive, Units 1 & 2 • Concord, Ontario Canada L4K 1X7 • 905-738-9600 • Fax 905-738-1306
Heikensstraat 6 • B 9240 Zele, Belgium • 052 45.11.11 • Fax 052 45.09.93
Regent Business Ctr., Jubilee Rd. • Burgess Hill, Sussex RH15 9TL U.K. • 01444-871313 • Fax 01444-871317
Copyright © 2009 Magnetrol International, Incorporated. All rights reserved. Printed in the USA.
Performance specifications are effective with date of issue and are subject to change without notice.
Magnetrol & Magnetrol logotype and Echotel® are registered trademarks of Magnetrol International.
BULLETIN: 51-136.2
EFFECTIVE: September 2009
SUPERSEDES: August 2006