EMERSON-NETWORKPOWER NLP65MEDICAL

Embedded Power for
Business-Critical Continuity
Rev.02.28.07
NLP65_3300
1 of 4
NLP65 Medical
Series
Single, dual
and triple output
Total Power:
65W
Input Voltage: 85 - 264 VAC
# of Outputs: Single, dual,
triple
Special Features
• 85 VAC to 264 VAC
universal input range
• Harmonic current correction
as standard
• Maximum component height
1.26 inches
• UL, CSA and VDE
safety approvals
• Overvoltage and
short circuit protection
• 5 x 3 x 1.26 inch (127.0 x 76.2 x
32mm) footprint
• Available RoHS compliant
• 2 year warranty
Safety
VDE0705/EN60601-1/IEC1010
File No. 10401-3336-0156/32480
Licence No. 121949
UL1950 File No. E147937
CSA C22.2 No. 950
File No. LR41062C
The NLP65 Medical series is a 65 W universal input ac-dc power
supply with input harmonic current correction as standard. This
compact design, packed in a 5 x x3 inch card with a maximum
component height of 1.26 inches, is ideal for use in a variety of
medical, laboratory and dental applications such as centrifuges,
incubators and infusion pumps. The NLP65 Medical series
comprises of eight models in single, dual and triple output
configurations. Providing 65 Watts of continuous output power
with free air convection, the NLP65 Medical series will deliver 75
Watts output power with 20 CFM of air. The series, with full
medical safety approval to EN60601 and UL2601, is CE marked,
greatly accelerating design-in time and reducing system
compliance costs.
Embedded Power for
Business-Critical Continuity
Rev.02.28.07
NLP65_3300
2 of 4
Specifications
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
OUTPUT SPECIFICATIONS
Output power
EMC CHARACTERISTICS
Natural convection
Total regulation
65 W max.
See table
Rise time
At turn-on
Transient response
Main output
25% step
at 0.1 A/μs
Temperature coefficient
1.0 s, max.
5.0% max. dev.,
1ms recovery
to 1.0%
±0.02%/°C
Overvoltage protection
Main outputs
Short circuit protection
Cyclic operation
125%, ±10%
Yes
INPUT SPECIFICATIONS
Input voltage range
(See Note 2)
Universal input
Input frequency range
47-63 Hz
Input surge current
(cold start)
120 Vac
230 Vac
Safety ground
leakage current
264 Vac, 60 Hz
Input current
120 Vac
230 Vac
Input fuse
85-264 Vac
17 A max.
32 A max.
95 μA
1.05 A rms
0.51 A rms
Conducted emissions
Radiated emissions
ESD air
ESD contact
Surge
Fast transients
Radiated immunity
Conducted immunity
EN55022, FCC part 15
EN55022, FCC part 15
EN61000-4-2, level 3
EN61000-4-2, level 4
EN61000-4-5, level 3
EN61000-4-4, level 3
EN61000-4-3, level 3
EN61000-4-6, level 3
Level A
Level A
Perf. criteria 1
Perf. criteria 1
Perf. criteria 1
Perf. criteria 1
Perf. criteria 2
Perf. criteria 2
120 Vac, 60 Hz
16 ms @ 65 W
GENERAL SPECIFICATIONS
Hold-up time
Efficiency
120 Vac, 65 W
72% typ.
Isolation voltage
Input/output
Input/chassis
4000 Vac
1500 Vac
Switching frequency
Fixed
100 kHz, ±5 kHz
Approvals and
standards
EN60601, UL2601
CSA 22.2 No. 125
Weight
MTBF
283 g (10 oz)
MIL-HDBK-217F
150,000 hours
ENVIRONMENTAL SPECIFICATIONS
Thermal performance
250 Vac F 3.15 A
Operating ambient,
(See derating curve)
Non-operating
0 ¡C to 50 ¡C ambient,
convection cooled
50 ¡C to 70 ¡C ambient,
convection cooled
Relative humidity
Non-condensing
Altitude
Operating
Non-operating
Vibration (See Note 5)
5-500 Hz
Shock
per MIL-STD-810E
0 °C to +70 °C
-40 °C to +85 °C
65 W
Derate to
50% load
5% to 95% RH
10,000 feet max.
30,000 feet max.
2.43 G rms approx.
516.4 Part IV
Embedded Power for
Business-Critical Continuity
Rev.02.28.07
NLP65_3300
3 of 4
Specifications Contd.
OUTPUT CURRENT
OUTPUT
VOLTAGE
MAX (1)
PEAK
2
3
4
5
6
TOTAL
MODEL
REGULATION (6)
NUMBER (11,12)
NLP65-9908J
+5 V
7A
9.1 A
8A
50 mV
±2.0%
+12 V
2.5 A
3.3 A
3A
150 mV
±5.0%
-12 V
0.5 A
0.81 A
1A
120 mV
±5.0%
+5 V
7A
9.1 A
8A
50 mV
±2.0%
+15 V
2.2 A
2.9 A
2.5 A
150 mV
±5.0%
-15 V
0.65 A
0.85 A
0.8 A
150 mV
±5.0%
+5 V
7A
9.1 A
8A
50 mV
±2.0%
+24 V
2A
2.6 A
2A
240 mV
±5.0%
NLP65-9910J
NLP65-9920J
+5 V
7A
9.1 A
8A
50 mV
±2.0%
+12 V
2.5 A
3.3 A
3A
150 mV
±5.0%
+5 V
10 A
13 A
12 A
50 mV
±2.0%
NLP65-9905J
+12 V
5.4 A
7A
6.5 A
120 mV
±2.0%
NLP65-9912J
+15 V
4.4 A
5.7 A
5.3 A
150 mV
±2.0%
NLP65-9915J
+24 V
2.7 A
3.5 A
3.5 A
240 mV
±2.0%
NLP65-9924J
Notes
1
RIPPLE (4)
FAN (10)
Natural convection cooling. Models NLP65-9929J, NLP65-9908J,
NLP65-9910J must not exceed 62.5 Watts continuous output power with
natural convection. Model NLP65-9920J not to exceed 65 Watts continuous
output power with natural convection.
When the input voltage is less than 90 Vac the operating temperature range is
0 ° C to +40 °C. The ripple and regulation specifications may not be met.
Peak output current lasting less than 60 seconds with duty cycle less than 5%.
During peak loading, output voltage may exceed total regulation limits.
Figure is peak-to-peak for convection power rating. Output noise
measurements are made across a 20MHz bandwidth using a 6 inch twisted
pair, terminated with a 10 μF electrolytic capacitor and a 0.1 μF ceramic
capacitor.
Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G rms 5
Hz to 500 Hz.
To maintain stated regulation then:
for single output units
I ³ 0.2 A I max.
for multiple output units
0.25 ² I(A)/I(B) ² 5, for I(A) ³ 0.2 A I(A) max.
NLP65-9929J
For optimum reliability, no part of the heatsink should exceed 120 °C, and
no semiconductor case temperature should exceed 130 °C.
8 CAUTION: Allow a minimum of 1 second after disconnecting line power when
making thermal measurements.
9 This product is only for inclusion by professional installers within other
equipment and must not be operated as a stand alone product.
10 Maximum continuous output power for all multiple output models must not
exceed 75 Watts with 20 CFM forced air cooling at 50 °C.
11 The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant.
TSE RoHS 5/6 (non Pb-free) compliant versions may be available on special
request, please contact your local sales representative for details.
12 NOTICE: Some models do not support all options. Please contact your local
Artesyn representative or use the on-line model number search tool at
http://www.artesyn.com/powergroup/products.htm to find a suitable
alternative.
7
Embedded Power for
Business-Critical Continuity
Rev.02.28.07
NLP65_3300
4 of 4
Americas
5.00
(127.00)
4.550
(115.57)
J1
3 N
0.225
(5.715)
Europe (UK)
J3
1
1 L
3.00
(76.20)
6
2.550
(64.770)
T1
J2
ALL DIMENSIONS IN INCHES (mm)
Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX
United Kingdom
Telephone: +44 (0) 1384 842 211
Facsimile: +44 (0) 1384 843 355
Asia (HK)
C19
Mounting
Hole Diameter
0.156 (4.00)
5810 Van Allen Way
Carlsbad, CA 92008
USA
Telephone: +1 760 930 4600
Facsimile: +1 760 930 0698
MAXIMUM COMPONENT
HEIGHT 1.26 (32.00)
TOLERANCE: .xxx ± 0.005"
.xx ± 0.02"
16th - 17th Floors, Lu Plaza
2 Wing Yip Street, Kwun Tong
Kowloon, Hong Kong
Telephone: +852 2176 3333
Facsimile: +852 2176 3888
For global contact, visit:
www.powerconversion.com
[email protected]
While every precaution has been taken to ensure
accuracy and completeness in this literature, Emerson
Network Power assumes no responsibility, and disclaims
all liability for damages resulting from use of this
information or for any errors or omissions.
Emerson Network Power.
The global leader in enabling
business-critical continuity.
AC Power
Connectivity
DC Power
Embedded Power
Inbound Power
Integrated Cabinet Solutions
Outside Plant
Precision Cooling
Site Monitoring and Services
EmersonNetworkPower. com
Emerson Network Power and the Emerson
Network Power logo are trademarks and
service marks of Emerson Electric Co.
©2007 Emerson Electric Co.