Embedded Power for Business-Critical Continuity Rev.09.10.07 NLP250 Medical Series 1 of 4 NLP250 Medical Series RoHS Single output Total Power: 250W Input Voltage: 85 - 264VAC # of Outputs: Single Special Features • Active PFC and EN61000-3-2 compliant • 250 W on main channel with forced air • Low profile fits 1U applications • U-Channel for maximum thermal performance • Optional cover (CJ suffix) • 5 V standby output • 12 V fan output • Integrated control and monitoring features • Compliance to EN55022-B conducted noise standard • RoHS compliant • 2 year warranty Electrical Specifications Input Input voltage range Universal input Input frequency range Input surge current 85-264 Vac 47-63 Hz 264 Vac (cold start) 40 A max. Safety ground leakage current 264 Vac, 50 Hz 150 μA Input current 120 Vac @ 250 W 230 Vac @ 250 W 2.78 A rms 1.36 A rms Input fuse UL/IEC127 T6.3 AH, 250 Vac In live and neutral Maximum power 200 LFM fored air 250 LFM with cover 250 watts Total regulation (line and load) Turn-on delay Main output Auxiliary outputs @ 120 Vac Input ±2.0% ±5.0% 2.0 s max. Transient response Main output 50-100% step at 0.5 A/μs 5.0% or 250 mV max. dev., 1 ms max. recovery to 1% Output Safety Temperature coefficient VDE0750/EN60950 IEC950/IEC60601-1 File No. 1040100-3336-0210 Overvoltage protection Main output 115%, ±5% Short circuit protection Minimum output current Cyclic operation Singles Continuous 0A Auxiliary outputs (See Note 8) 5 Vsb 12 V (fan) 5 V @ 1.0 A 12 V @ 0.3 A UL60601-1 File No. E135734 Certificate No. 40014041 CB Ref DE1-32468 ±0.02%/°C All specifications are typical at nominal input, full load at 25°C unless otherwise stated Embedded Power for Business-Critical Continuity Rev.09.10.07 NLP250 Medical Series 2 of 4 EMC Characteristics (5) Conducted emissions EN55022, FCC part 15 Level B Harmonic current correction EN61000-3-2 Compliant ESD air EN61000-4-2 Level 3 ESD contact EN61000-4-2 Level 3 Radiated immunity EN61000-4-3 Level 3 Fast transients Surge EN61000-4-4 EN61000-4-5 Level 3 Level 3 Conducted immunity EN61000-4-6 Level 3 Hold-up time 85 Vac @ 60 Hz 20 ms @ 250 W Efficiency 115 Vac @ 250 W 230 Vac @ 250 W 84% typ. 86% typ. Isolation voltage Input/output Input/chassis 4000 Vac 2000 Vac General Specifications Safety approvals (see note 6) UL/cUL UL60601-1, VDE EN60601-1 CAN/CSA22.2 No. 601-1 Weight 650g (22oz) MTBF (@25º C) Telcordia SR-332 MIL-HDBK-217F 317,000 hours min. 158,000 hours min. Environmental Specifications Thermal performance Operating ambient, 0° C to +70 °C (See derating curve) Non-operating -40 °C to +85 °C 0 °C to 50 °C ambient, 250 W 200 LFM forced air 250 LFM with cover 0 °C to 50 °C ambient, 175 W 0 °C to 40 °C with cover convection cooled 50 °C to 70 °C ambient, Derate linearly convection cooled to 50% load Relative humidity Non-condensing 5-95% RH Altitude Operating 10,000 feet max. Non-operating 30,000 feet max. Vibration (See Note 7) 5-500 Hz 2.4 G rms peak Shock per MIL-STD-810E 516.4 Part IV Embedded Power for Business-Critical Continuity Rev 09.10.07 NLP250 Medical Series 3 of 4 Ordering Information Output Current Output Voltage Min Max (free air) (1,4) Max (forced air) (2,4) 12 V 0A 14.6 A 21 A 24 V 0A 7.3 A 10.5 A Notes 1 2 3 4 Free air convection. Maximum continuous output power not to exceed 175 W. Refer to Figure 1 for the derating curve. 200 LFM (250 LFM with cover) forced air cooling from the longer side. Maximum continuous output power not to exceed 250 W. Figure is peak-to-peak for room temperature rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated with a 10 μF tantalum capacitor and a 0.1 μF ceramic capacitor. CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements. For optimum reliability no part of the heatsink should exceed 115 °C and no semi-conductor case temperature should exceed 120 °C. Total Model Regulation Numbers (9,10) 120 mV ±2.0% NLP250N-99S12J 240 mV ±2.0% NLP250N-99S24J Ripple (3) 5 No external filtering required during conducted emissions testing but some applications may require additional filtering to achieve system compliance. Compliance with radiated EMI specifications may require mounting in a suitable enclosure. 6 This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product. 7 Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G 8 5 V sb (standby) output is available whenever AC is present, regardless of remote ON/OFF signal status. 12 V (fan) present when main output is present. 9 The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. "CJ" suffix indicates covered RoHS version. 10 NOTICE: Some models do not support all options. Please contact your local Artesyn representative or use the on-line model number search tool at http://www.artesyn.com/powergroup/products.htm to find a suitable alternative. Figure 2: Mechanical Drawing CUSTOMER MOUNTING HOLES MAX. SCREW PENETRATION 0.08IN 6-32 1 J4 J2 1 5.850 (148.59) 1 0.25 (6.35) 7.000 [177.80] M3.0 6.250 (158.75) M3.0 4 PLACES 6-32 4 PLACES 4.700 (119.38) 1.500 [38.10] 5.500 (139.70) BOTTOM VIEW 2.500 (63.50) JP1 VR7 4.000 (101.60) J3 1.140 (28.96) 0.750 (19.05) J1 G L N Embedded Power for Business-Critical Continuity Figure 1b: Derating Curve With Cover Output Power (Watts) Figure 1: Derating Curve Output Power (Watts) DERATING CURVE Output Power (Watts) DERATING CURVE Output Power (Watts) WITH COVER All Models Forced Air 250 W 250 W 24 V Free Air 200 LFM (1.0 m/s) FORCED AIR COOLING 24 V Free Air 12 V Free Air 175 W 125 W 125 W 87 W 75 W NATURAL FREE AIR CONVECTION COOLING NATURAL FREE AIR CONVECTION COOLING 0W 0C All Models Forced Air 250 LFM (1.25 m/s) FORCED AIR COOLING 12 V Free Air 175 W Rev.09.10.07 NLP250 Medical Series 4 of 4 75 W 0W 10 C 20 C 30 C 40 C 50 C 60 C 70 C 0C 10 C 20 C 30 C 40 C 50 C 60 C 70 C Connector and Mating Connector Types Connector Type Mating Connector Type Molex 09-52-4054 (5239 series) or equivalent with Molex 08-52-0072 (2478 series) or equivalent crimp terminals J1 Molex 09-65-2058 (5273 series)void pins 2 and 4 or equivalent J2 Molex 22-01-3027 (2695 series) or equivalent with Molex 22-23-2021 (6373 series) or equivalent Molex 08-50-01113 (2759 series) or equivalent crimp terminals J3 Molex terminal block 387007504 or equivalent J4 Molex 22-01-3097 (2695 series) or equivalent with Molex 22-23-2091 (6373 series) or equivalent Molex 08-50-0113 (2759 series) or equivalent crimp terminals J1 PIN CONNECTIONS Pin 1 Ground/Earth Pin 2 Live Pin 3 Neutral Pin 1 Pin 2 Pin 3 Pin 4 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Pin 1 Pin 2 Terminal block contains #6-32 screw with clamp washer suitable for wire size 12-22 awg (0.5-2.5 mm2). Max Torque tp 1.36 Nm (12 in.lb) J2 PIN CONNECTIONS +12 V Fan Voltage SGND Return J3 PIN CONNECTIONS Vo +Main Output Vo +Main Output RTN Main Return RTN Main Return J4 PIN CONNECTIONS +S +Vo Remote Sense -S -Vo Remote Sense LS Load Share Signal PS OFF Remote ON/OFF signal NO PS ON Remote ON/OFF signal NC SGND Signal Common PW OK Power Good 5 Vsb Stand-by Voltage DC OK DC Power Good Signal Americas 5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698 Europe (UK) Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355 Asia (HK) 16th - 17th Floors, Lu Plaza 2 Wing Yip Street, Kwun Tong Kowloon, Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888 For global contact, visit: www.powerconversion.com [email protected] While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Power Inbound Power Integrated Cabinet Solutions Outside Plant Precision Cooling Site Monitoring and Services EmersonNetworkPower. com Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. ©2007 Emerson Electric Co.