EMERSON-NETWORKPOWER NLP250MEDICAL

Embedded Power for
Business-Critical Continuity
Rev.09.10.07
NLP250 Medical Series
1 of 4
NLP250 Medical
Series
RoHS
Single output
Total Power:
250W
Input Voltage: 85 - 264VAC
# of Outputs: Single
Special Features
• Active PFC and
EN61000-3-2 compliant
• 250 W on main channel with
forced air
• Low profile fits 1U applications
• U-Channel for maximum
thermal performance
• Optional cover (CJ suffix)
• 5 V standby output
• 12 V fan output
• Integrated control and monitoring features
• Compliance to EN55022-B
conducted noise standard
• RoHS compliant
• 2 year warranty
Electrical Specifications
Input
Input voltage range
Universal input
Input frequency range
Input surge current
85-264 Vac
47-63 Hz
264 Vac (cold start)
40 A max.
Safety ground leakage current 264 Vac, 50 Hz
150 μA
Input current
120 Vac @ 250 W
230 Vac @ 250 W
2.78 A rms
1.36 A rms
Input fuse
UL/IEC127
T6.3 AH, 250 Vac In live and neutral
Maximum power
200 LFM fored air
250 LFM with cover
250 watts
Total regulation
(line and load)
Turn-on delay
Main output
Auxiliary outputs
@ 120 Vac Input
±2.0%
±5.0%
2.0 s max.
Transient response
Main output
50-100%
step at 0.5 A/μs
5.0% or 250 mV
max. dev., 1 ms max.
recovery to 1%
Output
Safety
Temperature coefficient
VDE0750/EN60950
IEC950/IEC60601-1
File No. 1040100-3336-0210
Overvoltage protection
Main output
115%, ±5%
Short circuit protection
Minimum output current
Cyclic operation
Singles
Continuous
0A
Auxiliary outputs
(See Note 8)
5 Vsb
12 V (fan)
5 V @ 1.0 A
12 V @ 0.3 A
UL60601-1
File No. E135734
Certificate No. 40014041
CB Ref DE1-32468
±0.02%/°C
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
Embedded Power for
Business-Critical Continuity
Rev.09.10.07
NLP250 Medical Series
2 of 4
EMC Characteristics (5)
Conducted emissions
EN55022, FCC part 15
Level B
Harmonic current correction EN61000-3-2
Compliant
ESD air
EN61000-4-2
Level 3
ESD contact
EN61000-4-2
Level 3
Radiated immunity
EN61000-4-3
Level 3
Fast transients
Surge
EN61000-4-4
EN61000-4-5
Level 3
Level 3
Conducted immunity
EN61000-4-6
Level 3
Hold-up time
85 Vac @ 60 Hz
20 ms @ 250 W
Efficiency
115 Vac @ 250 W
230 Vac @ 250 W
84% typ.
86% typ.
Isolation voltage
Input/output
Input/chassis
4000 Vac
2000 Vac
General Specifications
Safety approvals (see note 6) UL/cUL UL60601-1, VDE EN60601-1
CAN/CSA22.2 No. 601-1
Weight
650g (22oz)
MTBF (@25º C)
Telcordia SR-332
MIL-HDBK-217F
317,000 hours min.
158,000 hours min.
Environmental Specifications
Thermal performance
Operating ambient,
0° C to +70 °C
(See derating curve)
Non-operating
-40 °C to +85 °C
0 °C to 50 °C ambient,
250 W
200 LFM forced air
250 LFM with cover
0 °C to 50 °C ambient,
175 W
0 °C to 40 °C with cover
convection cooled
50 °C to 70 °C ambient,
Derate linearly
convection cooled
to 50% load
Relative humidity
Non-condensing
5-95% RH
Altitude
Operating
10,000 feet max.
Non-operating
30,000 feet max.
Vibration (See Note 7)
5-500 Hz
2.4 G rms peak
Shock
per MIL-STD-810E
516.4 Part IV
Embedded Power for
Business-Critical Continuity
Rev 09.10.07
NLP250 Medical Series
3 of 4
Ordering Information
Output Current
Output
Voltage
Min
Max (free air) (1,4)
Max (forced air) (2,4)
12 V
0A
14.6 A
21 A
24 V
0A
7.3 A
10.5 A
Notes
1
2
3
4
Free air convection. Maximum continuous output power not to exceed 175 W.
Refer to Figure 1 for the derating curve.
200 LFM (250 LFM with cover) forced air cooling from the longer side.
Maximum continuous output power not to exceed 250 W.
Figure is peak-to-peak for room temperature rating. Output noise
measurements are made across a 20 MHz bandwidth using a 6 inch twisted
pair, terminated with a 10 μF tantalum capacitor and a 0.1 μF ceramic
capacitor.
CAUTION: Allow a minimum of 1 second after disconnecting line power when
making thermal measurements. For optimum reliability no part of the heatsink
should exceed 115 °C and no semi-conductor case temperature should exceed
120 °C.
Total
Model
Regulation
Numbers (9,10)
120 mV
±2.0%
NLP250N-99S12J
240 mV
±2.0%
NLP250N-99S24J
Ripple (3)
5
No external filtering required during conducted emissions testing but some
applications may require additional filtering to achieve system compliance.
Compliance with radiated EMI specifications may require mounting in a
suitable enclosure.
6 This product is only for inclusion by professional installers within other
equipment and must not be operated as a stand alone product.
7 Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G
8 5 V sb (standby) output is available whenever AC is present, regardless of
remote ON/OFF signal status. 12 V (fan) present when main output is present.
9 The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant.
"CJ" suffix indicates covered RoHS version.
10 NOTICE: Some models do not support all options. Please contact your local
Artesyn representative or use the on-line model number search tool at
http://www.artesyn.com/powergroup/products.htm to find a suitable
alternative.
Figure 2: Mechanical Drawing
CUSTOMER MOUNTING HOLES
MAX. SCREW PENETRATION 0.08IN
6-32
1
J4
J2
1
5.850 (148.59)
1
0.25 (6.35)
7.000 [177.80]
M3.0
6.250 (158.75)
M3.0
4 PLACES
6-32
4 PLACES
4.700 (119.38)
1.500 [38.10]
5.500 (139.70)
BOTTOM VIEW
2.500 (63.50)
JP1
VR7
4.000 (101.60)
J3
1.140
(28.96)
0.750 (19.05)
J1
G
L
N
Embedded Power for
Business-Critical Continuity
Figure 1b: Derating Curve With Cover
Output Power (Watts)
Figure 1: Derating Curve
Output Power (Watts)
DERATING CURVE Output Power (Watts)
DERATING CURVE Output Power (Watts)
WITH COVER
All Models
Forced Air
250 W
250 W
24 V Free Air
200 LFM (1.0 m/s)
FORCED AIR COOLING
24 V Free Air
12 V Free Air
175 W
125 W
125 W
87 W
75 W
NATURAL FREE AIR
CONVECTION COOLING
NATURAL FREE AIR
CONVECTION COOLING
0W
0C
All Models
Forced Air
250 LFM (1.25 m/s)
FORCED AIR COOLING
12 V Free Air
175 W
Rev.09.10.07
NLP250 Medical Series
4 of 4
75 W
0W
10 C
20 C
30 C
40 C
50 C
60 C
70 C
0C
10 C
20 C
30 C
40 C
50 C
60 C
70 C
Connector and Mating Connector Types
Connector
Type
Mating Connector Type
Molex 09-52-4054 (5239 series) or equivalent with
Molex 08-52-0072 (2478 series) or equivalent crimp
terminals
J1
Molex 09-65-2058 (5273 series)void pins 2
and 4 or equivalent
J2
Molex 22-01-3027 (2695 series) or equivalent with
Molex 22-23-2021 (6373 series) or equivalent Molex 08-50-01113 (2759 series) or equivalent crimp
terminals
J3
Molex terminal block 387007504 or
equivalent
J4
Molex 22-01-3097 (2695 series) or equivalent with
Molex 22-23-2091 (6373 series) or equivalent Molex 08-50-0113 (2759 series) or equivalent crimp
terminals
J1 PIN CONNECTIONS
Pin 1
Ground/Earth
Pin 2
Live
Pin 3
Neutral
Pin 1
Pin 2
Pin 3
Pin 4
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
Pin 9
Pin 1
Pin 2
Terminal block contains #6-32 screw with clamp
washer suitable for wire size 12-22 awg (0.5-2.5 mm2).
Max Torque tp 1.36 Nm (12 in.lb)
J2 PIN CONNECTIONS
+12 V
Fan Voltage
SGND
Return
J3 PIN CONNECTIONS
Vo
+Main Output
Vo
+Main Output
RTN
Main Return
RTN
Main Return
J4 PIN CONNECTIONS
+S
+Vo Remote Sense
-S
-Vo Remote Sense
LS
Load Share Signal
PS OFF
Remote ON/OFF signal NO
PS ON
Remote ON/OFF signal NC
SGND
Signal Common
PW OK
Power Good
5 Vsb
Stand-by Voltage
DC OK
DC Power Good Signal
Americas
5810 Van Allen Way
Carlsbad, CA 92008
USA
Telephone: +1 760 930 4600
Facsimile: +1 760 930 0698
Europe (UK)
Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX
United Kingdom
Telephone: +44 (0) 1384 842 211
Facsimile: +44 (0) 1384 843 355
Asia (HK)
16th - 17th Floors, Lu Plaza
2 Wing Yip Street, Kwun Tong
Kowloon, Hong Kong
Telephone: +852 2176 3333
Facsimile: +852 2176 3888
For global contact, visit:
www.powerconversion.com
[email protected]
While every precaution has been taken to ensure
accuracy and completeness in this literature, Emerson
Network Power assumes no responsibility, and disclaims
all liability for damages resulting from use of this
information or for any errors or omissions.
Emerson Network Power.
The global leader in enabling
business-critical continuity.
AC Power
Connectivity
DC Power
Embedded Power
Inbound Power
Integrated Cabinet Solutions
Outside Plant
Precision Cooling
Site Monitoring and Services
EmersonNetworkPower. com
Emerson Network Power and the Emerson
Network Power logo are trademarks and
service marks of Emerson Electric Co.
©2007 Emerson Electric Co.